Therapeutic equipment I

Slides:



Advertisements
Similar presentations
Appendix E Pacemakers Gail Walraven, Basic Arrhythmias, Seventh Edition ©2011 by Pearson Education, Inc., Upper Saddle River, NJ.
Advertisements

EE 4BD4 Lecture 20 Therapeutic Stimulation 1. 2 Weak and Strong Stimulation Recording Amplifier AC Coupled (bad choice) 3.
Current And Resistance
Chapter 17 Current Electricity. Conductors  Conductors are materials in which the electric charges move freely Copper, aluminum and silver are good conductors.
Chapter 11 The Cardiovascular System
Chapter 13-Webster Therapeutic and Prosthetic Devices
Current and Resistance
Pacemakers and Implantable Cardioverter-Defibrillators
Lecture 4 Defibrillator
Problems of creating physiologically matched artificial cardiac pacemakers Sergei Ovsjanski Tallinn University of Technology 2009.
Pacemaker Therapy and the Conducting System of the Heart By: Tom Kerrigan.
Pacemaker & its Classification
Artificial Pacemaker: The Heart
Pacemakers and Implanted Defibrillators Mike Harlan.
Single Chamber Temporary Pacing Operations & Troubleshooting.
Figure 13.1 Block diagram of an asynchronous cardiac pacemaker
Lecture 5 Cardioverter& Pacemakers
Cardiovascular System
Lecture 8 Electrical energy EMF Resistors in series and parallel
HAZARD OF MEDICAL INSTRUMENT by Mohd Yusof Baharuddin.
Cells in Series and Parallel (page 310) Cells in Series and Parallel Dry cells can be connected together into two basic types of circuits: series.
Neuroprosthetics Week 6 Design Issues. Neuroprosthetic Devices Three basic types Three basic types 1. Use neural signals as an input, possibly to control.
ELECTRODES FOR STIMULATION. Cardiac Pacemakers and Defibrillators  The heart is a natural example of a critical Bioelectromagnetic system.  An electrocardiogram.
Heart Rhythms: Normal or Abnormal (Arrhythmias) Anatomy & Physiology L2 and L3.
EE 4BD4 Lecture 24 Defibrillator Biomedical Device Technology: Principles and Design, Charles C. Thomas Publisher 2007 Anthony K. Chan 1.
Copyright © 2004 Lippincott Williams & Wilkins Chapter 13 The Heart.
Recap on the cardiovascular system and its response to exercise.
NEED FOR CARDIAC PACEMAKER Rhythmic beating of heart originates from SA node If SA node doesn’t function properly results in decreased heart rate & change.
PACEMAKER BATTERIES. Introduction -pacemaker  Small electrical device that runs on batteries.  Watches electrical signal of our heart and provide the.
Cardiovascular Physiology
EE 4BD4 Lecture 21 Cardiac Pacing.
Instructions for using this template. Remember this is Jeopardy, so where I have written “Answer” this is the prompt the students will see, and where.
CARDIOVASCULAR SYSTEM PHYSIOLOGY. HEART ACTIONS A cardiac cycle is a complete heartbeat During a cardiac cycle, the pressure in the heart chambers rises.
EE 4BD4 Lecture 22 Electrical Safety Electrode on Skin 4.
Heart. The Cardiovascular System The heart beats about 100,000 times/day, 35 million times/year In a resting adult the heart rate is about 75 beats/minute.
E LECTRONIC P ACEMAKER Dancho Grigorios B070049EC.
SINUSOIDAL CURRENT Dr. Amal Abd El Baky, LOGANATHAN CHANDRASEKAR 353 RHPT – 1435 – 1436H – 1 st SEM.
Introduction to the Nerve/Muscle Physiology Labs The digital oscilloscope, EMG, EKG.
Basics of Pacemaker Functioning
BME 312 BIOMEDICAL INSTRUMENTATION II LECTURER: ALİ IŞIN LECTURE NOTE 1 CARDIAC PACEMAKERS BME 312-BMI II-L1- ALİ IŞIN 2015 FACULTY OF ENGINEERING DEPARTMENT.
| rejuveindiameditour.com.
CARDIAC PACEMAKER Ms. Saranya N 27-Feb-18 Cardiac Pacemaker.
Basics of Medical Equipment Lecture 6
Action potential.
The Cardiovascular System
Chapter 13: The Heart.
Temporary Pacemakers.
Signal Generators Term 8.
Biomedical Instrumentation
Pacemaker II Lecture (6).
Senior Science Bionics
Hemodialysis Lecture 3.
PACEMAKER Yoga Yuniadi
Journal#2: Damage to the semilunar valve on the right side of the heart would affect blood flow to which vessel? Objective: Explain the events of the.
ELECTRICITY WITHIN THE BODY
Maintenance schedule for different equipment and their records in a hospital By Mr. Rahul kumar.
Iontophoresis Vineela.U 08B21A0538 CSE.
CARDIAC PACEMAKER A cardiac pacemaker is an electric stimulator that produces electric pulses that are conducted to electrodes normally located within.
EKGs and Pacemakers Cooper University Hospital
CARDIAC DEFIBRILLATORS
Diathermy and its safe use
17.5 Resistivity Expected: RL/A
CARDIAC DEFIBRILLATORS
Chapter 11 The Cardiovascular System
Automated External Defibrillation
Journal#2: Damage to the semilunar valve on the right side of the heart would affect blood flow to which vessel? Objective: Explain the events of the.
The heart..
Presentation transcript:

Therapeutic equipment I Lecture (5)

Electric stimulator A wide variety of electric stimulators is used in patient care and research. They range from very low current such as the cardiac pacemaker, to high current single pulse stimulator such as defibrillators.

Examples of electric stimulators Cardiac pacemaker Bladder stimulator Muscle stimulator defibrillators

Cardiac pacemaker The cardiac pacemaker is an electric stimulator that produces periodic electric pulses that are conducted to electrodes located on the surface of the heart (epicardium), within the heart muscle ( the myocardium), or within the cavity of the heart or the lining of the heart (endocardium).

The stimulus that conducted to the heart causes it to contract, this effect can be used in disease states in which the heart is not stimulated at a proper rate on its own. The principal pathologic conditions in which cardiac pacemakers are applied are known as heart block.

Types of cardiac pacemaker 1. Asynchronous/Fixed Rate. 2. Synchronous/Demand. 3. Single/Dual Chamber Sequential (A & V). 4.Programmable/non-programmable.

Asynchronous pacemaker

Power supply: provide the energy required for the operation of the pacemaker. Lithium-iodine cell is the current standard battery Advantages: (1) Long life – 4 to 10 years (2) Output voltage decreases gradually with time making sudden battery failure unlikely

The oscillator establishes the pulse rate for the pacemaker, this in turn, control the pulse output circuit that provide the stimulating pulse to the heart. This pulse is conducted along lead wires to the cardiac electrodes.

Another component of the overall construction that is the package itself. The package should be compatible and well tolerated by the body. It must provide the necessary protection to the circuit components, where the body is a corrosive environment, so the package must be designed to operate well in this environment, while occupying minimal volume and mass. Titanium and stainless steel are frequently used for the package.

Timing circuit: The asynchronous pacemaker represents the simplest kind of pacemaker, because it provides a train of stimulus pulses at a constant rate regardless of the functioning of the heart. A free running oscillator is all that is required for the timing pulse in such a system.

Output circuit: The pulse output circuit produces the actual electric stimulus that is applied to the heart. At each trigger from the timing circuit, the output circuit generates an electric stimulus pulse that has been optimized for stimulating the myocardium through the electrode system. Constant voltage or constant current amplitude pulses are the two usual types of stimuli produced by the output circuit.

Constant voltage amplitude pulses are typically in the range of 5 Constant voltage amplitude pulses are typically in the range of 5.0 to 5.5V with a duration of 500 to 600 micro-sec. Constant current amplitude pulses are typically in a range of 8 to 10 mA with pulse duration ranging from 1.0 to 1.2 mS. Rates for asynchronous pacemaker range from 70 to 90 beats per min. whereas pacemakers that are not fixed rate typically achieve rates ranging from 60 to 150 beats per min.

Implantation procedure The procedure is usually done under local anaesthesia The pulse generator is implanted under the skin The leads are inserted using x-ray control, via a vein found in this area, and positioned in the appropriate right sided heart chamber The leads are tested before the pulse generator is attached