Electron-beam lithography with the Raith EBPG

Slides:



Advertisements
Similar presentations
Lecture 11. Microscopy. Optical or light microscopy involves passing visible light transmitted through or reflected from the sample through a single or.
Advertisements

Microscope One or more lense that makes an enlarged image of an object. Occular lens Body Tube Arm Objective Lens Stage Stage Clips Base Diaphragm Light.
Components of Initial Eprobe Startup Electron Gun –turning on beam, setting beam current, acceleration, saturation level Focusing lenses –Permit focussed.
Fixed Beam Moving Stage using a stitching free exposure strategy.
Optimizing power to the detectors To avoid dark-current errors and bit-noise errors in your data, all four PSD detectors need a high SUM reading. However.
Scanning Electron Microscope د. عمر عبد القادر.
Introduction to scanning electron microscopy
EBL: Focus & height rev 0, 6/6/08. Electron gun ZrO/W emitter Suppressor First anode Second anode Acceleration electrodes Ground anode First alignment.
Introduction to Operation of the XL- 30 Scanning Electron Microscope Tutorial # 2 Introduction to Operation and Imaging.
How to Make a Simple DC Motor
Operation of the S4700 FESEM
Types of cameras, parts of the camera
Lab 3 - Centering. Centering; or the smart way to align centered optical elements and systems This lab will make use of concepts used in the previous.
Peripherals and Storage Looking at: Scanners Printers Why do we need storage devices anyway? What are magnetic disks? How do magnetic disks physically.
Vacuum tests on scintillating fibres. Table of content The Aim of the vacuum tests on the scintillating fibre plane Equipments used in the vacuum tests.
Sketchup Tutorial Designing for Construction. The Parts of a Wall.
Scanning Electron Microscope (SEM)
JEOL JBX-9300FS Electron Beam Lithography System Training
Sphere Standards and Standard Spheres Dr. Richard Young Optronic Laboratories, Inc.
Page 1 Workshop for Electron Beam Lithography System JBX-6300FS By Nelson LI 13 November 2009 Nanoelectronics Fabrication Facility.
Scanning Electron Microscope (SEM)
NANO 225 Micro/NanoFabrication Electron Microscopes 1.
Transmission Electron Microscope Basic premise of a TEM is to project a magnified image of the specimen onto a fluorescent screen where it can be viewed.
Proximity Effect in Electron Beam Lithography
Electron beam machining (EBM) – MM461 Dr. Dermot Brabazon Sch. Of Mech. and Manu. Eng. Dublin City University.
Hybrid panel heated plate study Currently the curing of the ASIC glue for the hybrid panels are done at room temperature. This is a study to determine.
Lithography in the Top Down Method New Concepts Lithography In the Top-Down Process New Concepts Learning Objectives –To identify issues in current photolithography.
Transmission Electron Microscope
Speaker: Shiuan-Li Lin Advisor : Sheng-Lung Huang
Best Practice T-Scan5 Version T-Scan 5 vs. TS50-A PropertiesTS50-AT-Scan 5 Range51 – 119mm (stand- off 80mm / total 68mm) 94 – 194mm (stand-off.
Copyright 2007 by Raith GmbH Adjust and Align Stage Adjustment & Writefield Alignment.
SEM Imaging - Magnification and Deflection System
Scanning Electron Microscope Eee-Jay Rodriguez. The Structure of the Microscope Inside Outside.
Prepared By: Ms. Carolyn R. Magat
Lab 3 - Centering.
Introduction to scanning electron microscopy
Pinhole Camera.
An introduction to the Compound Light Microscope (This is a note check, for p. 3) You must make sure you have these correct for studying them later.
Wisconsin Center for Applied Microelectronics
Introduction to scanning electron microscopy
Camera Basics.
LayoutBeamer, from GenISys GmbH
Laboratory equipment Lecture (3).
Electron-beam lithography with the Raith EBPG
Electron-beam lithography with the Raith EBPG
Electron-beam lithography with the Raith EBPG
LumiCal mechanical design, integration with LDC and laser alignment
NANO 230 Micro/NanoFabrication
Electron-beam lithography with the Raith EBPG
Electron-beam lithography with the Raith EBPG
Electron-beam lithography with the Raith EBPG
Living with e-beam drift Part 1: The e-beam is not broken
Microscope Basics T. Trimpe
Parts of the Compound Light Microscope
Getting the least from your e-beam
The Study of Mirrors and Lenses
Electron-beam lithography with the Raith EBPG
Advanced e-beam lithography overview
Electron-beam lithography with the Raith EBPG
Do Now(1) How many: Cm in 4 meters? Decimeters in 5 meters?
Basics E-beam lithography
Light Microscope: – light passes through one or more lenses to produce an enlarged image of a specimen compound light microscope – light microscope that.
Fundamentals of Data Representation
Light Microscope: – light passes through one or more lenses to produce an enlarged image of a specimen compound light microscope – light microscope that.
LITHOGRAPHY Lithography is the process of imprinting a geometric pattern from a mask onto a thin layer of material called a resist which is a radiation.
SEM Imaging - Magnification and Deflection System
Microscope Basics T. Trimpe
Controlling e-beam writing order to avoid stitching and drift problems
MACRO & CLOSE-UP PHOTOGRAPHY
Light Microscope: – light passes through one or more lenses to produce an enlarged image of a specimen compound light microscope – light microscope that.
Presentation transcript:

Electron-beam lithography with the Raith EBPG Part 3: Hardware M. Rooks, Yale University

100 kV accelerating potential electrons of course EBPG 5000+ System Hardware 100 kV accelerating potential electrons of course 20 bit main field DAC Digital to Analog Converter* 14 bit trapezium DAC for filling shapes 50 MHz maximum clock minimum dwell time 1/50 ms 6 inch stage 15 cm for you communists Two-stage magnetic deflection * You should know the meaning of “DAC” and “LSB”. Look it up.

Electron Source At the top of the column is the electron source, a thermal field emitter. Electrons are drawn out of the sharp tungsten tip by an extraction potential around 2 kV. The emitter pin is kept hot so that ZrO will flow down the surface, thereby lowering the work function and producing a larger current. Keeping the tip hot also makes the source more stable. This is exactly the same type of electron source commonly found in SEMs and TEMs. (Older systems used hot LaB6 emitters, but we do not speak of that.)

Electron Optics Electrons are generated from a field emitter in the gun, then focused by an electrostatic lens (C1) before accelerating through the anode. The magnetic lens (C2) focuses the beam to a crossover in the blanking cell, which is simply a plate which can be charged up to turn the beam off. Placing this blanking plate at the crossover allows the beam to turn off without shifting the image. The third lens (C3) is the objective, focusing the electrons to a spot on the sample. Inside C3 is a smaller, faster coil for rapid tweaks to the focus. The beam is automatically refocused every time the sample moves. Just like a SEM, the column contains coils to correct astigmatism, also adjusted automatically. C1 C2 C3

Stage Control The position of the stage is measured with a laser interferometer, with an accuracy below one nanometer. The interferometer uses a He-Ne laser and Zeeman splitting to create a GHz range beat frequency at the receivers. This frequency shifts as the stage moves, and so the Doppler effect can be used to measure the stage speed. Speed is integrated to find the stage position. Motors on the stage move in a sloppy manner, while the laser signal is used to deflect the beam electronically, thereby compensating for any jerky mechanical motion or mechanical vibration.

Main Field Deflecton Errors The EBPG uses the laser stage to automatically calibrate and compensate for gain, rotation, keystone and pincushion errors. Additionally, the focus and stigmation are dynamically corrected over the deflection field. If you ever wonder why the big system is better than a converted SEM, this is the answer. X, Y gain errors X, Y axis rotation errors Field keystone errors Pincusion and third-order distortion

Electron Optics – Bottom Stigmation coils & fast focus More sigmation & fast focus coils Secondary electron detectors – four unbiased scintillators on light pipes, with photomultipliers Electron Optics – Bottom Directly above the sample stage are the fast focus and stigmation coils. Secondary electrons from the sample (usually a silicon wafer) are detected by four scintillators on the ends of light-pipes, coupled to photomultiplier tubes. These detectors are similar to those in SEMs, but without the usual 300V bias drawing in the electrons. A bias on the ends of the detectors would distort the writing field. Unfortunately, without the bias these detectors are not very efficient. Other e-beam systems detect secondary electrons using a diode mounted under the objective lens, or with a scintillator built into the objective lens. It would be nice to have these more efficient detectors, but you can’t have everything.

System reads change in spot position on Dual Diode detector Laser Height Sensor Every time the stage moves, a height sensor corrects the focus and deflection gain. System reads change in spot position on Dual Diode detector Signals from Dual-diode detector determines distance moved by laser beam. Laser projects spot onto surface of substrate. Reflected laser beam Substrate Laser spot movement due to change in height of substrate surface. Laser spot reflected onto detector. If the laser beam hits the wafer’s edge or hits a metal fixture, then the height reading will fail. For this reason, patterns should be placed at least 1mm from any edge – especially if you are using a fixture for a small substrate.

Substrate Height Correction As the substrate height varies, the system must correct both focus and field size. Correcting the field size is critical, since this determines field stitching accuracy. You can see now why smaller field block sizes lead to smaller stitching errors. On the EBPG you should keep the block size less than 700 mm so that field stitching errors will be less than 50 nm. The maximum block size is equal to the maximum field, 1 mm. But it would be a mistake to use such a large deflection.

Errors caused by beam deflection Substrate Surface Focal Plain Objective lens & aperture Spot defocused and astigmatic when deflected Beam focused to form round spot at Field center. When the beam is deflected, the EBPG system corrects the focus, astigmatism, and field size. Dynamic focus, astigmatism, field gain, and height adjustment are all calibrated automatically before each exposure.

Exposure Strategy Software breaks a pattern into exposure fields (“blocks”) so that a large pattern can be stitched together by moving the stage. Inside each field the shapes are broken into simple parallelograms. Inside each parallelogram the beam paints the shape with a serpentine raster. The slower 20-bit main deflector positions the beam in the field, then the fast 14-bit trapezium deflector fills in each shape.

EBPG Sample Holders Top left: holder for 5 inch mask plates. Top right: wafer holder. Bottom left: two two-inch wafers (we do not have this one). Bottom right: piece holder. This is a bottom-reference piece holder which we do not have at Yale. This holder requires manual trimming to level the substrate and to move it into the correct plane. The next slide shows a better design. At Yale we have holders for 3, 4, and 6 inch wafers, as well as adapters for smaller substrates.

Adapters For Small Pieces These fixtures fit in place of a 4 inch wafer. The small substrate will be at the proper focus height, determined by its top surface. This is a better design than the one on the previous slide, because any junk on the back of the substrate will not tilt the sample, nor move it out of the focus plane. These fixtures are built at Yale, and so it is easy to make holders customized for your weird samples. Currently, we have fixtures with window sizes of 8mm, 10mm, 20mm, 38mm, and 2 inch (50mm). Your substrate must have one dimension at least 2mm larger than the window size. sample

Substrate fixturing ideas that do not work File under “It seemed like a good idea at the time” these bad ideas: Gluing the piece to a big wafer with PMMA, photoresist, or glue. Terrible idea – the solvent will not completely dry under the piece. In the vacuum system the solvent will boil, and the piece will pop off. Using vacuum grease to hold the piece. This is less stupid, but you will have a hard time getting the grease off the substrate, and you cannot put vacuum grease in the developer. Building an adapter with a “V” shaped window. Sorry – the angle will always be wrong, and the adapter will cover alignment marks, etc. Building an adapter with adjustable top rails. This idea is too complicated and too easily broken. We’ve been there, done that, and it doesn’t work.

Substrate fixturing ideas that sometimes work Clipping the wafer to a metal puck. This does work sometimes, but it requires a lot of fussing. The puck has to be shimmed to the correct plane, and the back of the wafer must be cleaned to avoid tilting. If the clip pushes down on a delicate spot, the substrate can easily break. Taping the substrate to a large wafer, with carbon tape. This does work, but the previous comments about shimming and cleaning also apply. In addition to fussing with shims, you will get carbon tape goo on the substrate. Also there is a risk of breaking the piece when removing the carbon tape. Using crystalbond wax to attach the little piece to a wafer. Yes, we have done this with diamonds, but it’s tricky. The substrate is not grounded, and so it must be coated with gold (on top of the resist). The wax can go into IPA/water or TMAH developer, and it can later be removed in acetone. The wafer/wax/diamond can go from the e-beam to developer, and then to the evaporator or etcher. The tricky bit is that liftoff in acetone will also lift off the little piece. Raith’s own butt-end clips (seen a few slides back) also work, after a fashion. Unfortunately, brittle materials like GaAs and InP tend to break when pushed from the side. You’ll also have the fun of trimming the height.

Small Substrates Are Bad Using substrates smaller than 1 cm is stupid for a number of reasons: Resist edge beak ruins the exposure within 1mm of the edge. Piece-part fixtures will cover another 2mm of the stubstrate. The laser height sensor will not operate closer than 1mm to the window, which means that you will lose 4 mm in one dimension and 2 mm in the other. That’s quite a lot of lost real estate on a 10mm piece. Alignment of a multi-layer device will be very difficult, because the piece must be pre-aligned to within 0.2 degrees. That means your alignment marks must be on-axis within 35 mm over 10mm. That’s not fun at all. If you are trying to save money by chopping up an expensive substrate, then think about how much e-beam time you are wasting by misaligning the substrate. One hour of ebeam time is $100. How much did you save? It takes 6 minutes to vent and 6 minutes to pump the load lock, as well as ~ 10 minutes to find your alignment marks. Therefore each piece has an overhead cost around $37. How much did you really save by chopping up a $50 wafer? Do the math.

Running the EBPG – preparing a pattern CAD e.g. with “layout” Output GDS format Running the EBPG – preparing a pattern Start by designing a pattern, then convert it to EBPG format with the program Beamer. Use cjob to set up the exposure, specifying the exposure dose and pattern placement on the substrate. Finally, coat the substrate with resist, put it in the EBPG, and start the exposure. pattern conversion with “beamer” Output GPF format pattern check with “cview” exposure setup with “cjob” Output job script exposure on EBPG with “beams”

Running the EBPG – a simple exposure Cjob produces a script that runs your exposure, but this is not required. You could expose a pattern with a simple set of commands, as shown here. Just type them into a terminal window. Complex jobs use the same commands, with more elaborate scripts created by cjob. pg info arc beam list beam files (that contain lens settings) pg arc restore beam 10na_300um_2 set the lenses for 10nA beam current mcur measure the beam current pg select pattern abc.gpf choose a pattern pg set resist 800.0 set the dose pg adjust ebpg calibration pg move position 75mm,75mm move somewhere pg expose pattern expose!

End of part 3 Now you should proceed to the quiz.