CONCLUSION INTRODUCTION RESULTS METHODS AND MATERIALS

Slides:



Advertisements
Similar presentations
Patient information extraction in digitized X-ray imagery Hsien-Huang P. Wu Department of Electrical Engineering, National Yunlin University of Science.
Advertisements

University of Minho School of Engineering Institute of Polymers and Composites Uma Escola a Reinventar o Futuro – Semana da Escola de Engenharia - 24 a.
RapidArc plan verification using ArcCHECK™
The Tomotherapy Experience at Advocate Good Samaritan Hospital
Photon Beam Monitor-Unit Calculations
Background:  IMRT has become the choice of treatment for disease sites that require critical structure sparing such as head and neck cancer.  It has.
The TrueBeam System ™ Clinic Name Presenter’s name Clinic location
Alamance Regional Cancer Center 2012
Real-time tumor tracking with preprogrammed dynamic MLC motion and adaptive dose-rate regulation B.Y Yi, S. Han-Oh, F. Lerma, B. Berman, C. Yu.
Lotte Verbunt Investigation of leaf positioning accuracy of two types of Siemens MLCs making use of an EPID.
What Happens in the days before treatment ?
Simulating Potential Layouts for a Proton Therapy Treatment Center Stuart Price-University of Maryland Bruce Golden- University of Maryland Edward Wasil-
Evaluation of New Pre-Treatment In-Air Patient Specific QA Software for TomoTherapy Treatments Lydia L. Handsfield¹, Quan Chen¹, Kai Ding¹, Wendel Renner²,
Quality Control in Radiation Therapy, A New Concept: Dosimetry Check
The days before Treatment
Comparison of Rectal Dose Volume Histograms for Definitive Prostate Radiotherapy Among Stereotactic Radiotherapy, IMRT, and 3D-CRT Techniques Author(s):
H Ariyaratne1,2, H Chesham2, J Pettingell2, K Sikora2, R Alonzi1,2
Diabetes mellitus and the incidence and time to onset of oxaliplatin-induced peripheral sensory neuropathy in patients with colorectal cancer: A pooled.
Test of the proposed method Introduction CCD Controller CCD Illuminator gel Filter 585nm Assembling the phantom before its irradiation. The phantom, ready.
AUTHORS (ALL): Huang, Xiaoyan 1, 2 ; Kuan, K M 2 ; Xiao, G L 2 ; Tsao, S Y 3, 2 ; Qiu, X B 2 ; Ng, K 2. INSTITUTIONS (ALL): 1. Radiation Oncology, Sun.
Simplified IMRT Plans Can Be Delivered with Conventional Jaws Ping Xia, Ph.D. University of California San Francisco.
1 4D: Adaptive Radiotherapy & Tomotherapy Bhudatt Paliwal, PhD Professor Departments of Human Oncology & Medical Physics University of Wisconsin Madison.
Surface dose prediction and verification for IMRT plans using line dose profiles † Ronald E. Berg, † Michael S. Gossman and ‡ Stephen J. Klash † Erlanger.
IMRT QA Plan Site 5%/3mm3%/3mm2%/2mm 0% noise1% noise2% noise0% noise1% noise2% noise0% noise1% noise2% noise HN
Institute for Advanced Radiation Oncology
Integrating the Health Care Enterprise- Radiation Oncology Use Case: In Vivo Patient Dosimetry Editor: Juan Carlos Celi - IBA Reviewer: Zheng Chang – Duke.
Dosimetric discrepancies due to positional errors in MLC movement during stereotactic lung VMAT A. Abbas, T. Karan, S. Kim, D. J. Moseley, M. M. Taremi,
1 A Comprehensive Study on the Heterogeneity Dose Calculation Accuracy in IMRT using an Anthropomorphic Thorax Phantom S Davidson 1, R Popple 2, G Ibbott.
The Answer to Your Dosimetry Needs
1 Radiotherapy, hadrontherapy and treatment planning systems. Faiza Bourhaleb INFN-Torino University Med 1er-Morocco  Radiotherapy  Optimization techniques.
Introduction The use of analgesics, sedatives, and paralytic agents are an important tool to help decrease pain and anxiety while improving the quality.
Data Mining BY JEMINI ISLAM. Data Mining Outline: What is data mining? Why use data mining? How does data mining work The process of data mining Tools.
2005/12/021 Fast Image Retrieval Using Low Frequency DCT Coefficients Dept. of Computer Engineering Tatung University Presenter: Yo-Ping Huang ( 黃有評 )
Somvilai Mayurasakorn, MD. Division of Therapeutic Radiology and Oncology, Faculty of Medicine, Chiang Mai University Somvilai Mayurasakorn, MD. Division.
Optimization of Volumetric Modulated Arc Therapy (VMAT) Planning Strategy Using Ring-shaped ROI for Localized Prostate cancer Kentaro Ishii, Masako Hosono,
1. Create an IMRT plan on the Cadplan/Helios system. When selecting the LMC options, select the “Extract Portal Image with” option and use 0 MU. This creates.
The Effects of Small Field Dosimetry on the Biological Models Used In Evaluating IMRT Dose Distributions Gene Cardarelli,PhD, MPH.
References : [1] Hurkmans CW, et al Set-up verification using portal imaging; review of current clinical practice. Radiother. Oncol. 2001;58: [2]
Treatment Chart Record of patients radiation therapy history. Must contain: History and diagnosis Rationale for treatment Treatment plan Consent Documentation.
Mathematical Derivation of Probability
Constructing a Predictor to Identify Drug and Adverse Event Pairs
CT Multi-Slice CT.
NCT: Gaining Medical Insights and Enhancing Care for Cancer Patients with SAP HANA® Organization National Center for Tumor Diseases (NCT) Heidelberg, part.
You Zhang, Jeffrey Meyer, Joubin Nasehi Tehrani, Jing Wang
Kasey Etreni BSc., MRT(T), RTT, CTIC
A. Sánchez-Reyes. , N. López Vilanova+, A. Vila. , M. A. Duch+, A
Blake R. Barker, MD, Shannon A
INTRODUCTION RESULTS DATA ANALYSIS DISCUSSION METHODS REFERENCES
Introduction and Objectives
Enhanced Recovery After Surgery (ERAS) clinical pathway for patients undergoing pancreatic surgery decreases hospital length of stay   Hayden P. Kirby,
Quality of treatment plans and accuracy of in vivo portal dosimetry in hybrid intensity- modulated radiation therapy and volumetric modulated arc therapy.
Duration of Therapy of Colony Stimulating Factors in Oncology
Template Matching Can Accurately Track Tumor Evaluation of Dose Calculation of RayStation Planning System in Heterogeneous Media Huijun Xu, Byongyong Yi,
Courtesy from Dr. McNutt
CONTACT Catalina A. Riley
Implementation of Object Spot Avoidance in Proton Pencil Beam Treatment on Whole Breast with Implant Metal Injector Peng Wang, PhD, DABR, Karla Leach,
Fig. 4. Percentage of passing rate between clinical and 544 plans.
Reducing Treatment Time and MUs by using Dynamic Conformal Arc Therapy for SBRT Breath-Hold Patients Timothy Miller, Sebastian Nunez Albermann, Besil Raju,
Insert tables Insert figure
Volumetric Modulated Arc Therapy (VMAT) versus Intensity Modulated Radiation Therapy (IMRT) for Anal Carcinoma Heather Ortega, BSRT(T), CMD, Kerry Hibbitts,
Insert tables Insert graphs Insert figure
L. A. den Otter. , R. M. Anakotta. , M. Dieters. , C. T. Muijs. , S
Tolerances.
Data and Interoperability:
Chapter 17 Intensity-Modulated Radiation Therapy
Toktobaeva B, Karymbaeva S Drug Information Centre Kyrgyzstan
TOLERANCES.
GHG meeting at ESTRO36 May, 2017
Rectal toxicity for patients treated with proton beam radiation diagnosed with prostate cancer using free rectal water, water filled rectal balloon or.
Average Dose-Volume Ratio
Presentation transcript:

CONCLUSION INTRODUCTION RESULTS METHODS AND MATERIALS Comparison of Compensator-IMRT and Segmental MLC-IMRT Techniques: A Retrospective Study on Treatment Time and Monitor Units in Clinical Application S Chang*, K Deschesne, L Levinson, L Potter, K Eljabaly, University of North Carolina, Chapel Hill, North Carolina INTRODUCTION Compensators have been successfully used clinically for dose-optimized IMRT treatment at the University of North Carolina at Chapel Hill since 1996. Today this valuable alternative IMRT delivery technique has been implemented in many cancer centers. There is a common conception that the manual compensator-IMRT treatment is much slower than the automated MLC-based IMRT delivery. Our years of experience in both the compensator-based and the segmental MLC-based IMRT (Siemens) indicated otherwise. In this retrospective study we compared the total monitor unit and the treatment delivery time in daily treatment using data stored in IMPAC R&V system for 150 randomly selected clinical IMRT cases. We also analyzed two components of the treatment time: the total monitor units per IMRT treatment and the time elapsed between treatment fields for both type of IMRT treatments. CONCLUSION Retrospective study of 150 clinical cases shows that the treatment time defined as time elapsed between the first field beam ON and the last field beam OFF for compensator-IMRT treatment is generally 50 to 100% faster than for Siemens segmental-IMRT treatment (Fig. 2). The average time elapsed between two fields (when therapists go in the treatment room and exchange compensators for compensator-IMRT cases) is no more than segmental MLC-IMRT treatments (Fig. 1). The compensator-IMRT in general requires similar or fewer monitor units compared to the MLC-IMRT technique (Fig. 3). The study is based on older generation Siemens accelerators, which is reportedly ~30% slower than newer generation accelerators for IMRT. RESULTS MLC-IMRT Compensator-IMRT (1-50nm in diameter, ~1-10mm long tube) Figure 1 (above) shows the average time elapsed between two treatment fields in five days for 57 IMRT treatments recorded in IMPAC R&V system. The data is presented individually for different treatment sites (H&N, prostate, or other) and the number of manual table kick-outs. The data variation is significant which is likely due to the specific level of care each patient needed. However, the data shows clearly that, contrary to common belief, the manual compensator-IMRT treatments did not use more time than the automated MLC-IMRT treatment between treatment fields in real clinical situations. METHODS AND MATERIALS One hundred fifty (150) randomly-selected IMRT cases treated in our department since 2001 were used in the retrospective study. Most of the 7-9 fields per treatment cases, however, are treated in the last two years. The IMRT treatments were either compensator-IMRT or segmental MLC-IMRT on one of our three Siemens accelerators (2 Primus and 1 MD2). IMRT treatment planning was done on our in-house TPS PLanUNC using the same dose optimization engine for both IMRT delivery techniques. For compensator-IMRT, the high-resolution intensity maps are directly converted into compensator files for fabrication; for segmental MLC-IMRT the maps are first truncated into discrete maps and then MLC segments are generated using IMMAXX algorithm from Siemens. The time record in LANTIS/IMPAC record & verify system for each of the IMRT patients in 5 daily treatments are analyzed and the average time is used. Data from days when port films are taken are excluded in the analysis. The time data is extracted from IMPAC database using a Crystal Report program that gives the starting time of each field/segment with time resolution of seconds. The finishing time for each field/segment is computed using the given treatment monitor units used and the monitor unit rate. The treatment time, defined as the time elapsed between the beam-ON of the first field/segment and the beam-OFF of the last field/segment of the treatment (single-isocenter only), is compared between the compensator-IMRT and segmental MLC-IMRT treatments. We also analyzed the time elapsed between the treatment fields - time for MLC shaping, gantry motion, compensator exchange, table kick-out, patient specific care, and other tasks therapists need to do between treatment fields in clinical treatment. Figure 2 (above) shows the five-day average treatment time, defined as the time elapsed between the first field/segment beam-ON and the last field/segment beam-OFF. Some (not all) of the treatments with “table kick” field(s) are labeled. Each data point is a five-day average of a clinical IMRT treatment. The data show that compensator-IMRT is about 50 to more than 100% faster than segmental MLC-IMRT in actual clinical application. Furthermore, the increase in treatment time with the number of fields in the treatment for compensator-IMRT treatments is much less than for segmental MLC-IMRT treatments. The data include variation in fractional dose and MU rate. Q: Is the automated segmental MLC-IMRT faster than the manual compensator-IMRT clinical treatment? A: No, this retrospective study showed otherwise. Figure 3 (above) shows the total monitor units (MUs) per IMRT treatment as a function of number of fields in the IMRT treatment. The data shows that the MU difference between compensator-IMRT and segmental MLC-IMRT is not significant. Although the trend is for MLC-IMRT treatments to require more MUs, compensator-IMRT treatment (one example in the 7-field group) can occasionally require more MUs than an MLC-IMRT plan. * sha_chang@med.unc.edu