EUSTAS Round-Robin Testing of Steviol glycosides

Slides:



Advertisements
Similar presentations
Instrumental Analysis
Advertisements

CHEMISTRY ANALYTICAL CHEMISTRY Fall
Chapter 9 Combining Reactions and Mole Calculations.
Chapter 9 Combining Reactions and Mole Calculations.
Ch 6: Good Titrations.
Chem. 31 – 4/15 Lecture. Announcements I Exam 2 – Hope to have graded by next Monday Lab Reports –AA report now due Monday, 4/20 (best to try to complete.
Chem. 31 – 4/8 Lecture. Announcements I Exam 2 – Monday –Covering Ch. 6 (topics since exam 1), 7, 8-1, 17, and parts of 22 (up to and including retention.
Fundamental Techniques and Measurements  Mass Measurements  Volume Measurements  Preparation of a solution of known concentration  UV-Visible Spectrophotometer.
Fundamental Techniques and Measurements
Mobile Phase pH Analyte pK a Shift Lecture 4 Yuri Kazakevich Seton Hall University.
Chem. 31 – 2/18 Lecture. Announcements Turn in AP1.2 Quiz today Exam 1 coming up (1 week from next Monday) Today’s Lecture –Chapter 4 Material Calibration.
SOLUTION PREP. AND BEER’S LAW Experiment #1. What is this experiment about? This experiment has 2 parts to it. They are as follows: 1.How to make solutions.
CALIBRATION METHODS.
EUSTAS Quality Assurance for Stevia Raw Products Presented by Peter Grosser MedHerbs, Aunelstrasse 70, D Wiesbaden, Germany Tel ,
Validation of Analytical Method
INTRODUCTION TO TITRIMETRY. Most common types of titrations : acid-base titrations oxidation-reduction titrations complex formation precipitation reactions.
Simultaneous quantification of bergenin, catechin, and gallic acid from Bergenia ciliata and Bergenia ligulata by using thin-layer chromatography 张 慧.
The following minimum specified ranges should be considered: Drug substance or a finished (drug) product 80 to 120 % of the test concentration Content.
Chem. 31 – 9/23 Lecture Guest Lecture Dr. Roy Dixon.
EUSTAS Round-Robin Testing of Steviol glycosides Prof. Jan M.C. Geuns Lab Functional Biology, KULeuven Kasteelpark Arenberg 31, MB Heverlee-Leuven.
Chem. 231 – 2/18 Lecture. Announcements Set 2 Homework – Due Wednesday Quiz 2 – Next Monday Set 1 Labs –should be switching instruments today (or after.
Quality Assurance How do you know your results are correct? How confident are you?
CALIBRATION METHODS. For many analytical techniques, we need to evaluate the response of the unknown sample against the responses of a set of standards.
HARRIS’s Ch Supplement Zumdahl’s Chapter 15.
Rachel Martin Displacement and Density. Introduction Animals are dosed by being given a specific volume of a test item (or control) formulation.
Solutions used in medical laboratory. O Solution is a homogeneous mixture of two or more substances. O Solute is the dissolved substance, whereas solvent.
HIGH PERFORMANCE LIQUID CHROMATOGRAPHY (HPLC). HIGH PERFORMANCE LIQUID CHROMATOGRAPHY High Performance Liquid Chromatography (HPLC) is one of the most.
Validation Defination Establishing documentary evidence which provides a high degree of assurance that specification process will consistently produce.
Stability Indicating Assays for Dose Formulation Stability Testing Teresa Fuller GSK.
Lab Skills Recap Biotech II. Metrology Vocabulary Unit of measurement Accuracy Precision Standards Calibration Verification Traceability Tolerance Errors.
7 장 적정 Stirring bar One method in volumetric analysis is titration In titration: - substance to be analysed is known as the analyte - the solution added.
1Ben ConstanceCTF3 working meeting – 09/01/2012 Known issues Inconsistency between BPMs and BPIs Response of BPIs is non-linear along the pulse Note –
Physiochemical properties of drugs Using the Sirius T3 to make measurements.
Lecture 8 Peak Parameters and Quantitative chromatography
Chem. 133 – 5/3 Lecture. Announcements Lab – Term Project Progress Report Due Today – Last Assignments: Term Project Poster and Peer Review Grading (Friday,
Instrumental Analysis (I) HPLC Tutorial 8. Graded presentation Students in groups of 4-5 individuals are asked to prepare a presentation (weight=5% of.
Chem. 31 – 6/13 Lecture. Announcements I Pipet and Buret Calibration Lab Report Due Quiz and Homework Returned in Lab Exam 1 on Thursday –Will cover material.
Chapter 5 Quality Assurance and
Accumulation pattern of steviol glycosides in Stevia rebaudiana Bertoni and its stimulation by red light Stijn Ceunen Laboratory of Functional Biology.
Art PowerPoints Harris: Quantitative Chemical Analysis, Eight Edition CHAPTER 05: QUALITY ASSURANCE AND CALIBRATION METHODS.
Chapter 13 Titrations in Analytical Chemistry. Titration methods are based on determining the quantity of a reagent of known concentration that is required.
Solutions used in medical laboratory
Martin-Luther University Halle-Wittenberg, Germany
Organic Chemistry Lab 315 Fall, 2016.
Introduction Results Aim of the study Methods Conclusion References
Chromatography- TLC & HPLC
Labware.
Internal Standard method (experiment lab-4)
Chem. 31 – 9/25 Lecture.
Fundamental Techniques and Measurements
High Performance Liquid Chromatography HPLC
Quantitative Gas Chromatography
A generic platform approach for performing HPLC assays
Chem. 31 – 10/23 Lecture.
iCAP OES Analysis of Trace Elements in Hair
Stoichiometry: Quantitative Information About Chemical Reactions
What is this experiment about?
Chapter 5 Quality Assurance and Calibration Methods
Chem. 133 – 5/4 Lecture.
Chapter 10: The Mole.
Gel Filtration Chromatography.
Fundamental Techniques and Measurements
Titration of Mixtures of Acids Using a Strong Base For two acids to titrate separately, the ka of the stronger one should be at least 104 times greater.
Analysis of an Acid Base Titration Curve the Gran Plot
Chapter 3 Mass Relations in Chemistry; Stoichiometry
Cu2+ + 4NH3 → Cu(NH3)42+ (deep blue)
EXP. NO. 6 Acid Base Titration
Analysis of an Acid Base Titration Curve the Gran Plot
Global Discovery Chemistry, 11 August 2017
Darrell Clinton and H. Dorota Inerowicz Purdue University
Presentation transcript:

EUSTAS Round-Robin Testing of Steviol glycosides Prof. Jan M.C. Geuns Lab Functional Biology, KULeuven Kasteelpark Arenberg 31, MB 2436 3001 Heverlee-Leuven BELGIUM Jan.Geuns@bio.kuleuven.be

Abbreviations SV: steviol, hence: SV gly and SV glu ST: stevioside Reb A – G: rebaudioside A - G Rub: rubusoside DulA: dulcoside A

Problems Different steviol glycosides (10) Different molecular weights Difference of a few % is very important! Validation: each step in analysis should be validated Very pure standards are required of (all) steviol glycosides Often an UV detector is used at 210 nm Often absorption spectra taken in EtOH

Problems Extinction coefficients of all compounds are not known Methods used so far were not validated, and are often based on the assumption that extinction coefficients are the same for all compounds If extinction coefficients are the same, it follows that the slopes of calibration curves should be the same (plotted as mM conc), eg. ST (804), RebA (966), Rub (642)

Proof of similar slopes

Problems Nearly everybody claims that their method is the best one and is accurate.....

Problems Nearly everybody claims that their method is the best one and is accurate..... So do we !

Extinction coefficients Extinction coefficients measured in water, EtOH, 80% AcCN and 35% AcCN Measured at 31 and 62 µM λmax = 205 (EtOH) 196 (H2O) 199 (80% AcCN) 197 (35%AcCN) Molar Extinction coefficient  = A/l.c

Extinction coefficients ε-values in H2O and AcCN mixtures much larger than in EtOH (± 7500 > ± 4000) ε-values measured at λmax much larger than at λ210 This is also reflected in slopes of calibration curves

Calibration Stevioside 190 and 210 nm

Problems Often percentual composition is taken as percentage purity! These are not related at all! Proof: measurement of SVgly in presence/absence of 90% glucose!

1 mg/mL Measurement in water

1 mg/mL plus 10 mg glu/mL

Measurement in presence of impurities (claim: >95% purity) Sample 505171 Ip0906 Nv2060301 Loss on drying 2.3 % 2.8% 5% Stevioside Reb A Reb C Dulcoside A Other 31.29% 14.67 4.89 1.63 0.65 52.44% 23.93 7.21 2.95 0.98 56.44 28.22 6.51 3.26 Total = Purity 53.13% 87.51% 97.68%

Examples of % composition

Examples of % composition

Examples of % composition Percentual composition was given Calculation: peak areas without correction for MW! Different MW were not taken into account!

Examples of % composition

Examples of % composition; Claim: >96% RebA

Same sample analysed

Same sample, claim >96% purity

Impurities in samples (C18 column) Always inject blanks (below)

Round-robin testing 10 labs Each lab: use of own standards and methods 2 unknown samples sample 1: 96.2% purity sample 2: 80:20 dilution of sample 1 with NaHCO3

AIM of Round-robin testing - check as many parameters as possible - drying process (NaHCO3) - peak integration - calculations - weighing........ - try to explain the differences in analyses - how to improve analysis?

Round-robin testing Most labs: NH2 columns, a few C18, 1 SILIC; 250 x 4.6 mm; 5µm; UV 205 or 210 nm Solvents: - AcCN:H2O or AcCN:diluted NH4OAc or phosphoric acid (isocratic or gradient) - C18: 2 x Alltech Alltima with gradient AcCN:0.1 mM phosphoric acid

Round-robin testing In theory, both HPLC methods should be equivalent and give similar results after proper calibration. Use of only 1 early eluting standard for calibration combined with an isocratic solvent, might give an underestimation of slowly eluting compounds (due to flattening of peaks, partly disappearing in base line).

Sample 1, gradient: 25 min

Sample 1, isocratic 33%; 60 min

Difference gradient vs. isocratic

Difference gradient vs. isocratic Later peaks are systematically smaller than early eluting peaks Combination of “disappearance in baseline” and greater signal due to longer time in detector, as demonstrated by a chromatogram at 1 mL/min and 0.8 mL/min

Flows of 1 and 0.8 mL/min

Conclusion flow rate Variation in flow rate (even a few %) might produce huge errors in quantification!

Round-robin testing: Sample 1

Sample 1 Only 3 labs analysed 8 or 9 compounds Only 2 labs found purity > 95% After correction for compounds not analysed: only 3 labs purity >95%. However, lab 3 excessive values for RebC and DulA

Sample 2: Weight loss 20 % NaHCO3 100 mg: 80 mg sample 1 + 20 mg NaHCO3 Heating at 105°C, several hours: 2 NaHCO3 Na2CO3 + CO2 + H2O loss of 36.9 %! 100 mg sample 2: contains 80 mg “wet” sample 1. Wt loss is 4.8 mg (6 %) 20 mg NaHCO3: loss of 7.38 mg Total: 4.8 + 7.38 = 12.18 mg or 12.18 %

Sample 2: Weight loss Reported weight loss: between 4.9 and 12.7% Insufficient drying! Drying should be done to constant weight. Weak point in JECFA recommendation (2 h at 105°C, wt loss< 6%). Moisture calibration of standard and sample: possible eg. RebA. What about all other compounds: hygroscopic properties???

Purity of sample 2 Amount of SV gly in sample 2: purity of sample 1= 96.2% or 90.4% before correction for weight loss (= 6 %) 100 mg sample 2 contains 80 % of sample 1 or 90.4 x 80 % = 72.32 mg. Theoretical purity of sample 2: 72.32/0.8782 = 82.35 mg or %. This is 85.6% of purity of sample 1.

Purity of sample 2

Purity of sample 2 Only 3 labs analysed 8 or 9 compounds Purity should be 85.6% of that of sample 1. Only 2 labs found about this value (85 and 87.3) and a wt loss of 12.1 and 12.4 %. Even after correction for compounds not analysed, purity reported by several labs was far below 82.35 %

Accuracy of measurement Only way to know what is the exact measurement Standard addition method: at least 3 known amounts are added of ultra-pure and dry standards Extrapolation gives intercept with Y-axis. This value should be the same as that of an unspiked sample.

Standard addition of Sample 1

Standard addition of Sample 1 Intercept gave a value of 1277.5 for RebA unspiked sample: value of 1309 Difference: 2.5 %, acceptable as expt. was done only once

Possible errors in (G)LP Purity of standards: absence of other SV gly drying to constant weight always validate quality of standards!!!

Purity of standards: ST: claim: 98.6% purity!

Purity of standards: % composition: ST only 94.97

Purity of standards: on dry wt. basis only 93. 54 % purity (wt loss: 3

Possible errors (in %) Calibration with impure standards leads to an overestimation of purity of samples! ST: claimed purity 98.6%, real purity: 93.54% Claimed Found (wet) Found (dry wt.) “100 %” vs. wet corr 98.6 90.4 93.54 (100 %) Δ 9 % Δ 5.4 % Δ 10.6 %

Conclusions HPLC method, NH2 vs. C18 less important than GLP! Purity of standards: utmost importance No other compounds present Moisture content To be validated by each lab Drying of standards and samples to constant weight Use of gradients might avoid underestimation of small peaks eluting at the end in isocratic HPLC

Conclusions Weighing process itself - Weigh enough sample/standard (at least 50 mg) on a sensitive balance - Weighing of solution (eg. mg/g) might decrease errors - if using automatic pipettes: do not trust them and calibrate regularly! and again and again.... check quality of tips

Conclusions Dissolution of sample! Check that sample is well dissolved Inject enough. Large peaks should be > 500 mV. This reduces relative standard errors of smaller peaks. Check base-line in peak integrations!

Conclusions Check accuracy by standard addition method! More round-robin tests are required to improve and unify the analysis of steviol glycosides. Cooperation between different companies would be good for the “Stevia Industry” as a whole.

Acknowledgements Hilde Verlinden, Bert Demarsin for their excellent help Peter Grosser, Medherbs, Wiesbaden, Germany for financial support My wife Christine for being a “Stevia Widow”