Characterization of the Xoft AXXENT™ X-ray Source

Slides:



Advertisements
Similar presentations
MOSFET Dosimeters for 50 kVp Electronic Brachytherapy with the Xoft Axxent Source Steve Axelrod, PhD Tom Rusch, PhD Xoft, Inc., Fremont, CA
Advertisements

Chapter 4 Radiation Dosimeters
Parameterizing a Geometry using the COMSOL Moving Mesh Feature
Lecture Notes Part 4 ET 483b Sequential Control and Data Acquisition
RAD309 Patient Dose.
The performance of Strip-Fiber EM Calorimeter response uniformity, spatial resolution The 7th ACFA Workshop on Physics and Detector at Future Linear Collider.
Ian Bailey Cockcroft Institute/ Lancaster University October 30 th, 2009 Baseline Positron Source Target Experiment Update.
Lic. Pedro Antonio Pérez.  To develop a simple and versatile dosimetric method capable of determining changes in matter (Xylenol Orange added Fricke.
E. Bavarnegin,Yaser Kasesaz, H. Khalafi
Emily Poon Frank Verhaegen March 6, 2006
At the position d max of maximum energy loss of radiation, the number of secondary ionizations products peaks which in turn maximizes the dose at that.
Performances of a pixel ionization chamber to monitor a voxel scan hadron beam A.Boriano 3, F.Bourhaleb 2,3, R. Cirio 3, M. Donetti 2,3, F. Marchetto 3,
Prepared by Mahmoud F. Elbatta Ahmad Alyazgi Amjad Alaskary Akram abo Twahina Roshdi tambora.
Dose Distribution and Scatter Analysis
Derek Liu E. Poon, M. Bazalova, B. Reniers, M. Evans
Linac Beam.
Dosimetric evaluation of a new design MOSFET detector Per H. Halvorsen* & Stephanie Parker University of North Carolina.
Fiber-Optic Dosimeter. Orbital Therapy’s HRD-6 High Resolution Fiber Optic Dosimeter is the first in a new class of radiation dosimeters with unprecedented.
Ye, Sung-Joon, Ph.D. Ove, Roger, M.D., Ph.D.; Shen, Sui, Ph.D.
Application of a 2-D ionization chamber array for dose verification of dynamic IMRT with a micro-MLC Fujio ARAKI, PhD 1, S. TAJIRI 2, H. TOMINAGA 2, K.
SCM 330 Ocean Discovery through Technology Area F GE.
Single Photon Emission Computed Tomography
Mitglied der Helmholtz-Gemeinschaft Calibration of the COSY-TOF STT & pp Elastic Analysis Sedigheh Jowzaee IKP Group Talk 11 July 2013.
Xoft Routine Per Article Manufacturing Testing of the Axxent ® Source S. Axelrod, T.W. Rusch, M. Powell; Xoft, Inc., Fremont, CA  Purpose: To select sources.
CMS WEEK – MARCH06 REVIEW OF MB4 COMMISSIONING DATA Giorgia Mila
CS TC 22 CT Basics CT Principle Preprocessing. 2 CT Basics CT principle preprocessing CS TC 22 Blockdiagram image processor.
Determining Radiation Intensity
FIRST RESULTS OF THE SILICON STRIP DETECTOR at STAR Jörg Reinnarth, Jonathan Bouchet, Lilian Martin, Jerome Baudot and the SSD teams in Nantes and Strasbourg.
If information seems to be missing, make any reasonable assumptions. 1.A target has an areal density of 2.3 g/cm 2 and a thickness of 0.8 inch. What is.
IAEA International Atomic Energy Agency RADIATION PROTECTION IN DIAGNOSTIC AND INTERVENTIONAL RADIOLOGY Part 19.04: Optimization of protection in Mammography.
ArgonneResult_ ppt1 Comparison of data and simulation of Argonne Beam Test July 10, 2004 Tsunefumi Mizuno
Basis of Mammography F. Milano Dept. Clinical Physiopathology University of Florence
The Effects of Small Field Dosimetry on the Biological Models Used In Evaluating IMRT Dose Distributions Gene Cardarelli,PhD, MPH.
EE 495 Modern Navigation Systems
Sanitary Engineering Lecture 7
Diagnostic Radiology II X-ray Tubes. Anode angle Anode angle defined as the angle of the target surface with respect to the central ray in the x-ray field.
SIMULATION CHECK LIST IMAGING PROTOCOL.
Digital Light Sources First introduced in 2001.
Mahadevappa Mahesh, MS, PhD, Nicholas Detorie, PhD, Randall W
Ch 36 Radiation Protection Design
Date of download: 10/14/2017 Copyright © ASME. All rights reserved.
Treatment Plan Validation of the Xoft AXXENT™ X-ray Source
6th European ALARA Network Workshop Madrid, oct. 23rd 2002
Dr: Mohamed Afifi By Lecturer Radiological Science
Absolute Dose Measurement
Influence of the grid size on the dosimetric characteristics of IMRT beams and on overall treatment plans G. Pittomvils1, L. Olteanu1, B. Vanderstraeten1,
A Brachytherapy Treatment Planning Software Based on Monte Carlo Simulations and Artificial Neural Network Algorithm Amir Moghadam.
Interferometric measurements on mirrors
HPS Collaboration meeting, JLAB, Nov 16, 2016
Ch 10. A System of Dosimetric Calculations
Ray Traces and Pinhole Images of the Xoft Axxent™ X-ray Source
Stability (fitted beam center)
© 2004 – 2011 PAVAC Industries Inc. All rights reserved
Lesson 10: Sensor and Transducer Electrical Characteristics
An Enabling Technology for Creating Sculpted Brachytherapy Dose Patterns With The Xoft Axxent™ System Steve Axelrod, PhD Don Pettibone, PhD Rob Neimeyer.
By: Vi Nhan Nguyen University of Arizona
Emily I. Bartle, Tara M. Urner, Siddharth S. Raju, Alexa L. Mattheyses 
A fixture designed to provide high precision, high throughput routine spatial characterization of the Xoft Axxent™ X-ray source Linda Kelley, Steve Axelrod,
Xoft Manufacturing Procedures
Direct Observation of Single MuB Polymers
Shiori Toba, Hiroyuki Iwamoto, Shinji Kamimura, Kazuhiro Oiwa 
Masataka Chiba, Makito Miyazaki, Shin’ichi Ishiwata 
Benjamin Scholl, Daniel E. Wilson, David Fitzpatrick  Neuron 
Benjamin Scholl, Daniel E. Wilson, David Fitzpatrick  Neuron 
Dosimetric Measurements
Hp(10) irradiations for personal dosimetry traceability
Computed Tomography (C.T)
Emily I. Bartle, Tara M. Urner, Siddharth S. Raju, Alexa L. Mattheyses 
Computed Tomography (C.T)
Gain measurements of Chromium GEM foils
Presentation transcript:

Characterization of the Xoft AXXENT™ X-ray Source Steve Axelrod, PhD Xoft, Inc. Fremont, CA

AXXENT X-Ray Source Miniature X-ray tube for therapeutic radiation treatment 2.2 mm diameter by 10 mm long Initial indication is breast brachytherapy Not yet cleared by the FDA Operates up to 50 kV at up to 300 µA beam current Up to 1 Gray/minute dose rate at Rx point Inside balloon applicator Mobile Controller unit User interface X-Ray power and safety interlocks Steps the source position

Family Photos Controller Balloon applicator X-Ray Source ass’y Source anode, activated

AXXENT X-Ray Source Characterization Quantification of X-Ray output via TG-43U1 parameterization for brachytherapy Terms for Source strength, dose ratio air/water, spatial distribution Spatial characterization includes polar and azimuthal angle distributions, and the depth-dose behavior

TG43-U1 Parameterization Dose rate Intensity terms Geometry term (1/r2) Depth-dose 2d anisotropy

Spatial Characterization Apparatus Custom apparatus developed to allow precision measurements of the dose distribution in 3 dimensions Ionization chamber encased in solid water case, immersed in water bath Detector assembly mounted to a linear stage, attached to a rotary stage, both under computer control Source position controlled by manual micrometer stages in 2 dimensions, fixed in third dimension.

Rotary and linear stages Picture of Apparatus Source catheter Support structure (sits on water tank) Rotary and linear stages Ion chamber Source location

Depth-Dose Measurements Depth dose (DD) measured in water phantom Air ionization chamber attached to precision linear stage Range from 10 to 70 mm from source center Positional accuracy ~± 100 μm DD characteristics dominated by point source 1/r2 effects Radial Dose Function (RDF) removes the 1/r2 Normalized to 1.0 at 10 mm from the source Remainder of change is due to attenuation in the medium Double exponential fit to data has residuals of ~0.5%

Depth-Dose and RDF Plots Radial dose function with 2 exponential fit Depth-dose curve

Polar Angle Distribution Polar distribution as a function of distance is the TG43 “2d Anisotropy Function” Normalized to 1.0 at 90 degrees Averaged over left/right sides in TG43 Measured using ion chamber positioned with a rotary stage in water phantom Precise location of source is critical to accuracy and reproducibility

Polar Anisotropy Data Data at 2, 4 and 6 cm averaged over ~20 sources Blue = 2, cm, red = 4 cm, green = 6 cm Polar plot of averaged 4 cm results

Azimuthal Angle Distribution Azimuthal angle goes around the source “equator” Ideally would like there to be no dependence Measured similarly to Polar angle, but with the source entering the apparatus at right angles Difficult measurement due to alignment considerations A “perfect” source, misaligned by just 100 μm would show a 1 – 2% effect, depending on distance Results plotted as “% difference from average” at each point

Azimuthal Angle Data Max difference < 5%, RMS difference < 2% % difference from average Azimuthal distributions for 10 sources, at 4 cm from source

Output Stability Measurements of output strength in Air Kerma and “dose rate constant” Λ provide overall scaling factors Strength similar to a new Ir-192 seed We look at stability of the source at short and long time scales Average and maximum deviations looked at over 10 second and 5 minute scales “Robustness” evaluated by cycling on and off repeatedly Look for consistent output each cycle

Stability and Robustness Data 10 on/off cycles plus 20 min steady running Need to get better data Max deviation drift analysis using data from steady running segment 1% drift at 10 minutes 0.13% drift at 10 seconds