Summary of the Compton-PET project

Slides:



Advertisements
Similar presentations
What are PET basics?.
Advertisements

Computer Simulation for Emission Tomography: Geant4 and GATE Xiao Han Aug
Single Planar Detector Monte Carlo Simulations Andrew Mather
X-Ray & γ-Ray Interactions with Matter
Advanced GAmma Tracking Array
Geant4 Low Energy Polarized Compton Processes Gerardo Depaola * Francesco Longo + Francesco Longo + * National University of Córdoba (Argentina) + University.
PET Design: Simulation Studies using GEANT4 and GATE - Status Report - Martin Göttlich DESY.
Andrej Studen, Karol Brzezinski, Enrico Chesi, Vladimir Cindro, Neal H. Clinthorne, Milan Grkovski, Borut Grošičar, Klaus Honscheid, S. S. Huh, Harris.
SmartPET: A Small Animal PET Demonstrator using HyperPure germanium Planar Detectors R. Cooper 1 st year PhD Student
June 6 th, 2011 N. Cartiglia 1 “Measurement of the pp inelastic cross section using pile-up events with the CMS detector” How to use pile-up.
Pulse Shape Analysis with Segmented Germanium Detector Xiang Liu, Max-Planck-Institut für Physik 1.Motivation 2.Pulse properties 3.Analysis procedure 4.Some.
Pair Spectrometer Design Optimization Pair Spectrometer Design Optimization A. Somov, Jefferson Lab GlueX Collaboration Meeting September
BME 560 Medical Imaging: X-ray, CT, and Nuclear Methods
Examples of GATE simulation Yun Dong. Different systems available in GATE Scanner: most generic system; Cylindrical PET: cylindrical geometry; CPET: simulate.
Signal Analysis and Processing David Scraggs. Overview Introduction to Position Resolution Induced Charges Wavelet Transform Future Work Discussion.
RESMDD'02 pCT: Hartmut F.-W. Sadrozinski, SCIPP INITIAL STUDIES on PROTON COMPUTED TOMOGRAPHY USING SILICON STRIP DETECTORS L. Johnson, B. Keeney, G. Ross,
Kirov A S, MSKCC Overview of Geant4 Use and Issues in Imaging: Emission Tomography (PET and SPECT) Assen S. Kirov Department of Medical Physics Memorial.
Simulations with MEGAlib Jau-Shian Liang Department of Physics, NTHU / SSL, UCB 2007/05/15.
Tracker Reconstruction SoftwarePerformance Review, Oct 16, 2002 Summary of Core “Performance Review” for TkrRecon How do we know the Tracking is working?
Planar scintigraphy produces two-dimensional images of three dimensional objects. It is handicapped by the superposition of active and nonactive layers.
RF background, analysis of MTA data & implications for MICE Rikard Sandström, Geneva University MICE Collaboration Meeting – Analysis session, October.
Radiation Detection and Measurement, JU, First Semester, (Saed Dababneh). 1 Spectrum if all energy is captured in detector. Allows identification.
The PEPPo e - & e + polarization measurements E. Fanchini On behalf of the PEPPo collaboration POSIPOL 2012 Zeuthen 4-6 September E. Fanchini -Posipol.
St. Petersburg State University. Department of Physics. Division of Computational Physics. COMPUTER SIMULATION OF CURRENT PRODUCED BY PULSE OF HARD RADIATION.
0 A Fast Time Incorporating Monte-Carlo Simulation of Wire Chamber Based Small Animal PET Scanners for Detector Scatter Correction M. Dawood 1, Don Vernekohl.
بسم الله الرحمن الرحيم ” وقل رب زدنى علماً “ صدق الله العظيم.
Fundamental Limits of Positron Emission Tomography
First Results from a Test Bench for Very High Resolution Small Animal PET Using Solid-State Detectors Klaus Honscheid for The CIMA Collaboration The Ohio.
BMI I FS05 – Class 4 “Nuclear Imaging: Math” Slide 1 Biomedical Imaging I Class 5 – Radionuclide Imaging (PET, SPECT), Part 3: Attenuation and Scatter.
Professor Brian F Hutton Institute of Nuclear Medicine University College London Emission Tomography Principles and Reconstruction.
Pedro Arce Introducción a GEANT4 1 GAMOS tutorial Compton Camera Exercises
Simulation of the energy response of  rays in CsI crystal arrays Thomas ZERGUERRAS EXL-R3B Collaboration Meeting, Orsay (France), 02/02/ /03/2006.
Increase in Photon Collection from a YAP:Ce Matrix Coupled to Wave Lenght Shifting Fibres N. Belcari a, A. Del Guerra a, A. Vaiano a, C. Damiani b, G.
Nuclear Medicine: Tomographic Imaging – SPECT, SPECT-CT and PET-CT Katrina Cockburn Nuclear Medicine Physicist.
Enhancing InBeam PET with single Photon (Compton) Detection CERN September 2nd 2008 VALENCIA GROUP, IFIMED José M. Benlloch (PET hardware, speaker) José.
1 NaI calibrationneutron observation NaI calibration and neutron observation during the charge exchange experiment 1.Improving the NaI energy resolution.
Accelerator Physics, JU, First Semester, (Saed Dababneh). 1 In the figure: Photoelectric suppressed. Single Compton (effect of crystal dimensions).
Validation of EM Part of Geant4
Event Analysis for the Gamma-ray Large Area Space Telescope Robin Morris, RIACS Johann Cohen-Tanugi SLAC.
PSD-V, London, September 1999 Two dimensional readout in a liquid xenon ionisation chamber V.Solovov, V.Chepel, A.Pereira, M.I.Lopes, R.Ferreira Marques,
P. Rodrigues, A. Trindade, L.Peralta, J. Varela GEANT4 Medical Applications at LIP GEANT4 Workshop, September – 4 October LIP – Lisbon.
Improved Hybrid PET Imaging Using Variable Focal Length Fan-Slat Collimators Thomas C. Rust and Dan J. Kadrmas, Ph.D. Medical Imaging Research Laboratory.
1 Nuclear Medicine SPECT and PET. 2 a good book! SR Cherry, JA Sorenson, ME Phelps Physics in Nuclear Medicine Saunders, 2012.
PET Imaging Positron Emission Tomography
Simulations in Medical Physics Y. TOUFIQUE*, R.CHERKAOUI EL MOURSLI*, M.KACI**, G.AMOROS**, *Université Mohammed V –Agdal, Faculté des Sciences de Rabat,
SIMULATION OF BACKGROUND REDUCTION TECHNIQUES FOR Ge DBD DETECTORS Héctor Gómez Maluenda. University of Zaragoza. GERDA/Majorana MC Meeting.
Status of the measurement of K L lifetime - Data sample (old): ~ 440 pb -1 ( ) - MC sample: ~125 pb -1 ( mk0 stream ) Selection: standard tag (|
Interactions of Ionizing Radiation
Dollan, Laihem, Lohse, Schälicke, Stahl 1 Monte Carlo based studies of polarized positrons source for the International Linear Collider (ILC)
Chapter-5 Positron emission tomography (PET)
+ Voxel Imaging Pizza Gianluca De Lorenzo. + Positron Emission Tomography April Gianluca De Lorenzo.
A Study of Reverse MC and Space Charge Effect Simulation with Geant4
Electron Polarization effects in Compton X/γ-ray Sources
GAMOS tutorial PET Exercises
Huagen Xu IKP: T. Randriamalala, J. Ritman and T. Stockmanns
Progress with Spin Tracking in GEANT4
Reconstructions with TOF for in-beam PET
Detailed simulations of a full-body RPC-PET scanner
HARPO, a Micromegas+GEM TPC for gamma polarimetry above 1 MeV
GAMOS tutorial PET Exercises
E166 Photon Asymmetry Simulation
RAD 254 Chapter 10 X-ray interaction with matter
Function and Structure in
Application of Nuclear Physics
Interaction of gamma rays with matter
V. Chepel, V. Domingos, R. Martins, A. Morozov, F. Neves and
Function and Structure in
MCP PET Simulation (7) – Pixelated X-tal
P. Rodrigues, A. Trindade, L.Peralta, J. Varela
COMPTON SCATTERING IN FORWARD DIRECTION
Presentation transcript:

Summary of the Compton-PET project V.Chepel LIP-Coimbra and Department of Physics, University of Coimbra Jornadas do LIP, Lisboa, 22-24 April, 2012 V. Chepel

Compton-PET stands for: “Feasibility study of using Compton scattering for medical imaging with positrons” (2008-2011) Jornadas do LIP, Lisboa, 22-24 April, 2012 V. Chepel

Motivation Scattering in the patient body: (~70 to 90%) Scattering in the detector: Probability of the first interaction to be a Compton: BGO – 53% LSO – 62% LXe – 79% LAr – 95% BGO – 78% LSO – 86% LXe – 95.6% LAr – 99.8% of coincidences involve scattered events Using these events for image reconstruction: improves statistics and/or allows reducing the dose administered to patient, but deteriorates resolution Jornadas do LIP, Lisboa, 22-24 April, 2012 V. Chepel

Possible solution: Compton tracing (E1, q1) (E2, q2) (E4) This is what we see (or not) in the detector: Reconstruction of the true sequence depends strongly on resolutions on E and q (i.e. position) p´g pg pg, p´g, pe are in the same plane pe Knowing electron track orientation can help Jornadas do LIP, Lisboa, 22-24 April, 2012 V. Chepel

Objective: explore what can be done with the scattered events Compton tracing : Compton electron track orientation: can we get some useful information on it ? If one can measure polarization of the g-par – would it help ? Do we really need to trace (i.e. reconstruct in detail) scattered g-rays ? Method: Monte Carlo and experimental How Geant deals with the electron binding ? (see talk by A.Mangiarotti on validation) Jornadas do LIP, Lisboa, 22-24 April, 2012 V. Chepel

1. Attempting to measure electron track orientation in LXe The idea: for 500 keV electrons 5.5 MeV a-s 511 keV g Jornadas do LIP, Lisboa, 22-24 April, 2012 V. Chepel

1. Attempting to measure electron track orientation in LXe 300 ns (expected ~25 ns) Neither field non-uniformity, nor diffusion in the gas explain such long rise time (unless diffusion in Xe vapour is much stronger than in non-saturated gas) LXe a-s, range 40 mm CHERNE2010, Coimbra, 7-9 June, 2010

2. Polarization of annihilation photons e+ + e-  g + g What if we could measure polarization of each g ? Any help with Compton tracing ? Any information about scattering in the patient body ? Jornadas do LIP, Lisboa, 22-24 April, 2012 V. Chepel

2. Compton scattering cross section for polarized photons Eg = 511 keV Maximum anisotropy = 5.5 (for   82º) Jornadas do LIP, Lisboa, 22-24 April, 2012 V. Chepel

2. MC of g-pairs with GEANT4: Point source in vacuum 2 - 1 Point source in of water (20cm phantom) Jornadas do LIP, Lisboa, 22-24 April, 2012 V. Chepel

3. Do we really need to do tracing ? Computation tool: Interactions (physics) Detector response Compton tracing and LOR definition Image reconstruction GEANT4 C++ Matlab CHERNE2010, Coimbra, 7-9 June, 2010

3. Choosing LOR: a “childish” approach CHERNE2010, Coimbra, 7-9 June, 2010

3. UTAH phantom simulated P r e l i m i n a r y Traditional PET: Single energy deposition in 350-650 keV Compton PET: Singles + doubles With total E= 350-650 keV Same contrast Statistics > x3 Detectors: LXe ring 80x5x5 cm3 Resolution 2 mm, dE/E=20% CHERNE2010, Coimbra, 7-9 June, 2010

3. Line source 4.5 cm off centre in 20 cm of water P r e l i m i n a r y Traditional PET: Scatter fraction  0.7 Compton PET: Scatter fraction  0.86 Detectors: LXe ring 80x5x5 cm3 Resolution 2 mm, dE/E=20% CHERNE2010, Coimbra, 7-9 June, 2010

Conclusions Compton electron tracking does not seem to be feasible for <=511 keV (at least with the readout we have tried). The effect of gamma-ray polarization on the observable distributions is small. Events involving Compton scattering in the detector can be recovered for the imaging with the advantage of higher image statistics and without jeopardizing the image quality. Complex Compton tracing does not to be necessary. Simple methods can be used for that. CHERNE2010, Coimbra, 7-9 June, 2010