Measurement of Flow and Volume

Slides:



Advertisements
Similar presentations
EE 4BD4 Lecture 26 Plethysmography 1. Definition Measure volume change over time Flow = dvolume/dt 2.
Advertisements

Lecture Notes Part 4 ET 483b Sequential Control and Data Acquisition
COMMUNICATION SYSTEM EEEB453 Chapter 3 (III) ANGLE MODULATION
Review 29:008 Exam 3. Ch. 12 Electrostatic Phenomena.
Alternating Current Circuits And Electromagnetic Waves Chapter 21.
Lock-in amplifiers Signals and noise Frequency dependence of noise Low frequency ~ 1 / f –example: temperature (0.1 Hz), pressure.
Chapter 8. Measurement of Flow and Volume of Blood John G. Webster
EE 5340, SMU Electrical Engineering Department, © Carlos E. Davila, Electrical Engineering Dept. Southern Methodist University slides can be viewed.
Monroe L. Weber-Shirk S chool of Civil and Environmental Engineering Closed Conduit Measurement Techniques  Pipeline systems  Transmission lines  Pipe.
Chapter 5 Part Three Blood Pressure and flow by Ibrhim AlMohimeed BMTS /3/2013.
Principles of Biomedical Systems & Devices / WEEK 10: Flow & Volume of Blood.
Self-Induction Transducers ~ AC Supply v ref Inductance Measuring Circuit x (Measurand) Ferromagnetic Target Object Coil is activated by the supply and.
Monroe L. Weber-Shirk S chool of Civil and Environmental Engineering Closed Conduit Measurement Techniques ä Pipeline systems ä pipe networks ä measurements.
1 Syafruddin Hasan. 2 Systems Concepts Automated systems that control flow rates or volume employ the following components: Automated systems that control.
1 Syafruddin Hasan. 2 Systems Concepts Automated systems that control flow rates or volume employ the following components: Automated systems that control.
DOPPLER SONOGRAPHY By Dr/ Dina Metwaly.
Lecture 6: Measurements of Inductance, Capacitance, Phase, and Frequency 1.
Lock-in amplifiers
EKT314/4 Electronic Instrumentation
EKT314/4 Electronic Instrumentation
© From J. G. Webster (ed.), Medical instrumentation: application and design. 3 rd ed. New York: John Wiley & Sons, Figure 8.1 Several methods of.
Lecture 2 Most basic facts from Electricity needed for understanding telecommunications Local transmission lines in the telephone system Decibels Signals.
Servos Servos are everywhere. Elements of servo System -- to be controlled Sensor or detector -- measure quantity to be controlled Reference -- desired.
생체계측 II Report # 송성진 Medical Instrumentation II.
Resident Categorical Course
EKT 451/4 SEM 2 Chapter 6 SENSOR & TRANSDUCER SHAIFUL NIZAM MOHYAR
Display of Motion & Doppler Ultrasound
MOCT(Magneto Optic Current Transduser)
Biomedical Electrodes, Sensors, and Transducers
Sonic / Ultrasonic Anemometers
Electromagnetic Blood Flow Meters Measures instantaneous pulsatile flow of blood Works based on the principle of electromagnetic induction The voltage.
As Unit 1 - Electricity What you need to know..  Current is the rate of flow of charged particles.  In metals these are conduction electrons, most electrons.
EKT 451 CHAPTER 6 Sensor & Transducers.
1 Transducers EKT 451 CHAPTER 6. 2 Definition of a Transducers   Transducer is any device that converts energy in one form to energy in another.  
Announcements Midterm Exam next Wednesday Exam starts at 6 PM, ~1 hr. Closed book, one page of notes Bring a calculator (not phone, computer, iPad, etc.)
Electric Pressure Transducer
Chapter 13 The Basic Elements and Phasors. Objectives Be able to add and subtract sinusoidal voltages or currents Use phasor format to add and subtract.
TERMINATIONS Copyright F. Canavero, R. Fantino Licensed to HDT - High Design Technology.
20 Overview current  magnetic field magnetic field  current
The rate of change of velocity.
Electromagnetic Induction
Biomedical Instruments Design Biopotential Amplifiers
Figure 3. 1 Op-amp equivalent circuit. The two inputs are 1 and  2
BME 312 BIOMEDICAL INSTRUMENTATION II LECTURER: ALİ IŞIN
CHAPTER 6 VOLTAGE REGULATOR Tulus Ikhsan Nasution.
Potential and Current Control
Variable Capacitance Transducers
Biomedical Instrumentation
MECH 373 Instrumentation and Measurements
Warm-up Why do loops of wire in a motor rotate?
Lock-in amplifiers
PART 3:GENERATION AND DETECTION OF ANGLE MODULATION

Chapter 31A - Electromagnetic Induction
Electromagnetic Induction
Figure 3. 1 Op-amp equivalent circuit. The two inputs are 1 and  2
Figure 8.1 Several methods of measuring cardiac output In the Fick method, the indicator is O2; consumption is measured by a spirometer. The arterial-venous.
Phys102 Lecture 18/19 Electromagnetic Induction and Faraday’s Law
UNIT 2 Magnetic Circuits
Amplifiers: A Bio amplifier is an electrophysiological device, a variation of the instrumentation amplifier, used to gather and increase the signal integrity.
Objectives: After completing this module, you should be able to:
Biomedical Instruments Design Biopotential Amplifiers 1.
Biopotential amplifiers
10.6 Fourier Transform Mass Spectrometry
SPECTROPHOTOMETRY Applied Chemistry.
8.5 Modulation of Signals basic idea and goals
Objectives: After completing this module, you should be able to:
Measurement of Motion.
C H A P T E R 11 A.C. Fundamentals.
Presentation transcript:

Measurement of Flow and Volume FYS 4250 Kap.8 Measurement of Flow and Volume of blood Primary measurements O2 concentration and other nutrition of the cells = Difficult -> Have to accept second class measurements like blood flow and volum = might be difficult -> third class measurements like blood flow (usually correlates adequately to blood flow) = might be difficult -> fourth class measurement of ECG (usually correlates adequately with blood pressure).

Fick’s method, cardiac Output (Continuous) Indicator-dilution method that uses continuous infusion Adv: nontoxic Disadv: Unpleasant for the patient, requires image guiding, must be placed in pulmonary artery due to different O2 concentrations in the atrium. Exhaled CO2 absorbed in a soda-lime canister -> O2 consumption is indicated directly by the net gas-flow rate. (Inaccurate). Inconvenient

Indicator-dilution method Rapid injection Dye Adv: Can be repeated, inert, harmless, economical Thermodilution adv: Even cheaper, Not necessary to puncture an artery (using the same catheter) T. Disadvantage: Inadequate mixing between injection and sampling site, Exchange of heat in blood vessels/heart chamber Heat exchange in catheter walls Calibrate with a dye dillution instrument

Elektromagnetic flowmeter Faraday’s law of induction Blood moves through a magnetic field, electromotoric field is induced B = magnetic flux L = Length between electrodes U = velocity of blood

Sensitivity of the electrodes Best measurements with electrodes outside the vessel Measures correctly for a uniform flow profile, otherwise only correct for average flow velocity Sources of error: Asymmetric velocity profile High velocity generates higher incr. Emfs = circulating currents in the transverse plane = varying drops in resistance within the conductive blood and tissue Conductivity varies with HCT Extravascular fluid has greater conductivity = shunting of flow signal Magnetic flux density is not uniform along the axos => circulating currents in axial direction

Electromagnetic flowmeter waveforms DC floweters are useless: Voltage across electrode interface is in series with the flow signal. Even with nonpolarizable electrodes, random drift of this voltage is of the same order as the flow signal -> impossible to separate ECG has a waveform and frequency content similar to the flow signal = interference In the frequency reange 0 to 30 Hz, noise in the amp. Is large = poor SNR Operating at 400 Hz. < gives bulky sensors, > stray capacitance problems Transformer voltage = Shaded loop in previous picture is not exactly parallel to B field -> induction of transformer voltage prop to dB/dt in the output voltage. (Many times larger than fllow voltage) Solution: Separate one elctrode in two in the axial direction=> tilt the shaded loop to exactly parallel to B field Measure when the transformer voltage is zero. (Undesired phase shifts may cause large errors and drifts Use Quadrature suppression circuit

Quadrature-suppression flowm. Magnitude of voltage from amplifier is detected by quadrature demodulator What is a demodulator? Conventional method of finding deviation from nominal freq = convert to phase and detect change. In quadr shift signal 90 deg at the center frequency, depending on the direction of deviation this shift is either greater or less than 90 deg -> Can be detected. (extracting amplitude and phase information from a carrier wave) Quad generator produce a signal proportional to the transformer voltage => balancing out transformer voltage at the input. (Reducing TFV by a factor 50) Low noise FET for amplifier input stagen, proper turns ratio on the step up transformer and full-wave demodulators = Excellent SNR Low pass filter => dc output

Flowmeter probe Probes usually platinum Not suitable on veins => they collapse

Ultrasonic doppler flowmeter Flow profiles Piezoelectric material (converts power from electric to acoustic form) Zirconate titanate most effective crystal Transducer has a finite diameter=>diffraction patterns as an apperture In the far field, the beam diverges, intensity is inversely proportional to the square of the distance Angle of beam divergense ø: sin Ø = 1.2 Landa/D (diameter) Lower spatial resolution in the far field To achieve near-field operation -> higher frequencies and larger transducers (f <> 2, 10 MHz)

Transducer systems (transit time) Delta t = the difference between upstream and downstream transit times Fd = doppler freqency shift Frequency to voltage converter -> quantitative output

Doppler flowmeter, sinewave Low pass filtered to produce an output proportional to the velocity of blood No sense of direction Sources of error: Carrier signal is larger than the doppler signal Some of the RF signal is directly coupled to the reciever through an electric field Some carrier signal travels a direct acoustic way to the reciever through side lobes Doppler signal = 0.1% of the received signal Not a single frequency: Cells move at different velocities A cell remains within the beam for a short time => time-gate function => band of frequencies Side lobes causes different frequency shifts Tumbling of cells and local velocities resulting from turbulence cause differend D. shifts

Quadrature-phase detector Phase shift network splits the carrier into two components that are in quadrature (90 deg apart)

Directional flowmeter waveforms If the blood flow is toward the transducer, doppler frequency is higher than the carrier frequency, doppler vector rotates counter clockwise. (Dashed waves) Sign of the phase will determine the direction Doppler frequency is in-phase but will affect the amplitude Important: The velocity is in the frequency, the direction in the phase

Termistor flowmeasurement Depend on convective cooling of a heated sensor (over blood temperature) => local velocity W (power)/delta T = a + b log u High sensitivity and reasonable resistance values Is direction insensitive

Termistorbridge velocity meter Two problems of constant current sensor circuit: Time constant of the sensor embedded in the probe is a few tenths of a second, too long for the desired frequency respons of 0 to 25 Hz Too achieve reasonable sensitivity at high velocities -> high sensor current -> lack of cooling when flow stops -> increased temp (5 deg) ->fibrin coats the sensor Constant temperature sensor overcomes problems: Ru is heated. Velocity increases -> Ru cools and resistance increases -> more positive voltage on the pos terminal -> increased vb -> increased bridge power -> increased Ru -> cooling is counteracted High-gain negative feedback -> bridge is always in balance -> constant temp at Ru High gain negative feedback divides sensor time constant by a factor equal to the loop gain = improved frequency response Varying blood temperature -> Rt (temperature compensating thermistor)

Plethysmography Measures changes in volume

Plethysmography curve

Impedance-plethysmography Blood volume changes/lung volume changes, impedance will change. Swansons model require three assumptions: Expansion of the arteries is uniform Resistivity of blood does not change Lines of current are parallel to arteries Frequency of about 100 kHz because: Desirable with a current greater than 1 mA to achieve adequate SNR. Low frequencies will cause an unpleasant shock -> frequencies above 20 kHz is used to avoid perception of the current High frequencies are used to decrease both skin-electrode impedance and motion changes in impedance Freq > 100 kHz will be vulnerable of low impedance stray capacitances

2- or 4-electrode Problems with 2 electrode: Current density is higher near the electrodes than elsewhere in the tissue Pulsations of blood in the tissue -> change skin-electrode impedance + tissue impedance. They are in series, impossible to determine the tissue impedance The current density is not uniform in the region of interest

4-electrode plethysmography Constant current Ideal situation -> constant current through Z. Practice -> shunting impedance Zi (stray and cable capacitance) Normally not a problem -> careful design can keep Zi high, and Z and Zi are close to 90 deg out of phase Zv same as Zi. Ideally high impedance of amplifier = 0 current. In practice, a small current ->

Photoplethysmography Absorption of light is modified by changes in volume of the vessels Adv: Simple Disadv: Poor measure of changes in volume, sensitive to motion artifact Tungsten lamp -> heat affecting measurements

Photoplethysmography