Sergey Vinogradov1,2,3, Elena Popova4, and Carsten Welsch2,3

Slides:



Advertisements
Similar presentations
Advanced GAmma Tracking Array
Advertisements

Chapter 8 Planar Scintigaraphy
Alexander Klyachko, IU Cyclotron, PTCOG51, Seoul, South Korea, May 17-19, 2012 A.V. Klyachko 1, D.F. Nichiporov 1, L. Coutinho 2, C.-W. Cheng 2, 3, M.
Medical Imaging Mohammad Dawood Department of Computer Science University of Münster Germany.
Optimisation of X-ray micro-tomography to perform low-dose imaging of highly-dosed gels P.M.Jenneson, E.C.Atkinson, P.Wai and S.J.Doran In 1993, Maryanski.
Simulations with MEGAlib Jau-Shian Liang Department of Physics, NTHU / SSL, UCB 2007/05/15.
Planar scintigraphy produces two-dimensional images of three dimensional objects. It is handicapped by the superposition of active and nonactive layers.
experimental platform
Pennazio F 1, Battistoni G 2, Bisogni MG 3, Cappucci F 2, Cerello P 1, Del Guerra A 3, Fiorina E 4,1, Giraudo G 1, Kraan A 3, Mairani A 5, Morrocchi M.
Applications of Geant4 in Proton Radiotherapy at the University of Texas M.D. Anderson Cancer Center Jerimy C. Polf Assistant Professor Department of Radiation.
G. Bartesaghi, 11° ICATPP, Como, 5-9 October 2009 MONTE CARLO SIMULATIONS ON NEUTRON TRANSPORT AND ABSORBED DOSE IN TISSUE-EQUIVALENT PHANTOMS EXPOSED.
Introduction Silicon Photomultipliers (SiPMs) are well recognized due to exceptional photon number and timing resolution. However, due to specific SiPM.
Fundamental Limits of Positron Emission Tomography
Possibilities of TOF measurements on NPI neutron generators Mitja Majerle Department of Nuclear Reactions Nuclear Physics Institute ASCR.
Jin Huang M.I.T. Hall A Analysis Workshop Dec 14, JLab.
Medical Accelerator F. Foppiano, M.G. Pia, M. Piergentili
HEALTH Novel MR-compatible PET detectors for simultaneous PET/MRI imaging FP7-HEALTH-2009-single-stage The focus should be to develop novel.
BMI I FS05 – Class 4 “Nuclear Imaging: Math” Slide 1 Biomedical Imaging I Class 5 – Radionuclide Imaging (PET, SPECT), Part 3: Attenuation and Scatter.
RICH Detectors for Particle ID 1 F. Sabatie Bochum Meeting CLAS12 JLab: Base configuration: 2 sectors ~1 m 2 photon detector per sector Extension.
Physics of carbon ions and principles of beam scanning G. Kraft Biophysik, GSI, Darmstadt, Germany PTCOG43 Educational Satellite Meeting: Principles of.
Latifa Elouadrhiri Jefferson Lab Hall B 12 GeV Upgrade Drift Chamber Review Jefferson Lab March 6- 8, 2007 CLAS12 Drift Chambers Simulation and Event Reconstruction.
Development of a Segmented Planar Germanium Imaging Detector
EAS Time Structures with ARGO-YBJ experiment 1 - INFN-CNAF, Bologna, Italy 2 - Università del Salento and INFN Lecce, Italy A.K Calabrese Melcarne 1, G.Marsella.
METAL DETECTORS: PERFORMANCES AND APPLICATIONS Oleksandr Okhrimenko Institute for Nuclear Research NAS of Ukraine, Kiev.
Welcome !! to the CAoPAC Workshop Carsten P. Welsch.
1 Nuclear Medicine SPECT and PET. 2 a good book! SR Cherry, JA Sorenson, ME Phelps Physics in Nuclear Medicine Saunders, 2012.
Peter Dendooven LaBr 3 and LYSO monolithic crystals coupled to photosensor arrays for TOF-PET Physics for Health in Europe Workshop February 2-4, 2010,
Nuclear Medicine Physics and Equipment 243 RAD 1 Dr. Abdo Mansour Assistant Professor of radiology
Feb. 3, 2007IFC meeting1 Beam test report Ph. Bruel on behalf of the beam test working group Gamma-ray Large Area Space Telescope.
Design of a tracking device for on-line dose monitoring in hadron therapy S. Muraro for INSIDE collaboration VCI
Manoj B. Jadhav Supervisor Prof. Raghava Varma I.I.T. Bombay PANDA Collaboration Meeting, PARIS – September 11, 2012.
Chapter-5 Positron emission tomography (PET)
1 The Scintillation Tile Hodoscope (SciTil) ● Motivation ● Event timing/ event building/ software trigger ● Conversion detection ● Charged particle TOF.
Characterisation of a pixellated CsI detector for the Distinguish Project. 1 Martin Jones The University of Liverpool
Brunel University London Field-off LiH Energy Loss Rhys Gardener CM45 – July 28th.
Fastest Data Processing in Image Reconstruction for Compton Camera Imaging
Solar gamma-ray and neutron registration capabilities of the GRIS instrument onboard the International Space Station Yu. A. Trofimov, Yu. D. Kotov, V.
Idee per lo sviluppo del Charge Identifier
EURADOS Working Group 9: Radiation Dosimetry in Radiotherapy
Advances in Imaging Instrumentation for Nuclear Cardiology
Development of a Compton Camera for online range monitoring
Very High Energy Electron for Radiotherapy Studies
Upgrade Tracker Simulation Studies
Closeout of VHEE17 Workshop
Diagnostics of FRIBs beam transport line
Cross-section Measurements of (n,xn) Threshold Reactions
Image quality and Performance Characteristics
Dr. Mikhail Runtso, Mr. Pavel Naumov,
Reconstructions with TOF for in-beam PET
Presentation of the health pole Philippe Laniece (IMNC)
Cockcroft Institute and The University of Liverpool
Reward-renewal model of SiPM response to arbitrary waveform optical signals of beam loss monitoring systems S. Vinogradov, L. Devlin, E. Nebot del Busto,
Summary of the Compton-PET project
OTR based measurements for ELI-NP Gamma Beam Source
Monte Carlo studies of the configuration of the charge identifier
SIGMA: a detector for γ-ray spectroscopy & imaging Dr Laura AGATA/GRETINA Collaboration Meeting
M. D. Anderson Cancer Center Houston, TX
Application of Nuclear Physics
Neutron Detection with MoNA LISA
MINOS: a new vertex tracker for in-flight γ-ray spectroscopy
On behalf of CEPC calorimeter working group
V. Chepel, V. Domingos, R. Martins, A. Morozov, F. Neves and
Neutron Detector Simulations for Fast Neutrons with GEANT4
First demonstration of portable Compton camera to visualize 223-Ra concentration for radionuclide therapy Kazuya Fujieda (Waseda University) J. Kataoka,
md-NUV PET project meeting
一种基于晶体间光分享原理的深度测量PET探测器
Neutron Detector Simulations for Fast Neutrons with GEANT4
N. Belcari, F. Attanasi, V. Rosso , A. Del Guerra
Overview of Beam Test Benoit, Ronaldo and Eduardo Nov 8 , 2005 thanks to Steve and Bill for helping to consolidate all the information.
N. Belcari, F. Attanasi, V. Rosso , A. Del Guerra
Presentation transcript:

Sergey Vinogradov1,2,3, Elena Popova4, and Carsten Welsch2,3 Feasibility of in-beam time-of-flight SPECT/PET gamma imaging based on Silicon Photomultipliers for high precision hadrontherapy Sergey Vinogradov1,2,3, Elena Popova4, and Carsten Welsch2,3 1 Lebedev Physical Institute of the Russian Academy of Sciences, Russia 2 University of Liverpool, Department of Physics, UK 3 Cockcroft Institute of Accelerator Science and Technology, UK 4 National Research Nuclear University (MEPhI), Russia Sergey Vinogradov TOF SPECT PET imaging MEDAMI-2016 Ajaccio, Corsica, France 03 May 2016 1

TOF SPECT/PET gamma imaging Outline Motivation In-beam gamma imaging modalities for hadrontherapy Needs and challenges TOF SPECT/PET gamma imaging New approach of prompt gamma timing (PGT) for dose range assessment Feasibility of combined TOF SPECT/PET modality Analysis of TOF SPECT signal detection Algorithm of TOF SPECT image reconstruction Discussion Harnessing SPECT and PET in one imager Anticipating TOF SPECT/PET benefits and drawbacks Moving forward: critical issues to be addressed (10 ps time resolution?!) Sergey Vinogradov TOF SPECT PET imaging MEDAMI-2016 Ajaccio, Corsica, France 03 May 2016 2

Gamma imaging modalities for hadron therapy verification Pro Contra (TOF) PET Clinically proven Good efficiency Low statistics of events Large background Short decay time Metabolic washout Limited in-beam geometry PGI Collimated cameras Compton cameras High yield of gammas Very low efficiency Scattering in collimator Complexity of collimation Low efficiency Overall complexity M. Dosanjh, ENLIGHT coordinator, in “ENLIGHT Highlights December 2015” “…Alberto Del Guerra (Pisa, Italy), who reviewed the mechanisms of PET imaging in detail, ending with a firm conclusion that a combination of prompt gamma and PET imaging are required for the best estimations of dose placement.” A. del Guerra, “The current status and challenges of detection and imaging in radiation therapy”, in ENLIGHT Meeting, Kraków 18-20 September 2015. Sergey Vinogradov TOF SPECT PET imaging MEDAMI-2016 Ajaccio, Corsica, France 03 May 2016 3

Prompt and annihilation gammas F. Pennazio et al., “A Study of Monitoring Performances with the INSIDE System”, 2015. K. Parodi, Hadron Therapy Workshop 2011 Can we deploy common gamma cameras as used in nuclear medicine? No. – G. Pauch, 2015 Sergey Vinogradov TOF SPECT PET imaging MEDAMI-2016 Ajaccio, Corsica, France 03 May 2016 4

Toward ideal imaging system K. Parodi, Hadron Therapy Workshop 2011 Sergey Vinogradov TOF SPECT PET imaging MEDAMI-2016 Ajaccio, Corsica, France 03 May 2016 5

+ = ? PET + PGI = ? W. Moses, NDIP 2008 Sergey Vinogradov TOF SPECT PET imaging MEDAMI-2016 Ajaccio, Corsica, France 03 May 2016 6

Many approaches and projects in the area In-beam TOF PET, SPECT, Compton @ENVISION project In-beam PET + Secondary charged particles profiler @INSIDE collaboration In beam TOF PET+dSiPM @TU Delft “Turning the physical advantages of protons into true clinical benefits with in-beam TOF-PET” TOF PGI (passive collimation) + dSiPM @TU Delft: “A time-of-flight prompt-gamma camera for real-time, in-situ monitoring of particle radiotherapy” J. Krimmer, et al., “Collimated prompt gamma TOF measurements with multi-slit multi-detector configurations”, JINST, 2015 G. Pausch, Dresden, 2015 Sergey Vinogradov TOF SPECT PET imaging MEDAMI-2016 Ajaccio, Corsica, France 03 May 2016 7

PET + Compton combination Contra: - Complexity - Space consuming design - Low efficiency in Compton mode - ...Multiple trade-offs… Sergey Vinogradov TOF SPECT PET imaging MEDAMI-2016 Ajaccio, Corsica, France 03 May 2016 8

Prompt Gamma Timing (PGT) @OncoRay: emission TOF distribution ~ spatial profile Knowledge of tp(xp) dependence is a key for reconstruction of spatial profile from time distribution C. Golnik et al, Range assessment in particle therapy based on prompt γ-ray timing measurements, Phys. Med. Biol. 59 (2014) 5399 Sergey Vinogradov TOF SPECT PET imaging MEDAMI-2016 Ajaccio, Corsica, France 03 May 2016 9

Latest results in PGT @OncoRay Sergey Vinogradov TOF SPECT PET imaging MEDAMI-2016 Ajaccio, Corsica, France 03 May 2016 10

Latest results in PGT @OncoRay with proton pencil-beam Sergey Vinogradov TOF SPECT PET imaging MEDAMI-2016 Ajaccio, Corsica, France 03 May 2016 11

TOF SPECT (PGT with detector array) + TOF PET Common scintillator–SiPM detector for prompt and annihilation γ In-spill TOF SPECT for prompt gammas Analytical image reconstruction using known TOF & PSF for any points Inter-spill and beam-off TOF PET – as usual High efficiency Simple design Resolvable trade-offs Compatible with CT K. Parodi, Hadron Therapy Workshop 2011 Sergey Vinogradov TOF SPECT PET imaging MEDAMI-2016 Ajaccio, Corsica, France 03 May 2016 12

Gamma emission and detection spatial distributions of pencil-beam profile h – distance to detector plane Proton beam Detector plane Object under test X axis Sergey Vinogradov TOF SPECT PET imaging MEDAMI-2016 Ajaccio, Corsica, France 03 May 2016 13

Analytical model: gamma emission time distribution with known dE/dx, t(x) Sergey Vinogradov TOF SPECT PET imaging MEDAMI-2016 Ajaccio, Corsica, France 03 May 2016 14

Analytical model: gamma detection time distribution D#1, -15cm D#2, 0cm D#3, 15cm D#4, 30cm h = 30cm Proton beam Object under test X axis D#0, start t=0 If energy deposition S(x,E0) and proton velocity V(x,E0) profiles are known, the gamma signal I(xd,t) in a planar detector configuration can be calculated analytically Imaging optics formalism Object: O(x,t) = S(x,E0)*δ(x-V(x,E0)∙t) Transfer function: T(x,t) = PSF(x,h)*δ(t-Td(x,h) Image: I(xd,t)= O(x,t)*T(x,t) Solution I(xd,t) is implicit function of t: TOF probabilistic formalism TOF is a sum of ind. random variables = =Tp-bunch + Tγ-emission + Tγ-detection σ2(TOF)= σ2(Tp)+ σ2(Te)+ σ2(Td) PDF(TOF)=PDF(Tp)*PDF(Te)*PDF(Td) Where PDF(Tp)*PDF(Te) = S(x,E0)*δ(x-V(x,E0)∙t) PDF(Td) = PSF(x,h)*δ(t-TOF(x,h) PDF(TOF) = <Nγ > I(x,t) Sergey Vinogradov TOF SPECT PET imaging MEDAMI-2016 Ajaccio, Corsica, France 03 May 2016 15

Analytical model: TOF detection distributions D#1, -15cm D#2, 0cm D#3, 15cm D#4, 30cm h = 30cm Proton beam Object under test X axis D#1, -15cm D#2, 0cm D#3, 15cm D#4, 30cm h = 30cm Proton beam Object under test X axis How to reconstruct dose range without apriori known t(x)? How to reconstruct dose profile without apriori known S(x)? D#0, start t=0 D#0, start t=0 Sergey Vinogradov TOF SPECT PET imaging MEDAMI-2016 Ajaccio, Corsica, France 03 May 2016 16

Analytical model: reconstruction algorithm with unknown S(x)=dE/dx and t(x) ALL detectors contribute to reconstruction at point x Sergey Vinogradov TOF SPECT PET imaging MEDAMI-2016 Ajaccio, Corsica, France 03 May 2016 17

TOF DOI monolithic scintillator + SiPM for best estimation of interaction point (x,y,z, t) H.T. van Dam S.Siefert D. Schaart ….. 2013 -2014 Sergey Vinogradov TOF SPECT PET imaging MEDAMI-2016 Ajaccio, Corsica, France 03 May 2016 18

Main challenge: timing uncertainty due to bunch width + detector time resolution “PGT method is not applicable at all clinical accelerators: only to those with a specific micro-time structure.” F. Hueso-González “Compton Camera and Prompt Gamma Ray Timing: Two Methods for In Vivo Range Assessment in Proton Therapy” Front. Oncol., 12 April 2016  Beam bunch time width: ~ 0.2 … 2 ns FWHM @ medical accelerators SiPM time resolution ~ 50 ps FWHM @4…5 MeV (assuming ~100 ps @511KeV) Gamma emission timing ~ 1 ps Sergey Vinogradov TOF SPECT PET imaging MEDAMI-2016 Ajaccio, Corsica, France 03 May 2016 19

Possible solutions from accelerator side? New dedicated accelerators with ps-short bunch time width? Laser-plasma wakefield accelerator? MiniLinac? CERN? MIT? Radiabeam? Existing medical accelerators A) Compressing bunch width from ns to ps scale ? B) Reducing number of hadrons per bunch to a single one? Serge Mathot with the first of the four modules that will make up the miniature accelerator (Image: Maximilien Brice/CERN) D#0, start t=0 Sergey Vinogradov TOF SPECT PET imaging MEDAMI-2016 Ajaccio, Corsica, France 03 May 2016 20

Deconvolution as a possible solution? high efficiency = high SNR Original image Standard deconvolution (50 iterat.) Correct sampling deconvolution (FWHM ~ 5 pixels) P. Magain, F. Courbin, S. Sohy, Deconvolution with Correct Sampling, Astrophys. J., V. 494, 1, pp. 472-477 (1998) Sergey Vinogradov TOF SPECT PET imaging MEDAMI-2016 Ajaccio, Corsica, France 03 May 2016 21

We propose TOF SPECT/PET imaging: very efficient and simple Summary We propose TOF SPECT/PET imaging: very efficient and simple Information on a prompt gamma emission profile is collected “free of charge”, it complements TOF PET and improves imaging Trade-offs in the dedicated detector design are minimal (vs. e.g. PET/Compton) thus it could be easily developed and optimized TOF SPECT/PET is fully compatible with CT Perfect alignment with the FAST initiative toward 10 ps timing TOF SPECT/PET seems to be a good starting point for initiative on dedicated medical (mini-) accelerators to enable development of new generation of inexpensive reliable compact hadron treatment facilities of high precision We are looking for a broad collaboration on TOF SPECT/PET Sergey Vinogradov TOF SPECT PET imaging MEDAMI-2016 Ajaccio, Corsica, France 03 May 2016 22

This work was supported in part by Acknowledgements This work was supported in part by EC under Grant Agreement 329100 “SiPM in-depth” Megagrant program of Russia under Grant Agreement 14.А12.31.0006. This participation at MEDAMI was supported in part by MEDAMI, Paul Lecoq COST FAST, Etiennette Auffray Sergey Vinogradov TOF SPECT PET imaging MEDAMI-2016 Ajaccio, Corsica, France 03 May 2016 23

Thank you for your attention The end Thank you for your attention Questions? Objections? Opinions? … Collaborations? Vin@lebedev.ru Sergey Vinogradov TOF SPECT PET imaging MEDAMI-2016 Ajaccio, Corsica, France 03 May 2016 24