SLE5000 HFO Ventilation.

Slides:



Advertisements
Similar presentations
Tidal volume delivery during high-frequency oscillatory ventilation in adults with acute respiratory distress syndrome RESUMENMÉTODOS CONCLUSIONES Carla.
Advertisements

Neonatal Mechanical Ventilation
3100B Theory of Operation and Controls. VIASYS Healthcare, Inc. 3100B Theory of Operation and Controls Approved for sale outside the US in 1998 for patients.
3100A Ventilator. VIASYS Healthcare, Inc. 3100A Ventilator Approved in 1991 for Neonatal Application for the treatment of all forms of respiratory failure.
Initial Ventilator Settings
Effect on Expiratory Flow
1 HFPV VDR4 ® Definition: “The VDR ® is classified as a pneumatically powered, pressure regulated, time cycled, high frequency flow interrupter.” - Delivers.
Theory of HFV.
1 Elsevier items and derived items © 2010 by Saunders, an imprint of Elsevier Inc. Chapter 19 Mechanical Ventilation of the Neonate and Pediatric Patient.
Waves.
3100B Theory of Operation and Controls. SensorMedics 3100B u Electrically powered, electronically controlled piston-diaphragm oscillator u Paw of 5 -
Answer the following questions based on the following figure.
1 HFPV VDR4 ® Definition: “The VDR ® is classified as a pneumatically powered, pressure regulated, time cycled, high frequency flow interrupter.” - Delivers.
Chapter 2 Basic Terms & Concepts of Mechanical Ventilation.
BY: NICOLE STEVENS.  Primary objective of mechanical ventilation is to support breathing until neonates own respiratory efforts are sufficient  First.
1 Elsevier items and derived items © 2010 by Saunders, an imprint of Elsevier Inc. Chapter 20 Neonatal and Pediatric High-Frequency Ventilation.
Ultrasound 10cm 2 Ultrasound (US) Therapeutic ultrasound –Uses acoustic energy, delivered at high frequencies for therapeutic purposes Electrical current.
Mechanical Ventilation Graphical Assessment
 Understand the dual control concept  Understand the pressure regulation mechanism in PRVC  Demonstration of PRVC  Settings and adjustment with Servo.
Several types of HFV  HFPPV  HFJV  HFOV. Principles of Oscillation Richard F. Kita BS, RRT, RCP Edited by Paula Lussier, CRT, NPS, RCP, BS.
“Top Twenty” Session Review for Mechanical Ventilation Concepts What you should remember from the Fall… RET 2264C-12.
How would you represent a transverse wave and a longitudinal wave using physics students as your medium? Discuss amongst yourselves… Be prepared to demonstrate!
FUNCTION GENERATOR.
Start with the airway pressure screen Step 1….determine the CPAP level This is the baseline position from which there is a downward deflection on, at.
Chapter 8 Introduction to Alternating Current and Voltage.
Pressure Regulated Volume Control Mode
Introduction to Basic Waveforms
Ventilator waveform analysis: often ignored bedside assessment
Ventilation Sam Petty Clinical Specialist Physiotherapist
Volume 90, Issue 5, Pages (November 1986)
Mechanical Ventilation in newborns
Magnitude of vacuum pressure for Flexercell at 30 cpm (A) and typical pressure waveforms for conventional mechanical ventilation at 0.5 Hz (CMV;B), high-frequency.
Volume 96, Issue 6, Pages (December 1989)
Mechanical Ventilator 1
PHYS 1443 – Section 003 Lecture #21
ALTERNATING CURRENT AND VOLTAGE
Advanced Modes of Mechanical Ventilation
Oscillation volumes (Vosc; white box) and transcutaneous carbon dioxide pressure (Ptc,CO2; grey box) levels before, during and after a 5-cmH2O increase.
Relationship of oxygenation, as reflected in the ratio of the arterial and alveolar oxygen tensions (a/A ratio), to the mean airway pressure (mPaw) applied.
Basic Concepts in Adult Mechanical Ventilation
Chapter 11 Alternating Current.
Placement of Aeroneb with HFOV
Harmonic Motion and Waves
The patient is being ventilated with 2 types of breaths.
The changes in peak flow and inspiratory time between a minimum rise time (first 2 breaths) and a maximum rise time (last 2 breaths), with the Servo-i.
Airway pressure and flow waveforms during constant flow volume control ventilation, illustrating the effect of an end-inspiratory breath-hold. Airway pressure.
Trigger pressure-time product (PTP) with zero pressure support, with no leak, medium leak, and large leak. Trigger pressure-time product (PTP) with zero.
Mechanical Ventilation: State of the Art
Auto Bag Jerome H. Santoyo Jessica S. Yeager Advisor: Richard Fries
Lung simulator diagram of airway pressure release ventilation (APRV): volume (yellow), lung pressure (white), and flow (orange)/time curve. Lung simulator.
Hanning and Rectangular Windows
High-frequency oscillation in adolescents
High-frequency oscillatory ventilation and an interventional lung assist device to treat hypoxaemia and hypercapnia  David M. , Heinrichs W.   British.
Practical assessment of respiratory mechanics
A) Experimental system to carry out a bench test of noninvasive ventilators. a) Experimental system to carry out a bench test of noninvasive ventilators.
Even though this patient is undergoing positive-pressure mechanical ventilation, the first 4 breaths have a relatively negative pressure (ie, pressure.
A: Pressure (green) and volume (black)/time curve in airway pressure release ventilation (APRV). A: Pressure (green) and volume (black)/time curve in airway.
The third breath has a negative deflection (ie, below PEEP) at the end of the mechanical breath (arrow A) associated with a flow increase (arrow B), indicating.
Simulated screenshot of flow starvation in volume control continuous mandatory ventilation. Simulated screenshot of flow starvation in volume control continuous.
Characteristics of a pressure-supported breath.
The peak flows (60 L/min) and flow patterns are the same for all the breaths. The peak flows (60 L/min) and flow patterns are the same for all the breaths.
During this tracing of 30 seconds, the ventilator displays that the patient rate is 16 breaths/min. During this tracing of 30 seconds, the ventilator displays.
Comparison of airway pressure release ventilation (APRV) (blue curve) and biphasic positive airway pressure (BIPAP) (black curve). Comparison of airway.
This tracing depicts 30 seconds of information.
Flow, airway pressure, and transversus abdominis electromyogram (EMG) waveforms from a mechanically ventilated patient with COPD receiving pressure-support.
Inspiratory load compensation responses before and after inspiratory muscle strength training (IMST) in the unweaned versus weaned subjects, with a 10.
Depiction of an expiratory flow curve.
The changes in peak flow and inspiratory time between a minimum rise time (first 2 breaths) and a maximum rise time (last 2 breaths), with the Servo-i.
Difference between mid-frequency ventilation (MFV), volume control continuous mandatory ventilation (VC-CMV), and pressure control CMV (PC-CMV) when frequency.
Representative waveforms for each of the devices tested from which the oscillatory f was counted. Representative waveforms for each of the devices tested.
Presentation transcript:

SLE5000 HFO Ventilation

SLE5000 HFO Ventilation

HFO Studies – 2014-15 The SLE5000 was the most powerful neonatal HFO ventilator in terms of volume delivery with performance similar to that of the Sensormedics 3100A New Generation Neonatal High Frequency Ventilators: Effect of Oscillatory Frequency and Working Principles on Performance Serge Grazioli MD, Oliver Karam MSc MD, and Peter C Rimensberger MD; Respir. Care 2015 The Sensormedics A and B and the SLE5000 delivered square waves; all the other oscillators generated sine waves Pressure and Flow Waveform Characteristics of Eight High-Frequency Oscillators Edward R. Harcourt, BSc; Jubal John, MD; Peter A. Dargaville, MD; Emanuela Zannin, PhD; Peter G. Davis, MD; David G. Tingay, PhD; PCCM 2014

Pressure and flow waveform characteristics of seven high frequency ventilators Figure 1. Representative Pao and V’ao waveforms during two oscillatory cycles for each HFV tested. Unique incisurae patterns were noted within the waveforms of the SM3100B and SLE5000. “The Sensormedics A and B and the SLE5000 delivered square waves; all the other oscillators generated sine waves.” Pressure and Flow Waveform Characteristics of Eight High-Frequency Oscillators Edward R. Harcourt, BSc; Jubal John, MD; Peter A. Dargaville, MD; Emanuela Zannin, PhD; Peter G. Davis, MD; David G. Tingay, PhD; PCCM 2014

Pressure and flow waveform characteristics (cont.)

8 Ventilators Pressure waveforms at 5Hz 10Hz 15Hz

Pressure and Flow Waveforms

Rapid rise in pressure (approx Rapid rise in pressure (approx. 1/100 s) from the SLE5000 and Sensormedics. Greater amplitude achieved by SLE5000 on same settings.

Period: 0.1s at 10Hz Period: 0.1s at 10Hz Rapid rise in pressure in approx. 1/100 s Slower rise in pressure in approx. 1/20 s The pressure wave created by the rapid rise in pressure from the SLE5000 contains much more energy than the slow rise in pressure from the other ventilators.

0.1s 0.1s Rapid rise in pressure in approx. 1/100 s Slower rise in pressure in approx. 1/30 s

0.1s 0.1s Rapid rise in pressure in approx. 1/100 s Slower rise in pressure in approx. 1/20 s

The settings for all ventilators were: Mean pressure: 10mbar; Amplitude: 30mbar (+/-15mbar) …thus, target pressures: -5mbar to +25mbar.  Target Amplitude -5 to +25mb Only the SLE achieved the target negative pressure of -5mbar (-7.2mbar at 10Hz). Even the Sensormedics 3100B failed to achieve a negative pressure. Amplitude 21.4 (21.7 to 0.3) in 1:2 i:e ratio 15.9 (17.0 to 1.1) in 1:1 i:e ratio

Greatest amplitude was achieved by the SLE5000 at 23.4 mbar. Ventilator I:E Ratio Peak Pressure Trough Pressure Amplitude % Target Amplitude Target Ventilator Settings 25.0 -5.0 30.0 Sensormedics 3100A 1:2 16.3 3.0 13.3 44% 1:1 16.5 0.7 15.8 53% Sensormedics 3100B 21.7 0.3 21.4 71% 17.0 1.1 15.9 SLE5000 16.2 -7.2 23.4 78% Fabian 15.6 15.3 51% Leonie+ 13.9 0.8 13.1 Sophiea 14.3 -0.3 14.6 49% VN500 12.8 43% BL8000b 10.7 2.2 8.5 28% Data from Table 1, Harcourt et al, PCCM 2014 Greatest amplitude was achieved by the SLE5000 at 23.4 mbar.

Amplitude maintained at higher frequencies Amplitude reduces at higher frequencies

Reducing the frequency to achieve greater amplitude (slower oscillation) risks over-distending the lung (according to Dr Peter Rimensberger speaking on HFO at EAPS 2014). At higher frequencies, the lung does not have time to stretch even at high amplitude oscillation around the appropriate mean pressure.

All the ventilators capable of conventional ventilation and HFO, except for the SLE5000, lost amplitude in the pressure waveform at 10 Hz and lost more pressure at 15Hz.

Rimensberger, Resp. Care 2015 Simulated Term infant 4kg: resistance 70cmH2O/L/s compliance 4mL/cmH2O Mean airway pressure 15 cm H2O , :E 1:1. “Note that, at a frequency of 15 Hz under full-term infant settings, the Babylog VN500 was not able to increase the pressure amplitude beyond 36 cm H2O (white marker)”

Rimensberger, Resp. Care 2015 “The SLE5000 was the most powerful neonatal HFO ventilator in terms of volume delivery with a performance similar to that of the Sensormedics 3100A.”