Motion in Radiotherapy Martijn Engelsman May as well have been called “Radiotherapy in motion” because motion-management has been, and still is, a rapidly evolving important part of Radiotherapy My feeling is that at least some of the previous presentations have gone pretty fast. Out of my own experience it is easy to assume basic understanding. Anecdote about treatment planning. I’ll try to be thorough and apologize to non-students and possibly even to students.
Contents What is motion ? Why is motion important ? Motion in practice Qualitative impact of motion Motion management Motion in charged particle therapy
What is motion ?
Motion in radiotherapy Aim of radiotherapy Deliver maximum dose to tumor cells and minimum dose to surrounding normal tissues “Motion” Anything that may lead to a mismatch between the intended and actual location of delivered radiation dose
Radiotherapy treatment process Diagnosis Patient immobilization Imaging (CT-scan) Target delineation Treatment plan design Treatment delivery (35 fractions) Patient follow-up
Why is motion important ?
PTV concept (1) (ICRU 50 and 62) High dose region PTV (Planning Target Volume): = 8 cm, V = 268 cm3 CTV (Clinical Target Volume): = 6 cm, V = 113 cm3 GTV (Gross Tumor Volume): = 5 cm, V = 65 cm3 Simplified (hej, patients are not square with round tumors, but as a physicist I’m allowed to shape reality into a comprehensive model), but the basis of radiotherapy treatment for the vast majority of patients
PTV concept (2) Margin from GTV to CTV Margin from CTV to PTV Typically 5 mm or patient and tumor specific Improved by: Better imaging Physician training Margin from CTV to PTV Typically 5 to 10 mm Tumor location specific Motion management Smart treatment planning GTV CTV PTV High Dose
Example source of motion 35 Fractions = 35 times patient setup For the vast majority of radiations are isocentric meaning that for each fraction the patient is positioned with respect to the treatment machine isocenter and then the treatment is delivered as a whole. During a treatment fraction, the patient will be irriated from several, static, angles. Many patients set-up to lasers only. Many exceptions and more sophisticated approaches, but for explaining types of motion lasers will visualize nicely. 35 fractions; never the same alignment twice. Patient skin is “loose” so markers can move. Patient can gain or loose weight resulting in “motion” The lasers have an inherent width The lasers may be half a mm displaced with respect to the iso-center of the linear accelerator. THAT’S WHY WE DO REGULAR QUALITY ASSURANCE www.pi-medical.gr
Sources of motion Patient setup Patient breathing / coughing Patient heart-beat Patient discomfort Target delineation inaccuracies Non-representative CT-scan Target deformation / growth / shrinkage Etc., etc. etc. I’ll discuss all of these in more detail within a few slides, this is just to give an idea. Patient setup: just imagine lining up to the lasers every fraction Target delineation: The physician uses the CT-scan to draw where he thinks is target. Inherently flawed approach Target deformation: Filled bladder / empty bladder. Gas in rectum, etc.
Subdivision of motion Systematic versus Random Inter-fractional versus Intra-fractional Treatment Preparation versus Treatment Execution Less commonly used
Systematic versus Random Same error for all fractions (possibly even all patients). Random Unpredictable. Day to day variations around a mean. Known but neither Breathing, heartbeat Relatively known means neither type. Breathing can furthermore be both systematic and random
Setup errors for three patients y Beam’s Eye View x Imagine I’m looking at the patient in the direction of the treatment beam. Center of patient tumor is supposed to be aligned with the axis origin.
Setup errors for a single patient Random (x) Random (y) Systematic (y) Systematic (x) Whenever there is random, there is also systematic.
Inter-fractional versus Intra-fractional Variation between fractions Intra-fractional Variation within a fraction Treatment preparation errors affect the whole treatment -> hence: systematic Treatment execution errors
Treatment preparation versus treatment execution Patient immobilization CT-scan Target delineation Treatment plan design Treatment delivery (35 fractions) Always systematic Systematic and/or random Treatment preparation Treatment execution
Motion in practice
Treatment preparation Target delineation Steenbakkers et al. Radiother Oncol. 2005; 77:182-90 Systematic Inter-fractional Treatment preparation Random Intra-fractional Treatment execution
Treatment preparation Patient setup x y Systematic Inter-fractional Treatment preparation Random Intra-fractional Treatment execution
Treatment preparation Target deformation / motion 1/3 Bladder Target Notice: variation in bladder shape due to bladder filling, may be different from day to day Bladder extends more downward in second scan Variation in rectum filling as well (both gasseous filling and non-gasseous filling) Note: overlap between bladder and target because of automatic expansion of the tumor volume Systematic Inter-fractional Treatment preparation Random Intra-fractional Treatment execution
Treatment preparation Target deformation / motion 2/3 Bladder Target Systematic Inter-fractional Treatment preparation Random Intra-fractional Treatment execution
Target deformation / motion 3/3 Patient immobilization CT-scan Target delineation Treatment plan design Treatment delivery (35 fractions)
Treatment preparation Breathing motion Movie by John Wolfgang Systematic Inter-fractional Treatment preparation Random Intra-fractional Treatment execution “ ”
Qualitative impact of motion
Importance of motion Breathing motion / heart beat Systematic errors Raise your hand to vote Breathing motion / heart beat Systematic errors Random errors Let’s “prove” it Most Least Almost least Of course, it depends on the magnitude of motion you can typically expect. But, without telling you magnitudes what do you think? I’ll give you magnitudes later
Simulation parameters (1) To enhance the visible effect of motion: High dose conformed to CTV GTV CTV PTV High Dose GTV CTV High Dose
Dose (% of prescribed dose) distance from beam axis (mm) Simulation parameters (2) -60 -50 -40 -30 -20 -10 10 20 30 40 50 60 70 80 90 100 95 % Dose (% of prescribed dose) distance from beam axis (mm) CTV Parallel opposed beams GTV CTV High Dose Direction of motion
DVH reduction into: Tumor Control Probability (TCP) Assumption: homogeneous irradiation of the CTV to 84 Gy results in a TCP = 50 %
Tumor motion and tumor control probability Amplitude of breathing motion (mm) Random setup errors (1SD) Systematic setup error TCP (%) 47.3 5 - 47.0 10 46.3 15 44.3 46.8 43.5 36.9 45.5 40.1 6.0 Typical motion:
Importance of motion Breathing motion / heart beat Systematic errors Therefore … Breathing motion / heart beat Systematic errors Random errors Most Least Almost least Of course, it depends on the magnitude of motion you can typically expect. But, without telling you magnitudes what do you think? I’ll give you magnitudes later
Why are systematic errors worse ? Random errors / breathing blurs the cumulative dose distribution dose Systematic errors shift the cumulative dose distribution CTV Slide by M. van Herk
In other words… Systematic errors Same part of the tumor always underdosed Random errors / Breathing motion / heart beat Multiple parts of the tumor underdosed part of the time, correctly dosed most of the time But don’t forget: Breathing motion and heart beat can have systematic effects on target delineation
Motion management
Radiotherapy treatment process Patient immobilization CT-scanning Target delineation Treatment plan design Treatment delivery
Patient immobilization Leg pillow Intra-cranial mask www.sinmed.com GTC frame Breast board www.sinmed.com www.massgeneral.og
Benefits of immobilization Reproducible patient setup Limits intra-fraction motion Patient spends 10-60 minutes on a not too comfortable treatment couch
Radiotherapy treatment process Patient immobilization CT-scanning Target delineation Treatment plan design Treatment delivery
CT-scanning Multiple CT-scans prior to treatment planning Reduces geometric miss compared to single CT-scan 4D-CT scanning Extent of breathing motion Determine representative tumor position See lecture “Advances in imaging for therapy”
Radiotherapy treatment process Patient immobilization CT-scanning Target delineation Treatment plan design Treatment delivery
Target delineation Multi-modality imaging CT-scan, MRI, PET, etc. Physician training and inter-collegial verification Improved drawing tools and auto-delineation
Radiotherapy treatment process Patient immobilization CT-scanning Target delineation Treatment plan design Treatment delivery
Treatment plan design Choice of beam angles e.g. parallel to target motion Smart treatment planning Robust optimization IMRT See, e.g., lecture “Optimization with motion and uncertainties”
Radiotherapy treatment process Patient immobilization CT-scanning Target delineation Treatment plan design Treatment delivery Large section on motion management in treatment delivery
Magnitude of motion in treatment delivery Systematic setup error Laser: S = 3 mm Bony anatomy: S = 2 mm Cone-beam CT: S = 1 mm Random setup errors s = 3 mm Breathing motion Up to 30 mm peak-to-peak Typically 10 mm peak-to-peak Tumor delineation See next slide Numbers express order of magnitude and are a little bit fuzzy because they sometimes express setup error of bony anatomy, sometimes of actual tumor location.
Tumor delineation 5? 22 Patients with lung cancer 11 Radiation oncologists from 5 institutions Comparison to median target surface Rad. Onc. # Mean volume (cm3) Mean distance (mm) Overall SD 1 36 -6.4 15.1 2 48 -3.7 11.6 3 53 -4.3 13.9 4 55 -2.4 7.0 5 58 -3.3 12.7 6 67 -1.6 10.0 7 69 -1.2 6.2 8 72 -1.0 6.6 9 76 -0.2 7.4 10 93 0.9 5.7 11 129 0.4 6.1 All 69 ( 25) -1.7 10.2 Perhaps it is 5 mm for any given point on other targets. But then we are talking about the GTV. Delineation of the CTV may be more error-prone because it is by nature non-visible. Naturally, better imaging helps a lot, e.g. PET 5? Steenbakkers et al. Radiother Oncol. 2005; 77:182-90
Motion management
Motion management for setup errors Portal imaging
Obtained from Treatment Planning System Portal imaging Obtained from Treatment Planning System Obtained in treatment room Digitally Reconstructed Radiograph
Setup protocol NAL-protocol (No Action Level) Portal imaging for first Nm fractions Calculate a single correction vector compared to markers for laser setup Daily imaging is the standard here at the proton center. Usually weekly imaging is performed with lasers used for de Boer HC, Heijmen BJ. Int J Radiat Oncol Biol Phys. 2001;50(5):1350-65 Lasers only
Motion management for breathing In treatment plan design Margin increase Overcompensating dose to margin Robust treatment planning See, e.g., lecture “Optimization with motion and uncertainties” Control patient breathing Breath-hold Gated radiotherapy
Breathing traces Trace Probability Density Function PDF = 1) 2) 3)
Margin increase
Effect of blurring on dose profile (conformal) Only a limited shift in 95% isodose level This is for a photon profile in lung. The shift would be larger for a steeper profile (e.g. due to IMRT)
Margin for breathing (conformal) 5 10 15
Margin for breathing (IMRT) Hypothetically Sharp Dose Distribution
Margin for breathing (IMRT) 5 10 15 IMRT
Breath hold
Control / stop patient breathing Exhale position most reproducible Inhale position most beneficial for sparing lung tissue
Breath hold techniques Voluntary breath hold Rosenzweig KE et al. The deep inspiration breath-hold technique in the treatment of inoperable non-small-cell lung cancer. Int J Radiat Oncol Biol Phys. 2000;48:81-7 Active Breathing Control (ABC) Wong JW et al. The use of active breathing control (ABC) to reduce margin for breathing motion. Int J Radiat Oncol Biol Phys. 1999;44:911-9 Abdominal press Negoro Y et al. The effectiveness of an immobilization device in conformal radiotherapy for lung tumor: reduction of respiratory tumor movement and evaluation of the daily setup accuracy. Int J Radiat Oncol Biol Phys. 2001;50:889-98 Usually, lung cancer patients have an impaired lung function and can not really hold there breath that easily.
Gating
Gated radiotherapy Gating window External or internal markers Usually 20% duty cycle Some residual motion
Gating benefits and drawbacks + Less straining for patient than breath-hold Increased treatment time Internal markers Direct visualization of tumor (surroundings) Invasive procedure / side effects of surgery External markers Limited burden for patient Doubtful correlation between marker and tumor position Intra-fractional Inter-fractional - + - + -
Motion in charged particle therapy
T. Bortfeld
Range sensitivity Spherical tumor in lung Paralell opposed - photons Single field - photons Single field - protons Displayed isodose levels: 50%, 80%, 95% and 100%
Range sensitivity Spherical tumor in lung Paralell opposed - photons Single field - photons Single field - protons Displayed isodose levels: 50%, 80%, 95% and 100%
Range sensitivity Spherical tumor in lung Paralell opposed - photons Single field - photons Single field - protons Displayed isodose levels: 50%, 80%, 95% and 100%
Dose-Volume Histogram (protons) PTV (static) CTV GTV CTV-GTV
SOBP Modulation Beam High-Density Structure Target Volume Critical Range Compensator Body Surface
Passive scattering system Aperture Range Compensator + = Lateral conformation Distal conformation
Smearing the range compensator High-Density Structure Target Volume Beam Critical Structure Range Compensator Body Surface
Smearing the range compensator High-Density Structure Target Volume Beam Critical Structure Range Compensator Body Surface
C D Smear Setup Error A B 10 C D B 10 C D Displayed isodose levels: 50%, 80%, 95% and 100%
Motion management in particle therapy Passive scattered particle therapy For setup errors and (possibly) breathing motion Lateral expansion of apertures Smearing of range compensators IMPT See, e.g., lecture “Optimization with motion and uncertainties”