MRI Hardware: An Overview for Clinicians

Slides:



Advertisements
Similar presentations
Physical Layer: Signals, Capacity, and Coding
Advertisements

Receiver Bandwidth affects signal-to-noise ratio (SNR)
Pulse Timing parameters & Weighting
Fund BioImag : Echo formation and spatial encoding 1.What makes the magnetic resonance signal spatially dependent ? 2.How is the position of.
Chapter3 Pulse-Echo Ultrasound Instrumentation
Assessment of tumoural ADC’s in rectal Tumours using Burst: New methodological Developments SJ Doran 1, ASK Dzik-Jurasz 2, J Wolber 2, C Domenig 1, MO.
Magnetic Resonance Imaging Lorenz Mitschang Physikalisch-Technische Bundesanstalt, 23 rd February 2009 I. Basic Concepts.
Fund BioImag : Echo formation and spatial encoding 1.What makes the magnetic resonance signal spatially dependent ? 2.How is the position of.
MRI Phillip W Patton, Ph.D..
Richard Wise FMRI Director +44(0)
BE 581 Lecture 3- Intro to MRI.
PHYSICS OF MAGNETIC RESONANCE
MR TRACKING METHODS Dr. Dan Gamliel, Dept. of Medical Physics,
1 NMR Instrumentation Modern Pulsed Fourier Transform NMR Spectrometer with superconducting magnet Why? Superconducting Magnet Computer Electronics Console.
Parameters and Trade-offs
ECE 4321: Computer Networks Chapter 3 Data Transmission.
Magnetic Resonance Imaging
Principles of MRI: Image Formation
The Basics of MRI The Basics of MRI. Current MRI technology displays images as multiple sets of gray tone images. Visualization and interpretation of.
Nuclear Magnetic Resonance
Resident Physics Lectures
NMR Spectroscopy Tuning / Matching Shimming and Lock.
Magnetic Resonance Imagining (MRI) Magnetic Fields
Chapter 5 How the spectrometer works Magnet Probe Transmitter SynthesizerReceiver ADC Pulse programmer Computer 5.1 The magnet: 1. Strength: 14.1 T for.
1 NMR Practical Aspects (I) Processes involving in a NMR study: 1. Sample Preparation/Considerations 2. Setting up Spectrometer 3. Setting up acquisition.
Department of Electronic Engineering City University of Hong Kong EE3900 Computer Networks Data Transmission Slide 1 Continuous & Discrete Signals.
The NMR spectrometer Magnet Probe Coils Transmitters
Basics of Magnetic Resonance Imaging
William Stallings Data and Computer Communications 7th Edition (Selected slides used for lectures at Bina Nusantara University) Data, Signal.
Magnetic Resonance Imaging Basic principles of MRI This lecture was taken from “Simply Physics” Click here to link to this site.
G Practical MRI 1 Receive Arrays.
DOPPLER SONOGRAPHY By Dr/ Dina Metwaly.
Tissue Contrast intrinsic factors –relative quantity of protons tissue proton density –relaxation properties of tissues T1 & T2 relaxation secondary factors.
Radiofrequency Pulse Shapes and Functions
MRI: an Introduction By Mohammad Ali Ahmadi Pajouh Amirkabir University of Technology Biomedical Eng. Dep.
Magnetic Resonance Imaging
RF Coils Used In MRI.
Medical Imaging Systems: MRI Image Formation
Chapter 5 Lecture 10 Spring Nonlinear Elements 1. A nonlinear resistance 2. A nonlinear reactance 3. A time varying element in you circuit or system.
Magnetic Resonance Imaging 4
Principles of MRI Physics and Engineering
Principles of Magnetic Resonance
The comparison of MRI imaging In 3.0 T & 1.5T By a.r.shoaie.
Magnetic Resonance Imaging
Seminar October, 2008 j. brnjas-kraljević. Imaging (MRI)  tomography technique  tomography technique – the volume image is built up by images of thin.
Medical Imaging Systems: MRI Image Formation
CE 4228 Data Communications and Networking
Basic of Magnetic Resonance Imaging Seong-Gi Kim Paul C. Lauterbur Chair in Imaging Research Professor of Radiology, Neurobiology and Bioengineering University.
September, 2003BME 1450 Introduction to NMR 1 Nuclear Magnetic Resonance (NMR) is a phenomenon discovered about 60 years ago.
Rad T 265 MRI Lecture. No Magnetic Field = No Net Magnetization Protons align with a magnetic field…
Pei-Ann Lin and PJ Velez December 13, 2011
Human Subjects in fMRI Research Credits: Robert Savoy, Ph.D. Franz Schmitt, Ph.D.
1 Introduction to Magnetic Resonance Imaging درس مقدمه ای بر فیزیک پزشکی Sahand University of Technology Faculty of Electrical Engineering M. Shamsi.
G Practical MRI 1 Gradients.
Active Detuning of Inductively Coupled Surface Coils Jolinda Smith Lewis Center for Neuroimaging University of Oregon.
Electromagnetic Spectrum
Ultra-low Field Nuclear Magnetic Resonance Measurements with SQUIDs
 This depends on a property of nuclei called spin.  Gyroscope: Principle: As long as its disc remains spinning rapidly the direction of the spin axis.
MAGNETIC RESONANCE IMAGING by PRADEEP V.EPAKAYAL. Mem.no L.
Ultrasound Physics Image Formation ‘97. Real-time Scanning Each pulse generates one line Except for multiple focal zones frame one frame consists of many.
Ultrasound Physics Image Formation ‘97.
FMRI data acquisition.
(Instrument part) Thanundon Kongnok M
Ultrasound Physics Image Formation ‘97.
Imaging Science Department, IC Jo Hajnal
MRI Pulse Sequences: IR, EPI, PC, 2D and 3D
Human Subjects in fMRI Research
(4)ELECTRONIC SUPPORT SYSTEM
Superconducting Magnets
Presentation transcript:

MRI Hardware: An Overview for Clinicians Richard G. Spencer, M.D., Ph.D. Nuclear Magnetic Resonance Unit National Institute on Aging, National Institutes of Health Baltimore, Maryland USA

ii) to give good answers Why go to a talk? i) to ask good questions ii) to give good answers

1. Magnet 2. Gradients 3. Transmitter 4. Receiver 5. Probe • Considerations • Reasonable questions • A few details • A few specifications

Magnet

Magnet Considerations • High field for high signal-to-noise ratio • Weight • Diameter and length (patient experience; field homogeneity) • Homogeneity (spatial); Field stability (temporal) • Configuration (access; patient experience) • Cryogenic efficiency (operating costs)

What type of magnet is it? Is it shielded? Field strength? Bore size? How stable is the field? What’s the field homogeneity?

Resistive Magnet Type Wire Solenoid • Field stability  need a very stable 10’s of kW power supply • Power requirements  B2  cooling requirements  0.2 T or less

Permanent Magnet Type • Excellent field stability • Configuration: open systems are available • No power consumption however... • Weight--can be enormous: iron 0.2 T whole body weighs about 25 tons • 0.2 T neodymium alloy ~ 5 tons • Homogeneity--can be a problem

Open Permanent Magnet System Siemens Viva--0.2 T

Superconducting Magnet Type • Required for high field systems • Homogeneous field • Stable field however... • Expensive • Quench phenomenon Siemens/Bruker Magnetom Allegra-3T, head only

• Polarization of atomic nuclei Field Strength • Polarization of atomic nuclei Larger field  More spin alignment  More signal from each pixel or voxel

increased at high field Trading Rules Signal-to-noise increased at high field Resolution Speed

More signal from each pixel means: Each pixel can be smaller

More signal per unit time means: Images can be acquired faster Cardiac MRI Functional MRI RN Berk, UCSD S Smith, Oxford

Field Strength in Spectroscopy • Greater spectral resolution 31P NMR Spectrum of Skeletal Muscle at 1.9 T 31P NMR Spectrum of Rat Heart at 9.4 T Baseline unresolved Baseline resolved PCr PCr -ATP -ATP

Field Strength Considerations However, at higher field: • increased chemical shift artifacts, e.g. fat/water • increased susceptibility artifacts • increased siting cost • increased initial cost

Field Strength Typical clinical systems: 0.2 T to 1.5 T to 3 T Whole-body: 3 T, 4 T, … , 8 T, 9.4 T available Ohio State, Bruker 8T Whole-body System Gradient-echo Images

Field Strength High field especially useful for fMRI and spectroscopy--less obviously so for standard imaging • Typical animal systems: 4.7 T, 7 T,…, 9.4 T, 11 T horizontal 9.4 T, 11 T vertical

• Bigger is better--e.g. “head only” fits only heads! Bore Size • Bigger is better--e.g. “head only” fits only heads! 600 mm magnet warm bore 570 mm magnet at shoulder incl. shim 360 mm gradient coil inner diameter 265 mm RF-headcoil inner diameter Siemens/Bruker Magnetom Allegra-3T, head only

Bore Size • Human i) Whole Body (typical: 100 cm) ii) Head only (typical: 80 cm) •Animal 15 cm, 20 cm, 30 cm, 40 cm, ... However: • Larger bore larger fringe field • Larger gradient sets can be slower • Cost

Magnetic Shielding: Containment of the Fringe Field • Fringe field: portion of the magnetic field that extends beyond the magnet bore • 5 G taken as maximum safe public exposure • Effect on e.g. pacemakers, steel tools, and magnetic cards • Effect from e.g. moving cars and elevators

Magnetic Shielding Considerations • Weight • Footprint • Expense including space Note: fringe field is 3D

Magnetic Shielding Options • Unshielded: • Lightest, cheapest However: largest field footprint, most expensive space • Passive shielding: ferromagnetic material placed outside magnet • Small field footprint: decrease by factor of 2 in all directions However: heaviest--10’s of tons of iron • Active shielding: electromagnetic counter-windings outside the main magnet coil • Similar field footprint as for passive shielding • Mild increase in weight vs unshielded However: highest magnet expense

Sample Magnet Specifications • Field stability: better than 0.1 ppm/hour drift Note: fat-water separation = 3.5 ppm • Field homogeneity*: better than 0.3 ppm over 22 cm and 5 ppm over 50 cm diameter spherical volume (DSV) *without room-temperature shims 1 ppm = 0.00001%

Gradients

Magnetic Field Gradients: required along all three axes Coils Magnetic Field—varies with position Field Strength I Distance I (current) direction indicated by

Operation of Gradients No gradient: Bz,Local = B0  z With gradient: Bz,Local = B0 + z Gz Spatial variation of the B field permits spatial mapping of spins: frequency  spatial position

Gradient Considerations • Want high in-plane resolution • Want narrow slice capability in 2D imaging • Want images which aren’t distorted • Want to be able to image quickly

What’s the gradient linearity? What’s the gradient strength? What’s the rise time of the gradients? Are the gradients actively shielded?

Two Different "Bandwidths" in MRI • Excitation Bandwidth of a radiofrequency pulse The pulse excites spins in this range of frequencies • Receiver Bandwidth The receiver can detect signals in this range of frequencies

Sampling during MRI signal acquisition Excitation BW 90 180 echo signal rf Gs GPE Detection BW Gread ADC Sampling time, t

Excitation Bandwidth and Gradient Strength

Excitation Bandwidth Excitation pulse Frequency band excited Fourier rf amplitude BW Short pulse: Broad band t (milliseconds)  (kilohertz) BW Long pulse: Narrow band Fourier rf amplitude t (milliseconds)  (kilohertz)

Pulse Bandwidth a.k.a. Excitation Bandwidth Longer duration pulses  • narrower excitation bandwidth however… • longer echo time—loss of signal from short T2 species • greater sample heating • relaxation effects during pulses Would like to be able to use short pulses and still have narrow slice

Slice gradient strength + Pulse bandwidth  Slice thickness  Gsz =  BWexcitation(Hz) 2 G/cm = 20 mT/meter Frequency   G 0.5 G/cm 2000 Hz Pulse Bandwidth Spatial dimension 2.5 mm Slice Thickness 1 cm Slice Thickness

Effect of Slice Gradient Strength on Slice Thickness  Slice Thickness

Receiver Bandwidth and Gradient Strength

MRI maps a frequency range to a spatial range  Gr FOV =  BWreceiver(Hz) Receiver BW=4,300 Hz Receiver BW=4,300 Hz Gr = 2 Gauss/cm Gr = 0.5 Gauss/cm  G  FOV = 2 cm FOV = 5 mm

Effect of Read Gradient Strength on In-plane Resolution

Gradient Eddy Currents Faraday’s Law  Binduced Bdesired Increased currents with more rapid switching

Binduced causes B-field distortions Gradient amplifier driving waveform Resulting gradient waveform Cure #1: Pre-emphasis Gradient amplifier driving waveform Resulting gradient waveform

Cure #2: Shielded Gradients Gradient Coils Magnet bore Gradient Shield Coils

Effect on Image Accuracy Gradient Linearity Effect on Image Accuracy Bz, Local = B0 + Gz(z) nonlinear Nonlinear gradient: geometric distortion Bz, Local = B0 + z Gz linear Linear gradient: nondistorted image z

Sample Gradient Specifications • Gradient strength: 2.5 G/cm (clinical) 4 G/cm, 8 G/cm 10-100 G/cm (animal) Increased gradient strength  higher resolution, narrower slices however: also increased heating, increased rise time (slower) • Gradient switching time (rise and fall time) depends upon inductance and driving voltage: 0.2 ms to rise to 2 G/cm Faster switching  better performance in rapid imaging sequences • Gradient linearity: 5% over 22 cm diameter spherical volume Better linearity  less image distortion

Transmitter Considerations • Need to uniformly excite large bandwidth • Require accurate shaped pulses (time, amplitude) • Desire easy-to-control power output • Frequency stability

What’s the transmitter power? What’s the linearity of the amplifier?

Transmitter Linearity Low power rf Input 1 volt Gain set to two Output 2 volts Low power rf Transmitter Gain set to two Input 2 volts Linear: Output 4 volts Nonlinear: Output = 3.5 volts

Transmitter linearity is important for accurate shaped pulses Low power input to transmitter amplifier High power output from transmitter amplifier ...and for calibrating pulses

What is a decibel? The dB scale expresses amplification or attenuation as the logarithm of a ratio: dB = 20 log10(A2/A1) A2/A1 dB 2 6 10 20 100 40

Sample Transmitter Specifications • Maximum output: 15 kW • Linear to within 1 dB over a range of 40 dB • Output stability of 0.1 dB over 10 ms pulse • Output stability of 0.1 dB pulse-to-pulse

Receiver Considerations Goal: Receive the microvolt NMR signal and convert it to a detectable echo/FID • Without corruption by noise • With faithful amplitude reproduction • With faithful frequency reproduction

What’s the digitizer resolution? What’s the receiver bandwidth?

Receiver Bandwidth The largest detectable frequency Suppose Gr = 1 G/cm FOV = 10 cm  Signal frequency range = 40 kHz Nyquist: have to sample at 80 kHz to accurately record frequencies

Receiver Digitization 12 bits: maximum amplitude ratio observable = 212 / 1= 4096 16 bits: maximum amplitude ratio observable = 216 / 1= 65,536 High digitization in imaging: • improved use of data from the periphery of k-space (low signal, but high-resolution information) High digitization in spectroscopy: • water suppression • low-concentration metabolites

Sample Receiver Specifications • Digitizer: 14 bit or greater • Receiver Bandwidth: 500 kHz ( 250 kHz) or greater • Preamp gain: 30 dB or greater • Preamp noise: 0.7 dB or less • System noise: 1.4 dB or less

Probe

Probe* Considerations i) Transmit coils ii) Receive coils Basic probe circuit CM: matching capacitor CT: tuning capacitor R: intrinsic resistance L: the coil • High SNR: high sensitivity, low noise • Good homogeneity for transmission and reception • Efficient power transmission to the sample *probe = rf coil

Probe Construction The “coil” component

How’s the rf inhomogeneity? How long’s the 90 degree pulse? What’s the coil geometry? What’s the Q? How’s the rf inhomogeneity?

Q also defines the sensitivity of the coil during reception Q is for Quality (“efficiency”) of a coil Q defines how much of the transmitted energy is delivered to the sample at the proper frequency Z Higher Q Lower Q  Q also defines the sensitivity of the coil during reception

One or more loops of wire form a surface coil Typical Surface Coil One or more loops of wire form a surface coil Bruker Biospec surface coil Siemens loop flex surface coil

Surface Coil Strength of B1 field Depth into patient ~ D D Advantages: • High B1 field: the coil is close to the sample • Low noise: sees only desired imaging region • Can be arranged into a phased array for greater spatial coverage Disadvantages: • Inhomogeneous transmission • Inhomogeneous reception Shaded images

Multiple loops form a solenoidal coil B1 B0 Advantages: • Homogeneous field • High SNR Disadvantages • Can’t orient with axis along the B0 field--the spins won’t flip

Multiple parallel wires form a birdcage resonator Birdcage Coil Multiple parallel wires form a birdcage resonator B1 Head coil; body coil Siemens Pathway MRI™ 1.5T Head Coil Advantages: • Much more homogeneous field • Intrinsically high SNR • Can run in quadrature mode for √ 2  SNR • Can use as transmit-only coil w/surface coil Disadvantages: • Complicated design and construction

Phased array coils •Maintain SNR advantages of surface coils Siemens Body Array •Maintain SNR advantages of surface coils •Large sensitive region

1. Magnet 2. Gradients 3. Transmitter 4. Receiver 5. Probe Basic understanding needed to specify, purchase, and operate an MRI system

Beach time! Magnetic Resonance Imaging Haacke, Brown, Thompson, Venkatesan Wiley, 1999 Beach time! In vivo NMR Spectroscopy de Graaf Wiley 1998 Electromagnetic Analysis and Design in MRI Jin CRC 1999 Biomedical MR Technology Chen and Hoult Adam Hilger, 1989