Results from the PROVIDE Trial

Slides:



Advertisements
Similar presentations
The MADIT II Trial Multicenter Autonomic Defibrillator Implantation Trial II Presented at the American College of Cardiology 51st Annual Scientific Session.
Advertisements

Multicenter Automatic Defibrillator Implantation Trial II
EP Testing and Use of Devices in Heart Failure HFSA 2010 Recommendations.
Cardiovascular Disaster in Hemodialysis patients
Myocardial Ischemia: An Underrated Cause of Sudden Cardiac Death?
MADIT Randomized Trial to Reduce Inappropriate Therapy (MADIT-RIT) Arthur J. Moss, MD for the MADIT-RIT Executive Committee AHA Late Breaking Trials November.
La stratificazione del rischio aritmico oltre la frazione di eiezione Milano 17 Aprile 2009 Prof. Luigi Padeletti Heart Failure & Co.
The Clinical Evaluation of Remote Notification to Reduce Time to Clinical Decision (CONNECT) Trial The Value of Remote Monitoring George H. Crossley, MD.
Prevention de la Mort Subite Treatment of Ventricular Arrhythmias and the Prevention of Sudden Cardiac Death S. Nasr, M.D. Clinical Cardiac Electrophysiologist.
Implantable Cardioverter Defibrillators to Prevent Sudden Cardiac Death: Background Frederick A. Masoudi, MD, MSPH Associate Professor of Medicine (Cardiology)
Outcome of Patients with Advanced Heart Failure who Receive Device- Based Therapy for Primary Prevention of Sudden Cardiac Death, G. Amit, N. Samniah,
PACT Clinical Trial Commentary Dr Eric Topol Chairman and Professor, Department of Cardiology Director of the Joseph J Jacobs Center for Thrombosis and.
Arrhythmias: The Good, the Bad and the Ugly
ICD FOR PRIMARY PREVENTION EVIDENCE REVIEW
Sudden death as co-morbidity in patients following vascular intervention Sudden death as co-morbidity in patients following vascular intervention Impact.
Sex Differences in Implantable Cardioverter Defibrillator (ICD) Implantation indications and outcomes Guy Amit, MD; Mahmoud Suleiman, MD; Mark Kazatsker,
Prognostic Value of Programmed Electrical Stimulation Among Implantable Cardioverter-Defibrillator Recipients Real-World Data from the Israeli National.
Heartland Cardiology Dr. John Dongas The Beat Goes On: Biventricular Devices.
Indications of ICD in 2010 Dr Mervat Aboulmaaty Professor of Cardiology Ain Shams University DAF 1 st EP course 2010.
Sudden Cardiac Death in Heart Failure Trial Presented at American College of Cardiology Scientific Sessions 2004 Presented by Dr. Gust H. Bardy SCD-HeFTSCD-HeFT.
Cardiac Arrhythmias in Coronary Heart Disease SIGN 94.
Mr. J is a 70 year old man with an ischemic cardiomyopathy who presents with class III CHF and significant dissatisfaction with his functional capacity.
Heart Failure Ben Starnes MD FACC Interventional Cardiology
EP Show – Aug 2003 ICDs – Primary prevention The EP Show: Which ICD for which patient? Part 2: Primary prevention Eric Prystowsky MD Director, Clinical.
AICD usage for primary prevention at Mercy Hospital: successes, challenges and next steps Mohammad Tahir PGY-3.
Renal function and clinical outcomes of patients undergoing ICD and CRTD implantation- Data from the Israeli ICD Registry Alon Eisen, Mahmoud Souleiman,
Clinical Characteristics and Outcomes of Elderly Patients Treated With ICD and CRTD in Real World Setting: Data From the Israeli ICD Registry M. Suleiman,
Author Disclosures Differences in Implantation-Related Adverse Events Between Men and Women Receiving ICD Therapy for Primary Prevention Differences in.
Presenter Disclosure Information John F. Beshai, MD RethinQ Trial Results Disclosures Information: The following relationships exist related to this presentation:
EP Show – April 2002 MADIT II The EP Show: MADIT II Eric Prystowsky MD Director, Clinical Electrophysiology Laboratory St Vincent Hospital The Care Group.
Disclosures Dr Scirica has received Dr Scirica has received honoraria for educational presentations (modest) from CV Therapeutics The study was supported.
Devices and the older patient with syncope Michael Gammage, Reader in Cardiovascular Medicine MHRA Committee for Safety of Devices.
Device and Antiarrhythmic Drugs: Advantages and Pitfalls Teresa Menendez Hood, M.D.
INTRODUCTION: INTRODUCTION: implantable cardioverter-defibrillators (ICDs) have clearly demonstrated to terminate an elevated percentage of sustained ventricular.
EP show – June 2004 EP show The EP show: Risk stratification for sudden death Eric Prystowsky MD Director, Clinical Electrophysiology Laboratory St Vincent.
An ICD for every CRT patient ?
EP Show – Dec 2003 ICDs – Primary prevention The EP Show: Guidelines and reimbursement at the crossroads: Primary prevention with ICDs Eric Prystowsky.
The Case for Rate Control: In the Management of Atrial Fibrillation Charles W. Clogston, M.D. Cardiologist CHI St. Vincent Heart Clinic Arkansas April.
PainFree SST Primary Manuscript Results James Coles, PhD Medtronic, Inc.
Shock Reduction History James Coles, PhD Medtronic, Inc.
Increased Patient Device Concerns But Not General Anxiety in Patients with a Secondary Indication for the ICD Susanne S. Pedersen, Professor of Cardiac.
CoRPS Center of Research on Psychology in Somatic diseases Poor health status in implantable cardioverter defibrillator patients: Shock, patient pre implantation.
Real World Performance of Rhythm Discrimination Algorithms in Hard-to-Treat Single Chamber ICD Patients James Coles, PhD Medtronic, Inc.
Author Disclosure Sex Differences in the Characteristics of Patients Receiving ICD Therapy for the Primary Prevention of Sudden Cardiac Death –Stacie L.
Ventricular Arrhythmias:A General Cardiologist’s Assessment of Therapies in 2004 C.Richard Conti M.D. MACC.
1 Is It Time to Redefine ICD Therapy ? The Importance of Reducing Shocks.
Wearable Cardioverter Defibrillators
Everolimus-eluting Bioresorbable Vascular Scaffolds in Patients with Coronary Artery Disease: ABSORB III Trial 2-Year Results Stephen G. Ellis, MD,
Sudden Cardiac Arrest Morhaf Ibrahim, MD, FHRS Electrophysiology.
Clinical Trial Commentary
Implantable Defibrillator Therapy Post Cardiac Arrest
BLOCK HF Study Biventricular versus Right Ventricular Pacing in Patients with Left Ventricular Dysfunction and Atrioventricular Block – Preliminary Results.
These slides highlight a presentation at the Late Breaking Trial Session of the American College of Cardiology 52nd Annual Scientific Sessions in Chicago,
Updates From NOTION: The First All-Comer TAVR Trial
Defibrillator in Acute Myocardial Infarction Trial
American College of Cardiology Presented by Dr. Stuart J. Connolly
Reactive ATP and MINERVA
ATHENA Trial Presented at Heart Rhythm 2008 in San Francisco, USA
The following slides highlight a presentation at the Late-Breaking Clinical Trials session of the American Heart Association Scientific Sessions, November.
The most common cause of death in North America is cardiac death and the most common cause of cardiac death is sudden death from ventricular arrhythmias.
The EP show: sudden death, part 1 Director
ICD’s: Current Roles and Evidence
MASTER Trial Inclusion Criteria
Volume 9, Issue 1, Pages (January 2012)
Univ. of Copenhagen, Denmark
The Synergy between Percutaneous Coronary Intervention with TAXUS and Cardiac Surgery: The SYNTAX Study One Year Results of the PCI and CABG Registries.
Avoiding Cardiovascular Events through COMbination Therapy in Patients LIving with Systolic Hypertension The First Outcomes Trial of Initial Therapy With.
The Synergy between Percutaneous Coronary Intervention with TAXUS and Cardiac Surgery: The SYNTAX Study One Year Results of the PCI and CABG Registries.
Systolic Heart failure treatment with the If inhibitor ivabradine Trial Effect of ivabradine on recurrent hospitalization for worsening heart failure:
The Heart Rhythm Society Meeting Presented by Dr. Johan De Sutter
Presentation transcript:

Results from the PROVIDE Trial Programming Implantable Cardioverter Defibrillators to Prolong Time to First Shock Results from the PROVIDE Trial Mohammad Saeed, MD Robert Styperek, MD Leo Polosajian, MD Ahmed Khan, MD Ibrahim Hanna, MD Joseph Alonso, MD Dionyssios Robotis, MD Curtis Neason, BS Linda Saberi, BS

Outline Key Conclusions Actual slides presented at HRS 2012

Key Conclusions By utilizing a combination of specific programming parameters, shock therapy can be significantly reduced without increasing syncope, in patients with ICD implanted for primary prevention. Majority of benefit was due to avoidance of inappropriate Shocks Appropriate shock rate was low and not different between groups There was no statistical difference between arrhythmic syncope among groups

Results from the PROVIDE Trial Programming Implantable Cardioverter Defibrillators to Prolong Time to First Shock Results from the PROVIDE Trial Mohammad Saeed, MD Robert Styperek, MD Leo Polosajian, MD Ahmed Khan, MD Ibrahim Hanna, MD Joseph Alonso, MD Dionyssios Robotis, MD Curtis Neason, BS Linda Saberi, BS

Financial Disclosures Mohammad Saeed, MD Clinical Investigator: Medtronic, Boston Scientific, St. Jude. Consultant/ Honoraria: Medtronic, Boston Scientific, St. Jude. The PROVIDE Trial was sponsored in full by St. Jude Medical.

Background SCD-HeFT1 and MADIT II2 have expanded the patient population eligible for ICDs to include primary prophylaxis of VT/VF. Nearly half of all shocks experienced by this patient population are inappropriate.3 Inappropriate shocks and other avoidable shocks can lead to ICD battery depletion, a decrease in the patient’s quality of life4,5,6 and may be detrimental to cardiac function. 1 Bardy GH et al. N Engl J Med, Jan 2005. 2 Moss AJ et al. N Engl J Med, 2002. 3 Wilkoff B et al. JACC, Aug 2008. 4 Schron EB et al. Circ, 2002. 5 Irvine JD et al. Am Heart J, Aug 2002. 6 Carroll DL et al. J Acute Crit Care, May 2005.

Background PainFREE Rx I1 & II2 showed that ATP is effective at terminating FVT in a broad ICD population. The PROVE trial3 demonstrated that ATP is highly effective at terminating VT safely in primary prevention population. EMPIRIC4 prescribed programming for all ICD recipients utilizing PF II programming and SVT discriminators. The PREPARE5 trial suggested that prolong detection interval (NID 30/40) in addition to ATP might reduce shocks in the primary prevention population. 1 Wathen MS et al. Circ, Aug 2001. 2 Wathen MS et al. Circ, Oct 2004. 3 Saeed M et al. JCE, Dec 2010. 4 Wilkoff BL et al. JACC, Jul 2006. 5 Wilkoff BL et al. JACC, Aug 2008.

Background However, PREPARE was non-randomized, used a historical control, and utilized programming changes limited only to the prolongation of detection intervals.1 Furthermore, programming an increased number of detection intervals in the VF detection zone may have contributed to the higher frequency of syncope in this study.1 Follow-up for PREPARE was 12 months. 1 1 Wilkoff BL, et al. JACC, Aug 2008.

Purpose To study the efficacy and safety of a strategic combination of programming parameters in order to reduce shocks: Higher detection rates Prolonged detection intervals Aggressive application of SVT discriminators Use of ATP therapy for FVT Patients receiving ICD for primary prevention of sudden cardiac death.

Methods Prospective, multi-center, randomized trial. Patients receiving a de novo ICD or CRT-D manufactured by St. Jude Medical for primary prevention indication. Visit schedule: Enrollment <30 days post implant and every 3 months until study conclusion. Patients with an inducible sustained VT <181 bpm during an electrophysiology test were excluded. All episodes were adjudicated by a panel of EPs to determine rhythm and success of therapy.

Methods Primary endpoint: Safety endpoint: Mean time to first shock Rate of arrhythmic syncope

Device Programming Control [MONITOR] 2x ATP Shocks (12 beats) Nominal SVTd Experiment 2x ATP Shocks (25 beats) 1x ATP (18 beats) (12 beats) Optimized SVTd 150 bpm 181 bpm 214 bpm 250 bpm

Device Programming Rationale for programming the Control Group Utilized zones and therapies from PROVE Trial One VT and VF zone along with monitor zone Very successful at terminating VT with ATP SVT discriminators were nominal settings except: All SVTd were turned ON SVTd Timeout was turned OFF [MONITOR] 2x ATP Shocks (12 beats) Nominal SVTd Control 150 bpm 181 bpm 214 bpm

Device Programming Rationale for programming the Experiment Group Zones and therapies based on PREPARE except: Detection intervals were tiered to minimize syncope Applied SVTd up to 214bpm (instead of 200 bpm) Attempted a second round of ATP between 182-214bpm SVT discriminators were based on literature Theuns et al.1 for DR and CRT-D settings Boriani et al.2 for SR settings 2x ATP Shocks (25 beats) 1x ATP (18 beats) (12 beats) Optimized SVTd Experiment 181 bpm 214 bpm 250 bpm 1 Theuns D et al. Heart Rhythm, Nov 2006. 2 Boriani G et al. PACE, Sep 2002.

SVT Discriminators DR/CRT-D Control Experiment SVT Discrimination Timeout OFF V<A, V=A If Any If All MD % Match 60% Stability Delta 80 ms 40 ms Onset Delta 100 ms 16% SR Control Experiment SVT Discrimination Timeout OFF VT Diagnosis If Any If 2 of 3 MD % Match 60% Interval Stability ON ON w/ SIH Stability Delta 80 ms 40 ms Onset Delta 100 ms 16%

Results All Patients (N=1670) Control (N=824) Experiment (N=846) P-value Follow-up Duration (days) , Mean, Range 548, 1-1019 531, 2-1005 565, 1-1019 0.002 Age (years), Mean ± SD 64 ± 13 64 ± 12 ns Gender, N 1219 (73%) 599 (73%) 620 (73%) NYHA Class I 79 (5%) 33 (4%) 46 (5%) II 520 (31%) 264 (32%) 256 (30%) III 782 (47%) 381 (46%) 401 (48%) IV 26 (2%) 8 (1%) 18 (2%) UNK 263 (16%) 138 (17%) 125 (15%) LVEF (%), Mean ± SD 27 ± 9 27 ± 10 QRS Dur (ms), Mean ± SD 124 ± 32 123 ± 31 124 ± 33 Cardiomyopathy Ischemic 1040 (62%) 510 (62%) 530 (63%) Non-Ischemic 630 (38%) 314 (38%) 316 (37%)

Results All Patients (N=1670) Control (N=824) Experiment (N=846) P-value Cardiac History Prior MI 817 (49%) 412 (50%) 412 (48%) ns Unstable Angina 180 (11%) 76 (9%) 104 (12%) 0.05 CABG 519 (31%) 258 (31%) 261 (31%) PTCA/Stents/Atherectomy 639 (38%) 313 (38%) 326 (39%) Other Medical Conditions Diabetes 600 (36%) 300 (36%) COPD 220 (13%) 115 (14%) 105 (12%) Hyperlipidemia 1005 (60%) 497 (60%) 508 (60%) Hypertension 1213 (73%) 607 (74%) 606 (72%) Peripheral Vascular Disease 150 (9%) 80 (10%) 70 (8%) Renal Disease 191 (11%) 105 (13%) 86 (10%) Valvular Disease 227 (14%) 112 (13%)

Results All Patients (N=1670) Control (N=824) Experiment (N=846) P-value Cardiac Medications, n (%) ACE Inhibitor 1015 (61%) 504 (61%) 511 (60%) ns Antiarrhythmic, Class I 19 (1%) 10 (1%) 9 (1%) Antiarrhythmic, Class III 155 (9%) 73 (9%) 82 (10%) ARB 305 (18%) 148 (18%) 157 (19%) Beta Blocker 1482 (89%) 734 (89%) 748 (88%) Calcium Channel Blocker 154 (9%) 76 (9%) 78 (9%) Diuretic 1092 (65%) 566 (69%) 526 (62%) 0.01 Nitrate 315 (19%) 159 (19%) 156 (18%) Aldosterone Inhibitor 295 (18%) 147 (17%) Anti-coagulant 491 (29%) 227 (28%) 264 (31%) Anti-platelet 1125 (67%) 564 (68%) 561 (66%) Arrhythmia History Atrial Fibrillation 443 (27%) 210 (25%) 233 (28%) Atrial Flutter 56 (3%) 27 (3%) 29 (3%) Sinus Tachycardia 54 (3%) 25 (3%) Atrial Tachycardia 13 (0.8%) 5 (0.6%) 8 (0.9%)

Device Type 22% 41% 37% p = ns 22% 40% 38%

Results: Primary Endpoint Time to First Shock : All-Cause 25 Control Experiment 18.7 % Patients 20 p = 0.0005 15 12.3 10 10.3 5 7.0 0.5 1.0 1.5 2.0 Years N at Risk Control 824 671 542 313 141 Experiment 846 729 599 392 190

Results: Primary Endpoint Time to First Shock - Appropriate Control Experiment 25 % Patients 20 15 p = 0.69 10 4.9 3.0 5 4.9 2.8 0.5 1.0 1.5 2.0 Years N at Risk Control 824 716 592 347 160 Experiment 846 749 625 406 196

Results: Primary Endpoint Time to First Shock - Inappropriate Control Experiment 25 % Patients 20 13.8 15 p < 0.0001 9.3 10 5 5.4 4.1 0.5 1.0 1.5 2.0 Years N at Risk Control 824 683 559 331 153 Experiment 846 743 616 411 199

Results: Safety Endpoint Time to First Arrhythmic Syncope 10 Control Experiment % Patients Number of Arrhythmic Syncopal Episodes Control 8 Experimental 14 5 p = 0.27 0.5 1.0 1.5 2.0 Years N at Risk Control 824 726 607 360 169 Experiment 846 759 635 415 201

Conclusions By utilizing a combination of specific programming parameters, shock therapy can be significantly reduced without increasing syncope, in patients with ICD implanted for primary prevention. Majority of benefit was due to avoidance of inappropriate Shocks Appropriate shock rate was low and not different between groups There was no statistical difference between arrhythmic syncope among groups

Recommendations The following programming parameters should be considered when implanting ICDs in primary prevention patients: Monitor zone OFF Lower rate cutoff for VT zone at 180 bpm Lower rate cutoff for VF zone at 250 bpm Prolonging the detection intervals SVTd ON with sustained rate timer OFF Empiric ATP for FVT up to 250bpm