A Brief Introduction to Tomographic Imaging

Slides:



Advertisements
Similar presentations
Computers and Computed Tomography
Advertisements

PHYSICS OF MAGNETIC RESONANCE
Nuclei With Spin Align in Magnetic Fields HoHo anti-parallel parallel Alignment Energy  E = h  H o Efficiency factor- nucleus ConstantsStrength of magnet.
MR TRACKING METHODS Dr. Dan Gamliel, Dept. of Medical Physics,
Foundations of Medical Ultrasonic Imaging
Gamma-Ray Spectra _ + The photomultiplier records the (UV) light emitted during electronic recombination in the scintillator. Therefore, the spectrum collected.
Chapter 8 Planar Scintigaraphy
Equations. More Organized Proof of The Central Slice Theorem.
MRI “Magnetic Resonance Imaging”. Nuclei with nuclear spin: elementary magnets Magnetic moment:  =magnetogyric ratio L=angular momentum.
Buxton & District Science Discussion Medical Scanners Marge Rose 16 th November 2012.
ECE 501 Introduction to BME
BME 560 Medical Imaging: X-ray, CT, and Nuclear Methods
Types of Radiation Interactions All or Nothing Many Small There is a finite probability per unit length that the radiation is absorbed. If not, there is.
Psy 8960, Fall ‘06 Introduction to MRI1 Introduction to MRI: NMR MRI - big picture –Neuroimaging alternatives –Goal: understanding neurall coding Electromagnetic.
lecture 5, Sampling and the Nyquist Condition Sampling Outline  FT of comb function  Sampling Nyquist Condition  sinc interpolation Truncation.
Computed Tomography. Introduced in 70’s Principle: Internal structures of an object can be reconstructed from multiple projections of the object.
Ultrasound Imaging Atam Dhawan.
Interactions with Matter
Introduction to Medical Imaging
Magnetic Resonance Imaging Basic principles of MRI This lecture was taken from “Simply Physics” Click here to link to this site.
LESSON 4 METO 621. The extinction law Consider a small element of an absorbing medium, ds, within the total medium s.
Planar scintigraphy produces two-dimensional images of three dimensional objects. It is handicapped by the superposition of active and nonactive layers.
Radiofrequency Pulse Shapes and Functions
lecture 2, linear imaging systems Linear Imaging Systems Example: The Pinhole camera Outline  General goals, definitions  Linear Imaging Systems.
Ultrasound Physics Sound is a mechanical, longitudinal wave that travels in a straight line Sound requires a medium through which to travel Ultrasound.
Medical Image Analysis Introduction Figures come from the textbook: Medical Image Analysis, Second Edition, by Atam P. Dhawan, IEEE Press, 2011.
Advanced Biomedical Imaging Dr. Azza Helal A. Prof. of Medical Physics Faculty of Medicine Alexandria University.
5.4.1 X-Rays. (a) describe the nature of X-rays Stowmarket Physics X-rays - nature Forms of electromagnetic radiation Short wavelength High frequency.
3/13/2009IB Physics HL 21 Ultrasound Medical Imaging Physics – IB Objectives I.2.7Describe the principles of the generation and the detection of ultrasound.
Doppler echocardiography & Magnetic Resonance Imaging.
X-Ray Medical Imaging Physics – IB Objectives
Introduction to Tomographic Imaging A Brief Introduction to Tomographic Imaging David G. Cory, NW Outline  General.
A-LEVEL PHYSICS 15/09/ :49 Medical Physics.
Interaction ionizing radiation with biological tissue. Basic dosimetry.
© Jimoid.com 2005 Imaging Basics A medical image is a 2D or 3D distribution of signals which represent properties of an object. The purpose of medical.
Medical Image Analysis Interaction of Electromagnetic Radiation with Matter in Medical Imaging Figures come from the textbook: Medical Image Analysis,
Interactions of EM Radiation with Matter
Properties of ElectroMagnetic Radiation (Light)
The Structure and Dynamics of Solids
MRI Physics: Spatial Encoding Anna Beaumont FRCR Part I Physics.
Magnetic Resonance Imaging Glenn Pierce, King’s College London, Department of Physics Introduction Edward Purcell and Felix Bloch were both awarded the.
Interaction of x-ray photons (and gamma ray photons) with matter.
Lecture 1: Magnetic Resonance
Medical Physics.
Interaction of Radiation with Matter
Presentation on.  There are many methods for measuring of fiber structure. Such as:  The absorption of infrared radiation  Raman scattering of light.
Magnetic Resonance Imaging
Medical physics AMMAR ALHASAN University of Central Florida
2.4.2 interaction of x-rays with matter
CLARIDGE CHAPTER 2.2; 3.2.
Introduction to medical imaging
Image quality and Performance Characteristics
Interaction of gamma rays with matter
MRI Physics in a Nutshell Christian Schwarzbauer
Department of radiology divided into
Application of Nuclear Physics
Medical Imaging Imagining Modalities.
Spatial Encoding: Sub mm from meter sized RF
Review of Ultrasonic Imaging
MRI Pulse Sequences: IR, EPI, PC, 2D and 3D
Interaction of gamma rays with matter
X Rays Medical Physics Notes.
Medical Physics Physics 421 Course Description:
Radiation Health Physics
Basic principles Geometry and historical development
Resident Physics Lectures
10.6 Fourier Transform Mass Spectrometry
Light.
الفيزياء الحيوية الطبية Medical Biophysics
(4)ELECTRONIC SUPPORT SYSTEM
Presentation transcript:

A Brief Introduction to Tomographic Imaging David G. Cory, NW14-2217 Dcory@mit.edu 253-3806 Outline  General Goals  Linear Imaging Systems  An Example, The Pin Hole Camera  Radiations and Their Interactions with Matter  Coherent vs. Incoherent Imaging  Length Scales  Contrasts  Photon Intensity Tomography  Magnetic Resonance Imaging 22.56 - Introduction to Tomographic Imaging

Imaging Definitions Object function - the real space description of the actual object. Resolution - the collected image is only an approximation of the actual object. The resolution describes how accurate the spatial mapping is. Distortions - describes any important non- linearities in the image. If there are no distortions, then the resolution is the same everywhere. Fuzziness - describes how well we have described the object we wish to image. Contrast - describes how clearly we can differentiate various parts of the object in the image. Signal to Noise ratio 22.56 - Introduction to Tomographic Imaging

Linear Imaging Systems If the blurring of the object function that is introduced by the imaging processes is spatially uniform, then the image may be described as a linear mapping of the object function. This mapping is, of course, at lower resolution; and the blurring is readily described as a convolution of the object function with a Point Spread Function. Image = object  Point Spread Function + noise The noise is an important consideration since it limits the usefulness of deconvolution procedures aimed at reversing the blurring effects of the image measurement. 22.56 - Introduction to Tomographic Imaging

An Example, the Pin-hole Camera One of the most familiar imaging devices is a pin-hole camera. The object is magnified and inverted. Magnification = -b/a. 22.56 - Introduction to Tomographic Imaging

An Example, the Pin-hole Camera 2 Notice, however, that the object function is also blurred due to the finite width of the pin-hole. The extent of blurring is to multiply each element of the source by the “source magnification factor” of (a+b)/a x diameter of the pin-hole. 22.56 - Introduction to Tomographic Imaging

Distortions of a Pin-hole Camera Even as simple a device as the pin-hole camera has distortions 1. Limited field of view due to the finite thickness of the screen. As the object becomes too large, the ray approaches the pin-hole too steeply to make it through. 22.56 - Introduction to Tomographic Imaging

Distortions of a Pin-hole Camera 2 Also, as the object moves off the center line, the shadow on the detector grows in area, (and the solid angle is decreased) so the image intensity is reduced. 22.56 - Introduction to Tomographic Imaging

22.56 - Introduction to Tomographic Imaging

Spatial Frequencies 22.56 - Introduction to Tomographic Imaging

Spatial Frequencies 2 When discussing linear imaging systems it is often useful to describe the measurement in terms of a mapping of Fourier components of the object function. 22.56 - Introduction to Tomographic Imaging

Transmission Tomography f y r x s object source detector Absorption imaging In absorption imaging, the integrated absorption along a column through the object is measured. An array of detectors therefore measures a ‘shadow profile’. 22.56 - Introduction to Tomographic Imaging

Projection Imaging Object Projections 22.56 - Introduction to Tomographic Imaging

Central Slice Theorem Consider a 2-dimensional example of an emission imaging system. O(x,y) is the object function, describing the source distribution. The projection data, is the line integral along the projection direction. The central slice theorem can be seen as a consequence of the separability of a 2-D Fourier Transform. The 1-D Transform of the projection is, The one-dimensional Fourier transformation of a projection obtained at an angle J, is the same as the radial slice taken through the two-dimensional Fourier domain of the object at the same angle. 22.56 - Introduction to Tomographic Imaging

Transmission Tomography f y r x s object source detector Absorption imaging In absorption imaging, the integrated absorption along a column through the object is measured. An array of detectors therefore measures a ‘shadow profile’. 22.56 - Introduction to Tomographic Imaging

Coherent vs. Incoherent Imaging In both cases the image is the result of the scattering of a field by the object. Incoherent - measure only the intensity fluctuations of this scatter. Usually frequencies are too high to permit convenient measures of the phase. Examples, light 1014 Hz X-rays 1018 Hz g-rays 1020 Hz A photograph is an incoherent image. Coherent - measure both the intensity and the phase of the scattered field. This is usually measured as a temporal evolution of the scattered field. The frequency of radiation is normally quite low to permit an accurate measure of the phase (such as microwaves). MRI is an example of coherent imaging. ---------------------------------------------------------------------------------------------------- Incoherent images are most readily measured by scanning a well collimated beam across the sample and observing the attenuation of that beam (this may be multiplexed with many detectors). Coherent images permit the characterization of the entire sample at once and with observation through a single detector element. A series of measurements are made for fields of varying frequency or direction. 22.56 - Introduction to Tomographic Imaging

Tissue Transparancy Ultrasound X-ray Radio-frequency damaging harmless 1mm 100mm 1cm 1m 100m Ultrasound X-ray Radio-frequency  1A 100A damaging harmless C-H bond energy Windows of transparency in imaging via sound and electromagnetic radiation. The vertical scale measures absorption in tissue. 22.56 - Introduction to Tomographic Imaging

X-ray Attenuation Coefficients 500 300 200 150 100 50 40 30 20 10 0.1 0.15 0.2 0.3 0.4 0.5 1.0 2 2.5 5 FAT MUSCLE BONE PHOTON ENERGY (kev) m/r (cm2/g) X-ray attenuation coefficients for muscle, fat, and bone, as a function of photon energy. 22.56 - Introduction to Tomographic Imaging

l1 < l2 Valence electrons l4 l3 l2 l1 Binding energy (KeV) Compton  0 0.6 5 34 Binding energy (KeV) 100 KeV incident photon l1 Photoelectron 66 KeV K L M N Characteristic X-ray 0.6 KeV (NM) 4.4 KeV (ML) 29 KeV (LK) 34 KeV Total Valence electrons Compton Electron (Ee-) q Angle of deflection Scattered Photon (Esc) l1 < l2 Incident (E0) 22.56 - Introduction to Tomographic Imaging

Attenuation Mechanisms m dependence Mechanism E Z Energy Range in Soft Tissue simple scatter photoelectric Compton pair production rises slowly with E falls slowly with E 1/E3 1/E Z2 Z3 independent above 20 MeV 30 keV-20 MeV 1-30 keV 1-20 keV 22.56 - Introduction to Tomographic Imaging

Attenuation Mechanisms 2 (log plot) total Compton Compton photoelectric simple scatter pair .01 .05 0.1 1 10 Photon energy (MeV) (log plot) .03 1.02 30 Attenuation mechanisms in water The optimum photon energy is about 30 keV (tube voltage 80-100 kV) where the photoelectric effect dominates. The Z3 dependence leads to good contrast: Zfat 5.9 Zmuscles 7.4 Zbone 13.9  Photoelectric attenuation from bone is about 11x that due to soft tissue, which is dominated by Compton scattering. 22.56 - Introduction to Tomographic Imaging

Photon Intensity Tomography X-ray CT SPECT PET measuring X-ray attenuation source distribution of radio-pharmaceuticals, gamma emitters source distribution of radio-pharmaceuticals, positron emitters anatomical information Yes No No beam definition collimators collimators coincidence detection 22.56 - Introduction to Tomographic Imaging

Photon Intensity Tomography 1 f y r x s object source detector Absorption imaging In absorption imaging, the integrated absorption along a column through the object is measured. An array of detectors therefore measures a ‘shadow profile’. 22.56 - Introduction to Tomographic Imaging

Photon Intensity Tomography 2 f y r x s object detector Emission imaging In emission imaging, the integrated emitter density is measured. 22.56 - Introduction to Tomographic Imaging

Reflection Imaging of Ultrasonic Waves Lateral position Grey level display amplitude modulations Time Ultrasound signal transducer Layers of tissue Ultrasonic beam Pulse of ultrasound Some questions of interest are: ultrasonic transmission, reflection, and scattering sample elasticity and interfaces sources/detectors safety of medical applications 22.56 - Introduction to Tomographic Imaging

Spatial and Temporal Limits in NMR MRI medical Mini-Imaging micro-imaging NMR-microscopy 1s 1ms 1nm 1A  1mm 1cm 1m Time Space absolute spatial measurements 22.56 - Introduction to Tomographic Imaging

The Bloch Equations w1 is the strength of an applied external resonant radio-frequency field. Dw is the precession frequency; it includes contributions from  variations in magnetic field strength (inhomogeneities),  applied magnetic field gradients,  chemical shifts (screening of the nucleus by surrounding electrons),  and coupling of spins to each other (the dynamics are more complicated than indicated by the Bloch equations however). 22.56 - Introduction to Tomographic Imaging

Spin Magnetization Gratings Grating - “a system of equidistant and parallel lines… to produce spectra by diffraction”. Spin magnetization grating - a periodic modulation of the phase (or amplitude) of the local spin magnetization vector superimposed on the spin density Spin Magnetization gratings may be created by spin evolution in a linearly increasing magnetic field. RF G k = 0 k u Mx My This produces a grating as a linear phase ramp, since motions are torques. Moire complex gratings are produced through a combination of RF and gradient pulses. The spatial frequency distribution of these are describable by a distribution of components, each at a given wave-number. 22.56 - Introduction to Tomographic Imaging

magnetic field strength Mz(t) My(t) time Mx(t) spatial offset precession frequency magnetic field strength spatial offset 2t t t = 0 precession angle 22.56 - Introduction to Tomographic Imaging

Mz(t) My(t) time Mx(t) Diagrams of the spin magnetization’s return to equilibrium after being aligned along the x-axis. In both pictures the evolution of a single bulk magnetization vector is being followed. The initial position is shown as the green vector at top, which spirals into the z-axis, the red vector. In the figure on the right, the three individual components of the magnetization are shown as a function of time. The NMR experiment measures the two transverse components, Mx and My. There are three motions, a precession about the z-axis, a decay of the transverse components and a slower growth along z towards the static equilibrium value. 22.56 - Introduction to Tomographic Imaging

22.56 - Introduction to Tomographic Imaging

22.56 - Introduction to Tomographic Imaging

G = 1 G/cm D = 310-5 cm2/s attenuation T2 = 0.1 s k attenuation 1.0 attenuation 0.8 0.6 0.4 0.2 k 500 1000 1500 2000 2500 attenuation attenuation 22.56 - Introduction to Tomographic Imaging