Bioengineering 280A Principles of Biomedical Imaging Fall Quarter 2005 MRI Lecture 5 Thomas Liu, BE280A, UCSD, Fall 2005.

Slides:



Advertisements
Similar presentations
Pulse Timing parameters & Weighting
Advertisements

Fund BioImag : Echo formation and spatial encoding 1.What makes the magnetic resonance signal spatially dependent ? 2.How is the position of.
MRI Physics 2: Contrasts and Protocols
Principles of the MRI Signal Contrast Mechanisms MR Image Formation John VanMeter, Ph.D. Center for Functional and Molecular Imaging Georgetown University.
Contrast T1 weighted – (MPRAGE-anatomical) T2 weighted – (fmri)
Magnetic Resonance Imaging
Imaging Sequences part I
Chapter 7: Gradient Echo Imaging Methods
Introduction to MRI Magnetic Resonance Imaging
PULSE SEQUENCES Emphasizing the differences among spin density, T1, and T2 relaxation time constants of the tissues is the key to the exquisite contrast.
Proton Spin In absence of a magnetic field, protons spin at random
MRI Phillip W Patton, Ph.D..
Richard Wise FMRI Director +44(0)
M R I Pulse Sequences Jerry Allison Ph.D..
BE 581 Lecture 3- Intro to MRI.
PHYSICS OF MAGNETIC RESONANCE
Institute for Biomedical Engineering EXCITE Afternoon Hands-On MRI Sessions: fMRI & DTI.
RF Pulse – generates a B 1 field that realigns the precessing spins in the low energy state In the case of a 90 o pulse the alignment is perpendicular.
MR Sequences and Techniques
MR TRACKING METHODS Dr. Dan Gamliel, Dept. of Medical Physics,
Statistical Parametric Mapping
Fund BioImag : MRI contrast mechanisms 1.What is the mechanism of T 2 * weighted MRI ? BOLD fMRI 2.How are spin echoes generated ? 3.What are.
Parameters and Trade-offs
Principles of MRI. Some terms: –Nuclear Magnetic Resonance (NMR) quantum property of protons energy absorbed when precession frequency matches radio frequency.
Relaxation Exponential time constants T1 T2 T2*
FMRI: Biological Basis and Experiment Design Lecture 10: The Dreaded Drop-out Spin echo review Field maps Through-slice dephasing © Melissa Tillery
Basics of Magnetic Resonance Imaging
Steady-state magnetization
MRI Physics I: Spins, Excitation, Relaxation
Tissue Contrast intrinsic factors –relative quantity of protons tissue proton density –relaxation properties of tissues T1 & T2 relaxation secondary factors.
Magnetic Resonance Imaging 4
Physics of Magnetic Resonance Chapter 12
Gradient echo pulse sequences
Imaging Sequences part II
MRI Physics: Just the Basics
Seminar October, 2008 j. brnjas-kraljević. Imaging (MRI)  tomography technique  tomography technique – the volume image is built up by images of thin.
Medical Imaging Systems: MRI Image Formation
Pulse sequences.
Chapter 6 Lecture Spin Echo Imaging Methods Mark D. Herbst, MD, PhD Two Main Types of MR methods –Spin Echo –uses RF pulse to get an echo –Gradient Echo.
Basics of MRI.
Basic Concept of MRI Chun Yuan. Magnetic Moment Magnetic dipole and magnetic moment Nuclei with an odd number of protons or neutrons have a net magnetic.
Contrast Mechanism and Pulse Sequences Allen W. Song Brain Imaging and Analysis Center Duke University.
G Practical MRI 1 Basic pulse sequences.
Contrast Mechanism and Pulse Sequences
Magnetic Resonance Imaging – Basic Principles –
MRI Physics Dr Mohamed El Safwany, MD.. MRI Magnetic Resonance Imaging Magnetic Resonance Imaging.
Magnetic Resonance Imaging
FMRI – Week 4 – Contrast Scott Huettel, Duke University MR Contrast FMRI Graduate Course (NBIO 381, PSY 362) Dr. Scott Huettel, Course Director.
Protons (hydrogen nuclei act like little magnets) MRI Collective Magnetic Moment of Protons (M 0 ) Each pixel is a glass of protons B 0 = 3T (not to scale)
V.G.Wimalasena Principal School of Radiography
MRI: Contrast Mechanisms and Pulse Sequences
Spinning Nucleus Produces Magnetic Moment
Magnetic Resonance Learning Objectives
Principles of MRI Physics and Engineering Allen W. Song Brain Imaging and Analysis Center Duke University.
Charged particle. Moving charge = current Associated magnetic field - B.
Chapter 5 Mark D. Herbst, M.D., Ph.D.. The MR Imaging Process Two major functions –Acquisition of RF signals –Reconstruction of images.
Parameters which can be “optimized” Functional Contrast Image signal to noise Hemodynamic Specificity Image quality (warping, dropout) Speed Resolution.
10 spring fMRI: theory & practice
FMRI data acquisition.
MRI Physics in a Nutshell Christian Schwarzbauer
بسم الله الرحمن الرحيم.
Where Mt is the magnetization at time = t, the time after the 90o pulse, Mmax is the maximum magnetization at full recovery. At a time = one T1, the signal.
Magnetic Resonance Imaging: Physical Principles
Physics of Magnetic Resonance Imaging
MRI Pulse Sequences: IR, EPI, PC, 2D and 3D
Magnetic Resonance Imaging
Introduction to MR Robert Frost
(4)ELECTRONIC SUPPORT SYSTEM
The echo time (TE) The echo time (TE) refers to the time between the application of the radiofrequency excitation pulse and the peak of the signal induced.
Introduction to MR Robert Frost
Presentation transcript:

Bioengineering 280A Principles of Biomedical Imaging Fall Quarter 2005 MRI Lecture 5 Thomas Liu, BE280A, UCSD, Fall 2005

Slice Selection RF Slice select gradient Gz(t) Slice refocusing gradient Gx(t) Gy(t) Thomas Liu, BE280A, UCSD, Fall 2005

Gradient Echo RF Slice select gradient Gz(t) Slice refocusing gradient Gx(t) Gy(t) Spins all in phase at kx=0 ADC Thomas Liu, BE280A, UCSD, Fall 2005

Static Inhomogeneities In the ideal situation, the static magnetic field is totally uniform and the reconstructed object is determined solely by the applied gradient fields. In reality, the magnet is not perfect and will not be totally uniform. Part of this can be addressed by additional coils called “shim” coils, and the process of making the field more uniform is called “shimming”. In the old days this was done manually, but modern magnets can do this automatically. In addition to magnet imperfections, most biological samples are inhomogeneous and this will lead to inhomogeneity in the field. This is because, each tissue has different magnetic properties and will distort the field. Thomas Liu, BE280A, UCSD, Fall 2005

Static Inhomogeneities The spatial nonuniformity in the field can be modeled by adding an additional term to our signal equation. The effect of this nonuniformity is to cause the spins to dephase with time and thus for the signal to decrease more rapidly. To first order this can be modeled as an additional decay term of the form Thomas Liu, BE280A, UCSD, Fall 2005

T2* decay The overall decay has the form. where Due to random motions of spins. Not reversible. Due to static inhomogeneities. Reversible with a spin-echo sequence. Thomas Liu, BE280A, UCSD, Fall 2005

T2* decay Gradient echo sequences exhibit T2* decay. RF Slice select gradient Gz(t) Slice refocusing gradient Gx(t) Gradient echo has exp(-TE/T2*) weighting Gy(t) ADC TE = echo time Thomas Liu, BE280A, UCSD, Fall 2005

Spin Echo   Discovered by Erwin Hahn in 1950. 180º The spin-echo can refocus the dephasing of spins due to static inhomogeneities. However, there will still be T2 dephasing due to random motion of spins. There is nothing that nuclear spins will not do for you, as long as you treat them as human beings. Erwin Hahn Thomas Liu, BE280A, UCSD, Fall 2005 Image: Larry Frank

Spin Echo   Phase at time  Phase after 180 pulse Phase at time 2 180º Phase at time  Phase after 180 pulse Phase at time 2 Thomas Liu, BE280A, UCSD, Fall 2005 Image: Larry Frank

Spin Echo Pulse Sequence 90 180 RF   Gz(t) Gx(t) Gy(t) ADC TE = echo time Thomas Liu, BE280A, UCSD, Fall 2005

Image Contrast Different tissues exhibit different relaxation rates, T1, T2, and T2*. In addition different tissues can have different densities of protons. By adjusting the pulse sequence, we can create contrast between the tissues. The most basic way of creating contrast is adjusting the two sequence parameters: TE (echo time) and TR (repetition time). Thomas Liu, BE280A, UCSD, Fall 2005

Saturation Recovery Sequence 90 TE 90 TE 90 TR TR Gradient Echo 180 180 90 90 90 TE TR Spin Echo Thomas Liu, BE280A, UCSD, Fall 2005

T1-Weighted Scans Make TE very short compared to either T2 or T2*. The resultant image has both proton and T1 weighting. Thomas Liu, BE280A, UCSD, Fall 2005

T2-Weighted Scans Make TR very long compared to T1 and use a spin-echo pulse sequence. The resultant image has both proton and T2 weighting. Thomas Liu, BE280A, UCSD, Fall 2005

Proton Density Weighted Scans Make TR very long compared to T1 and use a very short TE. The resultant image is proton density weighted. Thomas Liu, BE280A, UCSD, Fall 2005

Example T1-weighted Density-weighted T2-weighted Thomas Liu, BE280A, UCSD, Fall 2005

FLASH sequence  TE  TE  TR TR Gradient Echo Signal intensity is maximized at the Ernst Angle FLASH equation assumes no coherence from shot to shot. In practice this is achieved with RF spoiling. Thomas Liu, BE280A, UCSD, Fall 2005

FLASH/SPGR Thomas Liu, BE280A, UCSD, Fall 2005 GE Medical Systems 2003

Inversion Recovery 180 180 180 180 90 90 TI TE TR Intensity is zero when inversion time is Thomas Liu, BE280A, UCSD, Fall 2005

Inversion Recovery Thomas Liu, BE280A, UCSD, Fall 2005 GE Medical Systems 2003

Fast Spin Echo Thomas Liu, BE280A, UCSD, Fall 2005 GE Medical Systems 2003

Echoplanar Imaging Thomas Liu, BE280A, UCSD, Fall 2005 GE Medical Systems 2003