Electron Probe Microanalysis EPMA

Slides:



Advertisements
Similar presentations
Electron-Specimen Interactions
Advertisements

Rutherford Backscattering Spectrometry
ACADs (08-006) Covered Keywords Efficiency, LLD, CPM, DPM, relative efficiency, absolute efficiency, standard deviation, confidence, count time. Description.
First of all, do you know any methods to check chemical composition? Or how you know what is what? First of all, do you know any methods to check chemical.
Spectrum Identification & Artifacts Peak Identification.
Electron-Specimen Interactions
ED and WD X-ray Analysis
X-Ray Microanalysis – Precision and Sensitivity Recall… K-ratio Si = [I SiKα (unknown ) / I SiKα (std.) ] x CF CF relates concentration in std to pure.
Rutherford Backscattering Spectrometry
Qualitative, quantitative analysis and “standardless” analysis NON DESTRUCTIVE CHEMICAL ANALYSIS Notes by: Dr Ivan Gržetić, professor University of Belgrade.
Thin Film Quantitation of Chemistry and Thickness Using EPMA John Donovan Micro Analytical Facility CAMCOR (Characterization of Advanced Materials in Oregon)
COMPUTER-AIDED DESIGN The functionality of SolidWorks Simulation depends on which software Simulation product is used. The functionality of different producs.
Do it with electrons !. Microscopy Structure determines properties We have discussed crystal structure (x-ray diffraction) But consider now different.
Electron probe microanalysis - Scanning Electron Microscopy EPMA - SEM
Chapter 12 Atomic X-Ray Spectroscopy
EDS Energy Dispersive Spectroscopy
Quantitative Analysis Quantitative analysis using the electron microprobe involves measuring the intensities of X-ray lines generated from your unknown.
Adventures of an Igneous Petrologist in a Probe Lab John Fournelle University of Wisconsin Madison, Wisconsin.
Sphere Standards and Standard Spheres Dr. Richard Young Optronic Laboratories, Inc.
Version: November 16, 2009 EPMA of Particles Electron probe microanalysis EPMA.
Energy-Dispersive X-ray Microanalysis in the TEM Anthony J. Garratt-Reed Neil Rowlands.
The physics of electron backscatter diffraction Maarten Vos AMPL, RSPHYSSE, Australian National University, Acton 0200, Canberra Aimo Winkelmann Max Planck.
IRAD 2371 Week 3.  Very few detectors will count every interaction  Each detector will have its own counting efficiency  Eff=CPM/DPM  Can use efficiency.
COSIRES 2004 © Matej Mayer Bayesian Reconstruction of Surface Roughness and Depth Profiles M. Mayer 1, R. Fischer 1, S. Lindig 1, U. von Toussaint 1, R.
X-Ray Microanalysis – Precision and Sensitivity Recall… wt.fraction I = I SiKα (unknown) / I SiKα (pure std.) K-ratio I = [I SiKα (unknown ) / I SiKα (std.)
Electron probe microanalysis Low Voltage SEM Operation Modified 9/23/10.
Basics of Ion Beam Analysis
The probe Some material used from:. EPMA - electron probe microanalysis Probe signals.
The Significance of Refractive Index “K” Values LSMs, or, Why Long Wavelength Peak Markers Don’t Line Up Correctly John Fournelle* and K. Bolger*, J. Cook*,
Do it with electrons !. Microscopy Structure determines properties We have discussed crystal structure (x-ray diffraction) But consider now different.
Spatial Resolution and minimum detection
(Chapter 7) CHEM–E5225 Electron microscopy
Mohammed Zeeshan BT/PE/1601/ Microtexture: Electron Diffraction in the SEM Texture And Microstructure & Anisotropy.
Revision Tips – Key areas Final Countdown
DC Sputtering Disadvantage #1 Low secondary electron yield
Atoms, Molecules & Ions Isotopes and % Abundance
Lecture 4a Common Error in EPMA: Secondary Fluorescence from Outside Primary Excitation Volume John Fournelle, Ph.D. Department of Geoscience University.
Photons: Light Waves Behaving as Particles
Electron probe microanalysis EPMA
The Problem of Secondary Fluorescence in EPMA in the Application of the Ti-in-Zircon Geothermometer And the Utility of PENEPMA.
Electron probe microanalysis - Scanning Electron Microscopy EPMA - SEM
What is XPS? XPS (x-ray photoelectron spectroscopy) is also known as ESCA (electron spectroscopy for chemical analysis). XPS provides chemical information.
Ch. 2 Matter Earth Science.
A Metallurgical Comparison of Mack T-12 Rod Bearings. Lots U, V & W
SEM and EPMA Some Advanced Topics October 7, 2009.
Introduction to Atomic Spectroscopy
Electron probe microanalysis EPMA
Preface: What’s EPMA all about? How does Geology 777 work?
Electron probe microanalysis - Scanning Electron Microscopy EPMA - SEM
SEM and EPMA Some Advanced Topics Revised 2/17/2013.
Electron probe microanalysis EPMA
Electron Probe Microanalysis EPMA
Electronic Structure of Atoms
Electron probe microanalysis EPMA
Do it with electrons !.
Atomic Absorption Spectroscopy. Atomic absorption spectroscopy is based on the same principle as the flame test used in qualitative analysis.
The Material World What is Matter???.
GCSE Required Practical – Physics 1 – Specific Heat Capacity
Electron probe microanalysis EPMA
Electronics and Instrumentation Vidnyan Mahavidyalaya, Sangola.
AP Chem Turn in Shaving Cream Lab if you have not done so yet!
Nanocharacterization (II)
Electron probe microanalysis - Scanning Electron Microscopy EPMA - SEM
Scanning Electron Microscopy (SEM)
Electron probe microanalysis EPMA
SEM and EPMA Some Advanced Topics October 7, 2009.
Electron probe microanalysis
Lesson 4: Application to transport distributions
Computed Tomography (C.T)
Suppose we make a new compound containing of C,H, and N
Presentation transcript:

Electron Probe Microanalysis EPMA UW-Madison Geoscience 777 Electron Probe Microanalysis EPMA EDS Part 3: “Easy Errors” to make in Quant Analysis Table of Elemental Composition  Spectrum with X-ray Peaks Created 3/3/18

EDS  Spectrum  Wt% Elements EDS can easily give a perfect, 100 wt% total analysis But those numbers can easily be 100% wrong! The analyst must understand all of the assumptions which the software is making when it spits out a perfect-looking chemical composition.

Beware of these types of samples Rough surfaced bulk material Heterogenous materials (all geological materials; some/many manmade materials) Particles (Porous materials) (Thin films)

Critical Factors for Quant Analysis All samples must be polished perfectly flat/smooth Non-conductive materials must have conductive coating (?inside surfaces?) Samples must be perpendicular to the electron beam The sampled volume (=interaction volume) is homogeneous (=only ONE phase/composition)

Critical Factors for Quant Analysis Counts are acquired on BOTH unknowns and standards on the same instrument, under the same operating conditions, with the standards and samples all abiding by conditions 1-4 above: -- Same positions relative to beam & detector -- Same kV -- Same beam current, or measured directly Use a Faraday cup to measure beam current for all X-ray measurements

“Geometric effects” Proper quant EPMA assumes the ONLY difference between unknown and standard/s IS COMPOSITION NO other factors should influence the specimen’s measured X-rays, namely Specimen shape Specimen orientation relative to beam Specimen size

“Geometric effects for Bulk Specimens” These happen when the size and shape of the specimen change the interaction of electrons with material, changing the intensity of the X-rays produced within the specimen Change the path length PL of the emergent X-rays from the specimen, changing the extent of X-ray absorption so either more X-rays emerge (shorter PL) or less (longer PL)

“Geometric effects for Bulk Specimens” Rough surfaced materials

“Geometric effects for Bulk Specimens: X-ray production” When the sample is tilted, there are MORE backscattered elements, so less X-rays are possible to be generated, as these two DTSAII simulations show. Goldstein et al, 2018, SEMXRMA, p 382

“Geometric effects for Bulk Specimens: X-ray Absorption” Effect of non-flat surface normal to beam. Here shown as “surface roughness: but same principle for 3D samples with surfaces at various angles to the beam. Goldstein et al, 2018, SEMXRMA, p 383

“Geometric effects for Bulk Specimens: X-ray Absorption” There are two important points here: Generated X-ray intensity is reduced “on the way out” of the specimen by aborption, and the longer the path length, the greater the absorption. Not all X-rays are equally absorbed! Recall the Fe-Cr and Fe-Ni example. Same thing here: Si Ka of 1.740 keV will excite/ be absorbed by Al whose K edge is 1.559 keV—so Si is heavily reduced by aborption. Al Ka of 1.487 keV however cannot excite Si K edge of 1.838 keV so only gradually absorbed. Goldstein et al, 2018, SEMXRMA, p 383

“Geometric effects for Bulk Specimens: X-ray Absorption” Goldstein et al, 2018, SEMXRMA, p 389

“Geometric effects for Bulk Specimens: X-ray Absorption” Making the best of a bad situation: No good at all! A bit better, facing detector but high angle Best but compromised: flat, normal to beam, but possible secondary fluorescence off adjacent phase. Goldstein et al, 2018, SEMXRMA, p 394

EDS of particles There are lots of particles just crying out for attention: Particles of all sizes, shapes, dimensions: Solids In the Atmosphere In the Soil In fluids In manmade materials Nano-particles???

“Geometric effects for Particles” How do X-ray spectra from particles different from X-ray spectra of bulk samples of the same chemical composition? 3 different geometric effects:

Particles – 1 – Mass Effect Mass effect/error: electrons escape from sides of small particles if E0 is large enough, so quantitative analysis will be in error because different elements (with different binding energies) will be affected differently Goldstein et al, 1992, p. 479, 481

Particles – 1 – Mass Effect Goldstein et al, 2018,

Particles - 2 Absorption effect of non-flat upper surface: different path length from normal flat geometry and the “normal” way we do quantitative analysis is to use FLAT polished standards for calibration -- so we could have “too much” x-ray intensity for particles Goldstein et al, 1992, p. 479, 481

Particles - 3 Variable effect of geometry of trajectory between beam impact area on non-uniform surface and the location of the detector -- so we get different results from the same material, depending upon where we place the electron beam. Goldstein et al, 1992, p. 479, 481

EDS of particles … is easy…. Maybe too easy? --> And easy to make mistakes!

Results from 2006 777 student project on EDS of rough samples with VP-SEM The first column shows the actual chemical composition, followed by the average composition of 20 points on different grains, followed by the variation (=standard deviation) of those individual measurements

X-rays from the substrate may well end up in a particle spectrum.

Particles -4 This can be minimized by putting particles on a carbon tape, or on a thin support grid. Below are spectra of NIST K411 first only on carbon tape, then on a thin carbon film supported by a copper grid

Importance of beam placement-detector geometry Note the important differences mainly for the X-rays below 3.5 keV, here important Mg and Si. Also note the spurious Cu coming from either BSEs or X-rays off the side/edge of the K411 sphere.

Importance of beam placement-detector geometry Analytical errors (as relative to the reference true composition

“Overscanning” – A possible solution? Over the years, within the SEM-EDS community, it has been suggested that “overscanning” a rough, 3-dimensional particle, may “average” away the errors due to path length differences and the particular placement of the detector. However, the consensus of the authrs of Goldstein et al 2018 is that this is a error, at least in terms of expecting an accurate chemical analysis. It may be valuable for identifying the major elements present, but should not be taken to present an accurate chemical analysis.

Is the inside the same as the outside of the particle? Both natural and manmade particle (crystals; synthetic ceramic or metal spheres) may have one composition on the surface, but another inside! You must always be a little open to that possibility, and cross sectioning the materials may be an important sanity check (think gold-coated metal scams... wrong density...)

Goldstein et al 2018

UW- Madison Geology 777 The dangers of believing your own eyes... Or... Secondary Flourescence can bite you Here is a 3 phase sample in the Nb-Pd-Hf-Al system. The Pd3Hf and Pd2HfAl phases were thought to be free of Nb. Another lab reported 10 wt% Nb in what should have been Nb-free phases. John Perepezko asked me to investigate... Fournelle, Kim and Perepezko (2005)

It turned out that the other lab had used EDS ... at 30 kV. Why 30 kV? UW- Madison Geology 777 It turned out that the other lab had used EDS ... at 30 kV. Why 30 kV? Because the Nb La and Hf Ma lines they would have measured with interfered with by the adjacent Al Ka and Pd La lines... So they thought they’d solved the problem by cranking up the gun to 30 kV and going for the higher energy Nb and Hf Ka X-ray lines... Fournelle, Kim and Perepezko (2005)

And their totals were a perfect 100 wt% !!! UW- Madison Geology 777 And their totals were a perfect 100 wt% !!! Fournelle, Kim and Perepezko (2005)

The problem is... Secondary Fluorescence UW- Madison Geology 777 The problem is... Secondary Fluorescence I went and analyzed (WDS) the sample at 18 kV, using the Nb La line since it was easy to separate it from the Al Ka and the Pd La lines (which were too close in the EDS spectrum)…and found essentially zero Nb, with 100 wt% analytical totals. But not to stop there… Fournelle, Kim and Perepezko (2005)

UW- Madison Geology 777 I reproduced the phenomenon with WDS measurement of Nb Ka, using high (28 kV was the highest I dared go at that time) on the SX51 electron probe... And we see totals greater than ~100 wt% And the Nb being the main reason for the higher total…another reason for wanting always to see the analytical wt% total BEFORE any normalizing it done. Fournelle, Kim and Perepezko (2005)

Very briefly today Thin Films: a thin, deposited (sputter or vapor deposition) film on a thick substrate Thin films are commonly analyzed by EDS, with the operator assuming that they can simply “ignore” the substrate signal and “only look at the thin film elements and normalize them”... This can lead to significant errors, because the black box matrix correction DOESN’T KNOW IT IS A THIN FILM! So all the elements are “corrected” assuming ALL are in one homogeneous volume! We will examine this in an upcoming lab.

Very briefly today Porous (ceramic, insulator) materials This is perhaps the most difficult material to do EPMA on. Normalizing totals to 100 w% WITHOUT seeing the analytical totals hides the fundamental problems here: charging INSIDE the specimen, together with voids. Impossible? (but not if that is your job... There are always tricks the microanalyst can try...