ONLINE self-study Autoclave.

Slides:



Advertisements
Similar presentations
Part III: Safe Work Practices
Advertisements

Biological Hazards Routes of Entry
Biological Laboratory Safety. REHS Rutgers Environmental Health and Safety REHS supports Rutgers University by providing comprehensive and professional.
Module 6: Safety At the HIV Rapid Testing Site. Lab workersHealth workersCounselors 2 The Lab Quality System Process Control Quality Control & Specimen.
Autoclave Training Guide York University Department of Occupational Health and Safety Suite C, East Office Building, 4700 Keele St. Telephone :
DEPARTMENT OF HEALTH RESEARCH INSTITUTE FOR TROPICAL MEDICINE LABORATORY DECONTAMINATION and WASTE MANAGEMENT.
Annual Refresher Training. Preliminary Procedure in Glassware Washing Contaminated items. Return glassware that contains chemicals or contamination to.
CDC/NIH Vertebrate Animal Biosafety Level Criteria ► For all Animal Biosafety Levels (1 - 4)  IACUC Approval  Authorized access to facilities  Animal.
These guidelines are in place for the safety of the personnel and animals in the facility and the environment at large and have been reviewed and endorsed.
Prepared by; Miss Syazwani Mahmad Puzi School of Bioprocess Engineering UniMAP.
Bloodborne Pathogen Training Division of Facilities Construction and Management 2001.
Onsite Biomedical Waste Management Environmental Health & Safety Environmental Protection Program
Laboratory Biohazardous/Sharps Waste Management. Types of Biohazardous Waste Dry Solid- No pourable liquids! Dry Solid- No pourable liquids! –Contaminated.
Unit 1. Lab Safety video I TO PREVENT: __________________________________________ __________________________________________ exposure to __________________________________________.
Safe Handling of Infectious Waste Procedure Overview.
3.03 Understand support services
Infection Control AHS II Unit F. Standard Precautions Sometimes called “Universal” precautions Sometimes called “Universal” precautions Used to break.
PERSONAL PROTECTIVE EQUIPMENT
Autoclave Use and Safety
Micro Practical 3 PRINCIPLE AND WORKING OF AUTOCLAVE Dr. Shahzad Ali Assistant Professor Department of Wildlife and Ecology UVAS, Ravi Campus, Pattoki.
CSULB Department of Biological Sciences Common Use Equipment Training Autoclaves.
Autoclave Training Guide University of Windsor Chemical Control Centre Health and Safety B37 Essex Hall. Telephone : ext
Bloodborne Pathogens Occupational Safety and Health Course for Healthcare Professionals.
Lab Safety Procedures SCIENCE.
Autoclave. What is an Autoclave? Autoclave is a pressurized device designed to heat aqueous solutions above their boiling point at normal atmospheric.
8.02 Aseptic Techniques Implement aseptic technique to maintain equipment Images courtesy of google images.
Safety and Rules of the Lab
7. EMERGENCY RESPONSE RYERSON UNIVERSITY.
Heat Fire Working with or Heat or Fire When working with heat or fire, you will need to wear safety goggles. To move hot objects you will need to use.
 Dress Code Goggles – No contacts! Lab Coat, gloves Tie back long hair, loose sleeves, and necklaces. No sandals.
Presented by Steven P. Feltman Food Safety and Quality Specialist.
Bloodborne Pathogens Standard Precautions Unit 2.
Science Safety Rules! You must score 90 or above before you may participate in the laboratory science activities. If labs occur before your completion.
Safety and Rules of the Lab. Safety Symbols Know safety symbols They appear in your laboratory activities They will alert you to possible dangers They.
Clean UpEquipment/ Materials Safety Equipment Glass, Food, and Drink Accidents 100.
© 2006 Plano ISD, Plano, TX. Working with Heat Fire or.
Lab Specific Training June 14 th, Responsibilities completing all required safety training; reading the Lab Safety Plan and all lab standard operating.
SAFETY IN MEDICAL LAB.
Healthcare Workers Division of Risk Management State of Florida Loss Prevention Program.
May 7 th,  There’s a direct link to it from the Shimizu Lab Website or go to  All personnel are responsible for.
$1 Million $500,000 $250,000 $125,000 $64,000 $32,000 $16,000 $8,000 $4,000 $2,000 $1,000 $500 $300 $200 $100 Welcome.
Lab SAFETY. Laboratory Safety Rap! eo.php?video_id=
SAFETY IN THE LABORATORY 1-Don`t eat, smoke or drink inside the laboratory. 2-Don`t put any thing in your mouth such as pens, fingers…. 3-Don`t take.
CHAPTER 6 Microbiology-Related Procedures
Teachers are responsible for classroom safety. All of it.
CSULB Department of Biological Sciences Common Use Equipment Training
Chemical Sterilization
Laboratory safety rules Dalia Kamal Eldien Mohammed Practical NO (1)
Handling and Disposal of Infectious Wastes
Autoclaves Nadina Wiórkiewicz (2011), Small benchtop laboratory autoclave [photograph]. Retrieved from
SteriTite Instructions for use
Creating a Clean and Hygienic Foodservice Operation
Autoclave Training November 16, 2005 University of Kentucky Department of Biosafety.
Onsite Medical Waste Management and Bloodborne Pathogens
Autoclave Safety Prepared by Curt Speaker
3.03 Understand support services PP2
Laboratory equipment Lecture (5).
3.03 Understand support services
November 16, 2005 University of Kentucky Department of Biosafety
3.03 Understand support services
Infection Control Lesson 3: Surgical Asepsis
Lab Safety Unit 1 Lesson 2.
3.03 Understand support services
Bloodborne Pathogens Standard Precautions Unit 2.
3.03 Understand support services PP2
3.03 Understand support services
Sterile Technique.
Microbiology-Related Procedures
3.03 Understand support services
3.03 Understand support services
Presentation transcript:

ONLINE self-study Autoclave

Introduction North Carolina medical waste rules (15A NCAC 13 B .1200), require that "Regulated Medical Waste", defined as "blood and body fluids in individual containers greater than 20 ml, microbiological waste, and pathological waste,” be treated before disposal in order to render the waste nonhazardous. Autoclaving changes the biological character of the waste to reduce or eliminate its potential for causing disease. Additional autoclave information can be found at www.ehs.unc.edu under the Biological Safety link.

Proper Disposal of Biohazard Waste

Principles of Autoclave Operation Steam penetrates objects in the autoclave Condensation creates negative pressure and draws in additional steam Moist heat kills microorganisms via coagulation of proteins Two types of autoclaves Gravity Displacement Vacuum/Gravity Assisted

Gravity Displacement In a gravity displacement autoclave cold air escapes through the bottom of the chamber as steam displaces it from above. The valves should never be obstructed and the chamber must not be overfilled in order for this system to function efficiently.

A vacuum/gravity autoclave pulls pressure from the chamber. Vaccum/Gravity A vacuum/gravity autoclave pulls pressure from the chamber.

Types of Autoclaves Bulk Autoclaves Found throughout UNC-CH DLAM facilities. Each DLAM facility’s bulk autoclave may operate differently. Training on the use of each facilities bulk autoclave is to be conducted and documented by the facility supervisor.

Has an additional gauge Types of Autoclaves Tuttnauer Autoclaves Act as both a gravity displaced and vacuum assisted autoclave. Most autoclaves on campus perform both. Depends upon the cycle used to sterilize material. Some of these use “house” steam from the UNC Steam Plant. Others use “clean” steam (deionized water). Uses ‘house’ steam Has an additional gauge Uses deionized water/steam

Types of Autoclaves Other Manufacturers: NOTE: Even though autoclaves on UNC campus operate similarly in the context of mechanics, training is required in the use of each autoclave since not all autoclaves are programmed the same. Getinge Steris Castle

Preventative Maintenance Plug Screen/Drainer: In order to maintain the autoclave’s effectiveness: The plug screen or drainer should be removed with heat- resistant gloves, checked, and cleaned frequently to ensure that it is free of dirt, dust, or sediment which may collect and cause a clog. Look out for any sharps that may have become loose and caught in the plug screen/drainer. Plug screen/drainer location. Plug screen/drainer.

Preventative Maintenance Surfaces: The interior surfaces should be cleaned (according to manufacturer’s specifications) of any residues that collects over time. Interior of autoclave. Debris around drain inside autoclave.

Preventative Maintenance Surfaces: On a regular basis, gaskets, doors, shelves and walls should be visually inspected for residue buildup and wear. **NOTE If any problems are experienced with the autoclave contact Facilities Services (962-3456) immediately. Gasket Around Door Gasket Around Autoclave Jacket Melted plastic on a shelf from an un-autoclavable container.

Preventative Maintenance Air Filters: Some autoclaves have air filters in the exhaust located directly above to allow for the exhaust of steam when the autoclave is opened. Check the filter from time to time to ensure that it is not clogged. Contact the appropriate personnel to replace the filter and clean off any debris that has collected.

Pre-Autoclaving Procedures Before being autoclaved: Contaminated materials should remain in the laboratory. Biohazard waste containers should be removed for autoclaving when they are 2/3 full. Biohazard bags should be closed and placed within secondary containers (bins, trays). DO NOT USE REDDISH -ORANGE BIOHAZARDOUS BAGS AS THEY WILL MELT INCORRECT Un-autoclaved bio-hazardous bags left on the floor beside an autoclave. CORRECT Bio-hazardous waste stored in secondary containment

Pre-Autoclaving Procedures Indicator Tape (Heat Sensitive Tape): Indicator tape should be applied to the biohazard bag as it will aid in reducing the handling of the biohazard waste during removal. The indicator tape should be placed in an “X” pattern over the biohazard symbol. Indicator tape should change color after autoclaving (i.e. the word ‘Autoclaved’ appears on the tape after treatment.) Indicator tape is available through Fisher Scientific (www.fishersci.com) Indicator tape in an “X” pattern. Autoclaved (markings on tape) Not autoclaved (no markings on tape)

What can be autoclaved Surgical Instruments Glassware Plastic tubes and pipette tips Solutions and water Animal food and bedding Waste

Proper Autoclave Use Proper temperature and exposure time are critical in ensuring the reliability of the autoclaving process. These factors are dependent upon steam penetration to every part of the waste load. If all the air is not allowed to escape from the waste during the autoclave cycle, then steam will not replace the air. The autoclave user must be mindful to prevent the entrapment of air. (i.e. leaving the biowaste bag open inside the autoclave, removing the lid on sharps containers inside the autoclave.)

PPE for Autoclave Users Eye Protection Lab Coat, Buttoned Closed-toed Shoes Heat-resistant Gloves

Hazards Associated with Autoclaves High temperatures and presence of extremely hot water which is not vented creates potential for burns and scalding Inadequate decontamination allows for the potential of biological hazards and personnel and environmental contamination Large heavy doors and loading carriages also present an ergonomic and pinch hazard to employees.

Hazards Associated with Autoclaves How to prevent hazards associated with autoclave use: Read the owners manual as manufacturer recommendations vary. Make sure autoclave doors and gaskets are firmly locked into place before operating the autoclave. These interlocking mechanisms help to prevent a sudden release of high pressure steam. If the autoclave does not have interlocking mechanisms, take additional precautions to ensure the door is closed.

Hazards Associated with Autoclaves How to prevent hazards associated with autoclave use: Post a warning sign stating, “Hot Surfaces, Keep Away” on or next to the autoclave to remind people of the heat hazard as older autoclaves may have little or no heat shielding around its exposed sides. Never stack or store combustible near an autoclave. Never autoclave toxic chemicals including residual Trizol, bleach, etc., volatile, or radioactive material. Contact UNC EHS (962-5507) if you have Biohazardous waste that contain any of these materials.

Hazards Associated with Autoclaves How to prevent hazards associated with autoclave use : Remove solutions from the autoclave slowly and gently. Some solutions have the potential to boil over when moved or exposed to room temperature. Heat-resistant gloves, safety goggles and a faceshield, and a rubber apron must be worn when removing hot liquids from the autoclave. Clean up spills immediately. Report malfunctions or accidents to your supervisor immediately.

Autoclave: Performance Indicators How to know if autoclave is functioning correctly: Physical - Annual testing every two years by N.C. Dept. of Labor - Pressure, Temperature, Cycle times, recorded Chemical Heat sensitive autoclave tape - Not an indicator of successful sterilization, useful to keep track of autoclaved and unautoclaved items Biological - Tests ability of autoclave to sterilize effectively - Geoacillus stearothermophilus spore strips often used because they are resistant to steam sterilization.

Autoclave Sterilization Procedures- Liquid Materials The autoclave must have a ‘LIQUID’ setting that can be used for liquid materials. The settings for liquids run for a longer period at a lower temperature to minimize liquid evaporation and spills. Liquids should be placed in borosilicate (Kimax or Pyrex) or polypropylene containers for autoclaving and these containers should be filled to no more than 75% capacity. ****Add an additional 20 minutes to the liquid cycle to ensure maximum sterilization****

Autoclave Sterilization Procedures- Liquid Materials The caps/stoppers on the containers should be loosened. Never autoclave sealed containers of liquid as this could result in an explosion of superheated liquid. Liquid containers should be placed in a stainless steel or polypropylene tray with ¼ to ½ inch of water in the bottom on the tray. This tray should be placed on a shelf in the autoclave, but not on the bottom of the chamber. Polyethylene trays should NOT be used as they may melt and cause damage to the autoclave.

Autoclave Sterilization Procedures- Solid Materials Never overfill biohazard waste bags or the autoclave. Waste bag should be at least three fingers wide at the opening of the taped bag. Keep biohazard waste bags slightly opened to allow for proper steam penetration. Place packaged material in a secondary container Stainless steel tray or autoclavable polypropylene bin Add between 50-100 ml (¼ – ½ cup) of water to each bag of solid waste to promote steam penetration. Note: If there is naturally occurring water in the load, adding additional water is not necessary.

Loading the Autoclave Load material to allow efficient steam penetration (do not overfill the chamber) Do not allow material to touch the sides or top of the chamber Autoclave clean items and waste separately

Operating the Autoclave Be sure the autoclave is functioning properly before use Record information in User Log Check strainer and remove any debris Close door properly and securely Choose the correct conditions for your material Make sure door to autoclave room remains closed as this prevents the loss of negative air pressure, therefore preventing the release of odors.

Autoclave Explosion These pictures represent the results of an improperly closed autoclave door. http://www2.umdnj.edu/eohssweb/aiha/accidents/AutoclaveExplosion.htm

Autoclave Waste Decontamination Procedures Autoclaves are to be operated at 121°C (250°F) or higher for a minimum of 60 minutes. Some biohazard waste may be required to be autoclaved for a different set amount of time.

Autoclave Waste Decontamination Procedures Material Temperature Time Laundry 121°C (250° F) 30 Minutes Trash (Biohazard bags containing infectious waste.) 1 Hour Glassware Liquids Animals 8 Hours

Unloading the Autoclave Put on Personal Protective Equipment Allow the autoclave to completely finish cycle Pressure gauge must read zero Verify cycle conditions were met Open door slightly to allow steam to escape While slowly opening the autoclave door, make sure to keep head, face, and hands away from the opening. Verify that heat sensitive tape has changed color or word “autoclaved” has appeared. Allow contents to cool before removal

What’s Wrong With These Pictures? No secondary container (bin/tray) No autoclave tape

Autoclave Testing & Verification N.C. Medical Waste Rules require that: Autoclaves be monitored and tested weekly under conditions of full loading for effectiveness in killing microorganisms. This is achieved through the use of Biological Indicators such as Geobacillus stearothermophilus (with average spore populations of 104 to 106 organisms).

Autoclave Testing & Verification Use of a Biological Indicator: Tape the indicator to a brightly colored sheet of paper OR tie it to a long piece of string hanging out of bag so that it can be retrieved after the autoclaving cycle has completed. Run the autoclave cycle for 60 minutes at the designated decontamination cycle. Once the cycle has completed and the contents have cooled, don all proper PPE and remove the indicator from the waste bag.

Autoclave Testing & Verification Incubation of Biological Indicator: The autoclaved indicator and an un-autoclaved control indicator are then incubated as recommended by the manufacturer. Check for signs of growth at regular intervals during the incubation period (8, 12, 24, and 48 hours). There should be signs of growth on the controlled indicator or the test is considered to be invalid. If there is growth on the autoclaved indicator, then the material was not properly autoclaved.

Autoclave Testing & Verification Incubation of Biological Indicator: If growth is discovered on the autoclaved indicator: The autoclave process needs to be re-evaluated: Time, temperature, and autoclaving procedures should be reviewed. If a mechanical problem is determined to be the culprit, Facilities Services should be contacted immediately for repair (962-3456). The waste does not have to be held until the results of the testing confirm effectiveness. However, the autoclave should not be used until it is repaired. Once repaired, the first load run in the autoclave should be tested with a biological indicator to ensure that the autoclave is functioning properly.

Autoclave Testing & Verification Autoclave Testing Results: Autoclave Testing Log A log of each test should be maintained and kept near the autoclave. The log should include the following information: The type of indicator used. Date the test was conducted. Time the test was conducted. Result of the test. An autoclave testing log can be printed off the UNC EHS website and is available at the following link: http://ehs.unc.edu/ih/biological/autoclave.shtml

Autoclave Testing Log Maintain a copy of the current form at the autoclave station Maintain logs for 3 years

Ask for the Biological Safety Section Questions? Call 962-5507 Ask for the Biological Safety Section