Review of Ultrasonic Imaging

Slides:



Advertisements
Similar presentations
Chapter3 Pulse-Echo Ultrasound Instrumentation
Advertisements

* NEAR ZONE is the nearest area to the transducer & it always has the same diameter of the transducer, BUT if the transducer: Linear the near zone length.
Doppler Echocardiography
Doppler Echocardiography
Foundations of Medical Ultrasonic Imaging
Ultrasound machine knobology
Volume I Companion Presentation Frank R. Miele Pegasus Lectures, Inc.
Resident Physics Lectures
Hospital Physics Group
Pegasus Lectures, Inc. Volume II Companion Presentation Frank Miele Pegasus Lectures, Inc. Ultrasound Physics & Instrumentation 4 th Edition.
EE 5340/7340, SMU Electrical Engineering Department, © Carlos E. Davila, Electrical Engineering Dept. Southern Methodist University slides can be.
ECE 501 Introduction to BME
Ultrasonic Nonlinear Imaging- Tissue Harmonic Imaging.
Ultrasonic Nonlinear Imaging- Tissue Harmonic Imaging
Sparse And Low Rank Decomposition Based Batch Image Alignment for Speckle Reduction of Retinal OCT Images Ahmadreza Baghaie, Roshan M. D’Souza, Zeyun Yu.
Despeckle Filtering in Medical Ultrasound Imaging
THE ULTRASOUND IMAGE: GENERATION AND DISPLAY
Introduction of Medical Imaging Chun Yuan. Organization of the Course 8 Lectures (1.5 hours per lecture) – Introduction of medical imaging and MRI – Basic.
Ultrasound Physics Sound is a mechanical, longitudinal wave that travels in a straight line Sound requires a medium through which to travel Ultrasound.
Objective To study the effect of sub surface defects in surface roughness monitoring through ultrasonic flaw detector. To study the sizing of defects.
Ultrasonic Thermometry
生醫工程實驗 期中報告 Ultrasound Apr. 2’ 2007 第一組:郭昱昇 溫仁揚 溫昌懌 National Taiwan University Electrical Engineering Dept.
Ultrasound. Sound waves Sounds are mechanical disturbances that propagate through the medium Frequencies
Resident Categorical Course
1 Ultrasonic Elasticity Imaging. 2 Elasticity Imaging Image contrast is based on tissue elasticity (typically Young’s modulus or shear modulus).
Speckle Correlation Analysis1 Adaptive Imaging Preliminary: Speckle Correlation Analysis.
Adaptive Imaging Preliminary: Speckle Correlation Analysis.
Display of Motion & Doppler Ultrasound
Review of Ultrasonic Imaging
Doppler Ultrasound Dr Mohamed El Safwany, MD.. Introduction The Doppler Effect refers to the change in frequency that results when either the detector/observer.
Digital Beamforming. Beamforming Manipulation of transmit and receive apertures. Trade-off performance/cost to achieve: –Steer and focus the transmit.
Ultrasonic İmaging.
WEATHER SIGNALS Chapter 4 (Focus is on weather signals or echoes from radar resolution volumes filled with countless discrete scatterers---rain, insects,
Volume I Companion Presentation Frank R. Miele Pegasus Lectures, Inc.
Chapter 6: Real-Time Image Formation. display B, M, Doppler image processing Doppler processing digital receive beamformer system control keyboard beamformer.
Ultrasound Principle 超音波原理
Ultrasound Basis Michel Slama Amiens.
MEDT8002 Ultralyd bildediagnostikk Faglærer: Hans Torp Institutt for sirkulasjon og bildediagnostikk Hans Torp NTNU, Norway.
Artifacts Ultrasound Physics George David, M.S.
Performance Issues in Doppler Ultrasound 1. 2 Fundamental Tradeoffs In pulsed modes (PW and color), maximum velocity without aliasing is In pulsed modes,
PSF What is PSF 1way/2way PSFs PSFs for continious & discrete apertures.
TISSUE HARMONIC IMAGING (THI)
TISSUE HARMONIC IMAGING (THI)
TISSUE HARMONIC IMAGING NUR FASHIHA BINTI AZMAN A DIAGNOSTIC IMAGING AND RADIOTHRAPY /2.
NUR INANI BINTI MOHD RUZLAN A Tissue harmonic imaging is a technique in ultrasonography that provides images of better quality as compared to conventional.
1 Opto-Acoustic Imaging 台大電機系李百祺. 2 Conventional Ultrasonic Imaging Spatial resolution is mainly determined by frequency. Fabrication of high frequency.
TISSUE HARMONIC IMAGING (THI) Aimi Alwani bt Mat Nawi A
Ultrasound Physics Image Formation ‘97. Real-time Scanning Each pulse generates one line Except for multiple focal zones frame one frame consists of many.
Principles of Medical Ultrasound Zahra Kavehvash.
Chapter 10 Digital Signal and Image Processing
Ultrasound Physics Image Formation ‘97.
Doppler Spectral Analysis
Lecture on Ultrasonography
Sound in medicine Lect.10.
Single-element transducers properties
Advanced Imaging.
Ultrasound Physics Image Formation ‘97.
LESSON 2 IMAGING METHODS OF ACOUSTIC FIELDS
Optical Coherence Tomography
Spatially Varying Frequency Compounding of Ultrasound Images
MCQs (1) – Questions Ultrasound propagates through the medium as a transverse wave Ultrasound velocity is equal to the product of frequency and wavelength.
Final exam information
Volume I Companion Presentation Frank R. Miele Pegasus Lectures, Inc.
PARAMETRIC SUB-BOTTOM PROFILER: A NEW APPROACH FOR AN OLD PROBLEM
Digital Beamforming.
Medical Ultrasound: Image Optimisation
Ultrasound Despeckling for Contrast Enhancement
Chapter 12 Image Artifacts
Principles of Medical Ultrasound
Ultrasonic Imaging Using Contrast Agents
Presentation transcript:

Review of Ultrasonic Imaging

Clinical Values of Ultrasound Visualization of anatomical structures in real time. Review of Ultrasonic Imaging

Clinical Values of Ultrasound Detection of blood flows, including direction, velocity distribution, variance and energy. Review of Ultrasonic Imaging

Clinical Values of Ultrasound Estimation of mechanical properties such as strain, elasticity, attenuation, acoustic backscattering, …etc. Treatment of diseased tissue by hyperthermia. Treatment of stones by extracorporeal lithotripsy. Review of Ultrasonic Imaging

Characteristics of Ultrasonic Imaging Real-time. Reflection mode. Non-invasive. Access limited. Body type dependent. Review of Ultrasonic Imaging

Factors of Image Quality Spatial resolution. Contrast resolution. Temporal resolution. Uniformity. Sensitivity. Penetration. Review of Ultrasonic Imaging

Review of Ultrasonic Imaging Spatial Resolution Lateral and elevational : diffraction limited. Axial resolution : the width of the pulse. Given limited total bandwidth, there exists a tradeoff between axial and lateral/elevational resolutions. Z Y X 3D sample volume Review of Ultrasonic Imaging

Lateral Resolution (X) Diffraction limited. Fourier transform of the aperture function (in focus, CW). Determined by frequency, active aperture size and depth. Fixed transmit focus and dynamic receive focus. Review of Ultrasonic Imaging

Elevational Resolution (Y) Fixed lens (geometric focus). Determined by frequency, aperture size and depth. Geometric focus Review of Ultrasonic Imaging

Review of Ultrasonic Imaging Axial Resolution (Z) Pulse width (absolute bandwidth). System and transducer bandwidth. Transmit power. Attenuation consideration. Review of Ultrasonic Imaging

Review of Ultrasonic Imaging Contrast Resolution Contrast resolution is determined by both spatial resolution and speckle noise variations. Speckle comes from coherent interference of diffuse scatterers. In-coherent processing must be used to reduce speckle noise. There exists a tradeoff between contrast and spatial resolutions. Review of Ultrasonic Imaging

Review of Ultrasonic Imaging Contrast Resolution Contrast-to-Noise Ratio (CNR): On a log display I1 I2 A Review of Ultrasonic Imaging

Review of Ultrasonic Imaging Contrast Resolution Contrast resolution is primarily limited by speckle noise. Speckle is a multiplicative noise. On a logarithmic display, Review of Ultrasonic Imaging

Review of Ultrasonic Imaging Spatial vs. Contrast Speckle noise 4.34dB for true speckle, a figure of merit for detectability. CNR increases as speckle noise decreases, generally resulting in loss in spatial resolution. Both CNR and spatial resolution can be improved by reducing sample volume. Review of Ultrasonic Imaging

Speckle Reduction Techniques Must be done in-coherently. Spatial filtering, loss in spatial resolution. Compounding: Compound image of the same object with different speckle appearance. Better edge definition. Sub-optimal spatial resolution. Review of Ultrasonic Imaging

Frequency Compounding In-coherently adding images acquired at different frequencies. Loss in axial resolution. Maximal reduction is N1/2. f Review of Ultrasonic Imaging

Review of Ultrasonic Imaging Spatial Compounding In-coherently adding images from different angles. Loss in lateral resolution. Improved edge definition. Laterally or elevationally (with a 2D array). Maximal reduction is N1/2. Review of Ultrasonic Imaging

Review of Ultrasonic Imaging Temporal Resolution Temporal resolution is determined by acoustic frame rate. It is also related to spatial Nyquist criterion. Temporal resolution is fundamentally limited by sound velocity but can be improved by signal processing in some cases. There exists a tradeoff between temporal and spatial resolutions. Review of Ultrasonic Imaging

Review of Ultrasonic Imaging Increasing Frame Rate Smaller field of view. Reduced transmit line number: Spatial Nyquist criterion. Parallel beamformation. Review of Ultrasonic Imaging

Parallel Beamformation Simultaneously transmit multiple beams. Interference between beams, spatial ambiguity. t1/r1 t2/r2 Review of Ultrasonic Imaging

Parallel Beamformation Simultaneously receive multiple beams. Correlation between beams, spatial ambiguity. Require duplicate hardware (higher cost) or time sharing (reduced processing time and axial resolution). r1 r2 t Review of Ultrasonic Imaging

Review of Ultrasonic Imaging Image Uniformity Image uniformity is usually referred to as the variations of the system’s point spread function throughout the entire image. Factors of image uniformity include depth of field, pulse shapes and variations due to lateral displacements. To achieve image uniformity, a sophisticated imaging system is required. Review of Ultrasonic Imaging

Review of Ultrasonic Imaging Sensitivity Sensitivity is defined in the context of the detection of weak signals. Sensitivity is determined by transducer design and system’s dynamic range. Sensitivity is particularly important in Doppler imaging and can be improved by signal processing. Review of Ultrasonic Imaging

Review of Ultrasonic Imaging Penetration Penetration is determined by acoustic power delivered to the body on transmit and the dynamic range of the system on receive. The transmit power is regulated for safety reasons. Hence, penetration must be improved without exceeding regulations. Review of Ultrasonic Imaging