Sunday Case of the Day Physics (Case 1: MR)

Slides:



Advertisements
Similar presentations
Medical Image Reconstruction Topic 4: Motion Artifacts
Advertisements

Magnetic Resonance Imaging
Imaging Sequences part I
MRI Phillip W Patton, Ph.D..
In Chan Song, Ph.D. Seoul National University Hospital
M R I Pulse Sequences Jerry Allison Ph.D..
MR Sequences and Techniques
I MAGING D ATA E VALUATION AND A NALYTICS L AB (IDEAL) CS5540: Computational Techniques for Analyzing Clinical Data Lecture 15: Accelerated MRI Image Reconstruction.
Statistical Parametric Mapping
Clinical Evaluation of Fast T2-Corrected MR Spectroscopy Compared to Multi-Point 3D Dixon for Hepatic Lipid and Iron Quantification Puneet Sharma 1, Xiaodong.
Evaluation of Reconstruction Techniques
Figure 2. Signal level (left) degrades with slice offset and slice thickness when Z2 SEM is used in GradLoc imaging (ROI = FOV/2). To recover the full.
David A. Bluemke, M.D., Ph.D. Associate Professor, Clinical Director, MRI Departments of Radiology and Medicine Johns Hopkins University School of Medicine.
Chapter 9 Basic MRI I Mark D. Herbst, MD, PhD. Notice This lecture contained many drawings on the whiteboard, so get these from one of the other students.
Comparative Analysis of Continuous Table and Fixed table Acquisition Methods: Effects on Fat Suppression and Time Efficiency for Single-Shot T2-weighted.
Equipment Magnetic resonance imaging (MRI) scan requires the use of a very strong magnetic field. Unlike other devices used in radiology, MR imaging.
I MAGING D ATA E VALUATION AND A NALYTICS L AB (IDEAL) CS5540: Computational Techniques for Analyzing Clinical Data Lecture 17: Dynamic MRI Image Reconstruction.
Psy 8960, Fall ‘06 Parallel Imaging1 SMASH and SENSE High field advantage Pros and cons … But first, review of last few homework assignments.
Compressed Sensing for Chemical Shift-Based Water-Fat Separation Doneva M., Bornert P., Eggers H., Mertins A., Pauly J., and Lustig M., Magnetic Resonance.
Medical Imaging Systems: MRI Image Formation
vertebrae.
Imaging Sequences part II
M. D. A NDERSON Cancer Center Magnetic Resonance Imaging: 4.7 T, 40 cm Bruker Biospec James A Bankson, Ph.D. and John D Hazle, Ph.D. Small Animal Cancer.
Magnetic Resonance Electrical Impedance Mammography (MREIM) A new approach to breast cancer imaging Maria Kallergi, 1 John J. Heine, 2 and Ernest Wollin.
Medical Imaging Systems: MRI Image Formation
Partial Parallel imaging (PPI) in MR for faster imaging IMA Compressed Sensing June, 2007 Acknowledgement: NIH Grants 5RO1CA and 5P41RR008079, Pierre-Francois.
Contrast Mechanism and Pulse Sequences Allen W. Song Brain Imaging and Analysis Center Duke University.
Correcting for Center Frequency Variations in MRSI Data Using the Partially Suppressed Water Signal Lawrence P Panych, Ph.D., Joseph R Roebuck, Ph.D.,
Mitglied der Helmholtz-Gemeinschaft EPIK for TRIMAGE Seong Dae Yun 1 and N. Jon Shah 1,2 1 Institute of Neuroscience and Medicine 4 Forschungszentrum Jülich.
In Burst experiments, the read gradient is much larger then the other imaging gradients, hence, only translation along this gradient direction will cause.
Introduction Magnetic resonance (MR) imaging is recognised as offering potential benefits in the delineation of target volumes for radiotherapy (RT). For.
MR Compatibility Torques Artifacts Not Compatible Yes Yes MR Safe No
A CCELERATED V ARIABLE F LIP A NGLE T 1 M APPING VIA V IEW S HARING OF P SEUDO -R ANDOM S AMPLED H IGHER O RDER K-S PACE J.Su 1, M.Saranathan 1, and B.K.Rutt.
3D sequence MRI in the assessment of meniscofemoral and ligament lesions of the knee MA.Chaabouni,A.Daghfous, A.Ben Othman,L.Rezgui Marhoul Radiology departement.
Magnetic Resonance Learning Objectives
In vivo MRI of Fast Relaxing Spins Using a Swept Radiofrequency Djaudat Idiyatullin, Curt Corum, Jang-Yeon Park, Michael Garwood Center for Magnetic Resonance.
MRI Magnetic Resonance Imaging. Definition A non-ionizing technique with full three dimensional capabilities, excellent soft-tissue contrast, and high.
Receive Coil Arrays and Parallel Imaging for fMRI of the Human Brain
Substantial signal dropout was observed in arterial spin label (ASL) images, acquired at 3T in a young female patient three-months after mild traumatic.
SWIFT Dual Breast Imaging Sequence and Coil with Interleaved Adiabatic Fat Suppression high temporal and spatial resolution #3369 Computer 8 breast imaging.
Chapter 5 Mark D. Herbst, M.D., Ph.D.. The MR Imaging Process Two major functions –Acquisition of RF signals –Reconstruction of images.
He-Kwon Song, PhD Associate Professor & Co-investigator in the CMROI, Department of Radiology, University of Pennsylvania School of Medicine, Philadelphia,
FMRI data acquisition.
MRI ARTEFACTS.
Eric Lévêque, Benoit Hainaux,
Sunday Case of the Day Physics
7.3c. Post-Contrast Axial CT of the Brain
Imaging Science Department, IC Jo Hajnal
Sunday Case of the Day Physics
Monday Case of the Day Physics
Tuesday Case of the Day Physics History: A B MIP
Physics of Magnetic Resonance Imaging
When Should Abdominal Magnetic Resonance Imaging Be Used?
An Optimal Design Method for MRI Teardrop Gradient Waveforms
Parallel Imaging Artifacts in Body Magnetic Resonance Imaging
MRI Pulse Sequences: IR, EPI, PC, 2D and 3D
7.1b. Contrast coronal T1 Wtd MRI 7.1c. Contrast sagittal T1 Wtd MRI
Chapter 11 Lecture Mark D. Herbst, M.D., Ph.D..
Basic MRI I Mark D. Herbst, MD, PhD
When Should Abdominal Magnetic Resonance Imaging Be Used?
QUIZ 10: Optimizing the MRI of biliary network
Volume 11, Issue 2, Pages (April 2016)
3T-versus-7T DTI with 36 diffusion-encoding directions at b = 3000 s/mm2 and 2.0 × 2.0 × 2.0 mm isotropic voxel resolution. 3T-versus-7T DTI with 36 diffusion-encoding.
QUIZ 7: MR Enterography Technique
Without fat suppression techniques, fat and fluid both have a high-signal intensity on fluid-sensitive fast spin echo. This adult patient with a history.
Tuesday Case of the Day Physics History:
Figure 6a. (a-c) Axial T1-weighted GRE MR images (120/4
The echo time (TE) The echo time (TE) refers to the time between the application of the radiofrequency excitation pulse and the peak of the signal induced.
RF Coils in MRI By: Jamil Abdolmohammadi MSc of MRI
MRI basics for radiation oncologists
Presentation transcript:

Sunday Case of the Day Physics (Case 1: MR) Authors: Edward F. Jackson, PhD, and Donna M. Reeve, PhD Department of Imaging Physics, University of Texas M. D. Anderson Cancer Center, Houston. History: Abdominal images were acquired on a 1.5-T system with a body-array receive coil. Localizer and coronal SSFP-based images were obtained, followed by 2D axial FSE T2-weighted images. Following these acquisitions, an MR-compatible injector was connected, and pre- and postgadolinium axial T1-weighted images were obtained using a 3D fast gradient-echo sequence with slice interpolation, IR-prepped fat suppression, and SENSE-based parallel imaging with an acceleration factor of two. Selected images from the post-Gd axial T1-weighted series are shown below. (Similar artifacts were evident in the pre-Gd T1-weighted images.) What is the most likely cause of this particular artifact? Figure 1a. Representative T1-weighted images demonstrating artifacts not seen in axial T2-weighted images.

Findings: A more complete set of T1-weighted images (with artifact) and matching T2-weighted images without similar artifact are provided below. Figure 1b. Axial T1-weighted post-Gd images (top) and corresponding axial T2-weighted images (bottom).

Diagnosis: The artifacts are due to aliasing in the SENSE parallel imaging reconstruction. Signal from contrast agent in the coiled MR-compatible injector tubing that was, in this particular case, inappropriately positioned across the patient’s abdomen was aliased into the image.

Discussion: Parallel imaging techniques accelerate image acquisition rates by utilizing coil sensitivity modulation to encode the MR image and reduce the number of encoding steps. Sensitivity Encoding (SENSE) is one approach to parallel imaging that requires the acquisition of a coil sensitivity reference scan prior to a reduced-phase FOV imaging series. The reduced-phase FOV results in purposeful aliasing of the image in the phase-encode direction. During the reconstruction process, the sensitivity information from the multielement coil is used to “unwrap” the aliased images. Information outside the reconstructed FOV, however, commonly aliases into the FOV, as it cannot be appropriately unwrapped in the reconstruction process. Aliased Images Unaliased Image 1 2 3 4 SENSE: unfold Low-resolution fully encoded coil sensitivity map In this particular case, the coiled contrast agent delivery tube, resting on top of the anterior portion of the body array, gave rise to signal that was aliased into the anatomy because it was outside the field of view of the reconstructed image. The T2-weighted images and any images acquired without SENSE parallel imaging reconstruction did not demonstrate such artifacts.

Reference/Bibliography: Pruessmann KP, Weiger M, Scheidegger MB, Boesiger P. SENSE: sensitivity encoding for fast MRI. Magn Reson Med 1999;42:952–962.