Task Group 158 Measurement and Calculation of Doses Outside the Treatment Volume from External-beam Radiation Therapy Treatment Eric E Klein, Ph.D.

Slides:



Advertisements
Similar presentations
Chapter 4 Radiation Dosimeters
Advertisements

X-Rays In Medicine Noadswood Science, 2013.
Beam Therapy Equipment 2 Linear Accelerator Head.
Energy deposition and neutron background studies for a low energy proton therapy facility Roxana Rata*, Roger Barlow* * International Institute for Accelerator.
Photon Beam Monitor-Unit Calculations
Evaluation of the characteristics of TLD LiF:Mg.Ti-100 Powder: A Measure of Consistency Between Multiple Batches of Powder Paola Alvarez,Jose Francisco.
Radiation Exposure, Dose and Relative Biological Effectiveness in Medicine Background Image:
Proton Imaging and Fighting Cancer
MONTE CARLO RADIATION DOSE SIMULATIONS AND DOSIMETRY COMPARISON OF THE MODEL 6711 AND I BRACHYTHERAPY SOURCES Mark J. Rivard Department of Radiation.
G4NAMU GEANT4 North America Medical Users Organization (G4NAMU) GEANT4 in external beam therapy Harald Paganetti.
At the position d max of maximum energy loss of radiation, the number of secondary ionizations products peaks which in turn maximizes the dose at that.
Electron Beams: Physical Principles and Dosimetry
Dose Distribution and Scatter Analysis
Dosimetric evaluation of a new design MOSFET detector Per H. Halvorsen* & Stephanie Parker University of North Carolina.
Results The measured-to-predicted dose ratio criteria used by the RPC to credential institutions is , however for this work, a criteria of
TWO FIELD BREAST PLAN VS. OPTIMIZED CONFORMAL BREAST PLAN: COMPARISON OF PLAN PARAMETERS Authors: Borko Basarić, Ozren Čudić, Milan Teodorović, Borislava.
Introduction Ion recombination is approximately corrected for in the Task-Group-51 protocol by P ion, which is calculated by a two-voltage measurement.
External Beam Radiotherapy
Patient Plan Results: Table 3 shows the ratio of the Pinnacle TPS calculation to the DPM recalculation for the mean dose from selected regions of interest.
Test of the proposed method Introduction CCD Controller CCD Illuminator gel Filter 585nm Assembling the phantom before its irradiation. The phantom, ready.
David Followill, Ph.D. Radiological Physics Center
AUTHORS (ALL): Huang, Xiaoyan 1, 2 ; Kuan, K M 2 ; Xiao, G L 2 ; Tsao, S Y 3, 2 ; Qiu, X B 2 ; Ng, K 2. INSTITUTIONS (ALL): 1. Radiation Oncology, Sun.
Comparison of Clinical Parameters for Proton Therapy in the United States Paige Summers, MS.
Milestones M5 and M6 Mid-Term Review, October 2013M. Silari – ARDENT overview1 M5: comparison of detector technologies (WP1) M6: choice of detector technology.
CT physics and instrumentation
The RPC Proton Therapy Approval Process
JUSTIFICATION OF COMPUTERIZED TOMOGRAPHY EXAMINATIONS AND RADIATION RISKS IN EVERYDAY RADIOLOGICAL PRACTICE Darka R. Hadnađev, Olivera Nikolić, Sanja Stojanović.
Application of a 2-D ionization chamber array for dose verification of dynamic IMRT with a micro-MLC Fujio ARAKI, PhD 1, S. TAJIRI 2, H. TOMINAGA 2, K.
Institute for Advanced Radiation Oncology
Medical Accelerator F. Foppiano, M.G. Pia, M. Piergentili
Radiological Physics Center David Followill, Ph.D. and RPC Staff.
Permanent Interstitial Implants Ideal strategy to curatively manage small volume gynecologic malignancies Can deliver high cumulative radiation dose to.
ARDENT Advanced Radiation Dosimetry European Network Training initiative WP1: Gas Detectors S. Rollet.
1 Radiotherapy, hadrontherapy and treatment planning systems. Faiza Bourhaleb INFN-Torino University Med 1er-Morocco  Radiotherapy  Optimization techniques.
Araki F. Ikegami T. and Ishidoya T.
Introduction The Radiological Physics Center (RPC) anthropomorphic quality assurance (QA) phantom program is one tool the RPC uses to remotely audit institutions.
Improvement of the Monte Carlo Simulation Efficiency of a Proton Therapy Treatment Head Based on Proton Tracking Analysis and Geometry Simplifications.
Somvilai Mayurasakorn, MD. Division of Therapeutic Radiology and Oncology, Faculty of Medicine, Chiang Mai University Somvilai Mayurasakorn, MD. Division.
Optimization of Volumetric Modulated Arc Therapy (VMAT) Planning Strategy Using Ring-shaped ROI for Localized Prostate cancer Kentaro Ishii, Masako Hosono,
Measurements of the photon and neutron dose delivered to organs outside the radiation beams for 3DCRT and IMRT radiotherapy A. Kowalik 1, W. Jackowiak.
Adapting A Clinical Medical Accelerator For Primary Standard Dosimetry
Characterization of proton-activated implantable markers for proton range verification using PET J. Cho1, G. Ibbott1, M. Kerr1, R. A. Amos2, F. Stingo1,
RADIATION PROTECTION 04/12/2016.
CHAPTER 3 DOSE DETERMINATION FOR EXTERNAL BEAMS
Electron Beam Therapy.
Optimisation of Patient Protection for Radiography
The use of 4DCT images to optimize the Internal Target Volume in Radiotherapy  Nikos Giakoumakis, Brian Winey, Joseph Killoran, Tania Lingos, Laurence.
Dr: Mohamed Afifi By Lecturer Radiological Science
INTRODUCTION RESULTS DATA ANALYSIS DISCUSSION METHODS REFERENCES
Evaluation Of RTOG Guidelines For Monte Carlo Based Lung SBRT Planning
Pediatric Radiology By Dr. Wambani, J.S. Chief Radiologist
Development and characterization of the Detectorized Phantom for research in the field of spatial fractionated radiation therapy. D. Ramazanov, V. Pugatch,
A Brachytherapy Treatment Planning Software Based on Monte Carlo Simulations and Artificial Neural Network Algorithm Amir Moghadam.
Arghya Chattaraj, T. Palani Selvam, D. Datta
أجهزة العلاج الإشعاعي Clinical Radiation Generator
Implementation of Object Spot Avoidance in Proton Pencil Beam Treatment on Whole Breast with Implant Metal Injector Peng Wang, PhD, DABR, Karla Leach,
Fig. 4. Percentage of passing rate between clinical and 544 plans.
Insert tables Insert figure
Principles of Radiation Protection
Dosimetry of Alternative Techniques for Accelerated Partial Breast Irradiation Hanh Pham, B.S, CMD, Thanh Nguyen, BS, Christina Henson, MD, Salahuddin.
Development of Gas Electron Multiplier Detectors for Muon Tomography
Resident Physics Lectures (Year 1)
Ch 10. A System of Dosimetric Calculations
A Multi-Institutional Dosimetric Evaluation of Proton Versus Photon Stereotactic Body Radiation Therapy for Hepatocellular Carcinoma F. Khan, B. Nguyen,
Resident Physics Lectures (Year 1)
Sheng Yang a, Yen-Wan Hsueh Liu b
Hot and cold spots are common problems associated with planning:
GHG meeting at ESTRO36 May, 2017
Planning techniques of proton boost
Average Dose-Volume Ratio
Presentation transcript:

Task Group 158 Measurement and Calculation of Doses Outside the Treatment Volume from External-beam Radiation Therapy Treatment Eric E Klein, Ph.D

Task Group Members Stephen F. Kry, co-chair MD Anderson Cancer Center, Houston, TX Bryan Bednarz, co-chair, University of Wisconsin, Madison, WI Rebecca M. Howell, MD Anderson Cancer Center, Houston, TX Larry Dauer, Memorial Sloan-Kettering Cancer Center, New York NY David Followill, MD Anderson Cancer Center, Houston, TX Eric Klein, Brown University, Providence, RI Harald Paganetti, Massachusetts General Hospital and Harvard Medical School; Boston, MA Brian Wang, University of Louisville, Louisville, KY Cheng-Shie Wuu, Columbia University, New York, NY X. George Xu, Rensselaer Polytechnic Institute, Troy NY

Status of Report Measurement and Calculation of Doses Outside the Treatment Volume from External-beam Radiation Therapy Treatment Med Phys. 2017 Oct;44(10):e391-e429

Scope of Report This report aims to address the following charges as they pertain to non-target radiation: Highlight major concerns Provide a rough estimate of doses associated with different treatment approaches in clinical practice Discuss the uses of dosimeters and phantoms for measuring photon, electron, and neutron exposures Discuss the use of calculation techniques (including Monte Carlo) for dosimetric evaluations Highlight techniques that may be considered for reducing non-target doses Discuss dose reporting Make recommendations for clinical and research practice

Introduction to TG-158 Non-target dose? Out-of-field dose? Dose outside the PTV (no benefit to patient) Out-of-field dose? Dose outside any primary field border TG-158 addresses non-target dose, but primarily in a low-dose context In-field non-target dose Out of field non-target dose PTV: Target dose Out of field non-target dose

Why do we care? Low doses of radiation can be bad Second cancers Cardiac toxicity Fetal damage Implantable electronic devices Cataracts Skin dose (unique considerations not addressed in this report – see TG-176)

Where does the dose come from? Patient scatter Collimator Scatter Head leakage Neutrons Different properties than in-field radiation

So how much dose is there? Simple square fields Use TG-36 1995 What about newer techniques?

Chapter 3 of report IMRT; similar for Tomo, VMAT, FFF, SBRT, electron, proton, brachytherapy, imaging.

Caution These doses provide a range that is likely to be encountered There can be a lot of variability between individual treatments These are just rough guidelines You need to determine the dose for your own case Measurements and/or calculations

Techniques to Minimize Non-target Dose Use details from the report to discuss options and strategies to reduce non-target dose: Reducing the Target Volume (CTV and PTV) Treatment Modalities (protons, FFF, electrons) Treatment Energy (6X vs 18X) Photon Wedges (hard wedges vs dynamic/universal) MLC and Collimator Orientation Coplanar vs Non-coplanar Beams Jaw Tracking Patient Shielding Accelerator Shielding Imaging Dose Management

Peripheral Dose Contributions Characteristics Room scatter <=3% Close to field collimator contributes 20-40% Machine leakage is greatest contributor at distances >30 cm from CAX Characteristics Increases with increasing field size Minimal increase with decreasing depth No significant change with change in SSD

Reducing the target volume Reduce CTV Hodgkin’s Lymphoma – treat only the involved nodes rather than the whole mantle Very high impact on dose reduction Not usually a physics decision Reduce PTV (reduce margin size) Requires improved imaging/immobilization Reduces high and intermediate dose volumes Often needs additional imaging dose

FFF Out of field dose reduced for both IMRT and SBRT procedures using FFF Less head leakage, less collimator scatter Most benefit far from the treatment field, more modulated

Collimator orientation For tertiary MLC, align leaves along patient length (vs across patient) to achieve maximum shielding Dose reduced by 6-50% Easily implemented

Recommendations Needs are different between clinical care (e.g., measuring dose to a pacemaker) and a research study to evaluate relative risks between treatment approaches. Clinical care recommendations Research focused recommendations A more stringent set, particularly for neutrons

Clinical care rec. 1 Dose assessment should be based on needed precision to yield proper clinical management of the case. While an overestimate of the dose may be adequate, the magnitude of the error should nevertheless be known so as to ensure proper clinical management.

Clinical care rec. 2 Treatment planning systems show large errors as close as 3 cm from the field edge and doses calculated by the planning system should not be relied on beyond 3 cm from the field edge (or below the 5% isodose line). The mean organ dose can generally be trusted for cases where the majority of the organ is within the 5% isodose line.

Clinical care rec. 3 Photon measurements can be conducted with a wide range of dosimeters. Physicists must be aware of the nature of the radiation field and the response of the detector so as to at least know the magnitude of the sometimes large corrections that may be necessary.

Clinical Care rec 4. Because neutron measurements are prone to very large errors, TG-158 recommends that for clinical scenarios, data from the literature should be used to estimate patient doses rather than performing patient specific measurements. Surface doses from TG-158 can be used as conservative estimates of patient dose.

Clinical care rec. 5 and 6 For structures of interest within the IGRT imaging field, the imaging dose should be included in the total out-of-field dose. When indicated or possible, the non-target dose to the patient should be minimized following the approaches outlined in report.

Research recommendations Additional recommendations are presented on Neutron dosimetry Monte Carlo methods Dose reporting

Research Recommendations for Neutron Dosimetry Optimal neutron dosimetry consists of spectra measurements in air using, for example, Bonner spheres. Other measurement systems can provide reasonable and acceptable results. Monte Carlo is well suited for in-patient/phantom neutron dose calculations, after the model has been well validated against in-air neutron measurements.

Research Recommendations for Neutron Dosimetry Neutron measurements in-patient/phantom,, require that spatial variations in the neutron energy spectrum be accounted for in the response and calibration of the detector. Low-energy neutron detectors (e.g., standard Bonner spheres, track etch detectors, bubble detectors) should not be used in proton beams. Appropriate detectors include extended-range Bonner sphere systems, SWENDI detectors, and TEPCs.

Neutron Survey Instruments Neutron Survey Measurements Rem-Meter Detector assembly is a polyethylene cylinder, containing a BF3 proportional counter. At high Dose rates: Pulse pile-up and dead time occurs Only for use outside the field (≥ 1m) Bubble detectors for Neutron head leakage Note: too much pile-up if use Rem-meter for head leakage. Note: in many states, this is not necessary measurement because state will accept manufacturer documentation of neutron head leakage. Un-irradiated detector Bubbles form after neutron exposure

Monte Carlo Simulations For photon models, full modeling of the head is necessary beyond 15 cm from the field edge For situations with neutrons, head shielding and major structural components must be included Validation of the model must rely on out of field measurements (and/or neutrons), not in-field characteristics Phantom should match patient if possible, height is the most important parameter

Dose Reporting Relative merit of point dose, integral dose, effective dose, etc. for dose assessment How to manage neutron RBE (Q, wR, ambient dose equivalent, etc.) Effective dose should not be used when doses >2Gy are involved. Fun technical reading.

Use Diode Dosimetry Silicon semiconductor diodes Real-time measurement High sensitivity Good spatial resolution Small size Simple instrumentation No bias voltage Ruggedness Independence from temperature/air pressure changes

Or OSLD 7mm diameter x 0.2 mm thick disk Dose is nearly proportional to OSL signal Linear until ~ 3 Gy Some energy dependence Z ≈ 11.4 -Image courtesy of Landauer Some research as to could this be done during Jursinic (2007) -Figure from Jursinic (2007) -Figure from Jursinic (2007)

Do’s and Don’ts Do design techniques to reduce exposure Use tables from TG-158 Confer with Clinicians Do NOT use TP data !!