Wednesday Case of the Day

Slides:



Advertisements
Similar presentations
CT Scanning: Dosimetry and Artefacts
Advertisements

IMAGE QUALITY.
Computed Tomography Physics, Instrumentation, and Imaging
Key CT Parameters - What Are They Called and What Do They Mean?
Image Artifacts Chapter 8 Bushong.
Computed Tomography II
Spiral CT Bushong Chapter 5.
Seeram Chapter 13: Single Slice Spiral - Helical CT
NC HPS Meeting 10/18-19/2001 Boone, NC Recent Advances in CT Technology and Issues of CT Dosimetry T. Yoshizumi 1,2, M. Sarder 1, R. Reiman 1,2, E. Paulson.
1 Physics Testing for Performance Based Protocol version 3.0 (Now based on ACRIN 6678) QIBA Group 1C.
5 th September 2005 Paul Collins Computed Tomography Dosimetry Assessment of Effective Dose in Computed Tomography using an Anthropomorphic Phantom Paul.
How CT / MSCT works  Imageaquisition:. How CT / MSCT works.
David E. Hintenlang, Ph.D., DABR
Novel single-source high-pitch protocol for CT angiography of the aorta: comparison to high-pitch dual-source protocol in the context of TAVI planning.
IMAGE QUALITY NOISE LINEARITY CROSS-FIELD UNIFORMITY IMAGE ARTIFACTS.
BMME 560 & BME 590I Medical Imaging: X-ray, CT, and Nuclear Methods
Computed Tomography III
An algorithm for metal streaking artifact reduction in cone beam CT M. Bazalova 1,2, G. Landry 1, L. Beaulieu 3, F. Verhaegen 1,4 1-McGill University,
tomos = slice, graphein = to write
CT Quality Control for CT Scanners. Quality Control in CT A good idea? Yes Required for accreditation? Sometimes Improves image quality? Sometimes Depends.
Abstract ID 1196 CT volumetry of adrenal glands in normal subjects.
Computed Tomography Q & A
IAEA International Atomic Energy Agency Optimization of Protection in Computed Tomography (CT)-What can radiographers do? IAEA Regional Training Course.
QIBA CT Volumetrics - Cross-Platform Study (Group 1C) March 18, 2009 Interclinic Comparison of CT Volumetry Quantitative Imaging Biomarker Alliance.
Methods of Data Acquisition. General Methods of Data Acquisition Localizer scanning Step-and-shoot (axial) scanning Helical scanning.
Image Quality The capacity to define, measure, and assess image quality is a primarily responsibility of a CT Technologist.
Part No...., Module No....Lesson No
Somvilai Mayurasakorn, MD. Division of Therapeutic Radiology and Oncology, Faculty of Medicine, Chiang Mai University Somvilai Mayurasakorn, MD. Division.
Quality Assurance.
1 Physics Testing for Performance Based Protocol version 2.1 QIBA Group 1C.
Quality Control Bushong Chapter 10.
Nuclear Medicine Physics Jerry Allison, Ph.D. Department of Radiology Medical College of Georgia Image Artifacts in Nuclear Medicine Part B.
Monday Case of the Day History: Multiple examinations from a CT scanner using a standard helical scan protocol show diffuse hypointense regions of artifact.
CONVENTIONAL AND SPIRAL/HELICAL CT
History: A 61-year-old man with suspected thoracic aneurysm underwent ECG-gated chest CT angiogram. He was instructed to hold his breath during the exam.
Thursday Case of the Day The arrow indicates a region of high apparent FDG uptake posterior to the bladder. What is the cause of this artifact? a) attenuation.
Thursday Case of the Day The likely cause of the artifact is: A. Patient Motion B. Improper Cupping Correction C.Off-Focus Radiation combined with heel.
Wednesday Case of the Day History: 32-year-old female (approximately standard-sized female) undergoes CT scan of abdomen and pelvis after motor vehicle.
Sunday Case of the Day How does the presence of the object impact patient dose? A. Increases by 20% - 50% B. Increases by 3% - 5% C. No Change D. Decreases.
CT Multi-Slice CT.
45-year-old woman with BMI of 23.1 and chest circumference of 94.0 cm. Axial CT images obtained at 120 kV and 200 mAs show ascending aorta with image noise.
CT PRINCIPLES AYMAN OSAMA.
Sunday Case of the Day Physics
Computed Tomography Image Reconstruction
Monday Physics Case of the Day Answers (1 of 3) Paul E
Optimisation of Chest Computed Tomography Using a Phantom: Impact of mAs and Reconstruction Techniques on Image Quality C.S. Reis1; T. Faqir2; V. Harsaker3;
Sunday Case of the Day Physics
Monday Case of the Day Physics
Fig. 1. Images of 43-year-old woman with metastatic lung nodule from rectal cancer show round pulmonary nodule measuring 5 mm (arrows) in left basal lung.
Physics Testing for ACRIN 6678 Protocol version 3.0 QIBA Group 1C
Transaxial PET without attenuation correction.
Wednesday Case of the Day
Sunday Case of the Day Category Physics Question:
Wednesday Case of the Day
Figure Cervical spinal cord area assessment
Sample photon-counting detector images of low- and high-energy bins of a 70-year-old woman (section thickness, 2 mm; increment, 2 mm; window center, 45.
David J. Brenner, Maria A. Georgsson  Gastroenterology 
Evaluate the limit: {image} Choose the correct answer from the following:
Thursday Case of the Day
The effect of vertical centering and scout direction on automatic tube voltage selection in chest CT: a preliminary phantom study on two different CT.
CT ARTIFACTS.
Fig. 4 Pupil shape and image quality in the model sheep eye.
A) Axial, 5-mm slice thickness computed tomography (CT) image through the upper lobes at the level of the aortic arch. a) Axial, 5-mm slice thickness computed.
42-year-old male patient with follow-up neck CT for lymphoma at 70 kVp (A) and corresponding previous CT at 120 kVp (B). 42-year-old male patient with.
Daily Routines Math Book
Axial CT image through L5/S1 (3-mm section, 100 KVp, 50 mA).
A, Curved multiplanar reconstruction image obtained with a CT myelography shows the left hemicord with the small remnant of subarachnoid space running.
Given that {image} {image} Evaluate the limit: {image} Choose the correct answer from the following:
CT scans chosen for rWTH measures.
Presentation transcript:

Wednesday Case of the Day Category Physics Authors: Eric L. Gingold, PhD & Jaydev K. Dave, PhD Department of Radiology, Thomas Jefferson University, Philadelphia, PA 19107 Question: The ACR phantom images on the right are both nominal 5 mm reconstructions from the same scanner. However, the 0.5 mm spaced wire ramps (arrows) show a pronounced difference in slice thickness. What is the reason for this difference? Noise index mAs Slice sensitivity profile Bowtie compensation filter

Diagnosis: C. Slice sensitivity profile

Correct Answer: C. Different slice sensitivity profile The upper image was scanned in axial mode, and the lower image in helical mode. An important difference between axial and helical scanning is the shape of the slice sensitivity profile (SSP). The axial SSP is quite sharp, while a helical SSP is broad, and depends on the helical pitch, the interpolation and the reconstruction. From: Polacin et al, Evaluation of Section Sensitivity Profiles and Image Noise in Spiral CT, Radiol 185(1):29-35, 1992.

Additional Information (Vendor-Specific) GE CT scanners offer two types of helical reconstruction, called “Full” and “Plus” modes “Full mode” produces a thinner slice profile “Plus” mode has a 20% wider slice profile than “Full” but requires 15-20% less mA to achieve the same image noise The lower image in the COD question was reconstructed using the wider “Plus” mode. This explains the apparently larger slice width seen in the COD question (based on wire ramps observed in the ACR phantom)

References/Bibliography: Polacin et al, Evaluation of Section Sensitivity Profiles and Image Noise in Spiral CT, Radiol 185(1):29-35, 1992