Intoxilyzer 5000EN Factory Technical Specifications Rev 3.09
>0.450BrAC = Range Exceeded Error Message Majority of BrAC Results Intoxilyzer 5000EN Factory Technical Specifications RANGE: 0.005 to 0.450 BrAC grams/210 liters Majority of BrAC results are between 0.005 BrAC and 0.450 BrAC Results less than 0.005 BrAC = 0.000 BrAC on display Results greater than 0.450 Brac = Range Exceeded Error Message 0.005 BrAC 0.450 BrAC <0.005 BrAC = 0.000 BrAC >0.450BrAC = Range Exceeded Error Message Majority of BrAC Results Rev 3.09
Intoxilyzer 5000EN Factory Technical Specifications Accuracy: +3% or +0.003 BrAC which ever is greater Same accuracy as similar laboratory equipment Rev 3.09
Intoxilyzer 5000EN Factory Technical Specifications Precision: Standard Deviation of 0.003 BrAC or better If 10 tests of a known ethanol solution were run on the instrument, all results would be within 0.003 BrAC of each other Rev 3.09
Intoxilyzer 5000EN Factory Technical Specifications Test Time: Approximately 7 minutes For Colorado test sequence data entry time + (ACABAIA2ABACA) Rev 3.09
Intoxilyzer 5000EN Factory Technical Specifications Operating Temperature: 68 F to 86 F (20 C to 30 C) room temperature Keep room temperature at approximately 75 F Rev 3.09