Volume 11, Issue 1, Pages 167-168 (January 2014) Regular ventricular pacing inhibition due to electronic artifact emitted by pulse generator equipped with a self-adjusting sensitivity function Takashi Iizuka, MD, Yoshiaki Kaneko, MD, PhD, Tadashi Nakajima, MD, PhD, Masahiko Kurabayashi, MD, PhD Heart Rhythm Volume 11, Issue 1, Pages 167-168 (January 2014) DOI: 10.1016/j.hrthm.2013.10.031 Copyright © 2014 Heart Rhythm Society Terms and Conditions
Figure 1 Ambulatory ECG recording showing the absence of ventricular pacing stimuli tracking single nonconducted P wave (asterisk) at 13:57:52 h (A), 14:58:22 h (B), 16:59:20 h (C), and 01:03:11 h (D). The sinus rate was 55 bpm in A, 60 bpm in B, 55 bpm in C, and 60 bpm in D. Heart Rhythm 2014 11, 167-168DOI: (10.1016/j.hrthm.2013.10.031) Copyright © 2014 Heart Rhythm Society Terms and Conditions