M.J. Wood, E.S. Lin, J.P. Thompson  British Journal of Anaesthesia 

Slides:



Advertisements
Similar presentations
Measurement of static and dynamic compliance during mechanical ventilation. A tidal volume is delivered, causing a peak in airway pressure; dynamic compliance.
Advertisements

Assessing the performance of the Whisperflow® continuous positive airway pressure generator: a bench study  G.W. Glover, S.J. Fletcher  British Journal.
I. Galvin, G.B. Drummond, M. Nirmalan  British Journal of Anaesthesia 
M. Garnier, C. Quesnel, J. -P. Fulgencio, M. Degrain, G. Carteaux, F
Relationship of inspiratory and expiratory times to upper airway resistance during pulsatile needle cricothyrotomy ventilation with generic delivery circuit 
Increasing positive end-expiratory pressure (re-)improves intraoperative respiratory mechanics and lung ventilation after prone positioning  J Spaeth,
Is pulmonary resistance constant, within the range of tidal volume ventilation, in patients with ARDS?  G Mols, V Kessler, A Benzing, M Lichtwarck-Aschoff,
Dexmedetomidine vs remifentanil intravenous anaesthesia and spontaneous ventilation for airway foreign body removal in children  K.-Z. Chen, M. Ye, C.-B.
Ventilation with low tidal volumes during upper abdominal surgery does not improve postoperative lung function  T.A. Treschan, W. Kaisers, M.S. Schaefer,
C. Unzueta, G. Tusman, F. Suarez-Sipmann, S. Böhm, V. Moral 
Oxygen delivery through high-flow nasal cannulae increase end-expiratory lung volume and reduce respiratory rate in post-cardiac surgical patients  A.
I. Garutti, L. Puente-Maestu, J. Laso, R. Sevilla, A. Ferrando, I
Patient–ventilator interaction
T. N. Weingarten, F. X. Whalen, D. O. Warner, O. Gajic, G. J
The upper airway during anaesthesia
Forty years of closing volume
The effect of bi-level positive airway pressure mechanical ventilation on gas exchange during general anaesthesia  Yu G , Yang K , Baker A.B. , Young.
Double-lumen tubes and auto-PEEP during one-lung ventilation
Static versus dynamic respiratory mechanics for setting the ventilator
High-frequency oscillation in adolescents
Pleural mechanics and the pathophysiology of air leaks
Respiratory resistance during anaesthesia with isoflurane, sevoflurane, and desflurane: a randomized clinical trial  V. Nyktari, A. Papaioannou, N. Volakakis,
Anaesthetic conserving device AnaConDa®: dead space effect and significance for lung protective ventilation  L.W. Sturesson, M. Bodelsson, B. Jonson,
High-frequency oscillatory ventilation and an interventional lung assist device to treat hypoxaemia and hypercapnia  David M. , Heinrichs W.   British.
High PEEP in acute respiratory distress syndrome: quantitative evaluation between improved arterial oxygenation and decreased oxygen delivery  M. Chikhani,
Measurement of functional residual capacity by modified multiple breath nitrogen washout for spontaneously breathing and mechanically ventilated patients 
Comparison of pattern of breathing with other measures of induction of anaesthesia, using propofol, methohexital, and sevoflurane†  T.L. Strickland, G.B.
Perioperative assessment of regional ventilation during changing body positions and ventilation conditions by electrical impedance tomography  A. Ukere,
Influence of airway-occluding instruments on airway pressure during jet ventilation for rigid bronchoscopy  P. Biro, M. Layer, H.D. Becker, F. Herth,
Practical assessment of respiratory mechanics
Delivery of tidal volume from four anaesthesia ventilators during volume-controlled ventilation: a bench study†  G. Wallon, A. Bonnet, C. Guérin  British.
P.W. Buczkowski, F.N. Fombon, E.S. Lin, W.C. Russell, J.P. Thompson 
Improvement of lung mechanics by exogenous surfactant: effect of prior application of high positive end-expiratory pressure  A. Hartog, D. Gommers, J.J.
Respiratory impedance during weaning from mechanical ventilation in a mixed population of critically ill patients  J. Sellares, I. Acerbi, H. Loureiro,
Inflating LMA with normal saline: 14 yr after
U. Goebel, J. Haberstroh, K. Foerster, C. Dassow, H. -J. Priebe, J
L.M. Ferguson, G.B. Drummond  British Journal of Anaesthesia 
Recent advances in gas exchange measurement in intensive care patients
Protective ventilation in experimental acute respiratory distress syndrome after ventilator-induced lung injury: a randomized controlled trial  L. Uttman,
P. Michelet, A. Roch, D. Brousse, X. -B. D'Journo, F. Bregeon, D
Correction to the paper ‘Delivery of tidal volume from four anaesthesia ventilators during volume-controlled ventilation: a bench study’  M Hallbäck,
Effects of anaesthesia on paediatric lung function
New ventilators for the ICU—usefulness of lung performance reporting
H. J. Kim, H. -C. Lee, Y. S. Jung, J. Lee, J. J. Min, D. -M. Hong, E
Comparison of the effects of albumin 5%, hydroxyethyl starch 130/0
Use of an anaesthesia workstation barrier device to decrease contamination in a simulated operating room  S Hunter, D Katz, A Goldberg, H.-M. Lin, R Pasricha,
V. Uppal, G. Fletcher, J. Kinsella  British Journal of Anaesthesia 
B. Pearce, G.B. Drummond  British Journal of Anaesthesia 
Bispectral index is related to the spread of spinal sensory block in patients with combined spinal and general anaesthesia  R Iida, K Iwasaki, J Kato,
Tidal changes in PaO2 and their relationship to cyclical lung recruitment/derecruitment in a porcine lung injury model  D.C. Crockett, J.N. Cronin, N.
Accuracy of feedback-controlled oxygen delivery into a closed anaesthesia circuit for measurement of oxygen consumption†  A.W. Schindler, T.W.L. Scheeren,
Volume kinetics of glucose solutions given by intravenous infusion†
Delta down compared with delta pulse pressure as an indicator of volaemia during intracranial surgery  E. Deflandre, V. Bonhomme, P. Hans  British Journal.
An appropriate inspiratory flow pattern can enhance CO2 exchange, facilitating protective ventilation of healthy lungs  L.W. Sturesson, G. Malmkvist,
Jet or intensive care unit ventilator during simulated percutaneous transtracheal ventilation: a lung model study  Y.H. Liu, A.L. Wang, A.D. Marchese,
Pressure-controlled ventilation improves oxygenation during laparoscopic obesity surgery compared with volume-controlled ventilation  P. Cadi, T. Guenoun,
Β2-Adrenoceptor gene variants affect vasopressor requirements in patients after thoracic epidural anaesthesia  U.H. Frey, J. Karlik, F. Herbstreit, J.
P. Persson, O. Stenqvist, S. Lundin  British Journal of Anaesthesia 
M. Lichtwarck-Aschoff, A. Helmer, R. Kawati, M. Lattuada, U. H
High-frequency jet ventilation shortened the duration of gas embolization during laparoscopic liver resection in a porcine model  D. Fors, K. Eiriksson,
J Lee, S Gupta, C Price, A.P. Baranowski 
Chronic widespread pain, including fibromyalgia: a pathway for care developed by the British Pain Society  J. Lee, B. Ellis, C. Price, A.P. Baranowski 
Predictor of fluid responsiveness in the ‘grey zone’: augmented pulse pressure variation through a temporary increase in tidal volume  J.J. Min, N.-S.
Alveolar recruitment in acute lung injury
Remote measurement of the leak around the uncuffed tracheal tube: objective measurement and physical characteristics  A.J. Sims, K. Keltie, C.A. Reay,
Voltage changes in the lithium dilution cardiac output sensor after exposure to blood from horses given xylazine  T.D. Ambrisko, Y. Moens  British Journal.
S. Singaravelu, P. Barclay  British Journal of Anaesthesia 
R.L. Johnson, E.K. Cannon, C.B. Mantilla, D.A. Cook 
A recording of volume guarantee ventilation, running from left to right, showing 10 consecutive inflations. A recording of volume guarantee ventilation,
R. Parke, S. McGuinness, R. Dixon, A. Jull 
Presentation transcript:

Flow dynamics using high-frequency jet ventilation in a model of bronchopleural fistula†   M.J. Wood, E.S. Lin, J.P. Thompson  British Journal of Anaesthesia  Volume 112, Issue 2, Pages 355-366 (February 2014) DOI: 10.1093/bja/aet343 Copyright © 2014 The Author(s) Terms and Conditions

Fig 1 Schematic diagram of experimental set-up. British Journal of Anaesthesia 2014 112, 355-366DOI: (10.1093/bja/aet343) Copyright © 2014 The Author(s) Terms and Conditions

Fig 2 The effect of varying ventilator frequency on displayed and measured volumes in the intact artificial test lung (a) and cadaveric pig lung (b) models with inspiratory time (IT) 40% and driving pressure of 1.5 bar. All data presented as mean (sd). Each experiment repeated six times. Tidal volumes estimated by the jet ventilator, and directly measured entrained and expired volumes all decreased in an exponential manner as respiratory frequency increased in both artificial test lung (a) and cadaveric pig lung (b) models, with similar absolute values in both models. There was a marked difference between the measured expired volume and the tidal volume estimated by the HFJV ventilator over all the frequencies tested. In these intact lung models, there was no gas leak, which enables the injection pulse volume to be calculated (=VTE−ENT). The contribution of entrained volume becomes more important at lower frequencies and accounted for almost 50% of the measured expired tidal volume at frequencies <100 min−1. MACTV, tidal volume displayed by the high-frequency jet ventilator; ENT, measured entrained volume; VTE, measured expired volume; INJ PULSE, calculated injection pulse of HFJV; LEAK, measured leak volumes. British Journal of Anaesthesia 2014 112, 355-366DOI: (10.1093/bja/aet343) Copyright © 2014 The Author(s) Terms and Conditions

Fig 3 The effect of varying ventilator frequency on displayed and measured volumes in the porcine lung model, with a 2 and 10 mm BPF in proximal, middle, and distal positions. Data presented as mean (sd). Each experiment repeated six times. Tidal volumes estimated by the jet ventilator, and directly measured entrained and expired volumes all as respiratory frequency increased. Expired and displayed volumes were similar with a proximal, middle, and distal fistula (a–c). Entrained and leak volumes were relatively low with middle and distal fistulae (b and c), but comprised an increasing proportion of total tidal volume at low frequencies (a), despite expired volumes being similar to those with a middle and distal fistula, throughout the range of frequencies. With a 10 mm proximal fistula, no ventilation occurred and no measurements were possible. With middle and distal fistulae, there was a substantial decrease in expired tidal volume across all frequencies (d and e), with increasing leak volumes at low frequencies and entrained gas contributing an increasing proportion to total expired volume. Expired tidal volumes (VTE) were lower with a middle compared with a distal fistula. Note that the machine-displayed volumes are relatively constant irrespective of fistula size or position, and are much larger than the measured expired volumes, especially with a larger fistula at low frequencies. DP, driving pressure; MACTV, tidal volume displayed by the high-frequency jet ventilator; ENT, measured entrained volume; VTE, measured expired volume; INJ PULSE, calculated injection pulse of HFJV; LEAK, measured leak volumes. British Journal of Anaesthesia 2014 112, 355-366DOI: (10.1093/bja/aet343) Copyright © 2014 The Author(s) Terms and Conditions

Fig 4 The effect of altered driving pressure on 10 mm simulated distal (a–c) and proximal (d–f) fistula in the artificial test lung. Data presented as mean (sd). Each experiment repeated six times. Entrained, leak, and expired volumes all increased with increasing driving pressure in a distal fistula particularly at frequencies <150 min−1 (a–c). At a driving pressure of 2 bar with a distal fistula, expired volumes exceeded 400 ml at a frequency of <75 min−1 (c). With a proximal fistula (d–f), expired volumes increased at lower frequencies but were lower than with a distal fistula at lower frequencies. However, at a driving pressure >1.5 bar, leak volumes increased markedly and the entrained volumes exceeded the expired volumes for at 1.5 and 2.0 bar. Additional data for the middle fistula are available in Supplementary Figure S2. MACTV, tidal volume displayed by the high-frequency jet ventilator; ENT, measured entrained volume; VTE, measured expired volume; INJ PULSE, calculated injection pulse of HFJV; LEAK, measured leak volumes. British Journal of Anaesthesia 2014 112, 355-366DOI: (10.1093/bja/aet343) Copyright © 2014 The Author(s) Terms and Conditions

Fig 5 The effect of varying inspiratory time as a percentage of each cycle (IT %) on delivered and leak volumes in the artificial test lung with a 10 mm distal hole. Data presented as mean (sd). Each experiment repeated six times. There was little effect of alterations in IT on expired volumes at frequencies >200 min−1 with a 10 mm distal fistula (a–c). All volumes were increased at low frequencies with an IT of 40% compared with 20%. When IT was set at 60%, the proportion of entrained volume decreased compared with 20 and 40% IT, and the expired volumes were correspondingly reduced. At frequencies >250 min−1 and a 60% IT, the high airway pressures encountered caused the ventilator shut-off mechanism to halt ventilation and therefore no data were recorded. British Journal of Anaesthesia 2014 112, 355-366DOI: (10.1093/bja/aet343) Copyright © 2014 The Author(s) Terms and Conditions

Fig 6 The effect of varying IT (as a percentage of each cycle) on leak volumes in the artificial test lung at a proximal, middle, and distal sites, with a driving pressure of 1.5 bar. Data presented as mean (sd). Each experiment repeated six times. Changing IT had little effect on leak volumes with a distal fistula, particularly at frequencies >150 min−1 (a). However, leak volumes were highest with a 40% IT at frequencies <120 min−1 compared with 20 and 60% IT. A similar pattern occurred with a middle fistula, but the absolute leak volumes were greater, and increasing IT to 60% produced the highest leak volumes (b). With a proximal fistula (c), higher leak volumes occurred with all IT compared with middle and distal fistulae, but with an IT of 20% or 40%, remained relatively low at frequencies >200 min−1. British Journal of Anaesthesia 2014 112, 355-366DOI: (10.1093/bja/aet343) Copyright © 2014 The Author(s) Terms and Conditions