Ch 10. A System of Dosimetric Calculations

Slides:



Advertisements
Similar presentations
Image Reconstruction.
Advertisements

Beam Therapy Equipment 2 Linear Accelerator Head.
Pencil-Beam Redefinition Algorithm Robert Boyd, Ph.D.
The Tomotherapy Experience at Advocate Good Samaritan Hospital
Energy deposition and neutron background studies for a low energy proton therapy facility Roxana Rata*, Roger Barlow* * International Institute for Accelerator.
Photon Beam Monitor-Unit Calculations
Measurement of Absorbed Dose (6)
Sec.3 Practical application 临床处方剂量的计算.  What we should calculate?  Time or mu required to delivery proper dose for tumor  Dose distribution at the.
8.1 PRODUCTION AND CHARACTERISTICS OF X-RAYS
At the position d max of maximum energy loss of radiation, the number of secondary ionizations products peaks which in turn maximizes the dose at that.
Electron Beams: Physical Principles and Dosimetry
Chapter 5 Central Axis Depth Dose Calculations
Beam Therapy Equipment 3 Patient Treatment and Accessories.
Quality Control Rad T 110.
Dose Distribution and Scatter Analysis
Linac Beam.
Dosimetric evaluation of a new design MOSFET detector Per H. Halvorsen* & Stephanie Parker University of North Carolina.
Copyright © 2010 by Mosby, Inc., an affiliate of Elsevier Inc.
TWO FIELD BREAST PLAN VS. OPTIMIZED CONFORMAL BREAST PLAN: COMPARISON OF PLAN PARAMETERS Authors: Borko Basarić, Ozren Čudić, Milan Teodorović, Borislava.
Introduction Ion recombination is approximately corrected for in the Task-Group-51 protocol by P ion, which is calculated by a two-voltage measurement.
Научно-практический центр протонной лучевой терапии и радиохирургии (Москва-Дубна) A SYSTEM FOR MEASUREMENT OF A THERAPEUTIC PROTON BEAM DOSE DISTRIBUTION.
Radiographic Dosimetry
In vivo dosimetry Eirik Malinen Eva Stabell Bergstrand Dag Rune Olsen.
Application of a 2-D ionization chamber array for dose verification of dynamic IMRT with a micro-MLC Fujio ARAKI, PhD 1, S. TAJIRI 2, H. TOMINAGA 2, K.
Fundamental Dosimetry Quantities and Concepts: Review Introduction to Medical Physics III: Therapy Steve Kirsner, MS Department of Radiation Physics.
Medical Accelerator F. Foppiano, M.G. Pia, M. Piergentili
Araki F. Ikegami T. and Ishidoya T.
Radiation Quality Chapter 4. X-ray Intensity Intensity: the amount of energy present per unit time per unit area perpendicular to the beam direction at.
Part VIII:Medical Exposures in Radiotherapy Lecture 6: Determination of dose to the patient in Radiotherapy II IAEA Post Graduate Educational Course on.
بسم الله الرحمن الرحيم. Simulation of Photon Beam Interaction with the ZUBAL Head Phantom Presented by: Mostafa Hossni El-Soubki Plan-UNC-Image plotter.
Part VIII:Medical Exposures in Radiotherapy
Determining Radiation Intensity
Part No...., Module No....Lesson No
F. Foppiano, M.G. Pia, M. Piergentili
TLD POSTAL DOSE QUALITY AUDIT FOR 6MV AND 15MV PHOTON BEAMS IN RADIOTHERAPY CLINICAL PRACTICE Sonja Petkovska 1, Margarita Ginovska 2, Hristina Spasevska.
Interaction of x-ray photons (and gamma ray photons) with matter.
1 A protocol for the determination of absorbed dose from high energy photon and electron beams AAPM TG-21 Protocol (Med Phys 10: , 1983) 和信治癌中心醫院.
Commissioning of a commercial treatment planning system for IMAT and Dose Painting treatment delivery. G. Pittomvils 1,,L. Paelinck 1, F. Crop 2, W. De.
Monitor Unit Calculations for External Photon and Electron Beams
The Effects of Small Field Dosimetry on the Biological Models Used In Evaluating IMRT Dose Distributions Gene Cardarelli,PhD, MPH.
Rapid Arc Treatment Verification: post evaluation on Delta-4 and proposal of a new verification protocol G. Pittomvils 1,,L. Paelinck 1, T. Boterberg 1,
Date of download: 6/23/2016 Copyright © 2016 SPIE. All rights reserved. (a) Relative positions of the linear accelerator (LINAC) gantry, treatment region,
Adapting A Clinical Medical Accelerator For Primary Standard Dosimetry
AAPM TG-51 Protocol (Med Phys 26: , 1999)
Chapter 5 Central Axis Depth Dose Calculations. 2 Definition of Beam Geometry The accurate delivery of a radiation dose to patient depends on the precise.
SIMULATION CHECK LIST IMAGING PROTOCOL.
Mathematics for Computed Tomography
CHAPTER 3 DOSE DETERMINATION FOR EXTERNAL BEAMS
The physics of Radiation Therapy, pp
Electron Beam Therapy.
Khan, The Physics of Radiation Therapy, Ch7
Ch 7. Quality of X-ray Beams
Absolute Dose Measurement
A system of dosimetric calculations
Chapter-9 Dose Distribution And Scatter Analysis
A Brachytherapy Treatment Planning Software Based on Monte Carlo Simulations and Artificial Neural Network Algorithm Amir Moghadam.
Introduction Goal: Can we reconstruct the energy depositions of the proton in the brain if we are able to reconstruct the photons produced during this.
Above and below the diaphragm
Arghya Chattaraj, T. Palani Selvam, D. Datta
Resident Physics Lectures (Year 1)
Chapter 8 (Part 1) Measurement of Absorbed Dose
Resident Physics Lectures (Year 1)
Ch6 Measurement of Ionizing Radiation
Sheng Yang a, Yen-Wan Hsueh Liu b
Hot and cold spots are common problems associated with planning:
AL305-Physics II Meeting #2 Photon Beam Dosimetry
GRIDS PRESENTER-Dr.Vishwanath patil PG resident
Computed Tomography (C.T)
Antenna Theory By CONSTANTINE A.BALANIS Ch1.4, 2.1~2.2.4 O Yeon Jeong.
Principles and Practice of Radiation Therapy
Presentation transcript:

Ch 10. A System of Dosimetric Calculations The physics of Radiation Therapy, pp. 200 - 224 Ch 10. A System of Dosimetric Calculations

Introduction Dose Calculation Parameters Collimator Scatter Factor Phantom Scatter Factor Tissue-Phantom and Tissue-Maximum Ratios Practical Application Accelerator Calculations Irregular Fields Asymmetric Fields Other Practical Methods of Calculating Depth Dose Distribution Point Off-Axis Point Outside the Field

Introduction Limitation of using TARs, SAR & Percent depth dose for calculating absorbed dose in a patient Percent depth dose is suitable for SSD treatment technique. Tissue-air-ratios (TAR) suitable for SAD treatment technique, but limited to energies no higher than Co-60. Increase the size of chamber build-up cap Material of build-up cap is different from phantoms Overcome the limitation of the TAR Tissue-Phantom Ratio (TPR) Tissue-Maximum Ratio (TMR)

Dose Calculation Parameters Introduction Dose Calculation Parameters Collimator Scatter Factor Phantom Scatter Factor Tissue-Phantom and Tissue-Maximum Ratios Practical Application Accelerator Calculations Irregular Fields Asymmetric Fields Other Practical Methods of Calculating Depth Dose Distribution Point Off-Axis Point Outside the Field

Collimator Scatter Factor (Sc) The beam output measured in air depends on the field size Field size ↑; output ↑; collimator scatter ↑ “Output factor” Definition The ratio of the output in air for a given field to that for a reference field (10 x 10 cm) Direct measurement SAD AIR Reference field Build-up cap Reference field 1.0 ↓ Sc → Field Size

Phantom Scatter Factor (Sp) The change in scatter radiation originating in the phantom reference depth as the field side is change Definition The ratio of the dose rate for a given field at a reference depth (e.g. depth of Dmax) to the dose rate at the same depth of the reference field size (10 x 10 cm), with the same collimator opening Related to the change in the volume of the phantom irradiated

Phantom Scatter Factor (Sp) Indirect measuring Sp #1 through backscatter factor (BSF) BSF can be accurately measured for the photon beam (e.g. 60Co and 4 MV)

Phantom Scatter Factor (Sp) Indirect measuring Sp #2 Through total scatter factor (Sp) Contains both the collimator and phantom scatter SAD PHANTOM Reference field Reference depth Reference field 1.0 ↓ Sc,p → Field Size

Tissue-Phantom and Tissue-Maximum Ratios Definition of TPR The ratio of the dose at a given point in phantom to the dose at the same point at a fixed reference depth, usually 5 cm d t0 S × rd Dd Dt0

Tissue-Phantom and Tissue-Maximum Ratios Definition of TMR The ratio of the dose at a given point in phantom to the close at the same point at the reference depth of maximum dose Special case of TPR Adopted the point of central axis Dmax as a fixed reference depth

TMR data for 10 MV x-ray beams Properties of TMR Like TAR, TMR is independent of SSD, increases with energy and field size. TMR 3030 1010 00 Is caused entirely by the primary beam Depth in water TMR data for 10 MV x-ray beams

Tissue-Phantom and Tissue-Maximum Ratios Relationship between TMR and effective linear attenuation coefficient (μ) Obtain the effective linear attenuation coefficient (μ) Plotting μ(determined from TMR data) as a function field size Extrapolating it back to 0 × 0 field size

Tissue-Phantom and Tissue-Maximum Ratios Relationship between TMR and percent depth dose (P) Relationship between TMR and TAR

Tissue-Phantom and Tissue-Maximum Ratios

Scatter-Maximum Ratio (SMR) Definition The ratio of the scattered dose at a given point in phantom to the effective primary dose at the same point at the reference depth of maximum dose For Co-60 γ-rays SMRs are approximately at the same as SARs For higher energies SMRs should be calculated from TMR

Practical Application Introduction Dose Calculation Parameters Collimator Scatter Factor Phantom Scatter Factor Tissue-Phantom and Tissue-Maximum Ratios Practical Application Accelerator Calculations Irregular Fields Asymmetric Fields Other Practical Methods of Calculating Depth Dose Distribution Point Off-Axis Point Outside the Field

Practical Applications (accelerator calculations) Machine are usually calibrate to deliver 1 rad per MU At reference depth (t0) For a reference field size (10 × 10 cm) Source to calibration point distance of SCD

Practical Applications (accelerator calculations) SSD technique: Dose / MU = K dose rate under calibration conditions  Sc (rc)  Sp(r) field size changed  (SSD factor) distance (SSD) changed  PDD(d,r)/100 depth changed MU = TD / (Dose / MU)

Practical Applications (accelerator calculations ) example 1 SSD technique: Machine: 4 MV photons Calibration conditions: SSD = 100 cm, dmax = 1 cm, field size = 10  10 cm2. Calibration dose rate = 1 cGy / MU Treatment conditions: SSD = 100 cm, d = 10 cm, field size = 15  15 cm2, Sc(1515)=1.020, Sp(1515)=1.010, %DD=65.1, TD = 200 cGy. Dose/MU = 1  1.02  1.01  65.1 × 1 = 67.07 MU = 200  100 / 67.07 = 298

Practical Applications (accelerator calculations) example 2 SSD technique: Machine: 4 MV photons Calibration conditions: SSD = 100 cm, dmax = 1 cm, field size = 10  10 cm2. Calibration dose rate = 1 cGy / MU Treatment conditions: SSD = 120 cm, d = 10 cm, field size = 15  15 cm2, Sc(12.512.5)=1.010, Sp(1515)=1.010, %DD=66.7, TD = 200 cGy. Dose/MU = 1  1.01  1.01  [(100+1)/(120+1)]2  66.7 = 47.4 MU = 200  100 / 47.4 = 422

Practical Applications (accelerator calculations) SAD technique: Dose / MU = K dose rate under calibration conditions  Sc (rc)  Sp(rd) field size changed  (SAD factor) distance (SSD) changed  TMR(d,rd) depth changed MU = ID / (Dose / MU)

Practical Applications (accelerator calculation) example 3 SAD technique: Machine: 4 MV photons Calibration conditions: SCD = 100 cm, dmax = 1 cm, field size = 10  10 cm2. Calibration dose rate = 1 cGy / MU Treatment conditions: SAD = 100 cm, d = 8 cm, field size = 6  6 cm2, Sc(66)=0.970, Sp(66)=0.990, TMR(8, 66)=0.787, TD = 200 cGy. Dose/MU = 1  0.970  0.990  0.787 × 1 = 0.756 MU = 200 / 0.756 = 265

Irregular fields Calculation of percent depth dose for an irregular field Obtain average SMR by using Clarkson type integration SMR(d,rd) is then converted to TMR(d,rd) TMR(d,rd) may be converted into percent depth dose Final expression

Computer Program Data permanently stored in this computer program A table of SMRs as functions of radii of circular fields The off-axis ratios (Kp) The following data are provided for a particular patient Contour point outline of the irregular field The coordinate (x,y) of the point of calculation Reference point Patient measurements Patient thickness at various points of interest SSDs Source to film distance

As soon as a given area reaches its prescribed dose, it is shielded during subsequent treatments

can be assumed equal to symmetrical field Asymmetric Fields Independent jaw Field center positioned away from the true central axis of the beam Parameters changes after a field is collimated asymmetrically Collimator scatter Phantom scatter Off-axis beam quality Beam-flattening filters – greater beam hardening close to the central axis can be assumed equal to symmetrical field

Practical Applications (asymmetric fields) SSD technique: Dose / MU = K  Sc (rc)  Sp(r)  (SSD factor)  PDD(d,r)/100  OARd(x) SAD technique:  Sc (rc)  Sp(rd)  (SAD factor)  TMR(d,rd) MU = TD / (Dose / MU) MU = ID / (Dose / MU)

Asymmetric Fields OARd(x) off-axis ratio at depth d data are derived from cross-beam profiles measured at a number of depths in a phantom for the largest field available (e.g. 40 ×40 cm)

Other Practical Methods of Calculating Depth Dose Distribution Introduction Dose Calculation Parameters Collimator Scatter Factor Phantom Scatter Factor Tissue-Phantom and Tissue-Maximum Ratios Practical Application Accelerator Calculations Irregular Fields Asymmetric Fields Other Practical Methods of Calculating Depth Dose Distribution Point Off-Axis Point Outside the Field Point under a block

Irregular Fields Clarkson’s technique is not practical for routine manual or computerized calculation Time consuming Considerable amount of input data

Irregular Fields Approximate method Reasonably accurate calculations for most blocked field Effective field blocked field Sc Collimator field unblocked field, defined by the collimator Percent depth dose, TMR, Sp ×2 ×1

Point Off-Axis Clarkson’s technique is also not practical for manual calculation Day’s method PPD can be calculated at any point within the medium using the central axis data + a b c d P Q

Point Off-Axis The dose at depth d along the axis through Q 2a 2b 2c + a b c d P Q The dose at depth d along the axis through Q

Point Outside the Field ● a b c a c b

Points under a block a b c ● ‘t’ is the block transmission factor.

Thanks for your attention!