Tuesday Case of the Day Physics

Slides:



Advertisements
Similar presentations
S. Guilbaud Education Director School of Radiologic Technology
Advertisements

Technique Guidance Systems
Advanced Biomedical Imaging Lecture 4 Dr. Azza Helal A. Prof. of Medical Physics Faculty of Medicine Alexandria University.
Quality Control in Diagnostic Radiology
Quality Assurance and Digital Radiography
Dr. Mustafa Zuhair Mahmoud Mr.Ali B Alhailiy بسم الله الرحمن الرحيم 1.
QC/PACS Artifact Identification. Artifact Causes Image receptor/Hardware –Dirty –Foreign material –Failures Software –Algorithm –Enhancements/manipulations.
Technical Artifacts and Errors
 QC testing of screen speed should occur on acceptance and then yearly.  Evaluate first whether similar cassettes marked with the same relative speed.
Conventional and Computed Tomography
RADIATION PROTECTION IN DIAGNOSTIC AND INTERVENTIONAL RADIOLOGY
Factors affecting CT image RAD
ConcepTest 8.1Superposition a) b) c) d) If waves A and B are superposed (that is, their amplitudes are added) the resultant wave is.
Preparing Variable kVp Technique Charts By Prof. Stelmark.
Radiographic Equipment
More digital reading explaining LUT RT 244 Perry Sprawls, Ph.D. Professor Emeritus Department of Radiology Emory University School of.
The law of reflection: The law of refraction: Image formation
IFS prototype – PM3 LAM, 13/06/2003 Prototype testing at CRAL Tests at room temperature in the visible.
More digital 244 wk 12 Perry Sprawls, Ph.D. Professor Emeritus Department of Radiology Emory University School of Medicine Atlanta, GA,
Grid Performance By Prof. Stelmark.
IAEA International Atomic Energy Agency General Radiography Radiation Sources in medicine diagnostic Radiology Day 7 – Lecture 1(1)
Monday Case of the Day History: Multiple examinations from a CT scanner using a standard helical scan protocol show diffuse hypointense regions of artifact.
Monday Case of the Day History: During acceptance testing of a new general radiographic room, tube output measurements differed between the small and large.
Thursday Case of the Day What is the source of the artifact in the upper right-hand quadrant of the image? History: An AP view of the pelvis was acquired.
Substantial signal dropout was observed in arterial spin label (ASL) images, acquired at 3T in a young female patient three-months after mild traumatic.
Sunday Case of the Day History : Patient presented for screening mammogram. The radiologist noted that there “increased motion on both CC views” limiting.
Thursday Case of the Day The likely cause of the artifact is: A. Patient Motion B. Improper Cupping Correction C.Off-Focus Radiation combined with heel.
Tuesday Case of the Day What is the best solution to mitigate the artifact? A. There is no artifact, patient is wearing compression garment(s) which are.
Sunday Case of the Day How does the presence of the object impact patient dose? A. Increases by 20% - 50% B. Increases by 3% - 5% C. No Change D. Decreases.
Technique Guidance Systems By Prof. Stelmark. Anatomic Programming Anatomic programming, or anatomically programmed radiography (APR), refers to a radiographic.
Page 1 Physics Case of the Day - Thursday During routine QA testing of a computed tomography scanner, the physicist recorded the data below for the tube.
Diagnostic Radiology II X-ray Tubes. Anode angle Anode angle defined as the angle of the target surface with respect to the central ray in the x-ray field.
-42 Impact of signal non-repeatability on spectral CT images
THE X-RAY UNIT THE BASICS.
Electromagnetic Radiation Principles: Reflectance
Chapters 21 & 22.
Impact of Obesity on Medical Imaging and Image-Guided Intervention
Physics Case of the Day - Sunday
Diffraction FROM INTERFERENCE TO DIFFRACTION:
Sunday Case of the Day Physics
THE X-RAY UNIT THE BASICS.
Sunday Case of the Day Physics
Rad Tech 265.
Sunday Case of the Day Physics
Resident Physics Lectures
Tuesday Case of the Day Physics History: A B MIP
Quality Control Testing of Screen Speed
Sunday Case of the Day Category Physics Question:
Fluoroscopic Unit Thomas Edison 1896.
Resident Physics Lectures
12.1 – Characteristics of Lenses
Volume 88, Issue 4, Pages (April 2005)
Tuesday Physics Case of the Day
Monday Case of the Day Physics
QC And NEMA In The Nuclear Arena
Challenges with Digital Imaging
Wednesday Case of the Day
T. J. Okamoto (NAOJ/Kyoto Univ.)
Fraunhofer Diffraction
Resident Physics Lectures
Resident Physics Lectures
Resident Physics Lectures
Fluoroscopic Image Intensifier Image production
Artifacts and Errors.
Thursday Case of the Day
Fluoroscopic Image Intensifier Image Production
Intrinsic Uniformity QC Setup
Brightness Control Systems
Optimization of the fixed-flexion knee radiograph
RAD- PATHOLOGY TISSUE DENSITIES Asilah Al-Obeidani DEPARTMENT OF MEDICAL IMAGING.
Presentation transcript:

Tuesday Case of the Day Physics Authors: Laurence Parr, MS, DABR1 and Charles E. Willis, PhD, DABR2 1Department of Radiology, Naval Medical Center Portsmouth, VA 2U.T. M. D. Anderson Cancer Center, Houston, TX History: Radiolucent (dark) artifacts were noted by radiologists on images acquired using several mobile DR x-ray units. In addition, embedded accelerometers indicated the detectors may have been dropped. Normal window Narrow window Pediatric KUB 70 kVp 5 mAs Adult Chest 110 kVp 6 mAs Figure 2. Uniform exposure images showed numerous similarly sized bright and dark parallelogram-shaped defects. The likely cause of the artifact is: Detector damage Objects in the imaging path unrelated to the detector Out-of-date correction map Incorrectly obtained correction map Figure 1. Examples of the artifact (red arrows) are representative of those seen on all images.

Findings: The detector was recalibrated according to the manufacturer’s directions. 1. After recalibration, a uniform exposure image acquired without moving the detector showed no artifacts (Fig. 3). 2. A second uniform exposure image acquired after rotating the detector 180º showed numerous increased (dark) and decreased (light) signal artifacts (Fig. 4). Handle Edge Handle Edge Figure 3: Uniform Exposure – calibration orientation Figure 4: Uniform Exposure – rotated180º

Findings: (continued) 4. The correction map image (Fig. 5) showed focal regions of increased correction values that correlated with the increased signal artifacts (dark) on the rotated image (Fig. 6). 5. The decreased signal artifacts (light) did not correlate with the correction map. 6. Radial symmetry through the center of the 180º rotated image was noted between the dark and light artifacts (Fig. 6). Handle Edge Handle Edge Figure 5: Correction Map – rotated 180º Figure 6: Uniform Exposure – rotated180º

Diagnosis: B. Objects in the imaging path unrelated to the detector

Discussion: Reappearance of the artifacts after calibration when the detector was rotated indicates the source of the artifact is not fixed in position relative to the detector. If the source of the artifact had been fixed in position relative to the detector, recalibration of the gain map would compensate and the artifact would not reappear when the detector was rotated. Physical damage to the detector would have resulted in artifacts that maintained a fixed orientation with respect to the detector. Therefore, choice A. Detector damage is incorrect. Handle Edge This light artifact is an area of low exposure on the detector at the time of imaging where uniform exposure existed at the time of calibration. This dark artifact is over-correction at the time of imaging in an area of the detector where low exposure existed at the time of calibration. Recalibration only corrected the problem so long as the detector orientation was maintained. Therefore, choice C. Out-of-date correction map is incorrect. Calibration produced a gain correction map that was appropriate for the conditions of exposure, as demonstrated by the initial disappearance of the artifacts. Therefore, choice D. Incorrectly obtained correction map is incorrect. Several small lead filings were found on the collimator exit window of the x-ray generator. After the filings were removed and the detector was recalibrated, the artifacts no longer appeared in any detector orientation. Figure 7: Light artifacts are focal spot images from the inverse pinhole effect. Dark artifacts are over-compensation in the gain correction map where focal spot images were present during calibration.

Discussion: The artifacts are, in fact, images of the focal spot of the x-ray generator, by means of the inverse pinhole effect. The artifacts have the typical “double-banana” shape of a focal spot image. Figure 8: Slit camera image of the focal spot (Jones 2008). A small amount of high Z material in the x-ray beam path, such as the Pb filings in the collimator exit window, can produce the image of the focal spot in the image receptor plane. This artifact was first reported in diagnostic radiology by Poznanski (1969). Figure 9: Xeroradiograph at 60 kVp and 40 mAs of an inverse pinhole array. The inverse pinholes are tungsten from 75 -300 micron in diameter arrayed in ten rows from left to right on a sheet of lucite (Cowart 1976). Note the distortion of the focal spot projection as lateral distance increases from the central ray of the x-ray beam.

References/Bibliography: Cowart, RW. An investigation of the inverse pinhole camera. Thesis. University of Texas Health Science Center at Houston. Graduate School of Biomedical Sciences. Houston, TX. June 1976. 111 pages. Jones, AK. Personal communication. 2008. Poznanski, AK. Focal spot artefacts on breast radiographs. Radiology 92: 644. 1969.