New and Old Problems with Anesthesia Equipment BWH Anesthesia Grand Rounds July 9, 2003 © Copyright 2002-2003, James H Philip, All rights reserved.

Slides:



Advertisements
Similar presentations
Trouble Shooting (Mechanical Ventilation)
Advertisements

SECTION 11 LEVEL B SUIT OUT. SECTION 12 AIR MONITORING.
Next Home Visio2 troubleshooting guide Start the slide show then click Next.
Training Presentation
Positive pressure ventilation Ventilation of the lungs single most important and most effective step in cardiopulmonary resuscitation of the compromised.
Safety Requirements of the Anesthesia Workstation
Anesthesia Machine Presented by Gil Soto C.R.N.A.
INTERMEDIATE PRESSURE SYSTEM PRESENTED BY: Dr CHITTRA MODERATED BY: Dr JYOTI PATHANIA.
Flight 60 vs. HT Flight Medical Confidential 1.
Refrigerators.
Presented by: Dr. manoj kumar Moderated by: Dr. Aparna Sharma
Boussignac CPAP System
3.1.9 Student Book © 2004 Propane Education & Research CouncilPage Replacing Valves and Fittings in ASME Tanks Replacing ASME tank valves and fittings.
1 FUEL SYSTEM SERVICE ENGINE CLEAN FUEL TOOL MODEL RX 777.
Anesthesia Apparatus Checkout Recommendations
Inhaled Anesthetic Delivery Systems
1 DIESEL FUEL TOOL ON VEHICLE DIESEL FUEL SYSTEM SERVICE MODEL Rx HD 888 INSIDE FILTER.
Check Valves and Manufacturer-Recommended Leak Test
Next Home EcoCompact troubleshooting guide Start the slide show then click Next.
Clean Air Box Usage & Care CABs RAMs CO Monitors.
Clinical Conference 15 Dec 04
Suction Pumps.
Clinical Conference: Technology Rounds Biomedical Engineering Evelyn Fan, M.H.Sc., Clinical Engineer Biomedical Engineering November 2, 2005.
Regulators.
 The annular space. The clearance between the head of the float and the flow tube is known as the annular space.
© , James H Philip, all rights reserved
Technology Topics 2005 CA1 - CA40 focus BWH Anesthesia Clinical Conference August 31, 2005 James H. Philip, MEE, MD, CCE © Copyright 2005, James H Philip,
Clinical Conference Anesthesia Machines don ’ t fail, do they? James H. Philip M(E)E, MD Brigham and Women’s Hospital October 30, 2002 © Copyright 2002,
PART 3: Breathing Circuit
Parts of the Anesthetic Machine
ENGINE CLEAN OIL TOOL OIL RX • 999 TRAINING
Anesthesia Machine VAPORIZERS. Vaporizers Convert liquid anesthetic into a volatile inhalation agent Based on laws of physics You must memorize the chemical.
A). When the carrier gas is quickly switched to 100% nitrous oxide (B). the halothane concentration decreases to 3% within 8 to 10 seconds A new steady-state.
Dr. S. Parthasarathy MD., DA., DNB, MD (Acu), Dip. Diab. DCA, Dip. Software statistics PhD(physiology) Mahatma gandhi medical college and research institute,
Scavenging System A&A Pg 356  To reduce exposure to waste anesthetic gases  Can be active or passive scavenging system Active Scavenging System  Connected.
To Menu.
Trouble Shooting (Mechanical Ventilation) Arthur Sadhanandham Medical ICU, CMC.
SERVO-i WITH HELIOX OPTION
CENTRAL GAS SYSTEMS Without reference, identify principles relating to central gas systems as they pertain to facility and equipment interface with at.
Domina PLUS B WORKING DIAGRAMS TROUBLESHOOTING INTERNAL VIEWS WIRING DIAGRAMS INSTALLATION 03 PREVIOUS MODEL.
Waste Anesthetic Gases. The anesthetic gas and vapors that leak out into the surrounding room during medical and surgical procedures are considered waste.
Respiratory Safety Public Safety Murray State University.
EVAPORATIVE EMISSION CONTROL SYSTEMS
Spill and Overfill Prevention Alaska Spring 2004.
Without reference, identify principles about Anesthesia Units with at least 70 percent accuracy.
PART 3: Breathing Circuit
To Menu.
Anesthetic Problems and Emergencies A&A pg Why Do Problems Arise?  Human error  Equipment error  Adverse effects  Patient factors  Anesthetic.
Evaporative Emissions Control Systems (EVAP)
CHAPTER Fuel Pumps, Lines, and Filters 22 Copyright © 2016 by Pearson Education, Inc. All Rights Reserved ADVANCED ENGINE PERFORMANCE DIAGNOSIS Advanced.
Subacute Care Chapter 25 Subacute Care Care for Residents With Specific Needs Formerly cared for in Hospital Rehabilitation Complicated Respiratory Care.
Parts of the Anesthetic Machine A&A Pages
Automotive Engines: Theory and Servicing, 7/e By James D. Halderman Copyright © 2011, 2009, 2005, 2001, 1997 Pearson Education, Inc., Upper Saddle River,
STEP 1 Locate the transmission cooler line connection on the vehicle. Identify the correct adapters to be used by referring to the vehicle application.
7 A/C System Components, Operation, and Service.
Ventilators for Interns
Agent Monitoring For CEs and BMETs. Ready Anesthetic Agent Monitoring James H. Philip, M.E.(E.), M.D. Anesthesiologist and Director of Technology Assessment.
PART 3: Breathing System Parts
TXV Devices.
Simplified Instruction for use* Evita V300
Selector switch of the Datex-Ohmeda Aestiva 5 machine
Non-Rebreathing System
Opti-Purge® FAST INFLATION PURGE SYSTEM
Simplified Instruction for use* Evita V300
T-6B On-Board Oxygen Generation System (OBOGS)
SB Packing Automatic Downpuller Power Unit
Non-Rebreathing System
The Anaesthesia Machine
Automotive Technology Principles, Diagnosis, and Service
T-6B On-Board Oxygen Generation System (OBOGS)
Presentation transcript:

New and Old Problems with Anesthesia Equipment BWH Anesthesia Grand Rounds July 9, 2003 © Copyright 2002-2003, James H Philip, All rights reserved.

Equip Probs JHP Ready

If you stay you will learn how and why the following are true Suction can fail 11 different ways There are many ways a vaporizer can fail O-rings cause many failures New Tec7 vaporizer needs special filler adapter and must be pressed to fill Checkout includes Ventilator Checkout includes “Admit Patient” to reset alarms Use an HME to keep the circuit dry and ventilator and gas sampling working Gas Sample Filters are cracking. Save all that crack to be sent to Manufacturer Free-hosing with Forced Hot Air (e.g., Bair Hugger) is dangerous - don’t do it Draeger Fabius moos and can have negative (sucking) leaks A shorter checkout is possible - Should we post it? Should we do it? ECG Filter must not be set to Maximum if want to monitor S-T, T waves Problem Solving - Clinical vs Technical; Treat signs; Find & Treat Root Cause

Block on: Problem Solving in Anesthesia July 9, 2003 7 - 8 AM New & Old Problems with Anesthesia Equipment James Philip, M.D. 8 - 9 AM Hypotension George Topulos, M.D.

New and Old Problems with Anesthesia Equipment James H New and Old Problems with Anesthesia Equipment James H. Philip MEE MD CCE Anesthesiologist Medical Liaison for Partners Biomedical Engineering Director of Technology Assessment for BWH

Biomed Intro

Failed Suction Patient is a 29 yo F receiving general anesthesia for elective gynecologic surgery. Immediately after loss of consciousness, the patient vomits You place the Yankauer Suction in patient’s mouth and occlude the vent hole But, nothing happens What is going on?

Suction can fail in many ways DISS (Diameter Index Safety System) Connector Unscrewed at Wall

Suction can fail in many ways DISS (Diameter Index Safety System) Connector Unscrewed at Wall Vacuum Switched Off at DISS Connector

Canister problem Hose disconnect from canister Canister switched off Insert seated incorrectly Any port or hole open Hose disconnect from holder Tubing disconnect from canister Flap valve closed Canister Tilted Canister Full Hose kinked

Suction Solution Test Function Before Case During Case

New Vaporizers have New Failure Modes

Tec7 Vaporizer by Datex-Ohmeda Differences from previous White, not black Still Agent-specific for: Halothane, Enflurane, Isoflurane, Sevoflurane Wider flow range = .2 L - 15 LPM Non-spill system limits movement of liquid if tilted or inverted No periodic service required ( Tec 4 = 1 yr, Tec 5 = 3 yr, Tec 7 no) Easy-fil™ System 50% faster fill than previous (Tec 3,4,5) Low-leak fill Intuitive alignment (they say) Soldered sump and seal (not O-rings)

Press to overcome spring Easy Fil™ No Spill Press to overcome spring Auto Fill ? No

Which is filling as we watch? Desflurane Tec6 Semi-Automatic Fill (no patent pending)

Why no O-rings? O-rings fail Space Shuttle Forreger Anesthesia Machines O-Rings sealed a three-vaporizer turret Failed during drug administration Simultaneous delivery of multiple overdoses Last product Forreger ever made Desflurane bottle fill

The O Ring belongs here

The O-Ring is missing

This is filling with no O-ring

When you change a vaporizer

When you change a vaporizer You are expected to do a Negative Pressure Leak Test [See FDA Pre-Use Check Card attached to every Anesthesia Machine]

Negative Pressure Leak Test

Negative Pressure Bulb is in Drawer If not, Page your Anes Tech Negative Pressure Bulb is in Drawer

Condensed water vapor Makes Ventilators fail Makes Gas Monitors fail Use a Heat and Moisture Exchanger

Wet dome valve - Ohmeda Mod 2

Wet dome valve - Aestiva

Fabius flow sensor gets wet, too

Wet ET tube, dry circuit

Datex-Ohmeda Edith® is BWH present HME product

Gas Line Filters sometimes crack Save them if they do crack Page Anes Technician who will bring it to ORCSS Materials Management” to return to Manufacturer For analysis

Forced Hot Air warming Use a Blanket Do not use hose without Blanket FDA, ECRI, Manufacturers are appropriately concerned Blanket costs around $10.00 This is not the right place to save money

Fabius idiosynchrocies Moos like Cow Temporary problem which will be solved If reservoir bag is empty or has a leak Air will be drawn into breathing circuit during mechanical ventilation

Anesthesia Machine Checkout Important Will pick up many problems Common root cause of mishap is: Failure to Pre-Check Equipment

Check Ventilator Function as part of Check Out Ventilate reservoir bag Observe no volume loss Added recently to our list

Admit “New Patient” or re-boot Monitor and Ventilator Returns alarms and configuration to Site-Standard Settings (BWH) Failure to do this has led to several reported “events” Added recently to our list

Placard on every Anes Machine Should we also post abbreviated checkout?

Abbreviated Machine Check Out Calibrate Oxygen Monitor Check Machine for gas delivery Check Circuit for integrity Check Ventilator for function Check Exhaust for function

ECG Monitor filter must not be set to Maximum Monitor offers several bandwidth options All faithfully convey ST and T waves except Maximum Filter This keeps wave on screen but Distorts S-T segment and T waves ST and T are “low frequency” “Maximum” filter removes (ruins) these Bandwidth Definitions Diagnostic = .05 - 120 Hz Monitoring = .05 - 40 Hz Moderate = .05 - 25 Hz Maximum = .5 - 25 Hz - the only really bad one

Troubleshooting a “Clinical” Problem Phase 1 – Patient or Equipment Problem ? This differentiation is crucial Starting on the wrong track is a bad start Phase 2 – Treat Signs and seek Root Cause Phase 3 – Treat or correct Root Cause

Thank you

If you stayed you learned how and why the following are true Suction can fail 11 different ways There are many ways a vaporizer can fail O-rings cause many failures New Tec7 vaporizer needs special filler adapter and must be pressed to fill Checkout includes Ventilator Checkout includes “Admit Patient” to reset alarms Use an HME to keep the circuit dry and ventilator and gas sampling working Gas Sample Filters are cracking. Save all that crack to be sent to Manufacturer Free-hosing with Forced Hot Air (e.g., Bair Hugger) is dangerous - don’t do it Draeger Fabius moos and can have negative (sucking) leaks A shorter checkout is possible - Should we post it? Should we do it? ECG Filter must not be set to Maximum if want to monitor S-T, T waves Problem Solving - Clinical vs Technical; Treat signs; Find & Treat Root Cause

Requests from Anes Dept Negative Pressure tester in each room Techs do the negative pressure test when they change Vap Techs do circuit pressure test when they connect circuit Aquanaut in OR where used - likely Anes Machine Drawer 5% Sevo alarm default on RGM - Increase it if possible

End Here

New Resources - on Anes Server or CD From Anesthesia Clinical Bioengineering Elective Residents Brain Monitor Physiometrix PSA4000 - Maurizio Albala MD Brain Monitor Aspect BIS - Maurizio Albala MD GE (Marquette) 9500 Monitor - John Cabral MD Ohmeda RGM Monitor - Jason Ryan MD Ohmeda Tec5 Vaporizer in pieces at Sea Level & Altitude Jason Ryan MD Aestiva Check-Out - Morana Lasic MD Ventilators - Sascha Beutler MD Respiratory Mechanics Module - Eveline Ehrl MD Gas Sampling systems - Jasline Dhingra MD

End

Whom to Ask James H. Philip, MEE, MD Anesthesiologist and Director of Bioengineering (Anesthesia) Office PBBH - B Ground 732-7330 BB1615) jphilip@zeus.bwh.harvard.edu