HOW I DO CMR SCANNING SAFELY

Slides:



Advertisements
Similar presentations
MRI Safety Training.
Advertisements

CMR Patient Safety and Preparation Kraig V. Kissinger, BS,RT(R)(MR) Beth Israel Deaconess Medical Center Boston, Massachusetts.
YOUR GUIDE ON MRI.
RESEARCH MRI Safety Training
1 of 24 CCS&BS MR Safety SeminarCenter for Complex Systems and Brain Sciences Magnetic Resonance Safety Seminar: An Introduction to the Potential Hazards.
MRI Safety This slide show will educate all healthcare providers about the safe delivery of care in the magnetic resonance (MR) environment. This slide.
RESEARCH MRI Safety Training
MRI Safety. Magnet Type  The MRI magnet used at MCH is the superconducting type. Superconducting magnets use a special wire which loses all resistance.
MRI in patients with pacemakers DR PRADEEP SREEKUMAR.
Handling Liquid Cryogens Safely Tony Kent School of Physics and Astronomy.
Restricted © Siemens AG 2013 All rights reserved.siemens.co.uk/education Topic 15: A peep inside Siemens Education.
MRI Safety Jeopardy* ForcesProjectilesPotpourriHeatingScreening Because when you think MRI safety, think * Jeopardy is a trademark of Sony.
Equipment Magnetic resonance imaging (MRI) scan requires the use of a very strong magnetic field. Unlike other devices used in radiology, MR imaging.
MRI Safety Jeopardy* EnteringMagnetsPotpourriMetalScreening Because when you think MRI safety, think * Jeopardy is a trademark of Sony Digital.
Overview The scanner Dangers in around the scanner In case of an emergency Emergency numbers Emergency procedure Reporting adverse events Safety checklists.
FMRI: Biological Basis and Experiment Design Lecture 18: Physical practicalities Digression: analysis ICE9: Example for WA8 Safety limits –dB/dt –SAR –Acoustic.
MRI Physics 3: Hardware Douglas C. Noll Biomedical Engineering University of Michigan.
Magnetic Resonance Imaging Mary Holleboom ENGR 302 May 7, 2002.
MRI General Safety Rules. GENERAL RULES DO NOT ENTER the MRI Scan Room if the door is closed. Always see the MRI Technologist before entering room. Always.
THE MAGNETIC FIELD IS AT FULL POWER ALWAYS …. IT IS ALWAYS “ON”.
Magnetic Resonance Imaging
Introduction This course is designed to inform individuals of the necessity of Magnetic Resonance Imaging (MRI) safety. Any individuals that are in contact.
MRISC Organizational Chart Research and Graduate Studies Dr. James Tracy College of Medicine Fred de Beer, M.D. Director Dr. Charlie Smith Administrative.
MRI SAFETY FOR NON-MRI PERSONNEL.
MRI Safety Wendy Winkle
Biomagnetism By Dr. Amr A. Abd-Elghany. Magnetism Magnetism is a fundamental property of matter; it is generated by moving charges, usually electrons.
Institute of Neurological Sciences, Southern General Hospital, Glasgow
MRI Safety Course Module 1 –Patient Screening Learning Objective #1 The student will be able to describe in detail the patient screening process, contraindications.
DEVELOPMENT OF MRI PREPARED BY ZAINAL AHMAD KEU SAKINAH HASHIM KEU
Physical Science Chapter 8: Magnetism. A magnet is a device which attracts iron or other magnets, and produces a magnetic field around it’s body. The.
Magnetic Resonance Imaging
Nuclear Magnetic Resonance Imaging By Blake Sharin.
Diathermy Interactions with Implanted Leads and Implanted Systems with Leads Marian Kroen Office of Surveillance and Biometrics Issues Management Staff.
The UE directive 2004/40/EC and Magnetic Resonance Imaging a debate Thierry Metens Ph D Université Libre de Bruxelles, Hôpital Erasme Brain awareness week.
Overview of Neurostimulation
By Ass.Prof.Dr. Gehan Mosad
Radiation. Ionising Radiation Alpha Radiation Beta Radiation Gamma Rays X-Rays Neutrons.
MRI DRAFT Principles of MRI ► Sectional Anatomy ► Provides anatomic and physiologic info ► Non- invasive ► No radiation  Magnetic fields and.
MRI Safety Course Module 2- Emergency Situations.
Magnetic Resonance Imaging Dr Mohamed El Safwany, MD.
ELECTRODES FOR STIMULATION. Cardiac Pacemakers and Defibrillators  The heart is a natural example of a critical Bioelectromagnetic system.  An electrocardiogram.
MRI Safety Course Module 5- Magnetic Fields and Radiofrequency.
Human Subjects in fMRI Research Credits: Robert Savoy, Ph.D. Franz Schmitt, Ph.D.
Magnet Safety.
Pacemakers and MRI An outdated contraindication? Emily Myers Research in Allied Health East Tennessee State University.
Magnetic Resonance Imaging
Please Be Courteous To Others
Magnetic Resonance Imaging (MRI). The Components: A magnet which produces a very powerful uniform magnetic field. A magnet which produces a very powerful.
NMR Safety Powerful magnets. NMR Safety 2 NMR Safety Information. UMBC Dept. of Chemistry and Biochemistry. Accessed 7 Jan 2014.
 A medium that provides a means for transferring energy from one place to another  It is not expensive or difficult to control the risk of an electrical.
 This depends on a property of nuclei called spin.  Gyroscope: Principle: As long as its disc remains spinning rapidly the direction of the spin axis.
MAGNETIC RESONANCE IMAGING by PRADEEP V.EPAKAYAL. Mem.no L.
©ECRI XXVI. Imaging Equipment TRAINING SEMINAR ON MEDICAL DEVICE ACCIDENT INVESTIGATION for Kingdom of Saudi Arabia Saudi Food & Drug Authority.
Microwave Antennas and Propagation.
NHS Dumfries & Galloway MRI Unit
MRI Safety Awareness The purpose of this healthstream presentation is to educate the hospital staff that do not routinely come into the MRI environment.
Methodist Hospital for Surgery
Physical Science Chapter 8: Magnetism.
Biomedical Engineering Implants
Clinical engineering Lecture (8).
Jody Culham Brain and Mind Institute Department of Psychology
MRI Student orientation
How to set up a Cardiac MR Examination
Magnetic field safety October 2018.
Basics of MRI By Dr. Mohamed Samieh.
Human Subjects in fMRI Research
Center for Magnetic Resonance Research
Residential exposure to EMFs
Hoglund Brain Imaging Center
Presentation transcript:

HOW I DO CMR SCANNING SAFELY Elisabeth Burman Research Sister Royal Brompton Hospital, London UK This presentation is posted for members of scmr as an educational guide – it represents the views and practices of the author, and not necessarily those of SCMR. 2011 update: some updates added to terminology, MR conditional devices, and contrast with renal failure. JMoon

Magnetic Resonance One of the safest imaging modalities but there are still safety issues Advantages Non-invasive No ionizing radiation Repeat studies present no problem Fewer ethical problems imaging volunteers Why are there safety issues? Potential for accidents and even fatalities Damage to equipment / cost

Safety issues centre around the 3 electromagnetic fields and issues indirectly related Static magnetic field biological effects mechanical effects Gradient Field induced currents (PNS) auditory damage RF Field RF power deposition Burns Other issues contrast agents critically ill patients/resuscitation monitoring quench claustrophobia

Static magnetic field Measured in Gauss or Tesla (10,000G equivalent to 1T) 1.5 T is 30,000 x the strength of earth’s magnetic field

Static Magnetic Fields A superconducting magnet is always switched on! 1. Biological effects (potential risk) - exposure to static magnetic fields of up to 4T are not thought to be harmful Biological effects relevant to clinical imaging - distorted ECG (magnetohydrodynamic effect) - consider prudency with pregnancy 2. Mechanical effects (very real risk) - translational or attractive forces on metallic objects when brought into the field

Static Magnetic Field Biological effects Magnetohydrodynamic effect – augmented T wave Outside field 0.5 Tesla 1.5T

Magnetohydrodynamic effect seen as augmentation of T-wave Caused by the effect of the static magnetic field on moving blood (systole) as a conducting fluid. The gradient and RF fields also affect the configuration of the ECG Morphological ECG changes are therefore difficult to detect and diagnose, but rhythm is usually recognised Any concern regarding rhythm, remove patient from scanner and perform 12 lead ECG

Static Magnetic Field Pregnancy Patients 1st trimester – prudent to avoid CMR where possible 2nd and 3rd trimester – decision made on a risk versus benefit determination. For example if it avoids the patient being subjected to x-rays. Health Care Workers May enter MR scanning room regardless of trimester Should not remain in the room when scanner is operational, avoiding exposure to gradient and radiofrequency fields

Static Magnetic Field Mechanical effects Projectile or missile effect - the attractive forces exerted by the static magnetic field present the greatest potential for patient injury - objects will be pulled out of hands, pockets etc, and fly into magnet which has caused injury and death. Effect on ferromagnetic implants - electro-mechanical eg pacemakers - biomedical eg valves, stents

What is typically ferromagnetic? EQUIPMENT PERSONAL ITEMS (leave outside) Oxygen cylinders Keys, pens Wheelchairs Bleeps Trolleys Mobile phones IV stands Coins Monitoring equipment Stethoscopes Ventilators Scissors It is easy to forget objects, particularly when responding to an emergency! Remember also – magnetic strips will be wiped (credit cards)

Warnings to minimise accidents? OUTSIDE THE SCANNER ROOM Signage Barrier cords to scanner room Restricted access - swipe card / coded entry system INSIDE THE SCANNER ROOM (fringe field – 5 Gauss line) Older systems - passively shielded - extensive fringe field gave a warning Newer systems - actively shielded - small fringe field greater hazard because there is no gradual warning field strength increases hundreds of gauss in a few inches

Never be complacent – accidents do happen! Oxygen cylinder Infusion pump

Floor buffer

“Contraindications” to CMR* Implants & metal Electromechanical implants Cerebral aneurysm clips Metallic foreign body in the eye Shrapnel, bullets (in critical area) Ocular implants (containing metal) Swan-Ganz Pacemakers /ICD’s Pacing wires Cochlear implants Neurostimulators Hydrocephalus shunts Any device electrically or mechanically activated * In some circumstances MR has been performed despite “contraindications”, especially in specialist centres. Work is in progress to make some devices safe. For example, some cerebral aneurysm clips may now be scanned.

Implants Cerebral clips Foreign bodies modern clips are considered safe (titanium, elgiloy) older clips (martensic steel is highly ferrous) obtain operation notes with serial number of clip radiologist to consent patient seek advice from specialist centre Foreign bodies Maybe situated near vascular or nervous tissue If in doubt – x-ray

Pacemakers Highly specific guidelines are required General rule has been “absolute contra-indication” All 3 electromagnetic fields effect pacemaker function Non-pacemaker dependent patients have been scanned in specialist centres in exceptional cases Future: new pacemaker/lead designs may be MRI safe **2011 update: MR-conditional Medtronic pacemaker and leads now available

Potential effects of MRI on pacemakers excessive heating / induced currents in the pacemaker lead temporary or permanent modification of pacemaker function inappropriate sensing or triggering of the device movement of the pulse generator or lead pacing the heart

What can be safely scanned at 1. 5T What can be safely scanned at 1.5T? Lists of approved implants are available BIOMEDICAL IMPLANTS CARDIAC sternal wires heart valves & annuloplasty rings coronary stents epicardial wires (cut short and taped) other vascular stents, coils and filters If unsure, best to check at – check: www.mrisafety.com occlusion devices catheters (without guidewire) OTHER orthopaedic implants contraceptive devices dental implants Effects of implants on imaging - signal void / local distortion

Safety Checklist – comprehensive but concise Essential to know Important to know Cardiac pacemaker? Previous neurosurgery? Implants or metal in the body? Pregnant (prudent approach)? Drug patch with foil backing? Previous heart surgery? Diabetic or epileptic? Asthmatic or allergies? (in relation to contrast) Tattoos or permanent eye liner (iron oxide)? Essential preparation Removal of accessories - watch, jewellery (except wedding rings), body piercing rings, hearing aids, glasses, false teeth, artificial limbs and prostheses Removal of clothes containing metal eg zips, bras

Preparation of patient Remove watches / jewellery except gold wedding rings Remove hearing aids, false teeth, glasses, prostheses Remove all clothes except socks and underpants Patient gowns - no pockets - no metallic fastenings - ¾ length sleeves for IV access - wrap round for easy chest access Brass changing room keys Screen all accompanying personnel Check any suspicious item with small bar magnet

Gradient fields - Induced Currents Gradient fields induce an electric field and thus a current in the patient, potentially this can be of sufficient intensity in modern systems to produce a physiological response - peripheral nerve stimulation (PNS) - cardiac stimulation is not considered possible

Gradient fields – Auditory damage Hearing protection mandatory above 90dB time averaged for: patients staff remaining in the scanner room relatives accompanying children or patients

Radiofrequency (RF) fields Thermogenic effects - health & safety concern Physiological tissue heating response most of the transmitted RF power is transferred into heat within the patient’s tissue all MR systems have safety thresholds to avoid dangerous levels Patients with compromised thermoregularory systems are at greatest risk

Specific Absorption Rate (SAR) SAR is the RF power absorbed per unit mass of tissue (expressed in W/kg) complex function of numerous variables calculated by software from the average forward power passing into the RF transmitter coil and the body mass situated in the RF transmitting field Therefore an accurate patient weight is vital SAR increases 9 fold from 0.5T to 1.5T

Radiofrequency (RF) fields Potential for burns 1°, 2°, 3° burns have occurred in the past in patients undergoing MRI This is a result of excessive heat developing in the devices or objects ECG system is often the culprit Interventional MRI poses greater risk

Prevention of burns Electrodes - carbon fibre studs - placed close together ECG leads - carbon fibre - fibre optic - high impedance - short as possible (plaited if necessary) All conductive leads should be placed in a linear fashion coming out of bore of scanner Avoid crossed limbs where possible

RF Burn from non-Carbon Electrode

Other Issues Contrast agents Critically ill patients / resuscitation Monitoring and equipment related hazards Quench of superconducting magnets Claustrophobia

Contrast agents Possible reactions:- Contraindications all gadolinium chelates overall patient tolerance high incidence of adverse reactions very low Possible reactions:- headache, nausea and vomiting, pain if extravasation occurs anaphylaxis 1:100,000 Contraindications GFR<30ms/minute pregnancy (risk versus benefit ratio) breast feeding mothers infants < 2 years Caution asthmatics, history of previous reactions

Critically ill patients AICU and High Dependency Patients Assess patient the day before Lengthen and prime IV lines on ward Replace electrodes - carbon studs Lengthen chest drains Check intubation circuits for metal Educate nursing staff in preparing patient Ensure status of patient has not changed since referral Screen all accompanying personnel

Sedated patients Departmental ‘sedation protocol’ with anaesthetic approval Nil by mouth for 6 hrs (food & milk) 2 hrs water Medical examination Informed consent Tipping trolley, piped oxygen, full resuscitation equipment Responsible adult to collect from department and remain with for 24 hours

Resuscitation Alert cardiac arrest team Call for help BLS only can be performed in scanner room Keep MR Compatible ‘tipping trolley’ adjacent to scanner Patient must be moved to designated resuscitation area as quickly as possible Appoint non-clinical person to prevent access to scanner room Practice regular ‘cardiac arrest’ scenarios to identify potential problems

Monitoring Equipment must be MR Safe or MR Conditional MR Safe Device when used in the MR environment, has been demonstrated to present no additional risk to the patient BUT may affect the quality of diagnostic information MR Conditional The object may or may not be safe for the patient undergoing an MR procedure or an individual in the MR environment, depending on the specific conditions that are present. Ie can be used if certain device specific conditions met. See http://www.mrisafety.com/list.asp for definitions 2011: ‘MR conditional’ replaces ‘MR compatible’ as terminology

Quench Causes - physical Cryogens maintain the magnetic field - helium QUENCH = the liquid helium ‘boils off’ and becomes a gas Causes - physical - human error (accidental) - intervention (elective) Effects - ratio of gaseous to liquid helium 760:1 - should vent to the outside, but this can fail - pressure build up - asphyxiation/frostbite

Elective Quenching The magnet should only be quenched in two situations:- If someone is trapped to the scanner by a ferromagnetic object and is injured and/or distressed (eg O2 cylinder, piece of equipment) If there is a fire in the immediate vicinity on order to reduce risk to the Fire Brigade

Action to be taken in the event of a Quench Evacuate the room as quickly as possible Ensure the door is kept open during evacuation Close door after evacuation If trapped in room stay close to floor level Seek the advice of a senior physicist immediately Call scanner engineer

Claustrophobia Affects 5-10% of patients Causes restrictive dimensions of the interior of the magnet duration of the examination gradient coil induced noises the ambient condition within the imaging bore

Factors to reduce anxiety education and explanation trial visit to the department maintaining physical and verbal contact presence of a relative or friend use of a mirror, prism glasses good communication system alarm good light and ventilation music pleasant thoughts

Ensuring Safe Practice Safety Checklist – comprehensive but concise Restricted access to magnetic field area MR Compatible equipment – monitors, wheelchairs, syringe pumps, stethoscopes Awareness of medication common to patient area – contrast agents, stress agents Staff Induction Education and training of staff of all disciplines Risk Assessments Policies and Procedures – cardiac arrest, quench

This could be you!

References Shellock F.G. and Kanal E. (2001) - Magnetic Resonance Bioeffects, Safety, and Patient Management Health Protection Agency (HPA) (was National Radiological Protection Board -NRPB) International Electrical Committee (IEC) Medical and Healthcare products Regulatory Agency (MRHA) (was Medical Devices Agency - MDA) http://www.mrisafety.com (Shellock and Kanal) http://www.imrser.org/ http://www.magneticresonancesafetytesting.com/