Mobile Fluoroscopy Based on:

Slides:



Advertisements
Similar presentations
State of Michigan Computed Tomography Regulations
Advertisements

Veterinary Radiation Safety 4 Students Radiation Safety MA Law requires that persons who perform diagnostic radiology procedures on animal patients be.
Rad Protection in Fluoro & INTRO RHB regs
DESIGNING FOR RADIATION PROTECTION. TUBE HOUSING  REDUCES LEAKAGE TO LESS THAN 100 mR PER HOUR AT A DISTANCE OF ONE METER FROM HOUSING  One meter is.
CVT 102.  Time  Distance  Shielding.
Digital FLUORO Summer 2008.
Radiation Occupational exposures and protection A. H. Mehrparvar, MD Occupational Medicine department Yazd University of Medical Sciences.
Dental Assisting Radiology
Quality Control Rad T 110.
Introduction to Fluoroscopy & Radiation Safety
FLUOROSCOPY EQUIPMENT
Spot Filming Cinefluorography Fluoro Room Shielding Radiation Areas.
Fluoroscopy Q&A Robert L. Metzger, Ph.D..
RADIATION PROTECTION IN DIAGNOSTIC AND INTERVENTIONAL RADIOLOGY
RHB SYLLABUS on Fluoroscopy & Radiation Protection
Designing for Radiation Protection. Design Standards for Radiation Protection – Leakage Radiation.
Radiation Protection Procedures
Radiation Protection Procedures
Week 2 :Radiation Protection
Fluoroscopy. Real-time imaging Most general-purpose fluoroscopy systems use TV technology, operating at 30 frames/sec May be recorded (barium swallow.
Week 2: Radiographic Equipment
Protection1 Radiation Protection. Protection2 BENEFITS VS. ADVERSE EFFECTS.
فرهاد نعلینی. رادیولوژیست. دانشگاه علوم پزشکی کرمانشاه.
IAEA International Atomic Energy Agency Fluoroscopy Radiation Sources in medicine diagnostic Radiology Day 7 – Lecture 1(2)
IAEA International Atomic Energy Agency RADIATION PROTECTION IN DIAGNOSTIC AND INTERVENTIONAL RADIOLOGY Part 12.1 : Shielding and X-ray room design Practical.
Training Module Preparation for “AERB norms, licenses and signage”
RADIATION PROTECTION 04/12/2016.
Ch 36 Radiation Protection Design
radiation protection in dentistry
Introduction to and History of Fluoroscopy Radiation Protection
Pediatric Radiology By Dr. Wambani, J.S. Chief Radiologist
RADIATION PROTECTION IN DIAGNOSTIC AND INTERVENTIONAL RADIOLOGY
Do not adjust your set.
Chapter 13 Mgmt. of Imaging Personnel dose during Diag. X-Ray Proc.
Resident Physics Lectures
Rad T 110 Time, Distance, Shielding
Radiation Protection Procedures
Fluoroscopy: Dynamic Imaging Systems
Pediatric Fluoroscopy
Fluoroscopy Dynamic Image Recording Systems
Fluoroscopic Brightness Control Systems
Fluoroscopic Image Intensifier Image production
Mobile Fluoroscopy Based on:
Fluoroscopy Dynamic Image Recording Systems
Introduction to and History of Fluoroscopy Radiation Protection
Pediatric Fluoroscopy
High Level (Boost) Fluoroscopy
Fluoroscopy Static Image Recording Systems
Pima Medical Institute Online Education
Fluoroscopy – Viewing Systems TV Monitors
Radiation Protection in Fluoroscopy
Fluoroscopy – Viewing Systems Optical Mirrors TV Camera & TV Camera Tubes Charge Coupled Devices (CCD) Based on: Principles of Radiographic Imaging, 3rd.
Pediatric Fluoroscopy
Conducting the Fluoroscopic Examination & Basic Operational Procedures
Conducting the Fluoroscopic Examination & Basic Operational Procedures
Occupational Radiation Dose Management
Fluoroscopy Static Image Recording Systems
Gonadal Shielding & Radiation Safety
Pima Medical Institute Online Education
Fluoroscopic Image Quality Considerations
Digital Fluoroscopy PPT created by: Jed Miles, BSRS, RT(R), CRT-CA
It’s been a long road…….. When will it end?.
Occupational Radiation Dose Management
Fluoroscopic Image Intensifier Image Production
Fluoroscopy: Static Image Recording Systems
Mobile Fluoroscopy.
Fluoroscopic Image Quality Considerations
Occupational Radiation Dose Management
Brightness Control Systems
Phenotyping Facility X-ray Awareness Training
Presentation transcript:

Mobile Fluoroscopy Based on: Syllabus on Fluoroscopy Radiation Protection, 6th Rev. By: Radiologic Health Branch – Certification Unit PPT created by: Jed Miles, BSRS, RT(R), CRT-CA

Structural Provisions Structural shielding requirements for fluoroscopic units is for scattered radiation shielding and is generally minimal if any, since the entire x-ray beam is intercepted by the image receptor Certain regulatory provisions should be noted If mobile equipment is to be used in one location, shielding must be provided as for a fixed installation If mobile equipment is routinely used in operating rooms, appropriate structural shielding must be provided for these rooms Structural shielding = radiation attenuation material built into the walls of the room

Equipment Provisions (C-arm) Basic regulatory provisions regarding mobile fluoroscopic equipment Inherent provisions must be made so that the machine cannot be operated at a source-to-skin distance of less than 30cm (12 inches) Use spacer attached to tube/collimator assembly on c-arm Image intensification must be provided Conventional fluoroscopic screens not permitted It shall be impossible to energize the useful beam of a mobile fluoroscopic unit unless the entire beam is intercepted by the image receptor

Equipment Provisions (C-arm) Maximum permissible dose rate of 5 rads/min may not be exceeded as measured at 30 cm (12 inches) from input surface of fluoroscopic image assembly Personnel monitoring is required for all persons operating mobile x-ray equipment Protective aprons of a least 0.25 mm lead equivalent must be worn if one is likely to receive 5 millirads/hr or more (notice this one is rads/hour) When observance of motion is not necessary, use of an image storage video disc recorder with “electronic radiography” or “last image hold” will significantly reduce patient and operator exposure

Boost Position or High Level Output Control When a boost position or high level output control button is provided on a mobile fluoroscope to increase the maximum exposure level First: use “normal” fluoro mode to locate and center the area of interest Second: use boost mode only when it is necessary to achieve the quality of image required

What’s Next? Please close this PowerPoint presentation, and then click Pre-Test to continue.