Best Practices for Identification and Quantitation

Slides:



Advertisements
Similar presentations
Gas Chromatography.
Advertisements

Introducing AnalyzerPro. Chapter 1: Qualitative Analysis.
Gas Chromatography, GC L.O.:  Explain the term: retention time.  Interpret gas chromatograms in terms of retention times and the approximate proportions.
Chromatographic Process Provides the analyte transport. Immobile phase. Mixture of components dispersed in the mobile phase.
Fundamentals of Method 23 What You Always Wanted to Know About EMPC but were Afraid to Ask Office of Air Quality Planning and Standards Air Quality Assessment.
Pesticide screening LC-QTOF, Agilent. National Food Institute, Technical University of Denmark Disposition National Food Institute –EURL –NRL –Personale.
© 2009 Perkin Elmer As Speciation in Apple Juice Charles Schneider, Presenter Kenneth Neubauer, Co-Author PerkinElmer.
World Health Organization
Lecture 8. GC/MS.
Smart Templates for Chemical Identification in GCxGC-MS QingPing Tao 1, Stephen E. Reichenbach 2, Mingtian Ni 3, Arvind Visvanathan 2, Michael Kok 2, Luke.
1 Welcome! Mass Spectrometry meets Cheminformatics Tobias Kind and Julie Leary UC Davis Course 6: Concepts for GC-MS Class website: CHE Spring 2008.
Gas Chromatography And Mass Spectrometry
Introducing AnalyzerPro. Chapter 2: Quantitative Analysis.
Questions From HW.. 1. The Zn in a g sample of foot powder was titrated with mL of M EDTA (Y 4- ). Calculate the percent Zn in this.
Chemical Ideas 7.6 Chromatography. The general principle. Use – to separate and identify components of mixtures. Several different types - paper, thin.
Chromatographic separations Chapter 26 The “stuff” you do before you analyze a “complex” sample.
Organic Chemistry Lab 315 Fall, DUE DATES Today –At beginning of lab – Isolation of Caffeine Report –At end of lab -- copy of laboratory notebook.
Qualitative Data Analysis. 2 In This Section, We Will Discuss:  How to load data files.  How to use Signal Options for data display.  How to apply.
Raul Garcia-Sanchez Research Investigator: Dr. Paul R. Mahaffy Code 699, NASA Goddard Space Flight Center Research Mentor: Dr. Prabhakar Misra Department.
INTRODUCTION TO CHROMATOGRAPY
Outline EDTA EDTA Titration Acid Base Properties aY nomenclature
Novel Algorithms for the Quantification Confidence in Quantitative Proteomics with Stable Isotope Labeling* Novel Algorithms for the Quantification Confidence.
INF380 - Proteomics-101 INF380 – Proteomics Chapter 10 – Spectral Comparison Spectral comparison means that an experimental spectrum is compared to theoretical.
SoSe 2006 Alexander Semmler: HPLC 1 Alexander Semmler HPLC High pressure liquid chromatography High performance -‘‘-
7000A GC-QQQ Applications Pesticides in Foods (1).
By: Arafath and Nick. What is it  Chromatography is a technique that is used to separate the substances present in a mixture.  It is widely used to.
Introduction to Chromatography. Introduction Chromatography permit the scientist to separate closely related components of complex mixtures. In all chromatographic.
HPLC 1. Introduction 1.Introduction  INSTUMENTAL ANALYSIS  PRACTICAL 213 PHC  HPLC.
Lecture 8 Peak Parameters and Quantitative chromatography
DIA Method Design, Data Acquisition, and Assessment
EQUIPMENT and METHOD VALIDATION
Downloaded from کروماتوگرافی CHROMATOGRAPHY Downloaded from
이 장 우. 1. Introduction  HPLC-MS/MS methodology achieved its preferred status -Highly selective and effectively eliminated interference -Without.
QUALITATIVE Chemical Analysis
“Structure Elucidation”-Comprehensive Spectral Interpretation
Table 1. Quality Parameters Being Considered for Evaluation
Multi-Analyte LC-MS/MS Methods – Best Practice.
Determination of TKN by Subtraction using ASTM D and ASTM D
Metabolomics Data Analysis
An integrated GC-MS workflow solution for the determination of (semi)volatiles in drinking water and solid waste according to the U.S. EPA guidelines B.
Organic Chemistry Lab 315 Fall, 2016.
Overview Determining Identity Quantitative Analysis
Chemical Ideas 7.6 Chromatography.
Review of Basic Atomic Structure, Mass Spectrometry
Chapter 3 Forensic Laboratory Techniques
Figure S-3. RRT, RIA, FID examples
High Performance Liquid Chromatography HPLC
Quantitative Gas Chromatography
William Lipps Analytical & Measuring Instrument Division August 2017
Systems Medicine Automated Real-Time Quality Control of LC-MS Metabolomics Data: QC4Metabolomics stanstrup.github.io.
Janice L. Willey Senior Chemist
Strategies for Eliminating Interferences in Optical Emission Spectroscopy Best practices to optimize your method and correct for interferences to produce.
Added Monrovia and South Bend to the slide.
The Use of Advanced Instrumental Techniques to Address Emerging and Unique Circumstance Contaminants Charles Neslund, Scientific Officer, Eurofins Lancaster.
Update on ASTM and Standard Methods method development activities
HPLC detectors Pharmawiki 11/18/2018.
Mass Spectrometry.
Presentation Title NEMC 2018 Dale Walker, Bruce Quimby Agilent
A Field Analysis Technique for the Determination of PFAS Compounds
Intro to Chromatography CHM 480 Suroviec
The Practicality (Necessity) of Use of Clean-up Techniques for the Analysis of D/F and PCB Congeners Charles Neslund, Scientific Officer, Eurofins Lancaster.
An Introduction to Chromatographic Separations
A Holding Time Evaluation Study for the Analysis of PFAS in Aqueous Samples Charles Neslund, Scientific Officer, Eurofins Lancaster Laboratories Environmental,
Metabolomics: Preanalytical Variables
Best Practices for Calibration and Calibration Verification
Chapter 3 Forensic Laboratory Techniques
Aroclors, Homologs and Congeners – An Evaluation of the Options for PCB Analysis and a Comparison of the Interpretive Value Charles Neslund, Scientific.
Comprehensive, Non-Target Characterization of Environmental Exposome Samples Using GC×GC and High Resolution Time-of-Flight Mass Spectrometry James Carlson,
Operation manual of AI SIDA
Presentation transcript:

Best Practices for Identification and Quantitation Charles Neslund, Scientific Officer, Eurofins Lancaster Laboratories Environmental, LLC National Environmental Monitoring Conference, New Orleans, LA August 6-10, 2018 Providing comprehensive scientific resources to environmental clients worldwide.

Volatiles by GC/MS Four Programs – Multiple Methods RCRA – SW 846 Method 8260C SDWA – 524.2 CWA – 624.1 CERCLA – SOM02.4 2 2

Volatiles by GC/MS Several factors that are looked at in each method with respect to Identification and Quantitation. Retention time of target analyte Comparison of mass spectrum Characteristic ion Relative intensities of characteristic ions 3 3

Retention Time Method Criteria 8260 C Within 0.06 RRT units of the RRT of the Standard Component 524.2 (3,4) 3 SD of the mean RT of the compound in the calibration mixture 624.1 Within 10 sec of the RT of the analyte in the LCS at the beginning of the shift SOM02.4 Within 0.06 RRT units of the RRT of the corresponding CCV standard 4 4

Retention Time Retention Time Relative Retention Time Retention time (RT) is a measure of the time taken for a solute to pass through a chromatography column. It is calculated as the time from injection to detection. Relative Retention Time RRT is an expression of a sample’s retention time relative to the standard’s retention time.The use of the relative retention time (RRT) reduces the effects of some of the variables that can affect the retention time. RRT = Standard RT / Sample RT 5 5

Mass Spectrum Signals Method Criteria 8260 C Characteristic ions agree within 30% of ions in reference spectrum. Maximize in same scan or within one scan. Ions greater than 30%. 524.2 (3,4) Characteristic ions agree within 20% of ions in reference spectrum. Primary identification is based on RT. Ions greater that 30% 624.1 Characteristic ions agree within 50% to 200% of ions in reference spectrum. Maximize within the same two consecutive scans. Table 6 in method. SOM02.4 Characteristic ions agree within 20% of ions in reference spectrum. Within 0.06 RRT units of the RRT of the corresponding CCV standard 6 6

Reference Spectra Method Criteria 8260 C Must be generated by laboratory using conditions of the method 524.2 (3,4) Seems to allow use of lab generated reference spectra or database (NIST) 624.1 Allows the use of either lab generated reference spectra or database (NIST) SOM02.4 7 7

Reference Spectra The flexibility allowed between the methods in the generation of reference spectra has been a concern With shortened chromatographic runtimes and standard mixes containing many components, laboratory generated reference spectra are not always as “clean” as would be desirable Some initiatives to only use reference spectra that refer specifically to a commercial spectral database

Structural Isomers Method Criteria 8260 C Height of valley between peaks less that 50%, otherwise report as isomeric pairs 524.2 (3,4) Not addressed 624.1 SOM02.4 Must be “50 % resolved” 9 9

All methods recommend/require use of internal standard calibration Quantitation Method Criteria 8260 C EICP of primary/quant ion. ICAL < 20% use average RRF. Otherwise, calibration curve 524.2 (3,4) Linear or quadratic calibration curve. Weighting allowed. Cal stds must calculate within 30% 624.1 EICP of primary/quant ion. ICAL < 35% use average RRF. Otherwise, calibration curve SOM02.4 EICP of primary/quant ion. ICAL < 20-40%, depending on compound, use average RRF. Two failures allowed. All methods recommend/require use of internal standard calibration 10 10

Library Search - Unknowns Method Criteria 8260 C Guidelines given for qualitative identification of unknowns. Assume RRF of 1 for quantitation. 524.2 (3,4) Not specifically addressed 624.1 SOM02.4 Very prescriptive requirements for the generation of library search results including differentiating between alkane and non-alkane TICs 11 11

Conclusions Retention time matching is a critical part of the analysis, even though a mass spec application. All 4 methods use internal standards. The RRT is preferred. It would seem that there would be value in broadening the relative abundance criteria but specify the number of scans within which ions maximize. If we are not going to specify the requirements for generating reference spectra (RT resolution, etc.) then it would seem that commercial libraries should be used. Need some consensus on the quantitation model(s) to be used