Analyzing Biological and Organic Polymers by MALDI-TOF

Slides:



Advertisements
Similar presentations
Mass Spectrometry of Proteins and Peptides Electrospray Ionization(ESI) and Matrix Assisted Laser Desorption Ionization(MALDI) by Matt Fisher.
Advertisements

D e t e c t o r s f o r H P L C.
FC-MS from Teledyne Isco CombiFlash ® a Name You Can Rely On.
Sample clean-up (MALDI)
Matrix Assisted Desorption/Ionization Mass Spectrometry
Lecture 3 Ionisation techniques Gas Phase Ionisation Techniques : Chemical Ionisation.
12-1 Molecular Mass Spectroscopy Molecular structure Composition of mixtures Molecular mass spectra Ion Source Mass Spectrometers Applications.
Evaluation of Peptides in Wisconsin Beer Mckenna L. Missfeldt, Dr. Jennifer Grant, University of Wisconsin-Stout Abstract Matrix-assisted laser desorption/ionization.
Oligonucleotide analysis by Linear MALDI-TOF mass spectrometry Stephen J. Hattan , Kenneth C. Parker, Marvin L. Vestal; SimulTOF Systems, 60 Union Ave,
Sample spotting techniques Dried droplet Crushed crystal Thin layer Sandwich.
Introduction to MALDI-TOF MS
Analyzing Biological and Organic Polymers by MALDI-TOF Jonathan A. Karty, Ph.D.
Lecture 9. MALDI/TOF. Introduction Matrix-assisted laser desorption/ionization (MALDI) is a soft ionization technique used in mass spectrometry, allowing.
Electrospray and Atmospheric Pressure Chemical Ionization Quadrupole MS for Small Molecules Jonathan A. Karty, Ph.D.
Ionization Techniques
Mass Spectrometry in the Biosciences: Introduction to Mass Spectrometry and Its Uses in a Company Like Decode. Sigurður V. Smárason, Ph.D. New Technologies.
Proteomics The proteome is larger than the genome due to alternative splicing and protein modification. As we have said before we need to know All protein-protein.
PROTEIN IDENTIFICATION BY MASS SPECTROMETRY. OBJECTIVES To become familiar with matrix assisted laser desorption ionization-time of flight mass spectrometry.
Atomic Mass Spectrometry
 Amino acid analysis refers to the methodology used to determine the amino acid composition or content of proteins, peptides, and other pharmaceutical.
Matrix-Assisted Laser Desorption Ionization Time of Flight (MALDI TOF)
Chapter 8 – Mass Spectrometry. Mass Spectrometry The mass spectrometer can be used for: – Quantitative analysis – as a sophisticated and very sensitive.
GC Advantages 1. Very Large N (Very Long Columns) 2. No Packing Material (A=0) 3. Simple Mobile Phase (Compressed Gas) 4. Universal Detectors (FID) 5.
Introduction to Walk-Up Mass Spectrometry Jonathan A. Karty, Ph.D. July 21, 2008.
Tissue Imaging by 5 kHz High-Performance MALDI-TOF Poster Number TP191 Christina Vestal 1, Kenneth Parker 1, Kevin Hayden 1, George Mills 1, Marvin Vestal.
Based on ionization of gas phase molecule followed by analysis of the masses of the ions produced. The Mass Spectrum: Graph of ion intensity versus mass-to-charge.
Chemistry Topic: Atomic theory Subtopic : Mass Spectrometer.
INF380 - Proteomics-51 INF380 – Proteomics Chapter 5 – Fundamentals of Mass Spectrometry Mass spectrometry (MS) is used for measuring the mass-to-charge.
MALDI demonstration Frímann, Helgi, Long, Nanna. Sample preperation The sample of interest and matrix are dissolved in a volatile solvent The sample/matrix/solvent.
Chemistry 367L/392N Macromolecular Chemistry Lecture 8.
Mass Spectrometry Instrumentation: Ionization methods and analyzers.
Combining and Choosing Analytical Techniques Chapter 8.
LIQUID CHROMATOGRAPHY-MASS SPECTROMETRY
HPLC.
Separates charged atoms or molecules according to their mass-to-charge ratio Mass Spectrometry Frequently.
Chromatography and Instrumentation. Chromatography Separate Analyze Identify Purify Quantify Components Mixture Chromatography is used by scientists to:
Chem. 133 – 4/26 Lecture. Announcements Return graded quiz and additional problem Lab – Lab report deadlines (2:4 – Thursday) Today’s Lecture – Mass Spectrometry.
2014 생화학 실험 (1) 6주차 실험조교 : 류 지 연 Yonsei Proteome Research Center 산학협동관 421호
Chapter 29 Mass Spectrometry. 29 A Principles of mass spectrometry In the mass spectrometer, analyte molecules are converted to ions by applying energy.
What is Mass Spectrometry? Mass spectrometry could be considered as an analytical technique that involves the study in the gas phase of ionized molecules.
Yonsei Proteome Research Center Peptide Mass Finger-Printing Part II. MALDI-TOF 2013 생화학 실험 (1) 6 주차 자료 임종선 조교 내선 6625.
Peptide Mass Finger-Printing Part II. MALDI-TOF
Mass Spectrometry makes it possible to measure protein/peptide masses (actually mass/charge ratio) with great accuracy Major uses Protein and peptide identification.
Mass Spectrometry 101 (continued) Hackert - CH 370 / 387D
A guide for A level students KNOCKHARDY PUBLISHING
MALDI-TOF MS Bergdís Björk Bæringsdóttir Daníel Arnar Tómasson
Acknowledgements Slides and animations were made by Dr. Jon Karty Mass Spectrometry Facility Indiana University, Bloomington.
Chem. 133 – 4/13 Lecture.
Mass Spectrometry 101 Hackert - CH 370 / 387D
Time of Flight Analyzers
Mass Spectrometry Obaid M. Shaikh.
Starter Complete the quiz and hand in..
An Introduction to the ProteinChip® Reader
Anadarko Undergraduate Fellowships for Excellence in Energy Research
Mass Spectrometry Courtesy
S. Emonet, H.N. Shah, A. Cherkaoui, J. Schrenzel 
HPLC Equipment : Detectors
ROSINA, COSAC & PTOLEMY Reviewer: Heather Franz 2/15/16 – 2/17/16
Instrumental Chemistry
Time-of-Flight Mass Analyzers
Chapters 11&20 Mass Spectrometry 1 Mass Spectrometry
Schematic of the principles of mass spectrometry (MS).
Atomic Absorption Spectroscopy. Atomic absorption spectroscopy is based on the same principle as the flame test used in qualitative analysis.
S. Emonet, H.N. Shah, A. Cherkaoui, J. Schrenzel 
Molecular Mass Spectroscopy
Mass Spectrometry THE MAIN USE OF MS IN ORG CHEM IS:
Department of Chemistry St. Joseph’s College (Autonomous)
M.Prasad Naidu MSc Medical Biochemistry, Ph.D,.
A guide for A level students KNOCKHARDY PUBLISHING
CHROMATOGRAPHY.
Presentation transcript:

Analyzing Biological and Organic Polymers by MALDI-TOF Jonathan A. Karty, Ph.D.

Topics Covered Sample Requirements Instrument Overview General Instrument Use Instructions Tips and Tricks

What is the Bruker Autoflex III? Time-of-flight mass spectrometer Ions of given same kinetic energy, heavy ions travel slower than lighter ones Two modes of operation Linear Reflectron MALDI/LDI source 384 position target plate (~1 µL spot size) 355 nm Nd:YAG laser Can analyze positive or negative ions (same spot)

Autoflex III Picture

Matrix-Assisted Laser Desorption/Ionization (MALDI) Analyte is mixed with UV-absorbing matrix ~10,000:1 matrix:analyte ratio Analyte does not need to absorb laser A drop of this liquid is dried on a target Analyte incorporated into matrix crystals Spot is irradiated by a laser pulse Irradiated region sublimes, taking analyte with it Matrix is often promoted to the excited state Charges exchange between matrix and analyte in the plume (very fast <100 nsec) Ions are accelerated toward the detector

Image from http://www.noble.org/Plantbio/MS/iontech.maldi.html MALDI Diagram Image from http://www.noble.org/Plantbio/MS/iontech.maldi.html

MALDI Advantages Technique is relatively simple Volatilize and ionize labile molecules Imagine electron ionization on a protein MALDI creates very simple mass spectra Ions are usually (M+nH)n+ or (M-nH)n- Only 1-3 charge states are observed Usually 1 charge state for peptides < 3.5 kDa MALDI ideal for time-of-flight analyzers Theoretically unlimited mass range (100 kDa done here) MALDI is very rapid (<1 min/spot) Low sample consumption (1 µL) Wide array of matrices available for different analytes

Some Common MALDI Matrices

What Samples Can It Run? Biopolymers Organometallic complexes Peptides, proteins, DNA, RNA, oligosaccharides Organometallic complexes Organometallic salts work great Synthetic polymers Polymer need not be soluble in same solvent as matrix Molecules that photoionize upon irradiation by 355 nm laser Porphyrins

What Samples Can’t It Run? “Dirty” samples Significant concentration of involatiles Glycerol, urea, most buffers, many detergents Alkali metal salts can be quite problematic RNA/DNA analyses require extensive desalting Molecules with significant vapor pressures Instrument is held at ~10-7 torr Molecules that do not make stable ions in source Lack charge acceptor/donor site Cannot photoionize with Nd:YAG laser

MALDI Advantages Relatively gentle ionization technique Very high MW species can be ionized Molecule need not be volatile Very easy to get sub-picomole sensitivity Spectra are easy to interpret Positive or negative ions from same spot Wide array of matrices available

MALDI Disadvantages MALDI matrix cluster ions can obscure low m/z (<600) range Analyte must have very low vapor pressure Coupling MALDI with chromatography can be difficult Analytes that absorb the laser can be problematic Fluorescein-labeled peptides

Instrument Diagram 355 nm Nd:YAG laser Target Reflectron Linear Detector Lens Extraction Plate Reflector Detector Flight Tube Entrance

Linear Mode 355 nm Nd:YAG laser Target Reflectron Linear Detector Lens Linear mode is used for large (>3.5 kDa) molecules or exceedingly fragile species (oligosaccharides). It is capable of 4,000 resolving power @ 3.2 kDa (1,000 RP @ 12 kDa) Extraction Plate Reflector Detector Flight Tube Entrance

Reflectron Mode 355 nm Nd:YAG laser Target Reflectron Linear Detector Lens Extraction Plate Reflector Detector Reflectron mode is used for small species (<4 kDa) and is capable of 11,000 resolving power @ 3.2 kDa. Flight Tube Entrance

MALDI Example (ACTH 7-38+H)+ (ACTH 18-37+H)+ (Ubiq+2H)2+ (Ubiq+H)+ (Ins+H)+

MALDI Example I Continued

MALDI-TOF Example 2

General Sample Guidelines Purify analyte if possible Analyte should be 5 – 100 µM in concentration ZipTips can help purify dirty samples (C4 and C18 available in MSF) Use only volatile solvents/buffers MeOH, H2O, acetone, CH3CN, THF, CH2Cl2, C6H6 TFA, HOAc, formic acid, NH3, etc. Ionic strength < 20 mM (e.g. 0.1% v/v HOAc) If you need a detergent, 20 mM n-octylglucopyranoside can work No SDS, TWEEN, CHAPS, etc Need at least 2 µL

Sample Prep Tricks Ziptip to clean up dirty samples C18 for peptides < 3 kDa C4 for peptides/proteins > 3kDa Elute directly into matrix for added sensitivity ZipTip instructions on MSF website If CCA liquid turns yellow, pH is too high Spots from non-acidic CCA do not crystallize correctly Add a little 1% v/v or 10% v/v TFA to lower pH If sample needs base for solubility, try over-layer method Dissolve sample in NH3 or other volatile base Place 1 uL of sample on target, let dry completely Deposit 1 uL matrix over top of dried sample

Sample Prep Tricks 2 Non-aqueous over-layer Make 1 uL spot of matrix on plate, let dry Deposit small amount of sample in volatile solvent (e.g. CHCl3, acetone, CH2Cl2) You can even do internal calibration this way Put calibrants in matrix spot For better mass accuracy, let voltages stabilize 10-30 minutes before recording data

Hands-on Training Starts AFTER 11/7 Groups of no more than three One hour or so to complete No charge for first session After training, students must demonstrate competency by running their own samples prior to being granted after-hours access