Light Output measurement of BSO scintillator 2009/July/7th Tomoko Iwashita/Kenkichi Miyabayashi.

Slides:



Advertisements
Similar presentations
PREX main detector Piotr Decowski Aimee Shore Smith College Piotr Decowski Aimee Shore Smith College Kumar Krishna Jon Wexler Patrick Rogan Vireak Yim.
Advertisements

AS Physics Unit 3 Exam Questions Ks5 AS Physics AQA 2450 Mr D Powell.
NUCP 2371 Radiation Measurements II
Digital Imaging with Charge- coupled devices (CCDs)
What is Color? Color is related to the wavelength of light. If a color corresponds to one particular wavelength, this is called spectral color. =600 nm.
Development of Bialkali Transfer Photocathodes for Large Area Micro- Channel Plate Based Photo Detectors 1 Argonne National Laboratory, Argonne, IL 2 University.
Gamma-Ray Spectra _ + The photomultiplier records the (UV) light emitted during electronic recombination in the scintillator. Therefore, the spectrum collected.
1 Continuous Scintillator Slab with Microchannel Plate PMT for PET Heejong Kim 1, Chien-Min Kao 1, Chin-Tu Chen 1, Jean-Francois Genat 2, Fukun Tang 2,
Black body radiation BBR is the radiation emitted by a non-reflecting solid body. A perfect black body is one which absorbs all the radiations falling.
E. Nappi First results from measurements of optical properties of YAP:Ce crystals  Procurement of YAP:Ce crystals Price enquiry for 3.6x3.6x100.
R&D of Strip/Block Scintillators E.P.Jacosalem, S.Iba, N.Nakajima, H.Ono, A.L.Sanchez, A.M.Bacala & H.Miyata GLD Calorimeter Group 8 th ACFA Workshop on.
S. Bricola, A. Menegolli, M. Prata, M.C. Prata, G.L. Raselli, M. Rossella, C. Vignoli INFN and University of Pavia - Via Bassi, 6 – Pavia – Italy.
1 Cosmic Ray Test Stand with Scintillating Cells for Digital Hadron Calorimeter 06/23/2003 Kurt Francis - Northern Illinois University.
1.3 Cavity modes Axial modes λ = 2d / n ν = nc / 2d n = 2d / λ
Radiator study status Y. Horii, Y. Koga, N. Kiribe, … (Nagoya University, Japan) 1 B2GM, 6 th July.
Calibration of the 10inch PMT for IceCube Experiment 03UM1106 Kazuhiro Fujimoto A thesis submitted in partial fulfillment of the requirements of the degree.
Crystals and related topics J. Gerl, GSI NUSTAR Calorimeter Working Group Meeting June 17, 2005 Valencia.
Transducers Converts one type of energy into another. Light  Electrical (current, voltage, etc.) What characteristics should we look for in a transducer?
Gamma Spectroscopy HPT TVAN Technical Training
RF background, analysis of MTA data & implications for MICE Rikard Sandström, Geneva University MICE Collaboration Meeting – Analysis session, October.
Measurement of the absolute efficiency,
1 Light Collection  Once light is produced in a scintillator it must collected, transported, and coupled to some device that can convert it into an electrical.
Photon detection Visible or near-visible wavelengths
Report on SiPM Tests SiPM as a alternative photo detector to replace PMT. Qauntify basic characteristics Measure Energy, Timing resolution Develop simulation.
Measurement of 'intrinsic' properties of scintillating fibers H. Sakamoto Osaka University, Japan A.Sato M. Yoshida Y. Kuno Osaka Univ. K. Yoshimura KEK.
Experimental set-up Abstract Modeling of processes in the MCP PMT Timing and Cross-Talk Properties of BURLE Multi-Channel MCP PMTs S.Korpar a,b, R.Dolenec.
UV-Vis Absorption Spectroscopy
Electromagnetic Waves
Tests WLS - Readout axial PET - Bari - Janvier 2007 AXIAL PET - HPD AXIAL COORDINATE RECONSTRUCTION WITH WLS STRIPS
T. Sugitate / Hiroshima / PHX031 / Nov.01 The Photon Spectrometer for RHIC and beyond PbWO 4 Crystal Density 8.29 g/cm 3 Radiation length 0.89 cm Moliere.
TOP counter overview and issues K. Inami (Nagoya university) 2008/7/3-4 2 nd open meeting for proto-collaboration - Overview - Design - Performance - Prototype.
work for PID in Novosibirsk E.A.Kravchenko Budker INP, Novosibirsk.
Development of TOP counter for Super B factory K. Inami (Nagoya university) 2007/10/ th International Workshop on Ring Imaging Cherenkov Counters.
Systematic Studies of Small Scintillators for New Sampling Calorimeter E.P.Jacosalem, S.Iba, N.Nakajima, H.Ono, A.L.Sanchez, A.M.Bacala & H.Miyata GLD.
Development of a Gamma Camera Based on an 88 Array of LaBr3(Ce) Scintillator Pixels Coupled to a 64-channel Multi-anode PMT Hidetoshi Kubo, K.Hattori,
Abort Gap Monitoring Randy Thurman-Keup 6 / 8 / 2004 LARP Meeting.
X-RAY LIGHT SOURCE BY INVERSE COMPTON SCATTERING OF CSR FLS Mar. 6 Miho Shimada High Energy Research Accelerator Organization, KEK.
Reflection of Light How do objects interact with incoming light?
Current Status of Pepperpot Scintillator Screen
Gain stability and the LYSO beam radiation monitor measurements
Linearity Tests The laser induced high noise pulses on the APD rendering it useless. We tried several methods to shield and filter the noise. With a stabilized.
Lecture 3-Building a Detector (cont’d) George K. Parks Space Sciences Laboratory UC Berkeley, Berkeley, CA.
MCP-PET: Geant4 Simulation Geometry Implementations 1. Similar to the last report (polished surface). Scintillator : LSO, LaBr3 # of layers : 5 -> 4 Area.
PMT measurements in Antares Oleg Kalekin on behalf of Antares collaboration VLVnT 2011 Erlangen
Compilation of information about BSO option Kenkichi Miyabayashi 2009 July 7th Belle-II calorimeter group meeting.
A Brand new neutrino detector 「 SciBar 」 (2) Y. Takubo (Osaka) - Readout Electronics - Introduction Readout electronics Cosmic ray trigger modules Conclusion.
Photon Transport Monte Carlo September 27, 2004 Matthew Jones/Riei IshizikiPurdue University Overview Physical processes PMT and electronics response Some.
Status of NEWCHOD E.Guschin (INR), S.Kholodenko (IHEP), Yu.Kudenko (INR), I.Mannelli (Pisa), O.Mineev (INR), V.Obraztsov (IHEP), V.Semenov(IHEP), V.Sugonyaev.
CaMoO 4 at low temperature - LAAPD and PMTs - Woonku Kang, Jungil Lee, Eunju Jeon, Kyungju Ma, Yeongduk Kim Sejong University for KIMS collaboration Double.
RDMIS Radiation Detection and Medical Image Sensor lab. Composition Optimization of Polystyrene Based Plastic Scintillator Yewon Kim, Hyunjun Yoo, Chankyu.
Light and the Electromagnetic Spectrum
Light and the Electromagnetic Spectrum
Ultra fast SF57 based SAC M. Raggi Sapienza Università di Roma
Light How do we see?.
Recent progress of PureCsI+APD readout
Chapter 3. Components of Optical Instruments
PAN-2013: Radiation detectors
Geant4 Simulation :MCP PET
Simulation of optical processes in GEANT4
Practical Absorbance and Fluorescence Spectroscopy
Light and the Electromagnetic Spectrum
Department of Physics and Astronomy,
Deng Ziyan Jan 10-12, 2006 BESIII Collaboration Meeting
Basic Physics Processes in a Sodium Iodide (NaI) Calorimeter
Warm Up Set 15 Question Two rays of light, of wavelength 600nm, reflect from glass surfaces separated by 150nm. The rays are initially in phase. What.
Light and the Electromagnetic Spectrum
Light and Quantized Energy
HE instrument and in-orbit performance
Light and the Electromagnetic Spectrum
Light and the Electromagnetic Spectrum
Presentation transcript:

Light Output measurement of BSO scintillator 2009/July/7th Tomoko Iwashita/Kenkichi Miyabayashi

Outline Motivation – BSO is an alternative option; – Heavier and longer wavelength(~480nm). – Checked Light Output of small samples quickly. Set up ADC distribution – PureCsI – BSO Compared with PureCsI Summary and plans

Set up PMT : H3167(19mmφ) Borosilicate glass window Bialkali photocathode HV 1650V Scintillator PMT Radiation source 137Cs(662keV) AMPDivAMPATT DiscriG.G. Charge-sensitive ADC(LeCroy2249W) Width 500ns Veto 1ms Optical grease Threshold 20mV 30mV To match ADC range

Samples PureCsI 1φ×2cm BSO 1*1*2cm 3 BSO both ends mirror polish, 4 side surfaces are frosted. Wrapped by white Gore-TeX film.

ADC distributionof PureCsI Size : 1φ×2cm. (λ=330nm) Around photoelectric peak is magnified as right figure, together with fitted line by Gaussian. Photoelectric peak Compton scattering

ADC distribution of BSO Size : 1cm×1cm×2cm. (Purity %, λ=480nm) ATT 6dB : To adjust ADC range. Right figure is magnification around photoelectric peak. Compton scattering Photoelectric peak

To compare with PureCsI L.O. of BSO is two times as PureCsI. Sample L.O. (PureCsI equal 1 ) BSO SA, SB and No.1 are different samples; all purity %.. As for Pr:LuAG, see 2008 Dec. meeting slides BSO exhibits double effective L.O. because of longer wavelength!

Comments on sample dimension and surface finishing. Light collections for PureCsI(cylindrical)=BSO(1X1X2cm 3 ) assumed. In order to make sure, PureCsI (1X1X2cm 3 ) ordered, delivered in 3 weeks. Then answer becomes clearer.

Actually we experienced; L.O. of 2x2x2cm 3 BSO(all 6 surfaces mirror polished) exhibits ~ half of 1x1x2cm 3 (only both end surfaces are mirror polished) sample. – Though they are taken from the exactly same block. Thought to because of photons total reflection at its polished surfaces ~45 deg. Significant amount of photons are absorbed during travel inside crystal… x Common phenomena in crystal with n>sqrt(2).

Summary and plans Effective L.O. of BSO is 2 times as high as PureCsI because of longer wavelength. 1*1*2cm 3 PureCsI (Both ends mirror polish, 4 side surfaces are frosted) will be gotten soon. – To remove possible light collection difference. Radiation hardness test of BSO(and Pr:LuAG) small samples.

Back up

Typical Photocathode Spectral Response and Emission Spectrum of Scintillators PureCsI BSO From HAMAMATSU text