MRI ARTEFACTS PG Student : Dr Aditi Shah PG Guide: Dr Rajendra Chavan

Slides:



Advertisements
Similar presentations
Receiver Bandwidth affects signal-to-noise ratio (SNR)
Advertisements

Pulse Timing parameters & Weighting
Fund BioImag : Echo formation and spatial encoding 1.What makes the magnetic resonance signal spatially dependent ? 2.How is the position of.
Fund BioImag : Echo formation and spatial encoding 1.What makes the magnetic resonance signal spatially dependent ? 2.How is the position of.
Ari Borthakur, PhD Associate Director, Center for Magnetic Resonance & Optical Imaging Department of Radiology Perelman school of Medicine, University.
Principles of the MRI Signal Contrast Mechanisms MR Image Formation John VanMeter, Ph.D. Center for Functional and Molecular Imaging Georgetown University.
Imaging Sequences part I
Chapter 7: Gradient Echo Imaging Methods
Introduction to MRI Magnetic Resonance Imaging
MRI Phillip W Patton, Ph.D..
M R I Pulse Sequences Jerry Allison Ph.D..
BE 581 Lecture 3- Intro to MRI.
Magnetic Resonance Imaging - MRI
MR Sequences and Techniques
Statistical Parametric Mapping
Chapter 14: Artifacts Mark D. Herbst, MD, PhD. Artifact Something on the image that does not represent something in the patient –Many causes Equipment.
Parameters and Trade-offs
Time of flight Entry slice phenomina Intra-voxel dephasing
Basic Principles MRI related to Neuroimaging Xiaoping Hu Department of Biomedical Engineering Emory University/Georgia Tech
MRI Artifacts.
Chemical Shift To do with the different electron environments of protons Results in a shift of resonant frequency which affects the image e.g. fat (CH.
Functional Brain Signal Processing: EEG & fMRI Lesson 12 Kaushik Majumdar Indian Statistical Institute Bangalore Center M.Tech.
Encoding and Image Formation
Basics of Magnetic Resonance Imaging
Magnetic Resonance Imaging Basic principles of MRI This lecture was taken from “Simply Physics” Click here to link to this site.
Tissue Contrast intrinsic factors –relative quantity of protons tissue proton density –relaxation properties of tissues T1 & T2 relaxation secondary factors.
MRI: an Introduction By Mohammad Ali Ahmadi Pajouh Amirkabir University of Technology Biomedical Eng. Dep.
Magnetic Resonance Imaging 4
Principles of MRI Physics and Engineering
Principles of Magnetic Resonance
The comparison of MRI imaging In 3.0 T & 1.5T By a.r.shoaie.
Magnetic Resonance Imaging
Seminar October, 2008 j. brnjas-kraljević. Imaging (MRI)  tomography technique  tomography technique – the volume image is built up by images of thin.
Chapter 4 Mark D. Herbst, M.D., Ph.D.. Magnetization MRI depends on magnetization of tissues Temporary magnetization occurs when the hydrogen nuclei are.
Pulse sequences.
Contrast Mechanism and Pulse Sequences Allen W. Song Brain Imaging and Analysis Center Duke University.
September, 2003BME 1450 Introduction to NMR 1 Nuclear Magnetic Resonance (NMR) is a phenomenon discovered about 60 years ago.
Quiz In a 2D spin warp or FT MR scan, aliasing should only occur
Rad T 265 MRI Lecture. No Magnetic Field = No Net Magnetization Protons align with a magnetic field…
Contrast Mechanism and Pulse Sequences
MRI Physics Dr Mohamed El Safwany, MD.. MRI Magnetic Resonance Imaging Magnetic Resonance Imaging.
BIOE 220/RAD 220 REVIEW SESSION 6 March 5, What We’ll Cover Today General questions? Spinal cord anatomy review Fat in images T2* vs T2 decay Review.
1 Chemical Shift Image B shows a more marked amount of chemical shift, why can this be? Chemical shift is an artefact which occurs because fat and water.
Protons (hydrogen nuclei act like little magnets) MRI Collective Magnetic Moment of Protons (M 0 ) Each pixel is a glass of protons B 0 = 3T (not to scale)
Basic MRI Chapter 1 Lecture. Introduction MRI uses radio waves and a magnetic field to make images MRI uses radio waves and a magnetic field to make images.
Anna Beaumont FRCR Part I Physics
V.G.Wimalasena Principal School of Radiography
MRI: Contrast Mechanisms and Pulse Sequences
Principles of MRI Physics and Engineering Allen W. Song Brain Imaging and Analysis Center Duke University.
Charged particle. Moving charge = current Associated magnetic field - B.
 This depends on a property of nuclei called spin.  Gyroscope: Principle: As long as its disc remains spinning rapidly the direction of the spin axis.
CT and MR artifacts.
Artifacts and Suppression Techniques
MRI Artifacts Ray Ballinger, MD, PhD
FMRI data acquisition.
MRI ARTEFACTS.
Magnetic Resonance Imaging
Sunday Case of the Day Physics
MRI Physics in a Nutshell Christian Schwarzbauer
بسم الله الرحمن الرحيم.
Where Mt is the magnetization at time = t, the time after the 90o pulse, Mmax is the maximum magnetization at full recovery. At a time = one T1, the signal.
Magnetic Resonance Imaging: Physical Principles
Physics of Magnetic Resonance Imaging
An Optimal Design Method for MRI Teardrop Gradient Waveforms
Parallel Imaging Artifacts in Body Magnetic Resonance Imaging
Introduction to MR Robert Frost
(4)ELECTRONIC SUPPORT SYSTEM
The echo time (TE) The echo time (TE) refers to the time between the application of the radiofrequency excitation pulse and the peak of the signal induced.
T2 Relaxation Time T2 relaxation time is defined as the time needed to dephase up to 37% of the original value. T2 relaxation refers to the progressive.
Introduction to MR Robert Frost
Presentation transcript:

MRI ARTEFACTS PG Student : Dr Aditi Shah PG Guide: Dr Rajendra Chavan AK

Artefact is a feature appearing in an image that is not present in the original object. Artefacts – -reduce the image’s aesthetic quality -mask pathology leading to a missed diagnosis (false negative) -mimic pathology leading to overdiagnosis (false positive)

Some artefacts are irreversible, some can be completely eliminated, some can be reduced These can be classified as 1.patient related 2.machine related 3.signal processing dependant

GHOSTS / PHASE MISMAPPING Replication of something in the image. Motion is the important cause. Can arise from any structure that moves during acquisition for e.g. -chest wall during respiration -pulsatile movement of vessels -swallowing -eye movement occurs along phase encoding axis.

Remedy- Gating – ECG, pulse or respiratory gating Respiratory motion - breath holding, respiratory gating, ROPE - respiratory ordered phase recording. phase encoding axis swap- phase encoding direction is changed. presaturation-area producing motion artifacts is saturated with RF pulses before start of proper pulse sequence such that all signals from this area are eliminated. Gradient moment nulling-reduces ghosts caused by flowing nuclei moving along gradients by adjusting the gradient.

CHEMICAL SHIFT ARTEFACT Different chemical environment of protons in fat and water Artefact Appears at the interfaces of water and fat as a dark edge because frequency of precession is slightly different in these substances. Always occurs in frequency encoding direction. At 1T-147Hz, at .5-insignificant, at 1.5T-220Hz i.e. fat precesses 220 Hz less than water.

The amount of chemical shift is expressed in units known as ppm The amount of chemical shift is expressed in units known as ppm.of the main magnetic field strength. The actual dimention of this artefact depend on the FOV as this determines the size of each pixel. axis - frequency encoding axis.

Remedy- -using lower field strengths -reduce FOV -increase bandwidth -use fat saturation Advantage- chemical shift along frequency encoding axis forms the basis of MR spectroscopy.

CHEMICAL MISREGISTRATION This is due to precessional frequency difference between fat and water however, here, the fat and water are in phase at certain times and out of phase at others. So when fat and water are in phase their signals add constructively and when they are out of phase their signals cancel each other. This cancellation effect is known as chemical misregistration artefact,which cause a ring of dark signal around certain organs where fat and water interfaces occur within the same voxel.

More in GRE sequences than spin echo Remedy- Use a spin echo sequence Select a TE that generates an echo when fat and water are in phase. Advantage- In phase and out phase imaging

ALIASING / WRAPAROUND ARTEFACTS / BACK FOLDING It is artefact when anatomy that exists outside the FOV is mapped inside the FOV, if it is in close proximity to the receiver coil. It can occur in any axis -frequency wrap -phase wrap

Frequency wrap- is easier to correct Frequency wrap- is easier to correct. Low pass-filters are used to cut off the frequencies originating from outside FOV. Phase wrap- corrected by increasing FOV along phase encoding direction.

CROSS EXCITATION AND CROSS TALK RF excitation pulse is not exactly square hence nuclei in slices adjacent to the RF pulse may be excited by it and receive energy. This energy flips NMV of these nuclei into transverse plane so that they are saturated when they are excited by their own RF pulse. This phenomenon is called cross excitation.

CROSS TALK- The same effect is produced when energy is dissipated to nuclei from neighboring slice (due to spin lattice relaxation) as nuclei within selected slice relax after RF pulse is switched off. This is called as cross talk. REMEDY- Cross talk can never be eliminated. Cross excitation can be reduced by Increase interslice gap-At least 30% gap between the slices. Interleaved slices scanning- so that nuclei have time to relax. Using softwares to square off the RF pulses.

SHADING ARTEFACTS Image has uneven contrast with loss of signal intensity in one part of the image. Causes- -uneven excitation of nuclei within the patient due to RF pulses applied at flip angles other than 90 and 180 degree -abnormal loading of coil -coupling of coil -inhomogeneity of magnetic field

Remedy- -shimming to reduce the inhomogeneity of the magnetic field -load the coil correctly

MAGNETIC SUSCEPTIBILITY ATREFACT Magnetic susceptibility is ability of a substance to become magnetised. Normally most of the tissues has susceptibility value which fall in a fairly narrow range. Presence of paramagnetic material like Hb degradation products or metallic objects or tissue-air interphase causes MS differences. This causes dephasing at the interface of these tissues and signal loss. It is more prominent in GRE sequence than SE sequence because 180 degree pulse refocus the dephasing by reducing magnetic field inhomogeneities. axis - frequency and phase encoding.

Remedy- -use of SE sequence -remove all metals whenever possible Good effects- -used to diagnose haemorrhage -used to quantify liver and myocardial iron overload -forms basis of post-contrast T2* w MR perfusion studies.

STRAIGHT LINES AND ZIPPER LIKE ARTEFACT Zipper like artefact is a line with alternating bright and dark pixels propagating along the frequency encoding direction. Radiowaves from sources outside the MR machine (from inside or outside the shielded MR room) can cause pattern of straight lines. Door opened during imaging can cause this by interfering with inherently weak signal coming from the patient. Stimulated echo and spike in K-space can also cause this. axis - frequency encoding

Remedies- -RF leak should be located and corrected. -sources of RF inside the room should be removed -zipper artefacts can be eliminated by spoiler gradients.

TRUNCATION ARTEFACTS Also called edge, Gibbs and ringing artefacts. It is due to undersampling of the data so that interfaces of high and low signal are incorrectly represented on the image Seen as low intensity band running through high intensity area. Seen in phase encoding direction

Gibbs’ artefact Remedy- -increase the no. of phase encoding steps.

Shading artifact Produces difference in signal intensity across the imaging volume like loss of signal intensity in one part of the image. It is produced due to uneven excitation of nuclei due to RF pulses applied at flip angles other than 900 , Inhomogeneities in the main magnetic field, overflow of analog to digital converter. axis - phase and frequency encoding.

Point artefact- a bright spot of increased signal intensity in the center of the image. This is due to constant offset of DC voltage in the receiver coil which after Fourier transformation appear as a bright spot in the center of the image.

*CLASSIFICATION OF ARTEFACTS- PATIENT RELATED- MOTION artefacts, FLOW artefacts, magnetic SUSCEPTIBILITY ARTEFACTS. SIGNAL PROCESSING DEPENDANT- CHEMICAL shift ARTEFACTS, WRAP AROUND, GIBBS artefacts. MACHINE / HARDWARE RELATED- SHADING ARTEFACTS, STRAIGHT LINES.