One-year pivotal trial outcomes of the Nellix system for endovascular aneurysm sealing Jeffrey P. Carpenter, MD, Robert Cuff, MD, Clifford Buckley, MD, Christopher Healey, MD, Sajjad Hussain, MD, Michel M.P.J. Reijnen, MD, Jose Trani, MD, Dittmar Böckler, MD Journal of Vascular Surgery Volume 65, Issue 2, Pages 330-336.e4 (February 2017) DOI: 10.1016/j.jvs.2016.09.024 Copyright © 2016 Society for Vascular Surgery Terms and Conditions
Supplementary Fig 1 (online only) Freedom from major adverse events (MAEs) through 1 year. Kaplan-Meier estimate: 93.3% (standard error: 2.0%). The blue shaded region represents the 95% confidence limit band (87.9%, 96.3%) at 1 year. Values above the x-axis are the number of patients at risk at each follow-up interval. Journal of Vascular Surgery 2017 65, 330-336.e4DOI: (10.1016/j.jvs.2016.09.024) Copyright © 2016 Society for Vascular Surgery Terms and Conditions
Supplementary Fig 2 (online only) Freedom from all-cause mortality through 1 year. Kaplan-Meier estimate: 96.0% (standard error: 1.6%). The blue shaded region represents the 95% confidence limit band (91.3%, 98.2%) at 1 year. Values above the x-axis are the number of patients at risk at each follow-up interval. Journal of Vascular Surgery 2017 65, 330-336.e4DOI: (10.1016/j.jvs.2016.09.024) Copyright © 2016 Society for Vascular Surgery Terms and Conditions
Supplementary Fig 3 (online only) Freedom from abdominal aortic aneurysm (AAA)-related mortality through 1 year. Kaplan-Meier estimate: 98.7% (standard error: 0.9%). The blue shaded region represents the 95% confidence limit band (94.7%, 99.7%) at 1 year. Values above the x-axis are the number of patients at risk at each follow-up interval. Journal of Vascular Surgery 2017 65, 330-336.e4DOI: (10.1016/j.jvs.2016.09.024) Copyright © 2016 Society for Vascular Surgery Terms and Conditions
Supplementary Fig 4 (online only) Freedom from abdominal aortic aneurysm rupture through 1 year. Kaplan-Meier estimate: 99.3% (standard error: 0.7%). The blue shaded region represents the 95% confidence limit band (95.3%, 99.9%) at 1 year. Values above the x-axis are the number of patients at risk at each follow-up interval. Journal of Vascular Surgery 2017 65, 330-336.e4DOI: (10.1016/j.jvs.2016.09.024) Copyright © 2016 Society for Vascular Surgery Terms and Conditions
Supplementary Fig 5 (online only) Freedom from device-related secondary intervention through 1 year. Kaplan-Meier estimate: 96.6% (standard error: 1.5%). The blue shaded region represents the 95% confidence limit band (92.1%, 98.6%) at 1 year. Values above the x-axis are the number of patients at risk at each follow-up interval. Journal of Vascular Surgery 2017 65, 330-336.e4DOI: (10.1016/j.jvs.2016.09.024) Copyright © 2016 Society for Vascular Surgery Terms and Conditions
Supplementary Fig 6 (online only) Freedom from surgical conversion through 1 year. Kaplan-Meier estimate: 99.3% (standard error: 0.7%). The blue shaded region represents the 95% confidence limit band (95.3%, 99.9%) at 1 year. Values above the x-axis are the number of patients at risk at each follow-up interval. Journal of Vascular Surgery 2017 65, 330-336.e4DOI: (10.1016/j.jvs.2016.09.024) Copyright © 2016 Society for Vascular Surgery Terms and Conditions