Fluoroscopic Image Intensifier Image Production

Slides:



Advertisements
Similar presentations
Advanced Biomedical Imaging Lecture 3
Advertisements

Chapter 11 Prime Factors.
Formation Characteristics
Digital Radiography.
Fluoroscopy Equipment Operation
X-RAYS IN MOTION “Viewing dynamic studies of the human body”
X-ray radiography Seminar The "geometrical" unsharpness of X-ray images is limited by finite dimension of the source, i.e. the window on.
Fluoroscopy: Viewing Systems
Fluoroscopy Equipment
FLUOROSCOPY.
RAD 350 Chapt. 25 Fluoroscopy Fluoro’s primary function is to enable one to view the patient in “real time” with dynamic motion Invented by Thomas Edison.
DIGITAL FLUOROSCOPY.
Special Imaging Systems
F luoroscopy. 2 Fluoroscopy system 3 Different fluoroscopy systems  Remote control systems Not requiring the presence of medical specialists inside.
Fluoroscopy Intro to EQUIPMENT
FLUOROSCOPY EQUIPMENT
Spot Filming Cinefluorography Fluoro Room Shielding Radiation Areas.
INVERSE SQUARE LAW. The picture above demonstrates the typical x-ray tube used to produce a point source of x-rays. Then as radiation exits the tube it.
Fluoroscopy Equipment. Introduction Fluoro: is dynamic radiographic examination Fluoroscopy is primarily domain of the radiologist The role of radiographer.
Control room High resolution flat screen monitors CCD image intensifiers are easily identified by their shape A modern fluoroscopic suite.
Equipment Operation and Maintenance. Grid-controlled tube Designed to turn on and off rapidly, providing short precise exposures Use in fluoro and in.
Fluoroscopy Q&A Robert L. Metzger, Ph.D..
Appendix H: Chapter 40: Fluoroscopy
What is being tested? (kVp, mA, mAs, mR?) What is the purpose of the HVL test? If an x-ray tube does not pass the HVL test what is done to correct the.
Image Quality Q&A. RAPHEX Diagnostic Question 2001 D9: The imaging system which is best for visualizing small high contrast objects is: A. Computed tomography.
RAD 150 Fluoroscopy Fluoro’s primary function is to enable one to view the patient in “real time” with dynamic (motion)
RAD 354 Chapt. 25 Fluoroscopy Fluoro’s primary function is to enable one to view the patient in “real time” with dynamic motion.
Fluoroscopy. Real-time imaging Most general-purpose fluoroscopy systems use TV technology, operating at 30 frames/sec May be recorded (barium swallow.
Doses in Fluoroscopy factors influencing patient doses Colin Martin and David Sutton.
Fluroscopy and II’s. Fluroscopy Taking real time x-ray images Requires very sensitive detector to limit the radiation needed Image Intensifier (II) is.
Analog X-ray Imaging Recommended Book: Walter Huda, REVIEW OF RADIOLOGIC PHYSICS By: Maisa Alhassoun
Sergeo Guilbaud School of Radiologic Sciences
FLUOROSCOPY & IMAGE INTENSIFICATION Sergeo Guilbaud, RT School of Radiologic Sciences.
IAEA International Atomic Energy Agency Fluoroscopy Radiation Sources in medicine diagnostic Radiology Day 7 – Lecture 1(2)
Exposure Factors or Prime Factors
Diagnostic Radiology II X-ray Tubes. Anode angle Anode angle defined as the angle of the target surface with respect to the central ray in the x-ray field.
Chapt 29 Viewing the Digital Image
Resident Physics Lectures
Introduction to and History of Fluoroscopy Radiation Protection
Image quality and Performance Characteristics
Imaging System Components
Fluoroscopic Unit Thomas Edison 1896.
Equipment Operation and Maintenance
Fluoroscopy: Dynamic Imaging Systems
Resident Physics Lectures (Year 1)
Resident Physics Lectures
Pediatric Fluoroscopy
Fluoroscopy Dynamic Image Recording Systems
Fluoroscopic Brightness Control Systems
Fluoroscopic Image Intensifier Image production
Mobile Fluoroscopy Based on:
Mobile Fluoroscopy Based on:
Fluoroscopy Dynamic Image Recording Systems
Fluoroscopy: Viewing Systems
Resident Physics Lectures (Year 1)
Introduction to and History of Fluoroscopy Radiation Protection
Pediatric Fluoroscopy
Fluoroscopy Static Image Recording Systems
Fluoroscopy – Viewing Systems TV Monitors
Fluoroscopy – Viewing Systems Optical Mirrors TV Camera & TV Camera Tubes Charge Coupled Devices (CCD) Based on: Principles of Radiographic Imaging, 3rd.
Pediatric Fluoroscopy
Fluoroscopy – Viewing Systems TV Monitors
Fluoroscopy Static Image Recording Systems
Flouroscopic imaging.
Fluoroscopic Image Quality Considerations
Digital Fluoroscopy PPT created by: Jed Miles, BSRS, RT(R), CRT-CA
Fluoroscopy: Static Image Recording Systems
Mobile Fluoroscopy.
Fluoroscopic Image Quality Considerations
Brightness Control Systems
Presentation transcript:

Fluoroscopic Image Intensifier Image Production

Minification Gain Example Minification Gain = Input screen diameter2 Occurs as a result of the same number of electrons produced at the large input screen being compressed into an area of the smaller output screen Minification Gain = Input screen diameter2 Output screen diameter2 ^ is the symbol for squared if using MS Excel Typical output screen diameter is 1” (2.5cm) What is the minification gain if the input screen diameter is 12” and the output screen diameter is 1”? Example Minification Gain = 122 12 = 144 1 = 144

Flux Gain Measurement of increase in light photons due to conversion efficiency of the output screen Flux gain deals only with the gain accomplished by the electron to light conversion at the output screen This will cause a decrease in image quality exactly like an intensifying screen causing decreased resolution in diagnostic images due to the penumbral effect of individual phosphor crystals If the output phosphor produces 50 light photons for each electron that strikes it, the flux gain would be 50. Example

Total Brightness Gain = Minification gain x Flux gain Measurement of the multiplication of the light photons at the output phosphor compared to the incident radiation at the input phosphor Total Brightness Gain = Minification gain x Flux gain What is the total brightness gain from an II tube with a minification gain of 36 and a flux gain of 60? Example Total Brightness Gain = 36 x 60 = 2,160

Another Brightness Gain Measure Conversion factor = cd/m2 or (cd/m2/mR/sec) mR/sec Ratio of light intensity of output phosphor to input exposure rate Conversion factor = Intensity of output phosphor / mR / sec Candelas (cd) is a measurement of output phosphor light intensity (which is a unit of luminous intensity) Typical values of modern image intensifier systems are cd/m2 mR/sec 80-250 This represents a gain of 8,000 – 25,000 times

Brightness Gain Deterioration As much as 10% a year due to aging of input and output phosphors (Diagnostic intensifying screens also age in the same manner) Brightness gain can be evaluated by monitoring radiation dose required to obtain diagnostic image using a standardized fluoroscopic phantom When required dose increases to a significant amount, the image intensifier tube should be replaced

Magnification Increased voltage to electrostatic lenses results in increased electron deflection Causes focal point to move or shift toward input screen Results in reduced primary field of view being magnified prior to the striking the output screen Creates “analog electronic magnification” so to speak Electron flow of reduced primary field of view being magnified via a focal point shift prior to striking the output screen

input screen image cannot be viewed by output screen Magnification Tubes These tubes often termed: Multi- Dual- Triple- or Quad-field intensifiers Magnification image intensifiers are capable of 1.5 – 4.0 levels of magnification Resolution can be increased from ~ 4 lp/mm to 6 lp/mm When operating in magnification mode, electron image created at outer edge or periphery of input screen image cannot be viewed by output screen Therefore, primary photon beam should be collimated to useful viewing field of intensifier when operating in a magnification mode Minimal magnification used for most studies

Magnification Factor Example Magnification = Diameter of image intensifier input screen Diameter of input screen utilized during magnification Image intensification tubes described according to diameter of area of input screen utilized Magnification is in direct proportion to ratio of input screen to output screen diameters A 23/15cm dual-focus tube has a 9” (23cm) input screen when operating normally and uses a 6” (15cm) area when operating in a magnified mode. What is the magnification factor? Example Magnification = 9” / 6” = 1.5 magnification factor

Magnification Effects Minification gain reduced due to fewer photoelectrons being incident at the output phosphor as a result of magnification (Results in a dimmer image) To maintain brightness, tube mA must be increased to produce more x-rays Results in more x-rays per unit area present at input phosphor and, consequently, increased patient dose Magnification modes will result in increased: Spatial Resolution Contrast Patient Dose Rate

Please close this PowerPoint presentation, and continue the lesson. What’s Next? Please close this PowerPoint presentation, and continue the lesson. Presented by Based on: Principles of Radiographic Imaging, 4th Ed. By: R. Carlton & A. Adler Radiologic Science for Technologists, 8th Ed. By: S. Bushong Syllabus on Fluoroscopy Radiation Protection, 6th Rev. By: Radiologic Health Branch – Certification Unit