Differential Pressure Calibration Static pressure range up to 80MPa Model 5500 Differential Pressure Calibration Static pressure range up to 80MPa 5 Pa to 78MPa ΔP Accuracy ≤ 0,01% of ΔP reading Static pressure effect down to 10Pa Floating time longer 20 minutes Dual mounting post piston cylinder
Dual mounting system in one unit Differential Model 5500 Dual mounting system in one unit 1 set of weights and piston cylinder to tare (no calibration required) 1 set of mass and 1 piston cylinder to measure (calibrated and traceable) Oil lubricated piston for gas operation Trim weight down to 0.01g PCUs 0.5MPa/kg, 1MPa/kg, 2MPa/kg
Maximum Static Pressure (MPa) f = (PCU kn x Mass Set) resolution Model 5500 0,01g (2Pa) (5Pa) (10Pa) (20Pa) Accuracy = Static Pressure Effect + PCU Aef uncertainty Tell me more…