Intrinsic Uniformity QC Setup

Slides:



Advertisements
Similar presentations
Topic 8. Gamma Camera (II)
Advertisements

Instruments for Radiation Detection and Measurement Lab # 4.
Gamma Camera Quality Control
Experiences of using the Philips CardioMD Camera University Hospital Birmingham Foundation NHS Trust.
King Abdul-Aziz University Diagnostic Radiology Department MS.Nouf Al-Zahrani DR. Saddiq Jastniah Introduction to Nuclear Medicine 2 nd year.
Resident Physics Lectures
Chapter 8 Planar Scintigaraphy
W. Sautter Normal Line Normal Line ii rr ii rr Glass n = 1.5 Air n =1.0  r = angle of refraction  i = angle of incidence Light travels.
1. 2  Ray Imaging u Inject a pharmaceutical labelled with a radioactive nuclide  Detect where  -rays are coming from u Local concentration in patient.
Nuclear Medicine Spring 2009 FINAL. 2 NM Team Nuclear medicine MD Nuclear medicine MD Physicist Physicist Pharmacist Pharmacist Technologist Technologist.
Medical Imaging Mohammad Dawood Department of Computer Science University of Münster Germany.
Medical Imaging Mohammad Dawood Department of Computer Science University of Münster Germany.
QC And NEMA In The Nuclear Arena
BME 560 Medical Imaging: X-ray, CT, and Nuclear Methods
Gamma Camera Technology
Topics in Medical Physics Xiaoming Zheng, PhD. School of Dentistry and Health Sciences December 2009, Chengdu.
Lotte Verbunt Investigation of leaf positioning accuracy of two types of Siemens MLCs making use of an EPID.
Introduction to Medical Imaging
Instruments for Radiation Detection and Measurement
Two-Slit Interference, One Photon at a Time By: Kevin Masson.
RAD 466-L 8 by Dr. Halima Hawesa
Rice “Single-Pixel” CS Camera random pattern on DMD array DMD single photon detector image reconstruction object Richard Baraniuk, Kevin Kelly, and students.
Medical Image Analysis Dr. Mohammad Dawood Department of Computer Science University of Münster Germany.
Md.Kausher ahmed Electrical department. Biomedical engineering Code:6875.
Nuclear Medicine: Planar Imaging and the Gamma Camera Katrina Cockburn Nuclear Medicine Physicist.
Of Photons and Electrons Compton Effect, Pair Production and X-Rays.
Nuclear Medicine Physics
Nuclear Medicine Physics Jerry Allison, Ph.D. Department of Radiology Medical College of Georgia Image Artifacts in Nuclear Medicine Part A.
Quality Assurance.
Grids George David Associate Professor Department of Radiology
1 Nuclear Medicine SPECT and PET. 2 a good book! SR Cherry, JA Sorenson, ME Phelps Physics in Nuclear Medicine Saunders, 2012.
CAT scanners and gamma cameras Unit 16 Waves. Learning objectives Describe how CAT scanners can produce a much more detailed image than conventional X-rays.
Nuclear Medicine Physics and Equipment 243 RAD 1 Dr. Abdo Mansour Assistant Professor of radiology
Nuclear Medicine Physics Jerry Allison, Ph.D. Department of Radiology Medical College of Georgia Image Artifacts in Nuclear Medicine Part B.
Nuclear Medicine Instrumentation 242 NMT 1 Dr. Abdo Mansour Assistant Professor of radiology
Monday Case of the Day History: Multiple examinations from a CT scanner using a standard helical scan protocol show diffuse hypointense regions of artifact.
Monday Case of the Day History: During acceptance testing of a new general radiographic room, tube output measurements differed between the small and large.
Thursday Case of the Day What is the source of the artifact in the upper right-hand quadrant of the image? History: An AP view of the pelvis was acquired.
Chapter-2 The Planar Imaging Important points in chapter 2 (chapter 13 from the book) The gamma camera (the basic principles of the gamma camera) The types.
PET Imaging Positron Emission Tomography
Muhammad Musaddiq.
Case of the Day The likely cause of the poor image quality is: A. Incorrect collimators used B. Energy window incorrectly set over scatter below the 140.
Radiation detectors Ion chamber 2. Geiger Muller counter (GM).
Nuclear Medicine Physics
Nuclear Medicine Physics
Physics Case of the Day - Sunday
Sunday Case of the Day Physics
Quantitative Nuclear Medicine Imaging
Image quality and Performance Characteristics
Sunday Case of the Day Physics
Image quality and Performance Characteristics
Sunday Case of the Day Physics
Resident Physics Lectures
Gamma Camera.
CLRS 321: Nuclear Medicine Physics and Instrumentation I
Unit V: Introduction to Gamma Camera
Lecture 1: Development of the Analog Gamma Camera
Physics of Nuclear Medicine, SPECT and PET
CLRS 321 Nuclear Medicine Physics & Instrumentation I
QC And NEMA In The Nuclear Arena
Phases of the Moon The daily changes in the appearance of the moon as viewed from earth. It takes about 29.5 days for the moon to orbit the Earth The sun.
Tuesday Case of the Day Physics
Wednesday Case of the Day Physics
Lecture 1—Basic Principles of Scintillation
CLRS 321 Nuclear Medicine Physics and Instrumentation 1
Resident Physics Lectures
Function and Structure in
Gamma Camera Ilker Ozsahin Oct
Development of a Large Area Gamma-ray Detector
Remote sensing in meteorology
Presentation transcript:

Intrinsic Uniformity QC Setup Sunday Case of the Day Physics Patricia Svolos1, Matt Vanderhoek2 and Beth Harkness2 1McGovern Medical School, University of Texas Health Science Center at Houston, TX 2Henry Ford Health System, Detroit, MI Detector 1 Detector 2 Point Source Intrinsic Uniformity QC Setup PASS FAIL Extrinsic Uniformity Intrinsic History: Daily QC (Extrinsic Uniformity) of a dual head SPECT/CT system using a 153Gd rod source passes (uniformity ≤ 5%). The weekly QC (Intrinsic Uniformity) is performed on the same day using a 99mTc point source and fails for Detector 1 (uniformity ≥ 4%). What is the cause of the linear dark band artifact present on the left side of the Intrinsic Uniformity QC image for Detector 1? The collimator on Detector 1 was contaminated with Tc-99m. Bad linearity correction resulting in a linear band at the left side of the image. External source of radiation being partially blocked by the edge of the gamma camera. Band of defective photomultiplier tubes on the left side of the camera crystal.

Diagnosis: C. External source of radiation being partially blocked by the edge of the camera crystal. Intrinsic QC is performed without the collimators on the system. Contaminated waste in the trash was in the room during QC testing. The 180o orientation of the camera heads was such that Detector 1 faced down toward the floor while Detector 2 faced up toward the ceiling. Consequently, the contaminated trash (on the floor) irradiated the face of Detector 1 (yellow lines) but not the face of Detector 2. Photons were attenuated at certain angles by the leaded light rail on Detector 1, thus creating a linear “shadow” of the light rail on the edge of the acquired uniformity image. Detector 1 Detector 2 Detector 1 Detector 2 Light rail

Discussion: A. The collimator on Detector 1 was contaminated with Tc-99m Incorrect Collimators are used only during extrinsic uniformity QC. Collimator contamination of Detector 1 with Tc-99m would likely result in an artifact on the extrinsic uniformity QC image with a random pattern (looking more “splotchy”) rather than a linear dark band artifact. The acquired extrinsic uniformity image for Detector 1 was free of such artifacts.

Discussion: B. Bad linearity correction resulting in a linear band at the left side of the image Incorrect Artifacts due to bad linearity correction usually have a “tubular” pattern rather than linear. Image below shows a typical artifact pattern of all PMT being visible if linearity correction fails.

Discussion: D. Band of defective photomultiplier tubes on the left side of the camera crystal Incorrect It is unlikely that a set of defective photomultiplier tubes near the edge of the crystal would result in a linear dark band artifact. Image below shows a typical artifact due to bad photomultiplier tubes.

References/Bibliography: 1. Physics in Nuclear Medicine. Cherry S.R., Sorenson J.A., Phelps M.E.