“Closing the loop” on palpitations “Closing the loop” on palpitations. A report of unexpected communication between a loop recorder symptom marker and pacemaker Siva S. Ketha, MD, Rebecca N. Revell, RN, Thomas R. Flipse, MD, Fred M. Kusumoto, MD, FHRS, K.L. Venkatachalam, MD, FHRS HeartRhythm Case Reports Volume 3, Issue 8, Pages 377-379 (August 2017) DOI: 10.1016/j.hrcr.2017.04.003 Copyright © 2017 Heart Rhythm Society Terms and Conditions
Figure 1 Implantable loop recorder data at the time of a “symptom-marked event.” HeartRhythm Case Reports 2017 3, 377-379DOI: (10.1016/j.hrcr.2017.04.003) Copyright © 2017 Heart Rhythm Society Terms and Conditions
Figure 2 Device electrogram from the ADVISA pacemaker at the time of the same “symptom marked event” using the Patient Assistant Model PA96000. HeartRhythm Case Reports 2017 3, 377-379DOI: (10.1016/j.hrcr.2017.04.003) Copyright © 2017 Heart Rhythm Society Terms and Conditions