3E7_05151 wafer 11 IV curve Noise ~122e Source scans bad and independent on HV (10,20,30V): Only few pixels work with the correct charge collection
noise Noise is too low for almost all the pixels. The few that have higher noise than behave correctly in the source scan (see next) . To be checked systematically…
some pixels have the average at 30ToT Others to 5Tot Average ToT per pixel some pixels have the average at 30ToT Others to 5Tot Many do not respond
A good channel… A bad one
It is not noise Only few pixels have hits and the spectra is different.
In terms of charge
Very similar results for another device on the same wafer Noise is completely independent on the HV (always 127e) See next for correlation noise/good channels
3E7_06152 wafer 15 IV curve HV= 10V Noise ~230e
Noise Average noise 230 Distribution is reasonable
Hit map is more uniform
Average ToT Per pixel
And charge Pixel average (top) All hit entries bottom (higher peak to be better understood if due to single pixels or not)
Another chip: 3e11_07152 wafer 15 10 V 5V