Reference Membrane Location
Reference Membrane
Span Check Perform 8 minute count Compare mass value with ABS value ABS must be within 5% or BAM will log CAL error (D error code)
Calibration Test Screen CALIBRATION MODE REF MBRN: < COUNT (Io): 634000 COUNT (I): 556234 CAL MASS M: 0.801 mg/cm2 START STOP EXIT
Hands On Workshop – Reference Membrane Remove Reference Membrane Inspect Reference Membrane Reinstall Reference Membrane Inspect mechanical movement of Reference Membrane Perform Reference Membrane Count/Mass Test in TEST/CALIBRATE Screen Compare Mass value with ABS value