Radiometric Calibration Stuart F. Biggar, U of AZ Kurtis J. Thome, U of AZ Simon J. Hook, JPL.

Slides:



Advertisements
Similar presentations
Computer Vision Radiometry. Bahadir K. Gunturk2 Radiometry Radiometry is the part of image formation concerned with the relation among the amounts of.
Advertisements

Summary of Terra and Aqua MODIS Long-term Performance Jack Xiong 1, Brian Wenny 2, Sri Madhaven 2, Amit Angal 3, William Barnes 4, and Vincent Salomonson.
Tuz Gölü – Site Characteristics Larry Leigh 1, Dennis Helder 1, David Aaron 1 I. Behnert 2, A.Deadman 2, N.Fox 2 U. M. Leloğlu 3, H. Özen 3, Derek Griffith.
VIIRS Reflective Solar On-orbit Calibration and Performance Jack Xiong and Jim Butler Code 618.0, NASA/GSFC, Greenbelt, MD CLARREO SDT Meeting, NASA.
Polarization measurements for CLARREO
Brown Bag Lunch Lecture ABI Calibration
Radiometric Corrections
Clouds and the Earth’s Radiant Energy System NASA Langley Research Center / Atmospheric Sciences Methodology to compare GERB- CERES filtered radiances.
Orbits and Sensors Multispectral Sensors
Line scanners Chapter 6. Frame capture systems collect an image of a scene of one instant in time The scanner records a narrow swath perpendicular to.
Remote sensing in meteorology
Reflected Solar Calibration Demonstration System Update J. McCorkel K. Thome D. Jennings B. McAndrew J. Hair D. Rabin A. Daw National Aeronautics and Space.
K-Space Associates, Inc. kSA BandiT: Band-edge Thermometry.
OL 750 Measurement Systems
Satellites and instruments How RS works. This section More reflection Sensors / instruments and how they work.
ASIC3 WorkshopLandsdowne, VA May 16-18, 2006 J. Harder Page 1 Calibration Status of the Solar Irradiance Monitor (SIM) : The Present and the Future Jerald.
Modeling the imaging system Why? If a customer gives you specification of what they wish to see, in what environment the system should perform, you as.
NOAA Research and Operations Marine Optical Buoy Design Review July 18-19, 2006 Plan for calibration and maintenance of AHAB Uncertainty Budget: Laboratory.
Improving Uncertainties of Non-Contact Thermometry Measurements Mark Finch Fluke Calibration.
…….CT Physics - Continued V.G.WimalasenaPrincipal School of radiography.
6-1 EE/Ge 157b Week 6 EE/Ae 157 a Passive Microwave Sensing.
2010 CEOS Field Reflectance Intercomparisons Lessons Learned K. Thome 1, N. Fox 2 1 NASA/GSFC, 2 National Physical Laboratory.
Sergey Mekhontsev National Institute of Standards and Technology Optical Technology Division, Gaithersburg, MD Infrared Spectral Radiance Scale.
NA62 Gigatracker Working Group Meeting 2 February 2010 Massimiliano Fiorini CERN.
Transmittance Measurement Presented by Dr. Richard Young VP of Marketing & Science Optronic Laboratories, Inc.
STATUS REPORT OF FPC SPICA Task Force Meeting March 29, 2010 MATSUMOTO, Toshio (SNU)
ATMOSPHERIC CORRECTION – HYPERSPECTRAL DATA Course: Special Topics in Remote Sensing & GIS Mirza Muhammad Waqar Contact:
LECTURER PROF.Dr. DEMIR BAYKA AUTOMOTIVE ENGINEERING LABORATORY I.
October 29-30, 2001MEIDEX - Crew Tutorial - Calibration F - 1 MEIDEX – Crew Tutorial Calibration of IMC-201 Adam D. Devir, MEIDEX Payload Manager.
Sphere Standards and Standard Spheres Dr. Richard Young Optronic Laboratories, Inc.
Digital Image Processing GSP 216. Digital Image Processing Pre-Processing – Correcting for radiometric and geometric errors in data Image Rectification.
NIRSpec Operations Concept Michael Regan(STScI), Jeff Valenti (STScI) Wolfram Freduling(ECF), Harald Kuntschner(ECF), Robert Fosbury (ECF)
Slide 1 Implementation of algorithmic correction of stray light in a pushbroom hyperspectral sensor > Karim Lenhard > Implementation of algorithmic.
1 An Observatory for Ocean, Climate and Environment SAC-D/Aquarius HSC - Radiometric Calibration H Raimondo M Marenchino 7th SAC-D Aquarius Science Meeting.
TELECOMMUNICATIONS Dr. Hugh Blanton ENTC 4307/ENTC 5307.
Radiometric Correction and Image Enhancement Modifying digital numbers.
DECam Daily Flatfield Calibration DECam calibration workshop, TAMU April 20 th, 2009 Jean-Philippe Rheault, Texas A&M University.
7 elements of remote sensing process 1.Energy Source (A) 2.Radiation & Atmosphere (B) 3.Interaction with Targets (C) 4.Recording of Energy by Sensor (D)
CDE CDR, September 14, 2004 Your Position, Your Name 1 GATS AIM Science Team Meeting January 23-24, 2007 CIPS Calibration Review, Aimee Merkel, Bill McClintock.
ASTEROSEISMOLOGY and SEARCH for EXOPLANETS Vienna - September 18 th 2001 Optical performance characterization Point Spread Function (PSF) : Telescope +
Headwall Instrument Overview Laboratory Characterizations Geo-Location Field Characterization Data Product Description References.
1 Reflected Solar Calibration Demonstration System - SOLARIS K. Thome, D. Jennings, B. McAndrew, J. McCorkel, P. Thompson NASA/GSFC.
SNAP Calibration Program Steps to Spectrophotometric Calibration The SNAP (Supernova / Acceleration Probe) mission’s primary science.
A Method for Correcting for Telescope Spectral Transmission in the Geosynchronous Imaging Fourier Transform Spectrometer (GIFTS) John D. Elwell, Deron.
CALCON September B. Carol Johnson, Steven W. Brown and Howard W. Yoon National Institute of Standards and Technology Gaithersburg, MD Radiometric.
Radiometric Calibration of the ASTER Thermal Infrared (TIR) Subsystem Simon J. Hook NASA/JPL.
Electro-optical systems Sensor Resolution
# x pixels Geometry # Detector elements Detector Element Sizes Array Size Detector Element Sizes # Detector elements Pictorial diagram showing detector.
UNIT 2 – MODULE 5: Multispectral, Thermal & Hyperspectral Sensing
Testing of Highly Accurate Blackbodies Harri Latvakoski Mike Watson Shane Topham Mike Wojcik SDL/
Orbits and Sensors Multispectral Sensors. Satellite Orbits Orbital parameters can be tuned to produce particular, useful orbits Geostationary Sun synchronous.
1 CLARREO Advances in Reflected Solar Spectra Calibration Accuracy K. Thome 1, N. Fox 2, G. Kopp 3, J. McCorkel 1, P. Pilewskie 3 1 NASA/Goddard Space.
Digital Light Sources First introduced in 2001.
Pre-launch Characteristics and Calibration
NOAA VIIRS Team GIRO Implementation Updates
Lunar observation data set preparation
Lessons from MODIS Calibration and Characterization
MODIS and VIIRS Reflective Solar Bands Calibration Methodologies and
An Overview of MODIS Reflective Solar Bands Calibration and Performance Jack Xiong NASA / GSFC GRWG Web Meeting on Reference Instruments and Their Traceability.
Calibration and Performance MODIS Characterization Support Team (MCST)
MODIS Lunar Calibration Data Preparation and Results for GIRO Testing
San Diego Conference 2 August 2001
Image Stabilization System (ISS)
Using the Moon for Sensor Calibration Inter- comparisons
Instrument Considerations
Detective Quantum Efficiency Preliminary Design Review
S-NPP Visible Infrared Imaging Radiometer Suite (VIIRS) Lunar Calibration using GSICS Implementation of the ROLO model (GIRO) for Reflective Solar Bands.
MODIS L1B Data Product Uncertainty Index Jack Xiong (Xiaoxiong
Sensor calibration.
Remote sensing in meteorology
Presentation transcript:

Radiometric Calibration Stuart F. Biggar, U of AZ Kurtis J. Thome, U of AZ Simon J. Hook, JPL

Outline Preflight – Biggar –Characterization –Calibration –On-board calibration systems –Sensor artifacts In-flight VNIR and SWIR – Thome In-flight TIR - Hook

Calibration (absolute) Why do we calibrate? –Understand sensor performance –L1 data in physical units (at sensor) Radiance in Watts/m 2 -sr-micrometer (or similar) Determine sensor linearity Derived units –Reflectance –Geophysical units (temperature, etc) –Comparison to other sensors –Atmospheric correction –Reasonable looking imagery

Calibration (2) What do we need to be able to calibrate? –Stable instrument Over time With changing environment –Characterized instrument Spectral response Spatial response Noise Systematic errors –Image artifacts –Stray light –Stable calibration source

Characterization Piece part measurements –Filter transmittance –Lens and/or mirror transmittance and reflectance –Mirror reflectance and scatter (BRDF) –Lamp output, stability, longevity –Detector Quantum efficiency, spectral response, noise, temperature dependence

Filter spectral characteristics

Normalized Response

Characterization (2) Assembly level –Alignment –Relative spectral response (of telescope) –Noise – amplifier assembly for example –Field-of-view –Stray light

Characterization (3) Sensor level –Relative spectral response –Field-of-view –Modulation transfer function –Noise Random Coherent –Stray light

ASTER design features (1) Long linear array detectors with long strip filters in VNIR and SWIR – pushbroom –Good SNR due to longer integration time –Detector and filter uniformity issues –Many detectors (5000 in VNIR for example) Gain Offset Spectral response Large focal plane –Post detector electronic chains (amps, A/D, etc) –Stripes in images even of constant radiance scenes

ASTER L1A, bands 1and 13

ASTER design features (2) Short linear arrays (10 elements) with a scan mirror in the TIR sensor – whiskbroom –Shorter integration time –Larger pixels to improve the SNR –Limited number of detectors –Individual amplifiers –Stripes in images but repeated along track

ASTER design features (3) Three separate sensors –3 telescopes Individually pointed Alignment Rotation of images Focal plane distortions –3 manufacturers Different design practices Different measurement practices Different calibration methodologies and equipment

Radiometric calibration Preflight in the laboratory –Well characterized and controlled source –Reasonable but not normal environment –May (or may not) approximate normal imaging operation mode Scan mirror may not be operating Source is usually smaller angular extent than the earth –Uniform source is not a normal image

On-board calibrators VNIR –Lamp based source with monitors SWIR –Lamp based source with monitors TIR –Blackbody that can be heated

Philosophy Careful preflight calibration –Determine absolute response so we can convert from DN to radiance –At the same time, run the on-board systems to transfer a radiance value to them using the sensor (ASTER) –Determine noise (random and coherent) –Look for image artifacts –Use preflight values after launch Update after launch as sensors usually change –On-board –Vicarious (ground reference and moon)

Preflight source for VNIR/SWIR Large aperture integrating sphere –1 meter diameter sphere –Round output port –Multiple lamps Internal surrounding the port External through small ports –High reflectance coating on the sphere wall Barium sulfate (BaSO 4 ) Spectralon (sintered PTFE) Gold sandpaper (rough, diffuse surface)

40 BaSO 4 integrating sphere

Integrating sphere Advantages –Uniform radiance across port (one % or so) –Lambertian (radiance independent of direction) –Multiple output levels Change number of lamps (MELCO) Change voltage/current of all lamps (voltage for NEC sphere) Change aperture size between external lamp and sphere –Reasonably stable with good control of lamp (V or I) –Repeatable if temperature is controlled Disadvantages –Low output in blue relative to NIR and SWIR

TIR preflight cal source Large area blackbody –Measure the temperature precisely and accurately –Measure (or compute) emissivity –Compute radiance output Operated in a vacuum chamber with the sensor Vary temperature of source to get multiple radiance values –Absolute calibration –Linearity correction (output usually fit to something other than a gain and offset – quadratic is common)

Calibration chain NRLM (Japanese equivalent of NIST) calibrated a set of fixed point blackbody simulators Calibration transferred to portable, variable temperature blackbody simulators Calibration transferred to the ASTER calibration sources (Spheres and TIR calibration blackbody) VNIR and SWIR spheres were measured at NEC and MELCO with a set of transfer radiometers from NIST, NRLM, GSFC, and University of Arizona NRLM (now AIST) and NIST collaborate to ensure that their scales are consistent

Are preflight values usable? Sometimes … Launch may cause a shift in a sensor performance –Operating temperatures may be different –Something may have moved On-board lamps may change output –Convection inside the lamp may be different –Aging of the lamp Emissivity of the on-board blackbody may change Mirror reflectance may change Filters/detectors/amplifiers may change There may have been unexpected features

Unexpected features VNIR –On-board calibrator monitor is off-scale –Change in output is more than expected so dynamic range of A/D is too small SWIR –Unexpected stray light causes crosstalk –Present during preflight calibration and in all in-flight data

VNIR OBC VNIR OBC has two monitors for each of two redundant calibrators –One monitor diode at lamp –One monitor diode at output They track but at different rates We really want the output monitor to determine how well the on-board calibrator is working but we have only the lamp monitor on-scale

VNIR OBC output monitor NEC selected an expected output range based on preflight measurements, OPS experience with a similar calibrator, and desire to maximize resolution Output has fallen to below the lowest expected monitor output Telemetry value for monitor is now a flat line (signal – offset < zero)

SWIR crosstalk SWIR has multiple spectral bands –6 linear PtSi detector arrays –Spectral selection filters over the arrays –Not all light hitting the detector is absorbed –Light hitting between the detectors is reflected –Some reflected light is reflected back down by the filters Optical crosstalk

Stretched RGB of a Japanese Island, SWIR bands 4,5, & 9, 400x400 pixels

SWIR Crosstalk Present in all images –Preflight calibration –In-flight calibration –normal images Visible in images with strong contrast Not visible but present in others Currently NOT corrected for in normal processing of L1 data

Crosstalk correction If we know the amount of light leaking from one pixel to each other pixel in all the SWIR bands, we can correct for it Preflight data for MTF determination would probably contain much of the needed information but it was recorded only for the band under test –Scan a line source across the array in both directions –Scan a point (pixel size or less) in both directions It is possible to infer the correction from images with strong contrast (coast lines, islands, moon, and similar), however it is difficult and incomplete

SWIR correction Japanese team has developed a beta level, Windows (Win32) based, correction program. It does one scene at a time operating on a L1B HDF input file and writing a corrected L1B HDF file. US team is developing a program that is run as part of L2 processing. It starts with L1A data. For example, you will be able to order atmospherically-corrected, crosstalk-corrected surface reflectance at L2.

Corrected image, SWIR RGB with 4,5, & 9

SWIR Crosstalk Correction Qualitatively improves image Largest effect is band 4 into 5 and 9 However, any band should leak into all others (including itself) with the strongest effect on adjacent bands on the focal plane –Band 4 has higher typical radiance than the others –Japan correction has only band 4 into others There is interplay between crosstalk and water vapor absorption, especially in band 9

Conclusions (preflight) ASTER was calibrated preflight –VNIR accuracy was probably within spec –SWIR accuracy is poorer due to crosstalk –TIR was probably within spec Preflight calibration is probably not appropriate at this point –Change in sensor (VNIR and some TIR bands) –Crosstalk in SWIR Calibration is being updated