Air entrapment causing early inappropriate shocks in a patient with a subcutaneous cardioverter-defibrillator  Sing-Chien Yap, MD, PhD, Rohit E. Bhagwandien,

Slides:



Advertisements
Similar presentations
Implantation of the subcutaneous implantable cardioverter-defibrillator with retroperitoneal generator placement in a child with hypoplastic left heart.
Advertisements

Subcutaneous implantable cardioverter-defibrillator placement in a patient with a preexisting transvenous implantable cardioverter-defibrillator  Mohammad-Ali.
Inhibition of pacing in a dependent patient with an implantable cardioverter-defibrillator and a left ventricular assist device  Michele Murphy, MD, Timothy.
Volume 11, Issue 8, Pages (August 2014)
A new approach for implantation of a cardiac resynchronization therapy–defibrillator in a patient with bilateral pectoral neurostimulation devices  Saori.
Inhibition of pacing in a dependent patient with an implantable cardioverter-defibrillator and a left ventricular assist device  Michele Murphy, MD, Timothy.
Volume 8, Issue 12, Pages (December 2011)
Erroneous shock by an AED: Importance of obtaining AED tracing to prevent inappropriate ICD implantation  Brian J. Cross, MD, Mark S. Link, MD, FHRS 
Swimming pool saline chlorination units and implantable cardiac devices: A source for potentially fatal electromagnetic interference  John Wight, BS,
Inappropriate shock from delayed T-wave oversensing by a subcutaneous implantable cardioverter-defibrillator after septal myectomy for hypertrophic cardiomyopathy 
Recurrent commotio cordis: Déjà vu
Malignant transformation of a chronically infected implantable cardioverter-defibrillator pocket  Donna Kang, MD, Khuyen Do, MD, Jonathan Nattiv, MD,
Combination of a leadless pacemaker and subcutaneous defibrillator: First in-human report  Blandine Mondésert, MD, Marc Dubuc, MD, FHRS, Paul Khairy,
Successful management of wound dehiscence after implantation of a subcutaneous implantable cardioverter-defibrillator without device removal  Mitsuhiro.
Breast tissue expanders and implantable cardioverter-defibrillator: An unusual interaction  Malik Muhammad Khurram Sher Khan, MD, Malik Muhammad Humayun.
Unusual case of late dislodgment of a superior vena cava coil
T-wave oversensing during drug challenge test after subcutaneous implantable cardioverter-defibrillator implantation in a patient with Brugada syndrome 
Surgical and electrophysiological considerations in the management of a patient with a subcutaneous implantable cardioverter-defibrillator undergoing.
Inappropriate defibrillator shock during gynecologic electrosurgery
Anne Barnett, PA-C, Lee Eckhardt, MD, FHRS, Miguel A. Leal, MD, FHRS 
Respiratory rate trending as a cause for atrial lead noise: A first report in an implantable cardioverter-defibrillator patient  Isla McClelland, MD,
Remote magnetic navigation–guided ventricular tachycardia ablation with continuous- flow mechanical circulatory support  Anna M.E. Noten, MD, Nicolas M.
Recurrent idiopathic polymorphic ventricular tachycardia/ventricular fibrillation successfully treated by cardiac sympathetic denervation  Faisal Matto,
Darren C. Tsang, BS, Roger G. Carrillo, MD, MBA, FHRS 
Repeated molecular genetic analysis in Brugada syndrome revealed a novel disease- associated large deletion in the SCN5A gene  Anders Krogh Broendberg,
T-wave oversensing from postural changes: A rare cause of inappropriate shock from a subcutaneous defibrillator  Muhammad R. Afzal, MD, Auroa Badin, MD,
‘Runaway’ vagal nerve stimulator: a case of cyclic asystole
Failure of ICD therapy in lethal arrhythmogenic right ventricular cardiomyopathy type 5 caused by the TMEM43 p.Ser358Leu mutation  Kasper Aalbæk Kjærgaard,
“Closing the loop” on palpitations
Substernal implantation of a subcutaneous implantable cardioverter-defibrillator in a patient with preexisting Hemodialysis Reliable Outflow graft  Thomas.
Daniel P. Morin, MD, MPH, FACC, FHRS  HeartRhythm Case Reports 
Repeat inappropriate defibrillator discharges following defibrillator implantation. What is the mechanism and treatment?  Jeremy W. Docekal, MD, Mark.
Sing-Chien Yap, MD, PhD, Corstiaan A
Pneumopericardium after epicardial catheter ablation detected with “bruit de moulin”  Rintaro Hojo, MD, Seiji Fukamizu, MD, PhD, Marina Arai, MD, Dai Inagaki,
Volume 15, Issue 1, Pages (January 2018)
An unusual approach to intractable AVNRT in a pediatric patient
Imre Hunyor, MBBS, DPhil, Hany S. Abed, MBBS, PhD, Raymond W
Subcutaneous implantable cardioverter-defibrillator placement in a patient with a preexisting transvenous implantable cardioverter-defibrillator  Mohammad-Ali.
Right-sided subcutaneous implantable cardioverter-defibrillator placement in a patient with dextrocardia, tetralogy of Fallot, and conduction disease 
Managing cross talk between a subcutaneous implantable cardioverter-defibrillator and a dual-chamber unipolar pacemaker system  Konstantinos Kossidas,
Inappropriate shocks in a patient with a subcutaneous implantable cardioverter- defibrillator  Sana M. Al-Khatib, MD, MHS, FHRS, Sasmrita Belbase, NP,
Appropriate use of genetics in a young patient with atrioventricular block and family history of sudden cardiac death  Johnni Rudbeck-Resdal, MD, Jens.
Oversensing of an unexpected atrial flutter
Inappropriate subcutaneous implantable cardioverter-defibrillator therapy due to R-wave amplitude variation: Another challenge in device management  Syeda.
Graft-versus-host disease: A rare complication of device implantation
Daniel W. Nelson, BS, Caroline Vloka, BA, Jule Wetherbee, MD 
Progressive electrical remodeling in apical hypertrophic cardiomyopathy leading to implantable cardioverter-defibrillator sensing failure during ventricular.
Undetected ventricular fibrillation in a single-chamber implantable cardioverter- defibrillator: When the far-field channel sees more than the intraventricular.
Successful implantation of a subcutaneous cardiac defibrillator in a patient with a preexisting deep brain stimulator  Yousef Bader, MD, Jonathan Weinstock,
Changing place, changing future: Repositioning a subcutaneous implantable cardioverter-defibrillator can resolve inappropriate shocks secondary to myopotential.
Subcutaneous implantable cardioverter-defibrillator implantation in a patient with bilateral pectoral deep brain stimulators  Thor Tejada, MD, Faisal.
Impact of a novel algorithm designed to reduce T-wave oversensing with the subcutaneous defibrillator in a patient with type I Brugada electrocardiogram 
Abdominal implantable cardioverter-defibrillator placement in a patient requiring bilateral chest radiation therapy  Ankur A. Karnik, MD, FHRS, CCDS,
Serious case of the hiccups
The runaway defibrillator…A case of an implantable cardioverter-defibrillator that failed communication and deactivation with a magnet  Mahmoud Assaad,
Supraspinatus pocket: An alternative pacemaker location for patients with no prepectoral access  Guillaume Abehsira, MD, Françoise Hidden-Lucet, MD, Estelle.
Nothing inside the heart – Combining epicardial pacing with the S-ICD
Shock-induced right ventricular pacing failure caused by a short circuit: Uncommon but life-threatening complication of the Riata lead  Itsuro Morishima,
Implantation of the subcutaneous implantable cardioverter-defibrillator with retroperitoneal generator placement in a child with hypoplastic left heart.
Unusual fracture in a Durata lead with shock coil fragmentation and cable externalization  Toshiaki Sato, MD, Kyoko Soejima, MD, Hideaki Yoshino, MD,
Systemic infection due to subcutaneous implantable cardioverter-defibrillator implantation: Importance of early recognition and treatment of device pocket-related.
Nanostim leadless pacemaker system: A longer waiting period after active fixation may reduce unnecessary repositioning  Hiro Kawata, MD, PhD, Pranav M.
Philip M. Chang, MD, FHRS, CEPS, Rahul N. Doshi, MD, FHRS, FACC 
Inappropriate shocks by subcutaneous implantable cardioverter-defibrillator due to T- wave oversensing in hyperkalemia leading to ventricular fibrillation 
Cardiac implantable electronic device malfunction after deceleration injury without obvious chest trauma  Amr F. Barakat, MD, Brian Cross, MD, Jonathan.
Use of topical lidocaine in eliminating mechanically stimulated ventricular fibrillation in a patient with short QT syndrome  Maria J. Farag, BS, Joseph.
Ventricular tachycardia slower than the rate cut-off of a subcutaneous cardiac defibrillator sensed and successfully treated as a result of oversensing 
Mobile subcutaneous implantable cardioverter-defibrillator leads to oversensing and inappropriate shocks  Omid Kiamanesh, MD, Jacob M. Larsen, MD, PhD,
Orthostatic increase in defibrillation threshold leading to defibrillation failure and prolonged cardiac arrest in a sitting position: Lessons from a.
Substernal subcutaneous implantable cardioverter-defibrillator lead placement for the management of inappropriate shocks  Khuyen Do, MD, Chin C. Lee,
Presentation transcript:

Air entrapment causing early inappropriate shocks in a patient with a subcutaneous cardioverter-defibrillator  Sing-Chien Yap, MD, PhD, Rohit E. Bhagwandien, MD, Tamas Szili-Torok, MD, PhD, Dominic A.M.J. Theuns, PhD  HeartRhythm Case Reports  Volume 1, Issue 3, Pages 156-158 (May 2015) DOI: 10.1016/j.hrcr.2015.02.010 Copyright © 2015 Heart Rhythm Society Terms and Conditions

Figure 1 Device electrogram at the time of an inappropriate shock showing oversensing of low-amplitude signals with subsequent shock. The alternate vector was used for sensing. Note baseline wandering of the electrogram. HeartRhythm Case Reports 2015 1, 156-158DOI: (10.1016/j.hrcr.2015.02.010) Copyright © 2015 Heart Rhythm Society Terms and Conditions

Figure 2 Lateral chest radiography at day 0 (A, D), day 4 (B, E), and day 10 (C, F) after implantation. Note complete resolution of subcutaneous air (arrows) at the proximal sensing electrode during short-term follow-up. HeartRhythm Case Reports 2015 1, 156-158DOI: (10.1016/j.hrcr.2015.02.010) Copyright © 2015 Heart Rhythm Society Terms and Conditions