Creating a Veterinary Radiographic Technique Chart

Slides:



Advertisements
Similar presentations
Technical Aspects: the machine the image
Advertisements

EVERYTHING YOU WANTED TO KNOW BUT WERE AFRAID TO ASK!!!!
[Radiography] Technique - Exposure Factors
Formulating Technique Charts
Technique Guidance Systems
Ch. 14 Four Radiographic Qualities Ch
Chapter 11 Prime Factors.
4 Producing Quality Radiographs.
Advanced Biomedical Imaging Lecture 4 Dr. Azza Helal A. Prof. of Medical Physics Faculty of Medicine Alexandria University.
Quality Control in Diagnostic Radiology
Introduction to Abdominal Radiology
Radiographic Quality Chapter 5.
Topic 3 Selection of kV High kV scattered radiation is energetic. Angle of rays is same as those of primary beam and is not deflected very much. Follows.
Chapter 17 The Grid So far we have discussed how kVp, patient size and collimation impact scatter radiation. As the part size and kVp increase, scatter.
Chapter 19 & 20 Image Quality & Techniques
Analyzing the Image Density. Density Overall blackening of the image.
Technical Artifacts and Errors
Chapter 9 Film & Film Holders.
Veterinary Radiology Safety and Procedures. What is a radiograph? When an xray beam (a form of electromagnetic radiation) penetrates tissue to form a.
TIMERS.
Radiographic Technique Evaluation
Most of the images recorded during conventional radiography are obtained with film/screen combination image receptors. Which in lessens the patient dose.
Resident Physics Lectures Christensen, Chapter 6Filters George David Associate Professor Department of Radiology Medical College of Georgia.
Quality Control.
Mammography Questions and Answers
Exposure Factors or Prime Factors
Preparing Variable kVp Technique Charts By Prof. Stelmark.
Chapter 10 Exposure Mathematics. Exposure Mathematics u We have seen in Chapter Seven that there are several exposure factors that are predetermined before.
Tube Exposure Factors Math Technique Contrast and Density
Analog X-ray Imaging Recommended Book: Walter Huda, REVIEW OF RADIOLOGIC PHYSICS By: Maisa Alhassoun
TECHNIQUE CHARTS SERGEO GUILBAUD CLINICAL COORDINATOR.
 Make and type of x-ray machine  Incoming line voltage  Kilovoltage  Milliamperage and exposure time  Focal film distance/source-imagine distance.
Week Seven.
Radiology. The Cathode… Provides a source of electrons and directs the electrons towards to anode. The cathode has a coiled wire filament that emits electrons.
AP PHYSICS 1 SUMMER PACKET Table of Contents 1.What is Physics? 2.Scientific Method 3.Mathematics and Physics 4.Standards of Measurement 5.Metric System.
Exposure Factors or Prime Factors
Producing An Image RVT: Chapter 6. Learning Objectives: Chapter 6 Understand the 4 factors of radiographic exposure and how each impacts the production.
Technique Guidance Systems By Prof. Stelmark. Anatomic Programming Anatomic programming, or anatomically programmed radiography (APR), refers to a radiographic.
Exposure Calculations
RADIATION PROTECTION 04/12/2016.
EVERYTHING YOU WANTED TO KNOW BUT WERE AFRAID TO ASK!!!!
GENERAL EXPOSURE FACTORS
Chapters 21 & 22.
Impact of Obesity on Medical Imaging and Image-Guided Intervention
Avian Radiology Birds are very well suited to radiographic study.
Creating a Technique Chart
Optimisation of Patient Protection for Radiography
Protection of Patients
Producing An Image Lavin: Chapter 6 CTVT:
Rad Tech 265.
Equipment, Terminology and Positioning
Pediatric Radiology By Dr. Wambani, J.S. Chief Radiologist
Factors Effecting the Production
Exposure Factors Chapter 4
Lavin: Chapters 22 Chart on page 305
Quality Control Testing of Screen Speed
Abdominal Positioning
Developing a technique Chart Chapter 9
Comparing Exposure Systems
What makes a Quality Image
RECORDED DETAIL.
Artifacts and Errors.
Image Production and Evaluation
RECORDED DETAIL.
Brightness Control Systems
Ways to Check for Divisibility
Divisibility Rules Dividing by 2
GRIDS PRESENTER-Dr.Vishwanath patil PG resident
Different Modalities in Imaging: An Overview
Using the Rule Normal Quantile Plots
Presentation transcript:

Creating a Veterinary Radiographic Technique Chart Amy Crane, DVM

Step 1: Get your radiology house in order Have your processor serviced Be sure that you are using screens that are all the same age, speed, and manufacturer Use a dog that weighs about 40 pounds but is not overweight

Step 2: Select your mAs In veterinary medicine, patient motion is a great concern We like to use a high mAs and low kVp technique

Step 2: Select your mAs cont. The following are recommended mAs based on a par (medium, 200) speed intensifying screen –Tabletop (no grid) extremity: –Thorax: –Abdomen: –Spine: Important to note the difference between the mAs settings in regions!!!!!!!!!!!

Step 2: Select your mAs cont If your x-ray machine only has one mA station you just select your mA and pick the speed to give you the desired mAs. For example: if you have a 300mA station and you wanted to use a 5mAs for a thoracic technique chart you would pick –300mA x 1/120sec = 2.5mAs (extremity) –300mA x 1/60sec = 5mAs (thorax) –300mA x 1/40sec = 7.5mAs (abdomen) –300mA x 1/30sec = 10mAs (spine)

Step 2: Select your mAs cont If your x-ray machine has two mA stations I suggest that for extremity techniques you use the smaller of the mA stations (usually 100mA) and for all other applications you use the larger mA station (usually 300mA). In this case your settings would be as follows:

Step 2: Select your mAs cont –100mA x 1/40sec = 2.5mAs (extremity) –300mA x 1/60sec = 5mAs (thorax) –300mA x 1/40sec = 7.5mAs (abdomen) –300mA x 1/30sec = 10mAs (spine)

Step 2: Select your mAs cont *****The previous suggestions were for par (medium, 200) speed intensifying screens

Step 3: Select your initial kVp This step is to create a “perfect” radiograph To do that we need to find a kVp to go along with our suggested mAs setting

Step 3: Select your initial kVp To pick a starting kVp use Sante’s Rule

Step 3: Select your initial kVp If you are using a….. –5:1 grid add –8:1 grid add –12:1 grid add

Step 3: Select your initial kVp Most veterinary applications use an 8:1 grid If you know you are using a grid, but don’t know what kind of grid is under your table assume it is an 8:1 grid Remember, grids are not used in table top techniques If you are making a technique chart for a table top study use 0 for the grid factor

Step 4: Expose the Perfect Film The goal in this step is to obtain a radiograph with a “perfect” exposure The technique you use to generate this radiograph will be used to create the rest of the technique chart. Take your time and make this radiograph as good as it can be It may take 5 or 6 tries to get it right but the time you spend on this step will be worth it in the end

Step 4: Expose the Perfect Film To get started, take a radiograph with the mAs I recommended above for each particular study, and set your kVp according to Sante’s Rule For example: if I were creating an abdominal technique, using a 40 inch SID, an 8:1 grid, and my dog measures 15cm then I would expose a radiograph at 7.5mAs and 80kVp. Hopefully, that exposure will be in the ballpark and the radiograph will be fairly well exposed

Step 4: Expose the Perfect Film Unfortunately, this is not always the case first look at film density

Step 4: Expose the Perfect Film Once density is good now time to perfect contrast by changing the kVp

Step 5: Make the Technique Chart Now that you have a properly exposed radiograph, you are over the hump and the rest is easy To create a technique chart simply start with your “perfect exposure” and extrapolate to find the values of kVp for other measurements according to the following rules

Step 5: Make the Technique Chart Subtract 2 kVp from the original kVp for each cm decrease from the original measurement Add 2 kVp to the original kVp for each cm increase from the original measurement up to 80kVp Add 3 kVp for each cm increase that places the kvp above 80 up to 100 Add 4 kVp for each cm increase that places the kVp above 100

Step 5: Make the Technique Chart For example…let’s say that we started with a perfect exposure of 7.5mAs at 80kVp and our patient measured 15cm We would make an abdominal technique chart as follows

Repeat the above process to create these additional technique charts Step 6: Create a technique chart for each different study as abdomen, thorax, extremity, spine Repeat the above process to create these additional technique charts

Also remember every animal is different fat thin muscular/dense puppy/kitten exotic so settings (kVp) may need to be changed a little bit need to evaluate every film as it comes out