Corneal Biomechanics after Penetrating Keratoplasty and Descemet’s Stripping Automated Endothelial Keratoplasty Samer Hamada MRCOphth FRCS(Ed)1,3 Louai.

Slides:



Advertisements
Similar presentations
Biomechanical Properties of the Cornea in Normal- Tension Glaucoma Authors: Leonidas Traipe Ines Cayuqueo Fabiola Cerfogli Claudia Goya Allister Gibbons.
Advertisements

Statement of Commercial Interest: RDJ: None DRH: honoraria from Reichert for educational lectures on corneal biomechanics Multiple studies have shown ex.
Venting Incisions in DSAEK Patients: Is It an Absolute Necessity?
Use of a Novel Y- Suture Technique to Reduce Detachments in Descemet’s Stripping Automated Endothelial Keratoplasty (DSAEK) Habeeb Ahmad, MD Martin Heur,
Corneal biomechanical properties and their association with intraocular pressure (IOP) and central corneal thickness (CCT) Dimitrios P. Bessinis 1, Pandelis.
A Prospective Trial Evaluating Scleral Rebound Tonometry Sara Duke, MD, Andrew Logeman, Shuchi Patel, MD Loyola University Chicago, Department of Ophthalmology,
Recent Advances in IOP measurements
Comparison between Dynamic contour tonometry, Goldmann applanation tonometry and Ehlers-corrected Goldmann applanation tonometry in eyes after Laser In.
Roy E Lehman MD*, Samuel F Fulcher MD**
Influence of Keratoplasty Technique on Corneal Biomechanical Properties in Keratoconus Hans R Vellara, Noor Q Ali, Charles NJ McGhee, Dipika V Patel Department.
Biomechanical Properties of Healthy and Keratoconic Corneas Evaluated Using an Ultra-High-Speed Scheimpflug Camera nznec Department of Ophthalmology University.
E-Poster ALC Canedo, MD 2010 Rio de Janeiro CornealTomograhyandBiomechanics Study Group Study Group Comparative Study of Corneal Biomechanical Properties.
CR The Glaucoma Tonometer. What sets it apart? Reichert’s 7CR Auto Tonometer + Corneal Response Technology takes corneal biomechanical properties into.
Outcome of refractive surgery in keratoconus suspect eyes
Comparison of the Diaton Transpalpebral Tonometer Versus Goldmann Applanation R. S. Davidson 1 ; N. Faberowski 2 ; R. J. Noecker 3 ; M. Y. Kahook 1 1.
Variations in corneal biomechanical parameters and central corneal thickness during the menstrual cycle. David Zadok, MD, Yakov Goldich, MD, Yaniv Barkana,
A Prospective Trial Comparing Scleral Pneumotonometry to Goldmann Applanation Tonometry Sara Duke, MD, Usiwoma Abugo, BS, Shuchi Patel MD Loyola University.
Bryan Y Kim 1, Shintaro Kanayama MD PhD 1, Tueng T Shen MD PhD 1, Thomas E Gillette MD 2 1 University of Washington Department of Ophthalmology, 2 Eye.
Dr. K.S.SIDDHARTHAN Aravind Eye Hospital Coimbatore
Evaluation of Corneal Parameters and Spherical Aberration After DSAEK Measured with Pentacam System Orkun Muftuoglu, Pawan Prasher, R. Wayne Bowman, Steven.
TRANSPALPEBRAL TONOMETER APPLICATION DURING INTRAOCULAR PRESSURE EVALUATION IN THE PATIENTS WITH REFRACTION ANOMALY BEFORE AND AFTER KERATOPHOTOREFRACTIVE.
Correlation Between Corneal Biomechanics and Wavefront Aberrations Alejandro Rivera, MD Private Practice Mexico City The author has no financial interest.
Evaluation of Interface Reflectivity and Corneal Aberrations following DSAEK Hamid Khakshoor, MD Professor of Ophthalmology Mashhad University of Medical.
Relationship between Pachymetry and Ocular Response Analyzer Measurements Alejandro Rivera MD.
INTACS – PKP Comparison X X 180°-0.75 TransplantIntacs.
Case Report of Severe Haze After DSAEK
Descemet Stripping Automated Endothelial Keratoplasty in patients with Toxic anterior segment syndrome induced endothelial failure First and Presenting.
Variations in corneal biomechanical parameters and central corneal thickness during the menstrual cycle. Yakov Goldich, MD, David Zadok MD, Yaniv Barkana,
Confocal scan study of corneal grafts following Descemet-on versus Descemet-off deep anterior lamellar keratoplasty Zare M, Feizi S, Rezaie M, Hoseini.
Naomi R. Goldberg, MD PhD Kenneth J. Wolf, MD Eric J. Wolf, MD FACS The authors have no financial interest in the subject matter of this poster. Comparison.
E-Poster Rio de Janeiro Corneal Tomography and Biomechanics Study Group The study of the correlations of the optic disc size with Cornal Geometry and Biomechanical.
Ahmed El-Massry, M.D. Professor of Ophthalmology - Faculty of Medicine University of Alexandria Egypt long-Term Results of Corneal Biomechanical Changes.
Ocular Response Analyzer Waveform Analysis in the Ectatic Corneas: Correlation of the New Corneal Biomechanics Parameters and Severity of Keratoconus Kristin.
Deep Anterior Lamellar Keratoplasty (DALK) Vs Penetrating Keratoplasty (PK) in patients with Keratoconus (KC). Dr. K.S.SIDDHARTHAN Aravind Eye Hospital.
Jodhbir S Mehta, Donald Tan The Authors have a financial Interest in the Endoglide patent/royalty.
Etiological and decision making factors for repeat DSAEK or PK in cases of failed Descemet stripping automated endothelial keratoplasty First and Presenting.
Comparison of Endothelial Keratoplasty (EK) Visual Outcomes with OCT Derived Corneal Thickness Measurements Authors Abraham K. Sleem, MD Robert L. Schultze,
The authors have no conflicting interest in the subject matter of this poster. In Vivo Evaluation of DSAEK Interface with Scanning- Laser Confocal Microscopy.
Corneal Hysteresis Role of Corneal Biomechanics in Refractive Surgery Luis Rodriguez, MD Anny Villegas, MD Clínica de Córnea Centro Médico Docente La Trinidad.
Descemet’s Stripping Endothelial Keratoplasty (DSEK) in patients with prior Trabeculectomy or Tube shunt surgery. Thadani S.M. Fynn-Thompson N. Authors.
Mathematical Description of the Ocular Response Analyzer Applanation Signal in Post-LASIK Ectasia Mujtaba A. Qazi, M.D. 1,2 David A. Luce, Ph.D. 3 Jay.
BIOMECHANICAL PROPERTIES OF THE CORNEA IN PSEUDO-EXFOLIATION SYNDROME Aylin Kılıç Ertan, MD Özgür Bilen, MD Kudret Eye Hospital, Ankara, Turkey There is.
Corneal Biomechanical Changes Following Surface Keratorefractive Surgery Teeravee Hongyok, MD, Christopher J. Rapuano, MD, Ajoy Virdi, MBBS, Elisabeth.
CRF variations ( initial CRF – final CRF )
Characteristics of Primary Angle-Closure Glaucoma Patients with Normal Intraocular Pressure at the First Visit Won Hyuk Oh1, Bum Gi Kim1, Joo Hwa Lee2.
55th ASMR National Scientific Conference
Outcomes and Complications of DSEK in Eyes with Tube shunts or ACIOLs
JP Deva COMPARISON OF PRELIMINARY RESULTS FROM CROSSLINKED KERATOCONUS AND NORMAL CORNEAE USING ORA JP Deva
Influence of UVA-Riboflavin corneal collagen cross-linking on biomechanical properties of keratoconic eyes David Zadok MD, Yakov Goldich MD, Yaniv Barkana.
Date of download: 10/26/2017 Copyright © ASME. All rights reserved.
DSAEK using corneas with previous LASIK
Nienke Visser, Tos T.J.M. Berendschot, Rudy M.M.A. Nuijts
Endothelial keratoplasty in failed grafts
Kyoto Prefectural University of Medicine
Early Experience with Descemet’s Stripping Automated Endothelial Keratoplasty Combined with Phacoemulsification: Clinical and Refractive Outcome University.
R. S. Davidson1; N. Faberowski2; R. J. Noecker3; M. Y. Kahook1 1
Effects of corneal collagen crosslinking on Ocular Response Analyzer waveform-derived variables in keratoconus and post-LASIK ectasia Spatial treatment.
Hong A, Boehlke CS, Afshari NA, Kim T Duke University Medical Center
Long term Results of Endothelial Keratoplasty in patients with failed penetrating Keratoplasty Maoz D. Amiran, MD, Raneen Shehadeh-Mashor, MD, Marie Eve.
Keratometric Changes in Noninflammatory Corneal Ectasic Diseases After Intraocular Pressure Reduction Authors has no financial interest. Renata T Kashiwabuchi,
Associate Professor of Ophthalmology
Federal University of São Paulo, Brazil
Amit Patel, FRCOphth Massimo Busin, MD
Traumatic Wound Dehiscence After Corneal Keratoplasty
Role of HSV Infection in the Histopathology of Failed DSAEK
Yakov Goldich MD,Yaniv Barkana MD, Isaac Avni MD, David Zadok MD,
Anand K Shah MD1 Neda Shamie MD1 Paul Phillips MD1 Mark A Terry MD1,2*
I.J.E van der Meulen1, C.P. Nieuwendaal1,
Hyuck Jin Choi, Joo Youn Oh, Won Ryang Wee, Mee Kum Kim,
Jun Her MD. PhD, Sung-wook Kim MD, Jae-hwan Ahn MD
Presentation transcript:

Corneal Biomechanics after Penetrating Keratoplasty and Descemet’s Stripping Automated Endothelial Keratoplasty Samer Hamada MRCOphth FRCS(Ed)1,3 Louai Wehbeh MRCOphth3 Ben Royal Ophthalmic Physiologist1,2 Mohammed Laiquzzaman MBBS, PhD1,2 Sunil Shah FRCS(Ed) FRCOphth1,2,3 1. Birmingham Heartlands and Solihull Trust, Solihull UK 2. Anterior Eye Group (AEG), Neurosciences Research Institute, Aston University, Birmingham, UK 3. Birmingham and Midland Eye Centre, Birmingham, UK Financial Interest Disclosure: The authors have no financial interest in the subject matter of this poster

Corneal Hysteresis When air pulse is released it causes the cornea to move inwards past applanation and its shape becomes slightly concave but when the air puff shuts off the pressure decreases and the corneal shape tries to gain its normal shape, during this process the cornea again passes through an applanation phase - ‘outward applanation’. Theoretically these two pressures should be the same but this is not the case and this is described as Dynamic corneal Response (DCR), which is the resistance to applanation manifested by the corneal tissue due to its viscoelastic properties. The difference between these two pressures is termed as corneal hysteresis Corneal Hysteresis (CH) is believed to reflect the aggregate effects of thickness, rigidity, hydration and perhaps other factors yet to be determined. Corneal Resistance Factor (CRF) is believed to be a measure of the elastic properties of the cornea.

Measuring corneal hysteresis Various investigators have in the past tried to measure the ocular rigidity (elasticity) of the eye in order to assess the pathological processes affecting the corneal tissue Studies showed a reduced rigidity in keratoconic eyes, however there is not enough studies to describe corneal hysteresis post penetrating keratoplasty or posterior lamellar keratoplasty

Purpose   To compare corneal hysteresis (CH), corneal resistance factor (CRF), central corneal thickness (CCT), and intraocular pressure (IOP) following penetrating keratoplasty (PK) and Descemet’s Stripping Automated Endothelial Keratoplasty (DSAEK) and compare results with normal eyes using Ocular Response Analyser(ORA)

Methods Prospective comparative masked study of 34 consecutive post unilateral PKP eyes 28 consecutive post unilateral DSAEK eyes 166 normal eyes (as control) CH, CRF, and IOP of each eye was measured by ORA (Reichert’s non- contact tonometer with additional analysis software; the Ocular Response Analyser (ORA) [Reichert Ophthalmic Instruments, Buffalo, USA] CCT was measured with a hand held ultrasonic pachymeter (SP-2000, Tomey Corp, Aichi, Japan) in the mid-pupillary axis. Three readings were taken and the lowest value was used as the CCT.

Results Mean CH Normal eyes: 10.6 ± 2.0mmHg Post-PK eyes: 8.9±3.3mmHg Post-DSAEK eyes: 6.4 ± 2.9mmHg Mean CRF Normal eyes: 10.2 ± 2.0mmHg Post-PK eyes: 8.1±3.3mmHg Post-DSAEK eyes: 7.5 ± 2.7mmHg

CCT and IOP Mean CCT Normal eyes: 541.8±36.1µm Post-PK eyes: 556.0±69.2µm Post-DSAEK eyes: 632.5±57µm Mean IOP Normal eyes: 16.1± 3.1mmHg Post-PK eyes: 12.4± 2.9mmHg Post-DSAEK eyes: 17.2± 3.1mmHg

Results CCT was higher in post-DSAEK eyes compared with normal and post-PK eyes but not statistically significant (p>0.4, and>0.38) Correlation between CH/CCT, CRF/CCT, and IOP/CCT were statistically significant in normal eyes (p<0.0001) but not significant in the post-PK group (p>0.4, p>0.1 and p>0.2 respectively) and post-DSAEK group ( p>0.2, p>0.2, and p>0.1 respectively)

Discussion (1) Our results agree with previous findings that increased CCT can affect corneal biomechanical properties in normal eye, however this is not the case in grafted corneas (post-PKP, post DSAEK). CCT was higher after PKP/DSAEK but corneal hysteresis and resistance factor were reduced after corneal grafting

Discussion (2) Higher CCT did not mean higher levels of rigidity or elasticity. Mean higher CH and CRF in the normal eyes may be related to an intact cornea while the reduced CH and CRF, weak relationship between CH and CCT and CRF and CCT in post-PK/DSAEK eyes is postulated to be due to the weakened cornea because of the graft-host interface (i.e. Post-PK/DSAEK cornea may not achieve the same tensile strength after surgical intervention and thus its rigidity is compromised) We postulate that corneal structure is altered in post- PK/DSAEK eyes and that CH/CRF are moderately independent corneal properties.

Conclusion Biomechanical parameters were reduced post PK and DSAEK, possibly a result of altered corneal structure A larger study is underway to correlate corneal biomechanics to clinical outcomes after corneal transplantation