SVCC Respiratory Care Programs Analysis of ABG Samples.

Slides:



Advertisements
Similar presentations
The Respiratory Quotient (RQ)
Advertisements

OFFLINE COMPOSITION MEASURING SENSORS
A Novel Device for Plating Process Diagnostics
CHEM 213 Instrumental Analysis
Electrode measuring principle
MLAB 2401: C LINICAL C HEMISTRY K ERI B ROPHY -M ARTINEZ Designs in Instrumentation Electrochemistry.
RET 2414L Pulmonary Function Testing Module 6.0
Electrodes and Potentiometry
Lecture 15 CM1001.
Potentiometry. pH is a Unit of Measurement  pH = Power of Hydrogen (H + )  Defined as the Negative Logarithm of Hydrogen Ion Activity  pH = log (1/H.
Chem. 133 – 3/5 Lecture. Announcements Lab –Set 2 Period 2 Labs Set to Finish 3/10 –3/12 will be make up day (for both Period 1 and Period 2 labs) –Set.
Dr. Marc Madou, UCI, Winter 2012 Class V Potentiometric and Amperometric Sensors (I) Electrochemistry MAE-295.
IC Controls 865 ppb Dissolved Oxygen System pH / ORP Conductivity Dissolved Oxygen Chlorine Standards ppb Galvanic Dissolved Oxygen.
A CID -B ASES /G ASES IN BLOOD Under the supervision of : Dr. Malek Al – Qub.
ANALYTICAL CHEMISTRY CHEM 3811 CHAPTER 15 DR. AUGUSTINE OFORI AGYEMAN Assistant professor of chemistry Department of natural sciences Clayton state university.
Understanding Acid/Base Balance Marilyn F. Brown Broad Concepts: Body pH tolerated in very narrow range 7.35 – 7.45 pH is defined as H+ ion concentration.
Quality Assurance.
Equipment Application and Cleanliness
Quality Assurance in the clinical laboratory
Chemometrics Method comparison
Unit #7 - Basic Quality Control for the Clinical Laboratory
Acids and Bases Biotechnology I. Life Chemistry  Based on water  Cells contain 80-90% water  Proper pH essential to ALL living systems Plants cannot.
Quality Assurance.
Quality Assessment 2 Quality Control.
Interpretation of Blood Gases Chapter 7. Precise measurement of the acid-base balance of the lungs’ ability to oxygenate the blood and remove excess carbon.
Potentiometry and potentiometric measurements. potentiometer A device for measuring the potential of an electrochemical cell without drawing a current.
1 Radiometer Medical ApS, Åkandevej 21, DK-2700 Brønshøj, Tel: , RTC, December 2004 ABL800 FLEX Calibration.
Acid-Base Imbalance NRS What is pH? pH is the concentration of hydrogen (H+) ions The pH of blood indicates the net result of normal acid-base.
Getting an arterial blood gas sample
با نام و یاد خدا.
Automated CBC Parameters
Quality Control Lecture 5
ARTERIAL BLOOD GAS ANALYSIS Arnel Gerald Q. Jiao, MD, FPPS, FPAPP Pediatric Pulmonologist Philippine Children’s Medical Center.
1. Blood or Fluid Analysis Using carried out in central labs with problems of time delay costs and potential misslabeling Drive toward more bedside or.
MLAB 2401: Clinical Chemistry Keri Brophy-Martinez
Chapter 13 - Analytical Instrumentation 1 Chapter 14 Analytical Instrumentation.
Why do we need to do it? What are the basic tools?
RESPIRATORY MODULE. FAWAD AHMAD RANDHAWA MBBS ( King Edward Medical College) M.C.P.S; F.C.P.S. ( Medicine) F.C.P.S. ( Endocrinology) Assistant Professor.
Acid-Base Balance Disturbances
LECTURE 7: Electrochemistry. Types of electrodes and their using. ass. prof. Yeugenia B. Dmukhalska.
Electrodes and Potentiometry Introduction 1.)Potentiometry  Use of Electrodes to Measure Voltages that Provide Chemical Information - Various electrodes.
Outlines Introduction Body acidity has to be kept at a fairly constant level. Normal pH range within body fluids Normal pH is constantly.
Chem. 133 – 3/3 Lecture. Announcements Homework Set 2 (pass out) Grading –Working to get the Electronics labs and exam 1 graded by next Tuesday Lab –today.
PH THEORY What is it pH?What is it pH?/How is pH Measured?/What Equipment is Required to Measure pH?/How is a pH Measurement Device Calibrated?/Why is.
PH and pH meters Der Unfassbare (2006), “Ph-Meter” [photograph]. Retrieved from
PH meter Bushra mubarak. pH pH is defined as the negative logarithm of the hydrogen ion activity aH. Unit of measure which describes the degree of acidity.
Monitoring in Anesthesia Dr.Arkan Jaafar, M.D. Anesthesiologist,Medical college of Mosul.
Chapter 1: The Nature of Analytical Chemistry
Acid Base Balance B260 Fundamentals of Nursing. What is pH? pH is the concentration of hydrogen (H+) ions The pH of blood indicates the net result of.
In voltaic cells, oxidation takes place at the anode, yielding electrons that flow to the cathode, where reduction occurs. Section 1: Voltaic Cells K What.
Blood Gas Analysis and Analyzer (BME Lab 2) Palestine Polytechnic University Engineering & Technology Collage Electrical &Computer Department Biomedical.
Acid-Base Imbalance.
Electrochemistry MAE-212
Potentiometry and potentiometric measurements
Acid-Base Imbalance.
ABG Instrumentation Analytical Errors
Unit #6 - Basic Quality Control for the Clinical Laboratory
Ion-Selective Electrode (I.S.E.)
Quality Assurance in the clinical laboratory
Acid-Base Imbalance.
Rapid Lab Systems measuring principle
Acid-Base Imbalance.
Quality Assurance in the clinical laboratory
PH meter.
Acid base balance.
Acid base balance.
Quality Control Lecture 3
Quality Assessment The goal of laboratory analysis is to provide the accurate, reliable and timeliness result Quality assurance The overall program that.
Presentation transcript:

SVCC Respiratory Care Programs Analysis of ABG Samples

SVCC Respiratory Care Programs ABG Analysis, Introduction pH, PaCO 2, PaO 2 are measured directly by special electrodes contained in a device made for that purpose pH, PaCO 2, PaO 2 are measured directly by special electrodes contained in a device made for that purpose Other indirect measurements can be made or calculated from the above measurements i.e., HCO 3 -, O 2 Sat. Other indirect measurements can be made or calculated from the above measurements i.e., HCO 3 -, O 2 Sat.

SVCC Respiratory Care Programs pH pH electrode is constructed of two half cells, which develop an electrical potential when connected together pH electrode is constructed of two half cells, which develop an electrical potential when connected together Reference electrode maintains a constant potential and is bathed in KCl Reference electrode maintains a constant potential and is bathed in KCl Glass electrode (Sanz electrode) develops an electrical potential that is proportional to the amount of H+ present Glass electrode (Sanz electrode) develops an electrical potential that is proportional to the amount of H+ present

SVCC Respiratory Care Programs PaCO 2 Electrode (Severinghaus) Measures carbon dioxide tensions by allowing the CO 2 gas to undergo a chemical reaction that produces hydrogen ions Measures carbon dioxide tensions by allowing the CO 2 gas to undergo a chemical reaction that produces hydrogen ions The hydrogen ion concentration produced is directly proportional to the PCO 2 in contact with the membrane of the electrode The hydrogen ion concentration produced is directly proportional to the PCO 2 in contact with the membrane of the electrode Operates on the principle of electric potential between electrodes Operates on the principle of electric potential between electrodes CO 2 + H 2 O H 2 CO 3 H + + HCO 3 - CO 2 + H 2 O H 2 CO 3 H + + HCO 3 -

SVCC Respiratory Care Programs O 2 Electrode (Clark) Is a polarographic device that measures oxygen tensions by oxidation/reduction reactions, a chemical process that generates measurable electrical currents Is a polarographic device that measures oxygen tensions by oxidation/reduction reactions, a chemical process that generates measurable electrical currents Has platinum cathode and silver anode immersed in an electrolyte solution Has platinum cathode and silver anode immersed in an electrolyte solution Volume of O 2 will be directly proportional to the number of electrons used in the cathode rxn. and by measuring current is a measure of O 2 diffused across membrane Volume of O 2 will be directly proportional to the number of electrons used in the cathode rxn. and by measuring current is a measure of O 2 diffused across membrane

SVCC Respiratory Care Programs QA in Blood Gas Analysis ABG lab must be able to assure accurate and reliable results ABG lab must be able to assure accurate and reliable results The above is accomplished by applying protocols in 3 areas: - pre-analytic error - calibration - quality control The above is accomplished by applying protocols in 3 areas: - pre-analytic error - calibration - quality control

SVCC Respiratory Care Programs Pre-analytic Error All factors that cause variance in lab results prior to the sample arriving in the ABG lab. All factors that cause variance in lab results prior to the sample arriving in the ABG lab. 4 factors assoc. with signif. P. E. are: - air bubbles in sample 4 factors assoc. with signif. P. E. are: - air bubbles in sample - time delay (iced sample with more than 60 min. or uniced with more than 10 min.) - blood clots in sample - small sample size where excessive anticaogulation is suspect

SVCC Respiratory Care Programs Calibration Purpose is assure consistency Purpose is assure consistency Def.: the systemic standardization of the graduation of a quantitative measuring instrument Def.: the systemic standardization of the graduation of a quantitative measuring instrument Calibrating standards for blood gas analyzers should simulate the physical properties of blood and meet manuf. specs. Calibrating standards for blood gas analyzers should simulate the physical properties of blood and meet manuf. specs. When 2 standards are used ---> 2-point calibration, performed after 50 blood gases or at least every 8 hours When 2 standards are used ---> 2-point calibration, performed after 50 blood gases or at least every 8 hours

SVCC Respiratory Care Programs Calibration (contd) A one-point calibration is an adjustment of the electronic response of an electrode to a single standard and is performed more freq. than a 2 pt. cal., ideally prior to each sample analysis A one-point calibration is an adjustment of the electronic response of an electrode to a single standard and is performed more freq. than a 2 pt. cal., ideally prior to each sample analysis

SVCC Respiratory Care Programs pH Calibration Several types of buffer solutions are acceptable for pH electrode calibration Several types of buffer solutions are acceptable for pH electrode calibration Buffers are excellent standards because of their long shelf life and stability for days after being opened Buffers are excellent standards because of their long shelf life and stability for days after being opened 2-point cal. of pH electrode is done by setting the slope potentiometer to a low pH buffer (6.84) and setting the balance potentiometer with a near-normal buffer (7.384) with an accuracy of +/ point cal. of pH electrode is done by setting the slope potentiometer to a low pH buffer (6.84) and setting the balance potentiometer with a near-normal buffer (7.384) with an accuracy of +/

SVCC Respiratory Care Programs PaCO 2 Calibration PaCO 2 electrode may be calibrated by introducing a known gas concentration PaCO 2 electrode may be calibrated by introducing a known gas concentration Gases used are 5% and 10% CO 2, +/-.03% - 1-point cal. uses 5% standard to set the balance point - 2-point cal. uses 5% for balance point and 10% for slope point Gases used are 5% and 10% CO 2, +/-.03% - 1-point cal. uses 5% standard to set the balance point - 2-point cal. uses 5% for balance point and 10% for slope point PCO 2 = (BP - 47 mmHg) x %CO 2 PCO 2 = (BP - 47 mmHg) x %CO 2

SVCC Respiratory Care Programs PO 2 Calibration Properly calibrated PO2 electrodes perform within the manuf. stated accuracy in PaO 2 ranges below 150 mmHg but may vary 20% at 500 mmHg Properly calibrated PO2 electrodes perform within the manuf. stated accuracy in PaO 2 ranges below 150 mmHg but may vary 20% at 500 mmHg 0% oxygen is used for the slope point and 12% or 20% for the balance point 0% oxygen is used for the slope point and 12% or 20% for the balance point PO 2 = (BP - 47 mmHg) x % O 2 PO 2 = (BP - 47 mmHg) x % O 2

SVCC Respiratory Care Programs Quality Control Refers to a system that documents the accuracy and reliability of the blood gas measurements and is essential to assure accuracy in the blood gas lab Refers to a system that documents the accuracy and reliability of the blood gas measurements and is essential to assure accuracy in the blood gas lab Media available as blood gas controls include: Media available as blood gas controls include: - aqueous buffers - glycerin soltn. - human/animal serum and blood - artificial blood A QC system must ID problems and specify corrective action, document. of accept. oper. A QC system must ID problems and specify corrective action, document. of accept. oper.

SVCC Respiratory Care Programs QC (contd) Documentation of QC is usu. on Levy-Jennings Chart which shows measured results on the y axis versus time of measurement on the x axis Documentation of QC is usu. on Levy-Jennings Chart which shows measured results on the y axis versus time of measurement on the x axis SD is used to summarize a mass of data: the difference between a number in a data set and the mean of the data set is called a deviation. A deviation shows how much a number varies from the mean SD is used to summarize a mass of data: the difference between a number in a data set and the mean of the data set is called a deviation. A deviation shows how much a number varies from the mean

SVCC Respiratory Care Programs QC (contd) A properly functioning electrode that repeatedly analyzes a known value will produce results within a rel. small range, e.g., a PaCO 2 electrode that analyzes a 40 mmHg standard 100 times will produce results where 2/3 of the measurements are mmHg and nearly all measurements fall in range A properly functioning electrode that repeatedly analyzes a known value will produce results within a rel. small range, e.g., a PaCO 2 electrode that analyzes a 40 mmHg standard 100 times will produce results where 2/3 of the measurements are mmHg and nearly all measurements fall in range 95% of the control measurements should fall within 2 SD 95% of the control measurements should fall within 2 SD

SVCC Respiratory Care Programs QC (contd) Random errors indicates a value outside of 2 SD of the mean: a single random error has minor signif., but if number increased the machine and techniques must be evaluated Random errors indicates a value outside of 2 SD of the mean: a single random error has minor signif., but if number increased the machine and techniques must be evaluated Systematic errors is recurrent measurable deviation from the mean Systematic errors is recurrent measurable deviation from the mean Causes of systematic errors: Causes of systematic errors: - contaminated standard - variations in electrode temp. - inconsistent introduction of standard

SVCC Respiratory Care Programs QC (contd) Causes of systematic error (contd) - inconsistent calibration technique - change in QC standard storage or prep. - electrode problems, e.g., protein contamin., membrane malfunction, contamin. electrolyte, or electrical problems Causes of systematic error (contd) - inconsistent calibration technique - change in QC standard storage or prep. - electrode problems, e.g., protein contamin., membrane malfunction, contamin. electrolyte, or electrical problems

SVCC Respiratory Care Programs QC Levels Level 1 simulates a patient hypoventilating Level 1 simulates a patient hypoventilating Level 2 simulates a patient with normal ventilatory status Level 2 simulates a patient with normal ventilatory status Level 3 simulates a patient hyperventilating Level 3 simulates a patient hyperventilating