Double devices: Dysfunction or not? Sok-Sithikun Bun, MD, Decebal Gabriel Laţcu, MD, Ahmed Mostfa Wedn, MD, Nadir Saoudi, MD, FESC, FHRS HeartRhythm Case Reports Volume 4, Issue 7, Pages 278-280 (July 2018) DOI: 10.1016/j.hrcr.2018.03.006 Copyright © 2018 Heart Rhythm Society Terms and Conditions
Figure 1 Fluoroscopic image in anteroposterior projection showing the presence of the 2 batteries on the left side. HeartRhythm Case Reports 2018 4, 278-280DOI: (10.1016/j.hrcr.2018.03.006) Copyright © 2018 Heart Rhythm Society Terms and Conditions
Figure 2 Surface electrocardiogram showing the presence of a biventricular device pacing at 70 beats per minute and concomitant asynchronous pacing from an old abandoned dual-chamber battery at 65 beats per minute. HeartRhythm Case Reports 2018 4, 278-280DOI: (10.1016/j.hrcr.2018.03.006) Copyright © 2018 Heart Rhythm Society Terms and Conditions
Figure 3 Telecardiology transmission report showing a nonphysiological recording of abnormal short VV intervals. HeartRhythm Case Reports 2018 4, 278-280DOI: (10.1016/j.hrcr.2018.03.006) Copyright © 2018 Heart Rhythm Society Terms and Conditions