Cardiac implantable electronic device malfunction after deceleration injury without obvious chest trauma  Amr F. Barakat, MD, Brian Cross, MD, Jonathan.

Slides:



Advertisements
Similar presentations
Implantation of the subcutaneous implantable cardioverter-defibrillator with retroperitoneal generator placement in a child with hypoplastic left heart.
Advertisements

Pulmonary arterial pressure sensing in a patient with left ventricular assist device during ventricular arrhythmia  Jill R. Harris, MSN, ACNP-C, CNS,
Inhibition of pacing in a dependent patient with an implantable cardioverter-defibrillator and a left ventricular assist device  Michele Murphy, MD, Timothy.
Central venous catheter fracture during pacemaker lead extraction
MediGuide-assisted atrial flutter ablation in a patient with a HeartMate II left ventricular assist device  Hien Kiem Nguyen Thanh, MD, Jason Andrade,
A new approach for implantation of a cardiac resynchronization therapy–defibrillator in a patient with bilateral pectoral neurostimulation devices  Saori.
Inhibition of pacing in a dependent patient with an implantable cardioverter-defibrillator and a left ventricular assist device  Michele Murphy, MD, Timothy.
Central venous catheter fracture during pacemaker lead extraction
Erroneous shock by an AED: Importance of obtaining AED tracing to prevent inappropriate ICD implantation  Brian J. Cross, MD, Mark S. Link, MD, FHRS 
Implantation of a cardiac resynchronization therapy device using the anchor balloon technique in a patient with a tortuous coronary sinus branch  Yu Kumagai,
A unique case of pulmonary embolism presenting as a paroxysmal atrial tachycardia instigated only by recumbency and stooping  Nagesh Chopra, MD  HeartRhythm.
Swimming pool saline chlorination units and implantable cardiac devices: A source for potentially fatal electromagnetic interference  John Wight, BS,
Malignant transformation of a chronically infected implantable cardioverter-defibrillator pocket  Donna Kang, MD, Khuyen Do, MD, Jonathan Nattiv, MD,
A case of paroxysmal atrioventricular block–induced cardiac arrest
Combination of a leadless pacemaker and subcutaneous defibrillator: First in-human report  Blandine Mondésert, MD, Marc Dubuc, MD, FHRS, Paul Khairy,
To the Editor— Leftward on left anterior oblique is not always septal!
Is it bradycardia or something else causing symptoms?
Simultaneous lead extraction and vacuum-assisted vegetation removal
Ruchit Shah, MD, Vineet Kumar, MD, FHRS  HeartRhythm Case Reports 
Inappropriate defibrillator shock during gynecologic electrosurgery
Anne Barnett, PA-C, Lee Eckhardt, MD, FHRS, Miguel A. Leal, MD, FHRS 
Respiratory rate trending as a cause for atrial lead noise: A first report in an implantable cardioverter-defibrillator patient  Isla McClelland, MD,
Cardiac inflammation and ventricular tachycardia in Chagas disease
Inside-out abrasion and contained conductor cable externalization in a defibrillation lead with asymmetric conductor cable lumen distribution  Ernest.
Darren C. Tsang, BS, Roger G. Carrillo, MD, MBA, FHRS 
Increased threshold in nonselective His-bundle pacing suspected to be caused by amiodarone  JoonHyuk Kim, MD, Seth Goldbarg, MD, FHRS, Steven Leung, MD,
Left axillary active can positioning markedly reduces defibrillation threshold of a transvenous defibrillator failing to defibrillate at maximum output 
Pulmonary arterial pressure sensing in a patient with left ventricular assist device during ventricular arrhythmia  Jill R. Harris, MSN, ACNP-C, CNS,
T-wave oversensing due to left ventricle–only pacing in cardiac resynchronization therapy optimization algorithm  Philippe André, MD, MSc, Benoit Plourde,
Cardiac resynchronization therapy in coronary sinus atresia delivered using leadless endocardial pacing  Dewi E. Thomas, MD, MRCP, Nicholas M. Child,
Daniel P. Morin, MD, MPH, FACC, FHRS  HeartRhythm Case Reports 
Jeffrey Ho, MD, Jordan M. Prutkin, MD, MHS, FHRS 
Use of a cerebral protection device for the laser extraction of a pacemaker lead traversing a patent foramen ovale  James L. Harrison, MA, PhD, MRCP,
Cardiac contractility modulation therapy: Are there superresponders?
Termination of anticoagulation therapy at 45 days after concomitant atrial fibrillation catheter ablation and left atrial appendage occlusion resulting.
Christian Steinberg, MD, Danny Dvir, MD, Andrew D
Complete congenital heart block: A case of multilevel block
Expanding the electrical phenotype of NKX2-5 mutations: Ventricular tachycardia, atrial fibrillation, and complete heart block within one family  Simone.
Intravenous sotalol for conversion of atrial flutter in infants
Right-sided subcutaneous implantable cardioverter-defibrillator placement in a patient with dextrocardia, tetralogy of Fallot, and conduction disease 
Managing cross talk between a subcutaneous implantable cardioverter-defibrillator and a dual-chamber unipolar pacemaker system  Konstantinos Kossidas,
Inappropriate shocks in a patient with a subcutaneous implantable cardioverter- defibrillator  Sana M. Al-Khatib, MD, MHS, FHRS, Sasmrita Belbase, NP,
Transvenous approach to pacemaker lead implantation for sinus node dysfunction after extracardiac lateral tunnel Fontan conduit placement  Edward O’Leary,
Albert N. Nayeri, MD, Behzad B. Pavri, MD  HeartRhythm Case Reports 
Air entrapment causing early inappropriate shocks in a patient with a subcutaneous cardioverter-defibrillator  Sing-Chien Yap, MD, PhD, Rohit E. Bhagwandien,
Santosh K. Padala, MD, Kenneth A. Ellenbogen, MD, FHRS, Jayanthi N
Inappropriate subcutaneous implantable cardioverter-defibrillator therapy due to R-wave amplitude variation: Another challenge in device management  Syeda.
Daniel W. Nelson, BS, Caroline Vloka, BA, Jule Wetherbee, MD 
Progressive electrical remodeling in apical hypertrophic cardiomyopathy leading to implantable cardioverter-defibrillator sensing failure during ventricular.
Epicardial appendage ooze causing pericardial tamponade after left atrial appendage device implantation  Karen P. Phillips, MBBS, FHRS, Christopher Smith,
Undetected ventricular fibrillation in a single-chamber implantable cardioverter- defibrillator: When the far-field channel sees more than the intraventricular.
Successful implantation of a subcutaneous cardiac defibrillator in a patient with a preexisting deep brain stimulator  Yousef Bader, MD, Jonathan Weinstock,
Malfunction of an MRI-Conditional Pacemaker Following an MRI
Changing place, changing future: Repositioning a subcutaneous implantable cardioverter-defibrillator can resolve inappropriate shocks secondary to myopotential.
HeartRhythm Case Reports
Polymorphic ventricular tachycardia associated with an episode of reflex syncope: Is this the needle in the haystack?  Matthew A. Tester, Brooke C.D.
Abdominal implantable cardioverter-defibrillator placement in a patient requiring bilateral chest radiation therapy  Ankur A. Karnik, MD, FHRS, CCDS,
Serious case of the hiccups
The runaway defibrillator…A case of an implantable cardioverter-defibrillator that failed communication and deactivation with a magnet  Mahmoud Assaad,
Shock-induced right ventricular pacing failure caused by a short circuit: Uncommon but life-threatening complication of the Riata lead  Itsuro Morishima,
Implantation of the subcutaneous implantable cardioverter-defibrillator with retroperitoneal generator placement in a child with hypoplastic left heart.
HeartRhythm Case Reports
Abdul R. Mourad, MD, Michael H. Kim, MD, FHRS, Malcolm M
Philip M. Chang, MD, FHRS, CEPS, Rahul N. Doshi, MD, FHRS, FACC 
Double devices: Dysfunction or not?
Ventricular tachycardia slower than the rate cut-off of a subcutaneous cardiac defibrillator sensed and successfully treated as a result of oversensing 
Mobile subcutaneous implantable cardioverter-defibrillator leads to oversensing and inappropriate shocks  Omid Kiamanesh, MD, Jacob M. Larsen, MD, PhD,
A true case of wandering pacemaker
Unique clinical presentation and management of lead-stent abrasion
Substernal subcutaneous implantable cardioverter-defibrillator lead placement for the management of inappropriate shocks  Khuyen Do, MD, Chin C. Lee,
Presentation transcript:

Cardiac implantable electronic device malfunction after deceleration injury without obvious chest trauma  Amr F. Barakat, MD, Brian Cross, MD, Jonathan Wertz, MD, Samir Saba, MD, FHRS, Krishna Kancharla, MD  HeartRhythm Case Reports  Volume 5, Issue 5, Pages 285-287 (May 2019) DOI: 10.1016/j.hrcr.2019.02.003 Copyright © 2019 Heart Rhythm Society Terms and Conditions

Figure 1 Device interrogation strip showing high-amplitude nonphysiologic electrograms on the right ventricular lead channel (black arrows), triggering inappropriate antitachycardia pacing therapy (black double arrow). Atrial lead also displays nonphysiologic signals on this tracing. HeartRhythm Case Reports 2019 5, 285-287DOI: (10.1016/j.hrcr.2019.02.003) Copyright © 2019 Heart Rhythm Society Terms and Conditions

Figure 2 Device interrogation strips showing simultaneous abrupt increase in pace impedance in all 3 leads. A = atrial; LV = left ventricular; RV = right ventricular. HeartRhythm Case Reports 2019 5, 285-287DOI: (10.1016/j.hrcr.2019.02.003) Copyright © 2019 Heart Rhythm Society Terms and Conditions

Figure 3 A: Chest radiograph (lateral view) showing abnormal angulation between the device header and the body of the generator (black double arrows), with no evidence of fracture in any of the leads. B: Intraoperative images demonstrating the fractured device header (black arrow) in comparison to an intact one from the new generator (black star). HeartRhythm Case Reports 2019 5, 285-287DOI: (10.1016/j.hrcr.2019.02.003) Copyright © 2019 Heart Rhythm Society Terms and Conditions