50th Particle Measurement Programme Informal Working Group Meeting “kf influence on CPC calibration and counting efficiency“ 3. April 2019
kf influence on CPC calibration and counting efficiency AVL CPC 488; serial number 403 (Calibration Certificate according to ISO/IEG 17025 3. April 2019
kf influence on CPC calibration and counting efficiency TSI CPC Model 3792; serial number 3792153904 (Certificate of Calibration, Emery Oil) 3. April 2019
kf influence on CPC calibration and counting efficiency TSI SN3792153904 efficieny curve with kf (1.08) applied after calibration AVL SN403 efficieny curve with kf applied during calibration 3. April 2019
kf influence on CPC calibration and counting efficiency Application of kf has to be specified… PMP CPC counting efficiency should be within legal limits, taking into account all corrections 3. April 2019