ESTABLISHMENT OF A CORRECTION FACTOR FOR MONITORING THE EXTREMITIES OF INTERVENTIONAL FLUOROSCOPISTS Karen Colucci, Steve King, Mike Sheetz, Joe Och.

Slides:



Advertisements
Similar presentations
State of Michigan Computed Tomography Regulations
Advertisements

4/2003 Rev 2 II.3.6 – slide 1 of 18 Part IIQuantities and Measurements Module 3Principles of Radiation Detection and Measurement Session 6Photographic.
Radiation Safety & Compliance External Beam –Electrons (linacs) –Photons (linacs, Gamma Knife) –Protons (Summer 2013) Brachytherapy –Sealed Sources –Radiopharmaceuticals.
CVT 102.  Time  Distance  Shielding.
Alexander Brandl ERHS 630 Effective and Committed Effective Dose Environmental and Radiological Health Sciences.
ICRP Publication 117 Authors on behalf of ICRP M.M. Rehani, O. Ciraj-Bjelac, E. Vano, D.L. Miller, S. Walsh, B.D. Giordano, J. Persliden.
Certification of Central Venous Lines Georgia Health Sciences Medical Center Augusta, Georgia November 13, 2012.
Personal Radiation and Occupational Safety – CT Policy This policy is in both paper form in the lab’s compliance manual, and on a Power Point presentation.
Module 6 Exposure Control and Personnel Monitoring.
Radiation Protection - Eyes 4 Imaging Solutions, LLC Mark Struthers BBA, BSRT (R)(MR), CMRT, CIIP.
ANALYTICAL X-RAY SAFETY User Training Centre for Environmental Health, Safety and Security Management.
Chapter 3 Radiation Safety Vet Tech Institute. Radiation should be respected not Feared! Safety is Always important! Safety is Always important! Stray.
ERCP Q&A.
 Justification is the answer.  Dose limits are not applicable only recommened.
IAEA International Atomic Energy Agency Interventional Radiology Radiation Sources in medicine diagnostic Radiology Day 7 – Lecture 1(3)
MEASURING PEAK SKIN DOSE John H. O’Connell Adela Salame-Alfie, Ph.D. NYSDOH David F. Lewis, Ph.D. Poppy Pattanayak International Specialty Products.
BEIR VII Implications of the Report for the Future of Medical Imaging G. Donald Frey, Ph.D. Department of Radiology Medical University of South Carolina.
IC0023: Attire for Surgical and Procedural Areas
HEALTH PHYSICS TERMS l RAD (Gy) l mRad l R l mR l Rem l ALARA l NCRP.
Neutron Dose Per Fluence and Weighting Factors for Use at High Energy Accelerators (Submitted to Health Physics) J. Donald Cossairt, Ph.D., C.H.P. Kamran.
RADIATION SAFETY Mrs. Brinston. Introduction As a healthcare worker, you know that radiation is an important tool for detecting and treating diseases.
Development of Patient Dose SR Work Item l Estimation of patient or organ dose requires: Radiation beam characteristics that interact with patient Models.
PET/CT Workshop Presented at VCUHS by Jennifer Love and Mark Crosthwaite by Jennifer Love and Mark Crosthwaite.
RHB SYLLABUS on Fluoroscopy & Radiation Protection
Hand Hygiene Secret Shoppers. Hand Hygiene an infected or colonized body site on one patient, or after touching the patients’ environment, if hand hygiene.
Designing for Radiation Protection. Design Standards for Radiation Protection – Leakage Radiation.
Radiation Protection Procedures
Health and Safety Executive Health and Safety Executive Dose limitation – Estimated Impacts James Taylor OEWG Meeting11 November 2014.
Radiation Protection Procedures
Week 2 :Radiation Protection
1 2 3 Terminal Objective Don full O.R. attire and perform visual check of other personnel for attire upon entry/exit from the O.R.
OHIO RADIOGRAPHY EVENTS Michael Snee, Chief Bureau of Radiation Protection 1.
IAEA International Atomic Energy Agency Radiation protection of the public IAEA Regional Training Course on Radiation Protection of patients for Radiographers,
Program Manager’s Review Dosimetry Program Henry Tran (James Liu) September 19, 2005.
Radiation Dosimeters.
IAEA International Atomic Energy Agency Fluoroscopy Radiation Sources in medicine diagnostic Radiology Day 7 – Lecture 1(2)
Louisiana State University Radiation Safety Office
Training Module Preparation for “AERB norms, licenses and signage”
HEALTH CARE STATISTICS AND RESEARCH HEPR 410
Information for Workers Issued a Radiation Monitoring Badge
IC0023: Attire for Surgical and Procedural Areas
Radiation Protection Systems to Mitigate Occupational Health Risks
Infection Control Test 2
Patient & Occupational radiation dose management Chapter 37 & 38
Radiopharmaceutical Production
Medical Assistant Skills
From: Effect of C-Arm Year Model on Radiation Scatter
Radiation Badge Training
Monitoring of fetal radiation exposure during pregnancy
Get the Lead off Our Backs!
Radiation Protection Procedures
HEALTH PHYSICS.
Presented at VCUHS by Jennifer Love and Mark Crosthwaite
Fluoroscopy-guided versus CT-guided Lumbar Steroid Injections: Comparison of Radiation Exposure and Outcomes    The mean effective radiation dose for patients is lower for fluoro-scopically-guided.
Mobile Fluoroscopy Based on:
Mobile Fluoroscopy Based on:
High Level (Boost) Fluoroscopy
Pima Medical Institute Online Education
Monitoring Pregnant Technologist
Occupational Radiation Dose Management
Occupational Radiation Dose Management
REGULATORY IMPLICATIONS OF THE NEW DOSE LIMIT FOR LENS OF THE EYE
Mobile Fluoroscopy.
Occupational Radiation Dose Management
Scatter Cloud of the C-arm
Vascular Technology Lecture 35: Patient Care HHHoldorf
RC325C Personnel Protection during Fluoroscopic Procedures
THE BASIS FOR RADIATION PROTECTION
Radiation Badge Training
Presentation transcript:

ESTABLISHMENT OF A CORRECTION FACTOR FOR MONITORING THE EXTREMITIES OF INTERVENTIONAL FLUOROSCOPISTS Karen Colucci, Steve King, Mike Sheetz, Joe Och

REGULATIONS The NRC and DEP have established occupational dose limits for the extremities, specifically a shallow-dose equivalent of 50 rem (500 mSv) to the skin of any extremity. Furthermore it is required to monitor occupational exposure to radiation from radiation sources and to supply and require the use of individual monitoring devices by adults likely to receive, in 1 year from sources external to the body, a dose in excess of 10 percent of the limits, (50 mSv in the case of the extremities).

INFECTION CONTROL Wearing dosimetry devices can be a source of infection.

The Association of Operating Room Nurses, (AORN), recommends that personnel remove jewelry if working in the operating room setting.

The World Health Organization consensus recommendation is to strongly discourage the wearing of rings during health care.

Pennsylvania’s Institute for Health Care Improvement mandates the removal of rings, bracelets, and watches before beginning surgical hand scrubbing.

yearly mortality in excess of 100,000 due to hospital- acquired infections

NCRP NCRP 168 makes clear recommendations for the use of personnel dosimetry and urges monitoring the dose to the eye, the only language pertaining to extremity monitoring is that it is “recommended if high hand exposure is expected.” This document makes 31 recommendations regarding interventional fluoroscopy, 4 pertaining to dosimetry, but none focusing on wearing extremity monitors. |

CONVERSION FACTOR Use of a conversion factor for estimating dose is not uncommon in interventional Fluoroscopy. The Webster calculation is widely accepted for use in determining dose to the body, and lens dose values are often calculated from other readings.

Several papers have been published regarding the development of a conversion factor for fluoroscopy. In 2010, Miller et al used historical data to establish a conversion value of 2.8. In 2011, Martin et al found a range of values from 0.8 to 3.4, with a mean of 2.1. Also in 2011, Schueler et al, using phantom scatter measurements found a range of 1.3 to 2.7, with a mean of 1.7.

OUR DATA TABLE 1: Extremity/collar Conversion Factors . TABLE 1: Extremity/collar Conversion Factors based on review of historic dosimetry data Facility Range mean GEISINGER 1.4 – 2.9 1.8 UPMC 2.5 HERSHEY 1.9-2.0 2.0 LEHIGH 1.9-2.1

These were supplemented with direct measurements.

Methods A number of interventional fluoroscopy procedures were observed in order to familiarize ourselves with the processes involved Observations included fluoroscopist positions, c-arm positions, fluoroscopy settings, shield placement, etc

Observations PA-DEP inspectors visited catheterization lab for observations Detailed notes taken: Physician location in relation to patient and table Physician’s hands in relation to the patient and table Fluoroscopy machine details (technique, FOV, SID, time, crane angles, etc) Shielding use and location

RESULTS TABLE 2: Extremity/collar Conversion Factors based on phantom measurement Facility Range mean GHS 1.0 – 3.5 2.3  

CONCLUSION Based on the above findings, a conversion factor of 2.5 is indicated. We request that the PA DEP accept a conversion factor of 2.5 for estimating the dose to the hands when ring badges cannot or will not be used for individuals performing interventional fluoroscopic procedures