Electron probe microanalysis EPMA

Slides:



Advertisements
Similar presentations
Electron probe microanalysis EPMA Trace Element Analysis Mod 11/8/10.
Advertisements

ED and WD X-ray Analysis
X-Ray Microanalysis – Precision and Sensitivity Recall… K-ratio Si = [I SiKα (unknown ) / I SiKα (std.) ] x CF CF relates concentration in std to pure.
Qualitative, quantitative analysis and “standardless” analysis NON DESTRUCTIVE CHEMICAL ANALYSIS Notes by: Dr Ivan Gržetić, professor University of Belgrade.
Thin Film Quantitation of Chemistry and Thickness Using EPMA John Donovan Micro Analytical Facility CAMCOR (Characterization of Advanced Materials in Oregon)
STANDARD SCORES AND THE NORMAL DISTRIBUTION
Quantitative Analysis Quantitative analysis using the electron microprobe involves measuring the intensities of X-ray lines generated from your unknown.
Al, Mg, Si and Na Ka Peak Shifts in Common Silicate and Oxide Minerals: Relevance to Achieving the Goal of 1% Accuracy in EPMA John H. Fournelle Eugene.
Fundamentals of Data Analysis Lecture 9 Management of data sets and improving the precision of measurement.
Statistics and Quantitative Analysis Chemistry 321, Summer 2014.
By C. Kohn Waterford Agricultural Sciences.   A major concern in science is proving that what we have observed would occur again if we repeated the.
Pushing the Envelope for Beam Sensitive Samples and Trace Element Sensitivity and Accuracy John Donovan (541)
Energy-Dispersive X-ray Microanalysis in the TEM Anthony J. Garratt-Reed Neil Rowlands.
Detection Limits N no longer >> N B at low concentration What value of N-N B can be measured with statistical significance? Liebhafsky limit: Element is.
CALIBRATION METHODS. For many analytical techniques, we need to evaluate the response of the unknown sample against the responses of a set of standards.
Interferences: Pathological and Otherwise John Donovan Micro Analytical Facility CAMCOR (Center for Advanced Materials Characterization in ORegon) University.
Data Analysis and Presentation Chapter 5- Calibration Methods and Quality Assurance EXCEL – How To Do 1- least squares and linear calibration curve/function.
2015 Cameca Users Meeting Background characterization – must account for background shape Curvature Spectrometer efficiency Natural Bremsstrahlung shape.
UES Specimen Heterogeneity Analysis : Revisited F. Meisenkothen Air Force Research Laboratory | AFRL Materials Characterization Facility | MCF Operated.
X-Ray Microanalysis – Precision and Sensitivity Recall… wt.fraction I = I SiKα (unknown) / I SiKα (pure std.) K-ratio I = [I SiKα (unknown ) / I SiKα (std.)
Electron probe microanalysis Low Voltage SEM Operation Modified 9/23/10.
Instrumental Methods: Intro
The probe Some material used from:. EPMA - electron probe microanalysis Probe signals.
Measurements and Their Analysis. Introduction Note that in this chapter, we are talking about multiple measurements of the same quantity Numerical analysis.
The Significance of Refractive Index “K” Values LSMs, or, Why Long Wavelength Peak Markers Don’t Line Up Correctly John Fournelle* and K. Bolger*, J. Cook*,
Analysis of Fluorine in Fe-bearing phases: Problems and solutions John Fournelle Eugene Cameron Electron Microprobe Lab Department of Geology & Geophysics.
MAN versus Off-peak Background Measurements
Spatial Resolution and minimum detection
GOVT 201: Statistics for Political Science
An in-depth look at trace metallic impurity determination by ICP-OES
Step 1: Specify a null hypothesis
Physics 114: Lecture 13 Probability Tests & Linear Fitting
Linear Algebra Review.
DTC Quantitative Methods Bivariate Analysis: t-tests and Analysis of Variance (ANOVA) Thursday 20th February 2014  
Lecture 4a Common Error in EPMA: Secondary Fluorescence from Outside Primary Excitation Volume John Fournelle, Ph.D. Department of Geoscience University.
TYPES OF GRAPHS There are many different graphs that people can use when collecting Data. Line graphs, Scatter plots, Histograms, Box plots, bar graphs.
Electron probe microanalysis EPMA
The Problem of Secondary Fluorescence in EPMA in the Application of the Ti-in-Zircon Geothermometer And the Utility of PENEPMA.
Electron probe microanalysis - Scanning Electron Microscopy EPMA - SEM
Statistical Data Analysis - Lecture10 26/03/03
Instrumental Methods: Intro
Distribution of the Sample Means
iCAP OES Analysis of Trace Elements in Hair
Electron probe microanalysis EPMA
Preface: What’s EPMA all about? How does Geology 777 work?
By C. Kohn Waterford Agricultural Sciences
The Analysis of Soils and Waters in Accordance with U. S
Pamela Aaron, Dr. Charles Shearer, and Paul Burger
Introduction to Summary Statistics
Histograms: Earthquake Magnitudes
Electron probe microanalysis EPMA
ID1050– Quantitative & Qualitative Reasoning
Introduction to Summary Statistics
Chapter 5 Quality Assurance and Calibration Methods
Electron probe microanalysis EPMA
Electron Probe Microanalysis EPMA
Electron probe microanalysis EPMA
Using Statistical techniques in Geography
Electron Probe Microanalysis EPMA
Displaying and Summarizing Quantitative Data
Electron probe microanalysis EPMA
Electron probe microanalysis EPMA
Electron probe microanalysis EPMA
Electron probe microanalysis - Scanning Electron Microscopy EPMA - SEM
Electron probe microanalysis EPMA
Electron probe microanalysis
Chapter 9 Hypothesis Testing: Single Population
Computed Tomography (C.T)
Data Literacy Graphing and Statisitics
MGS 3100 Business Analysis Regression Feb 18, 2016
Presentation transcript:

Electron probe microanalysis EPMA Trace Element Analysis Mod 1/10/2016

What kind of confidence level should be place on such a number? What’s the point? What’s the minimum detection limit for a particular element – or said otherwise, at what point can be be sure that a small inflection above the surrounding background really is a peak? What kind of confidence level should be place on such a number?

“No” Zn ... but at what level of confidence? Definitions “Generally, WDS can achieve limits of detection of 100 ppm in favorable cases, with 10 ppm in ideal situations where there are no peak interferences and negligible matrix absorption.” (Goldstein et al., p. 341) Major >10 wt% Minor 1-10 wt% Trace <1 wt% “No” Zn ... but at what level of confidence?

Trace elements .... and trace elements In the real world, the definition of “trace element analysis” is sometimes broader than the strict quantitative analysis of ppm level elements in one microvolume (~micron interaction volume). Many individuals desire to use EPMA to tell them about the “distribution of trace elements” in their materials, e.g. where the 30 ppm of Pb is in a cast iron. There are two possibilities: the 30 ppm is spread uniformly throughout the material, or in fact most of the material has probably <1 ppm of Pb, but a small fraction of the volume has phases that have Pb at major element levels.The question is then, are they at least the size of the interaction volume and if so, where are they. Our discussion here will deal with all these aspects.

A little background-1 Interest in trace elements dove tails with the develop of techniques that could achieve better/quicker/cheaper/more precise/small volumes of said elements. From the 1960s on, geochemists and petrologists developed increasing interest in trace element partitioning between fluids/melts and minerals. The electron microprobe became the instrument of choice for characterizing the trace levels in doped experiments . There has been an interest in “trace elements” in certain minerals to assist in the search for ore bodies that contain said elements. A related research field is locating the naturally-occurring minerals that are responsible for certain levels of groundwater contamination (e.g. As).

A little background-2 Also… In both material science and geology, diffusion processes are studied, and EPMA is a prime technique. As you go further and further from the boundary between the two phases, the elements drop to trace element levels. But where do they drop below detection? Or how do you set up the EPMA conditions to reach down to a desired very low level?

How low can we go? The USNM olivine standard above (San Carlos, Mg.9 Fe.1SiO4) has a published Ca content of “<0.04 wt% (= 400 ppm). This scan was acquired at 20 keV, 30 nA, with 10 seconds per channel. Clearly there is a peak at the Ca Ka position (24 cps), somewhat above the background (~10 cps). At what point can we say with 99% confidence that there is a statistically significant peak?

MDL Equations - 1 The key concept here is minimum detection limit (MDL), i.e., what is the lowest concentration of the element present that is statistically above the background continuum level by 3 sigma (commonly accepted level). There are (at least) two equations used to define the MDL: the first* uses the Student t test values: where the detection limit CMDL is in wt%, CS=std wt %, bar NS=ave. peak cts on std, bar NSB=bkg cts on std, SC= std dev of measured values and n=number of data points the second†, which is probably more wider used, was developed by Ziebold (1967): where n=number of measurements, T=seconds per measurement, P=pure element count rate, P/B= for pure element, and a=matrix correction (a -factor or ZAF). * Goldstein et al., p. 500, equation 9.84 † Goldstein et al., p. 500, equation 9.85

MDL Equations - 2 There are several points to be made about these two equations: the first (Student t test) equation only works for the average of several measurements, since it uses SC, the average of several measurement. This calculation is useful in that special case. however, as many times as not, a specific area or region is only measured once (e.g., a linear traverse across a zoned crystal), and the second equation is the appropriate one to use. note in the second equation, the term P times P/B appears in the denominator. As P2/B increases, the MDL decreases (the lower, the better!). This P2/B term is called the ‘figure of merit” for trace element work. following some discussions with John Valley about the traditional (second) equation and how the peak and background used in it were from the pure element standard—not the unknown, I went back to first principles and derived the equation.

Deriving the MDL equation-1 1. We need to determine the precision of the background value for the unknown, i.e. for a given background value, how big is the statistical error bar (counting statistic, 3 sigmas) above it. So here, is it large as on the right below, or is it small as on the LEFT?

Deriving the MDL equation-2 2. Let us consider Ca Ka peak on our olivine. We measure the background and get 9 cts/sec. The 1 sigma value however is calculated from TOTAL counts, NOT count rate. So we must multiply 9 cts/sec by the time, 10 seconds, and we get 90 counts. 1s=Sq.Rt. of 90 =9.5 counts, so 3s =28.5. We’ll use 3 s for now, the 99.7% confidence level. Ergo, our MDL for Ca in the olivine is 29 counts above background, over 10 seconds (or if plotted on the wavescan where data are in cts/sec, it would be a value of 3 cps (the left purple marker). Note: we haven’t said one word about count rate on a standard, and we have figured out the minimum detection limit for Ca in our unknown -- though we don’t know what that mdl of 3 cts/sec translates to in ppm or wt%).

Deriving the MDL equation-3 3. However, we usually want to “translate” those raw counts into a more usable number, i.e. so many ppm. For that, we need some reference intensity counts for Ca Ka. We then count Ca ka peak and backgrounds for the same time (10 sec) on CaSiO3 (38.6 wt% Ca) and find a count rate of 6415 cts/sec on the peak and 16 cts/sec on the background. 4. So what is 2.9 cts/sec equal to in elemental wt%? We create a pseudo k-ratio where we take the statistical uncertainty of the background counts (square root, i.e. 1 sigma) divided by the Peak-Bkg of the standard counts on the element peak of interest and multiply by 3 (for 3 sigma, 99% confidence) and the ZAF of each and then by the composition C of the standard.The mdl will be in whatever units C is in.

Deriving the MDL equation-4 This is virtually the same result as the “single line” detection limit provided by Probe for Windows (0.015 wt%, shown on next slide), derived from the Ziebold equation. It would appear that the Ziebold equation is not exactly correct, for we must really be concerned with the background precision of the unknown, and the background level of the standard could be several times higher or lower. Going back and re-reading Ziebold, we find two interesting statements: that the equation “gives a measure of the detectability limit” and “there is more than one way to define a detectability limit”. Both are correct, and yes, the equation gives an approximation of the detection limit -- but not the limit per se.

MDL in olivine - single line Good totals Excellent stoichometry These are the single line detection limits, calculated with Ziebold’s equation (Goldstein, eq 9.85, p. 500)

MDL in olivine - average For a homogeneous sample, it is “legal” to add together all the counts, which gives greater precision and a lower detection limit, e.g. 110 ppm here for Ca at 99% ci. This is a handy chart that shows what kind of counting time would be required, under the same analytical conditions, to achieve a lower detection limit. For example, to get a mdl of 25 ppm, you’d need to count for ~5 minutes on the Ca peak and then background (for each of 10 spots). The current analysis here was a little over 1 minute per spot (thus, about 12 minutes for a mdl of 110 ppm. For 25 ppm, it would add an additional 100 minutes to the analysis time.

“Figure of Merit” Variation of figure-of-merit (P2/F) with accelerating voltage for various elements in different matrices. Scaled so 15 kV=1, no ZAF corrections. Variation of figure-of-merit with accelerating voltage, each element relative to its level at 15 kV. Detection limits (3 sigma) calculated for 100 sec counts, 50 kV, 450 nA. Data are all k-ratios, not ZAF corrected. Probers in Australia have much interest in pushing the lower limits of EPMA detection, for mineral exploration research.Utilizing extreme operating conditions (50 kV, 475 nA, 10 minute counts) they have achieved mdl’s below 5 ppm for some elements. They utilize a “figure of merit” of P2/B as a measure of how to achieve lower mdl (the higher the P2/B ). From our first principles derivation, we can see that the P comes from the standard, the B from the unknown. From Advances in Electron Microprobe Trace-Element Analysis by B.W. Robinson and J. Graham, 1992, ACEM-12

Keys to low detection levels Maximize counts by utilizing Highest currents feasible (concern: beam damage) Highest E0 as feasible (concern: increased penetration/range) Longer count times Correctly determine background locations Be 100% aware of n>1 peaks occurring in neighborhood Correct for unavoidable on-peak interferences within the matrix correction* *Donovan, Snyder and Rivers, 1993, An improved interference correction for trace element analysis, Microbeam Analysis, 2, 23-28.

Silicate Glass with 90 ppm P P Ka peak = red line position. But n=2 Ca Kb1,3 also very close! BIR =Icelandic Basalt with 9.5 wt% Ca (USGS reference material)

Silicate Glass with 90 ppm P n=1 Si Kb1,3 n=1 P Ka1,2 n=3 Fe Ka1,2 n=2 Ca Kb1,3 n=3 Ca Ka1,2

Backgrounds: traces can overlap traces Correct locating of background positions is particularly important in trace element work, as both first order and higher order peaks can cause incorrect assessment of background level. Here, scans of the 3 Caltech/MAS trace element glass standards are overlain. (Xe L edge present as a Xe gas sealed counter used.) From Carpenter, Counce, Kluk, and Nabelek, Characterization of Corning Standard Glasses 95IRV, 95IRW and 95IRX: NIST/MAS Workshop, April 2002.

Backgrounds ... holes Au La Probers in Australia, interested in detecting trace levels of gold in certain minerals, discovered a “hole in the background” about 200 sin theta units below the Au La position. (This scan was on SrTiO3, on the LIF crystal).

2 Examples Ti in quartz Monazite

Ti-in-quartz Multi-point Background Modelling

Ti-in-quartz Multi-point Background Modelling

Ti-in-quartz 2 important options: Aggregate (Sum Up) >1 spectrometer; Blank Correction

Ti-in-quartz After collecting  Assign Blank Correction

Ti on 4 spectrometers: no blank correction

Ti on 4 spectrometers: with blank correction

Ti: All 4 spectrometers aggregated; blank corrected

Monazite Important mineral for age-dating by EPMA But Low amounts of Pb; complicated background

Monazite

Monazite

Monazite

Monazite

Monazite

Monazite EPMA procedure Julien Allaz at University of Colorado has a detailed procedure… contact him Julien.allaz@colorado.edu

Best Trace Element EPMA needs: Strong, flexible, background modelling Interference correction within the matrix correction Ability to add up counts from many spectrometers Blank correction within the matrix correction

Standards: validating trace element procedure There is an issue of trace element accuracy on unknowns, where the standard for the element of interest was at a high level. Such a standard should be used for peaking the spectrometer and acquiring standard counts, but it is recommended that a secondary trace level standard be also analyzed to validate the procedure. Such secondary standards could be Synthetic glasses such as the Caltech/MAS 95IRV,W and X glasses; NIST glasses and metals; Ni-Cr diopside glass, etc. Minerals and glasses analyzed by ion probe

Comparison: Trace elements by WDS vs EDS WDS is clearly the better method for acquiring trace element data, by an order of magnitude or so compared to EDS. Goldstein et al, 1992, p. 501