Detecting and measuring atmospheric hazards in underground mines

Slides:



Advertisements
Similar presentations
Carbon Monoxide Monitoring
Advertisements

Mine Gases Atmospheric Testing C.4. Contributed by: Richard Black Industrial Scientific Corp.
CARBON MONOXIDE.
Sensor Operating Theory
Air Pollution By: JW Option E. What is an Air pollutant? What are the Sources of Air pollution? What are the effects of Air pollution? What are.
MINE GASES.
E1-Air Pollution! Heather Yin Period 3. Why Should I Care?! As humans populate the planet, we produce waste that is absorbed by our atmosphere which directly.
Air Pollution By: JW Option E. What is an Air pollutant? What are the Sources of Air pollution? What are the effects of Air pollution? What are.
Safety Training Presentation
Arson. The Common Accelerants  The most commonly used accelerants are petrol, kerosene, mineral turpentine and diesel.  These accelerants are generally.
November 2009 Principles of gas detection Slide 1 Confined Space Entry.
Topics for Today Pollution, Pollution! (Gaseous Pollutants)
W Each year over 200 people die from CO poisoning and over 5000 are injured. W CO causes more accidental poisonings than any other chemical in the U.
Mine Rescue Gas Detection
Basic Air Monitoring Concepts And Use Of The Altair 4 Gas Meter
Chapter 15 Chemical reactions.  Any material that can be burned to release thermal energy is called a fuel.  Most familiar fuels consist primarily of.
WAN AZIZ ZUL SHAHRIN HAIRI SYUKOR EXHAUST EMISSION POLLUTANTS EFFECT TO HUMAN HEALTH.
Gases Underground Stone and Limestone Mines Objectives 4 Identify mine gases 4 Describe the hazards of mine gases 4 Explain the effects of gas exposures.
Metering and CO.
MULTI-GAS MONITOR FOR CONFINED SPACE APPLICATIONS
WELCOME TO Gas Tester Training O2, LEL Monitoring
CONFINED SPACE We are going to start this class with a small quiz. The purpose of the quiz is to allow you to determine where you are and what you know.
Health Hazards of Gases A review of the toxicities of substances that exist in the gaseous state under normal conditions of use.
CO fact sheet-1 CO diffuses rapidly blood vessel membranes. Once it’s present in the bloodstream, CO binds to hemoglobin 200 times more readily than oxygen.
MINE GASES. Normal Air is made up of: Nitrogen 78% Oxygen 21% Trace Gases 1%
Carbon Monoxide Monitoring Presented by: Interscan Corporation.
1.
1 ENERGY AND ENVIRONMENT The conversion of energy from one form to another often affects the environment and the air we breathe in many ways, and thus.
Power Plant Engineering
CARBON MONOXIDE POISONING. Aviation Toxic Substance Aviation Fuel Combustion Aviation Fuel Combustion - Carbon Monoxide: One of the most common and toxic.
Grade 13 Air Pollution.
Hazards of Confined Spaces. INSTRUCTIONAL GOAL The participant till understand the various hazards associated with confined space entries and the need.
1Korea University of Technology and Education 2 Generation Common Rail VGT Variable Swirl 32Bit Computer Elec. Controlled EGR Flap C P F Electronically.
Bi-Weekly Safety Meeting November 29, The days of using canaries in mines as primitive gas detectors are long gone..they've been replaced by fixed.
Misuse of Combustible Gas Meters Supervisor Training 1/11/08.
AQ Data Collection and Monitoring Emissions. Learning Objectives After this activity, you should be able to: List the products of complete and incomplete.
Chapter 12 - Air SECTION 1 – WHAT CAUSES AIR POLLUTION.
2.2 Monitoring Air Quality and the Atmosphere GO C2Identify processes for Measuring the Quantity of Different Substances in the Environment and for Monitoring.
CHAPTER 6 Gases and Vapors. Learning Objectives Describe the various properties of the chemicals that are most important to the practice of industrial.
Air pollution OSH 320.
Ch. 18 Air Pollution.
ArsonAnalysis of Fire Arson is defined as purposely setting fire to a house, building or other property. Arson is the second leading cause of death by.
Confined Spaces in Construction Awareness Training
Activity 7: SDS Worksheet
Jim Vermeulen MSA Territory Manager
Air Pollution.
Chapter 20 Air Pollution.
Mine Gasses.
Combustion Reactions.
CO Poisoning Huron Hospital and Trauma Center
Montrose Fire Department
Benzene.
Carbon dioxide (CO2) and Carbon monoxide (CO)
Bob Henderson, BS, MBA GFG Instrumentation Inc. Ann Arbor, MI
Recent changes in confined space entry requirements
Carbon Monoxide.
Confined Space Hazards
Emission Control Systems
DALTON’S LAW Total pressure equals the sum of the partial gas pressures in a mix. TP = P1+P2+P3+P4 ETC To find Partial pressure multiply total pressure.
16.3 Categories of Air Pollutants
Air pollution Domina Petric, MD.
16/11/2018 Air Quality (OCR 21st Century) W Richards The Weald School.
LQ: How are pollutants formed?
All About the Air….
Catalytic Converter and Computer Controlled Emissions
Recent changes in the rules for confined space entry
Carbon Monoxide Poisoning
Gas Monitoring Basics Presented by Dan Klimek Sisk & Company.
Carbon Monoxide Monitoring
Major Air Pollutants.
Presentation transcript:

Detecting and measuring atmospheric hazards in underground mines Robert E. Henderson President GfG Instrumentation, Inc. 1194 Oak Valley Drive, Suite 20 Ann Arbor, Michigan, USA , 48108 E-mail: bhenderson@gfg-inc.com Website: www.goodforgas.com Mobile: +1-734-355-1291

Technical support and downloads www.goodforgas.com Application Notes, Technical Notes and Presentations: http://goodforgas.com/support/ #appnotes Complete AIHA PDC: Methods and Applications for Chemical Detection Real-Time http://goodforgas.com/wp- content/uploads/2014/01/PDC- 424-Methods-Applications- Chemical-Detection-in-Real- Time.pdf

Common atmospheric hazards Oxygen deficiency Oxygen enrichment Presence of toxic gases Presence of combustible gases

Oxygen (deficiency and enrichment) Any area that has an oxygen level of less then 19.5% by volume is considered to be oxygen deficient 19.5%

Composition of fresh air 78.1 % Nitrogen 20.9 % Oxygen 0.9 % Argon 0.1 % All other gases Water vapor CO2 Other trace gases

Partial pressure O2 vs. percent volume at varying altitudes Height Atm. Pressure PO2 Con. feet meters mmHg kPa % Vol   16,000 4,810 421.8 88.4 11.8 20.9 10,000 3,050 529.7 111.0 14.8 5,000 1,525 636.1 133.3 17.8 3,000 915 683.3 143.3 19.1 1,000 305 733.6 153.7 20.5 760.0 159.2 21.2 19.5% O2 at sea level = 18 kPa

Effects of oxygen at various concentrations > 23% Oxygen enrichment 20.90% Normal air concentration 19.50% Minimum “safe level” 16% First sign of anoxia appears 16 – 12% Breathing and pulse rate increase, muscular co-ordination is slightly impaired 14 – 10% Consciousness continuous; emotional upsets, abnormal fatigue upon exertion, disturbed respiration 10 – 6% Nausea and vomiting, inability to move freely and loss of consciousness may occur < 6% Convulsive movements and gasping occurs, respiration stops

Most O2 sensors have a “capillary pore” used to allow sensor to self-stabilize at new pressure O2 sensors with capillary pore are true percent by volume measurement devices Are able to self stabilize to changes in pressure due to: Barometric pressure Pressurized buildings Altitude Stabilization at new pressure is not instantaneous, may take 30 seconds or longer Capillary pore (located under external moisture barrier filter)

Effects of changes in pressure on O2 sensor readings Readings from instrument taken through negative pressure airlock at a nuclear generating station Readings recovered to above 19.5% O2 within 55 seconds. O2 sensor took 3.3 minutes to stabilize at 20.5% O2 Readings eventually reached 20.9% after about 10.5 minutes

O2 concentration in fresh air as function of temperature and humidity O2 in dry air = 20.95% O2 at 20°C and 25% RH ≈ 20.8% O2 at 40°C and 85% RH ≈ 19.7% O2 at 40°C and 90% RH ≈ 19.5% O2 at 40°C and 100% RH ≈ 19.4% What should you do if the instrument is used in high temp, high RH conditions? Use cylinder of “Zero Air “ to fresh air adjust the O2 sensor.

Oxygen Enrichment Proportionally increases rate of many chemical reactions Can cause ordinary combustible materials to become flammable or explosive Any area with an O2 level of more than 23.0% is dangerously enriched 23.0%

Heavy vehicle exhaust hazards Exhaust gas from diesel and gasoline engines primarily consists of nitrogen (N2), water vapor (H2O), and carbon dioxide (CO2) A relatively small part of the exhaust consists of toxic materials such as: Particulate contaminants (soot) Carbon monoxide (CO) from incomplete combustion Hydrocarbons from unburnt fuel VOCs from incomplete combustion Nitrogen oxides (NOx) Sulfur oxides (SOx) from fuel impurities

Diesel engine exhaust By nature, diesel engines run “lean” rather than rich Fuel must be injected, vaporized and heated in order for combustion to occur Diesel engines consume oxygen! When warm, produce very little CO Diesel engines produce: Carbon particulates (soot) Hydrocarbons (from unburned fuel) VOCs and other combustion by-products Often adsorbed on particles of soot, increasing danger of respirable particles carrying toxic VOCs deep into lungs Carbon dioxide (CO2) Oxides of nitrogen (NOx) Especially when the engine is running hot, or in reduced levels of O2 Oxides of sulfur (SOx) Especially when the fuel contains high concentrations of sulfur

An O2 reading lower than 20.9% indicates there is too much of some other gas present in the atmosphere

In this example as O2 reading drops CO concentration rises

Although O2 never dropped below 19 Although O2 never dropped below 19.5%, CO concentration reached alarm level more than once

CO2 (not CO) actually the primary contaminant replacing the O2 in the monitored atmosphere

Important to directly measure all the contaminants that can materially affect the atmosphere

Effect of displacing gas on oxygen concentration Be very cautious when using O2 concentration to estimate concentration of some other displacing gas Every 5% of displacing gas introduced into a confined space reduces O2 concentration by only about 1%

Measuring specific toxic exhaust gases Carbon dioxide (CO2) CO2 is not only an asphyxiant; it is a toxic gas! Carbon monoxide (CO) Sulfur oxides (SOx) Primary gas of interest is sulfur dioxide (SO2) Nitrogen oxides (NOx) Primary gases of interest are nitric oxide (NO) and nitrogen dioxide (NO2) Volatile organic chemical (VOC) vapors

CO2 Properties Present as a natural component in fresh air (approximately 350 ppm) Colorless Odorless Tasteless Heavier than air (density of 1.5 times that of fresh air) When released into enclosed space it tends settle to bottom Because of tendency to settle, as CO2 produced it can reach higher and higher concentrations

CO2 exposure symptoms Besides displacing oxygen in fresh air, high concentrations may worsen symptoms related to oxygen deficiency, and interfere with successful resuscitation Exposure Symptoms include Headaches Dizziness Shortness of breath Nausea Rapid or irregular pulse Depression of central nervous system

Even moderate concentrations of CO2 can produce symptoms 350 – 400 ppm Normal background concentration in outdoor ambient air 350 – 1,000 ppm Concentrations typical of occupied indoor spaces with good air exchange 1,000 – 2,000 ppm Complaints of drowsiness and poor air 2,000 – 5,000 ppm Headaches, sleepiness, and stagnant, stale, stuffy air. Poor concentration, loss of attention, increased heart rate and slight nausea may also be present >5,000 ppm Exposure may lead to serious oxygen deprivation resulting in permanent brain damage, coma and even death

CO2 is toxic contaminant with strictly defined exposure limits Standard / Country 8-hour Time Weighted Average 15-minute Short Term Exposure Limit USA NIOSH REL 5,000 ppm 30,000 ppm USA OSHA PEL None Listed ACGIH® TLV® 8-hour TWA limit is 5,000 ppm 15-minute STEL limit is 30,000 ppm. IDLH concentration is 40,000 ppm Concentration of 100,000 ppm leads to loss of consciousness within 15 minutes or less High altitude makes symptoms worse!

Non-dispersive infrared (NDIR) sensors Infrared sensors shine a beam of IR light through a sensing chamber Molecules present in the chamber absorb some of the IR The rest of the light is transmitted through the chamber without hindrance The presence of a chemical groups within the molecule determine the wavelengths where IR light is absorbed Non-dispersive IR sensors measure at a specific range of wavelengths associated with a particular gas or class of gases

2 4 Carbon monoxide Colorless Odorless Slightly lighter than air By-product of combustion Flammable (LEL is 12.5%) Toxic! 2 4

Carbon monoxide Bonds to hemoglobin in red blood cells Contaminated cells can’t transport O2 Chronic exposure at even low levels harmful, especially if there are underlying medical conditions, or when there are additional factors (such as heat stress)

Toxic effects of CO Toxic effects of carbon monoxide 25 ppm TLV exposure limit for 8 hours (TWA) 200 ppm Possible mild frontal headaches in 2-3 hours 400 ppm Frontal headaches and nausea after 1-2 hours. 800 ppm Headache, dizziness and nausea in 45 min. Collapse and possibly death in 2 hours 1,600 ppm Headache and dizziness in 20 min. Unconsciousness and danger of death in 2 hours 3,200 ppm Headache and dizziness in 5-10 min. Unconsciousness and danger of death 30 min. 6,400 ppm Headache and dizziness in 1-2 min. Unconsciousness and danger of death 10-15 min 12,800 ppm Unconsciousness immediately, danger of death in 1-3 min.

Exposure limits for carbon monoxide OSHA PEL: 50 ppm 8-hr. TWA NIOSH REL: 35 ppm 8-hr. TWA 200 ppm Ceiling TLV: 25 ppm 8-Hr. TWA 2 4

Substance-specific electrochemical sensors Electrochemical sensors are used to measure CO, SO2, NO, NO2 and many other gases Gas diffusing into sensor reacts at surface of the sensing electrode Sensing electrode made to catalyze a specific reaction Use of selective external filters further limits cross sensitivity

Typical electrochemical sensor detection reaction CO Sensor: Carbon monoxide is oxidized at the sensing electrode: 4CO + 4H2O 4CO2 + 8H+ + 8e- Oxygen (from the air) is reduced at the counter electrode: 2O2 + 8H+ + 8e- 4H2O Thus the overall reaction is: 4CO + 2O2 4CO2 2CF CO sensor signal output: 50 ± 20 nA / ppm CO

Exposure limits for NO2 US OSHA PEL: Ceiling = 5 ppm US NIOSH REL: 15 min. STEL = 1 ppm Old TLV: 8 hr. TWA = 3 ppm 5 min. STEL = 5 ppm New 2012 TLV 8 hr. TWA = 0.2 ppm

Suggested alarm settings for NO2 NIOSH: Low: 3.0 ppm High: 5.0 ppm STEL: 1.0 ppm TWA: 1.0 ppm TLV®: Low: 0.6 ppm High: 1.0 ppm STEL: 0.2 ppm TWA: 0.2 ppm

Exposure limits for sulfur dioxide SO2 OSHA PEL: TWA = 5.0 ppm NIOSH REL: TWA = 2.0 ppm STEL = 5.0 ppm Old TLV : TWA = 2 ppm STEL = 5 ppm New (2009) TLV: STEL = 0.25 ppm

Suggested alarm settings for SO2 NIOSH: Low: 2.0 ppm High: 5.0 ppm STEL: 5.0 ppm TWA: 2.0 ppm TLV®: Low: 0.75 ppm High: 1.25 ppm STEL: 0.25 ppm TWA: 0.25 ppm

Explosive or Flammable Atmospheres

Measuring combustible gases and vapors

Catalytic “Hot Bead” Combustible Sensor Detects combustible gas by catalytic oxidation When exposed to gas oxidation reaction causes the active (detector) bead to heat Requires oxygen to detect gas! D.C. voltage supply Output - + Compensator Detector Signal Trimming resistor R1 R2 VR1 +VS -VS

Catalytic Sensor Structure

Typical carbon number distribution in No. 2 Diesel Fuel (liquid) Less than 2% of molecules in diesel vapor are small enough to be measured by means of standard LEL sensor

Combustible sensor limitations Contaminant LEL (Vol %) Flashpoint Temp (ºF) OSHA PEL NIOSH REL TLV 5% LEL in PPM Acetone 2.5% -4ºF (-20 ºC) 1,000 PPM TWA 250 PPM TWA 500 PPM TWA; 750 PPM STEL 1250 PPM Diesel (No.2) vapor 0.6% 125ºF (51.7ºC) 15 PPM 300 PPM None Listed None Listed Ethanol 3.3% 55ºF (12.8 ºC) 1,000 PPM TWA 1000 PPM TWA 1000 PPM TWA 1,650 PPM Gasoline 1.3% -50ºF (-45.6ºC) None Listed None Listed 300 PPM TWA; 500 PPM STEL 650 PPM n-Hexane 1.1% -7ºF (-21.7 ºC) 500 PPM TWA 50 PPM TWA 50 PPM TWA 550 PPM Isopropyl alcohol 2.0% 53ºF (11.7ºC) 400 PPM TWA 400 PPM TWA; 500 PPM STEL 200 PPM TWA; 400 PPM STEL 1000 PPM Kerosene/ Jet Fuels 0.7% 100 – 162ºF (37.8 – 72.3ºC ) None Listed 100 mg/M3 TWA (approx. 14.4 PPM) 200 mg/M3 TWA (approx. 29 PPM) 350 PPM MEK 1.4% 16ºF (-8.9ºC) 200 PPM TWA 200 PPM TWA; 300 PPM STEL 200 PPM TWA; 300 PPM STEL 700 PPM Turpentine 0.8 95ºF (35ºC) 100 PPM TWA 100 PPM TWA 20 PPM TWA 400 PPM 100 PPM TWA; 150 PPM STEL Xylenes (o, m & p isomers) 0.9 – 1.1% 81 – 90ºF (27.3 – 32.3 ºC) 100 PPM TWA 100 PPM TWA; 150 STEL 450 – 550 PPM

Volatile organic compounds (VOCs) VOCs are organic chemicals or mixtures characterized by tendency to evaporate easily at room temperature Familiar VOCs include: Solvents Paint thinner Nail polish remover Gasoline Diesel Heating oil Kerosene Jet fuel Benzene Butadiene Hexane Toluene Xylene Many others

VOCs are measured with a photoionization detector (PID) sensor

Photoionization Detectors

Catalytic (CC) LEL vs. PID Sensors Catalytic LEL and photoionization detectors are complementary detection techniques Catalytic LEL sensors excellent for measurement of methane, propane, and other common combustible gases NOT detectable by PID PIDs detect large VOC and hydrocarbon molecules that are undetectable by catalytic sensors Best approach to VOC measurement is to use multi-sensor instrument capable of measuring all atmospheric hazards that may be potentially present

Instruments should be tested by exposure to calibration gas on a regular basis! Gas detection instruments can only keep miners safe when they are maintained and used properly Mining regulations in the USA, Canada, Europe and South Africa require that gas detectors must be tested at the mine by exposure to known concentration gas before each day’s use No exceptions!

Safest approach is to perform a bump test before each day’s use! At this mine, the only way miners can go through the turnstile is with an instrument that has been successfully tested by exposure to gas within the last 24 hours The personally assigned instrument automatically activates the turnstile release, which logs the ID and tracks miners entering and leaving the mine

Thank you for attending! Questions? Thank you for attending!